4. Press the two defibrillator paddles onto the contact electrode plates on the
front of the Analyzer.
5. Press the START softkey and initiate the defibrillator charge cycle.
6. As soon as the defibrillator reaches full charge, discharge it, noting the
time (in seconds) on the display. The maximum for the Analyzer is 60
seconds. After 60 seconds, the Analyzer displays OVER.

Pacemaker (Non-Invasive) Testing

1. From Main Menu 1, press more to access Main Menu 2. The following
displays:
2. Set the range switch to pacer. Otherwise the unit displays the following
message:
3. From Main Menu 2, press the PACE softkey to display the Pacer
Tests menu:
4. Connect the output from the pacer to the Analyzer, as shown in Figure
2-9.
2-22