PELV Test

The PELV (Protective Extra Low Voltage) test measures
the voltage on the PROBE PELV input when the idle
screen is being displayed.
To perform the PELV test, do the following:
1
Revert to the idle screen if it is not already
STOP
being displayed.
2
Connect the test probe to the tester PROBE
PELV terminal and connect the appliance to
a mains supply socket.
3
Apply the test probe to the test point.
4
Store the test result, if required.
MEM
28
The display can show the following specific information:
PELV FAIL result, the
PELV
30.0 V
threshold (25V) is exceeded.
PELV overrange.
PELV > 39.9 V
If the PELV threshold is not
230 V
exceeded the display shows
50 Hz
the mains voltage and
frequency (PASS result)
Tip:
To store a PELV FAIL result press
To store a PELV PASS result press
item "SAVE PELV pass result".
and
.
GO
MEM
and select menu
MEM