Atten AT7016 User Manual

Single trace 10mhz oscilloscope

Advertisement

Quick Links

ATTEN
AT7016 SINGLE TRACE
10MHz OSCILLOSCOPE
USER'S MANUAL

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the AT7016 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Atten AT7016

  • Page 1 ATTEN AT7016 SINGLE TRACE 10MHz OSCILLOSCOPE USER’S MANUAL...
  • Page 2: Table Of Contents

    AT7016 User Manual                                                                                                                                 CONTENTS ...
  • Page 3: Safety Precautions

    AT7016 User Manual 1. SAFETY PRECAUTION The AT7016 oscilloscope is easy to operate and highly reliable. It is an ideal instrument for research, production, education, and development. It is a compact and portable oscilloscope with a frequency bandwidth of 10MHz and sensitivity of 5mV/DIV-5V/DIV. Supplied as standard, with a switchable X10: X1 probe which extends the sensitivity up to 50V/div.
  • Page 4: Specifications

    AT7016 User Manual Do not place this instrument on an unstable or slanting surface . The dropping or turning over of this instrument could result in electrical shock, injury or fire. If this instrument has been dropped or its cover has been damaged, switch the power off, remove the power cord and call for repair. Do not allow any foreign matter such as metal or inflammable substance to get into the instrument via the air holes.
  • Page 5: Power Source

    AT7016 User Manual Accelerating Voltage 1200V Display Color Green 2.7 POWER SOURCE Voltage Range 110V±10% 220V±10% Frequency 50Hz ±2Hz Power Consumption 2.8 PHYSICAL FEATURES Weight Dimensions (H x W x D) 190 ×130 ×270mm 2.9 WORKING ENVIRONMENT Working temperature 5°C ~ 40°C Storage Environment °C ~60°C, 10~80%RH Working Altitude...
  • Page 6: Control And Indicators

    AT7016 User Manual 3. CONTROL AND INDICATORS 3.1 Control Panel Position 3.1.1 Front Panel Fig 3-1 3.1.2 Rear Panel Fig 3-2...
  • Page 7: Functions Of Control Switches

    AT7016 User Manual 3.2 FUNCTIONS OF CONTROL SWITCHES NO. SWITCHES FUNCTIONS POWER SWITCH Power on/off POWER LIGHT Lights when power on INTENSITY Controls brightness of display After obtaining appropriate brightness with INTENSITY, adjust FOCUS FOCUS for clearest line Provide symmetric square wave for 0.5V range, CALIBRATION frequency=1KHz.
  • Page 8: Operating Instructions

    AT7016 User Manual 3.3 OPERATING INSTRUCTIONS 3.3.1 VOLTAGE CHECKING AT7016 oscilloscope is set for 110V±10% , 220V ±10% voltage. Before connecting the unit to the mains supply, make sure that the correct voltage is being used. Incorrect mains voltage may cause damage to the instrument.
  • Page 9 AT7016 User Manual Fig 3-3 3.3.3 VERTICAL SYSTEM OPERATION (1) VOLTS/DIV switch should be turned to the correct position following the input signal range. Adjust position [16] to show the whole waveform within the available area. Adjust with VAR [17] if necessarily, trimming ratio is 2.5:1.
  • Page 10: Measurement

    AT7016 User Manual 3.3.6 SIGNAL CONNECT (1) Probe operation : Use 10:1 to 1:1 switchable probes. When using 10:1 probe, input impedance is 10MΩ 16 pF. While if 1:1 is used for observing a signal, input impedance is 1MΩ 30 pF. At this stage, please consider the affect of the probe in certain circuits.
  • Page 11 AT7016 User Manual LF COMP Fig 4-2 4.2 MEASUREMENT 4.2.1P-P Voltage Measurements Step : (1) Input signal to INPUT [19] terminal. (2) Setup VOLTS/DIV and observe waveform, set waveform display on the screen within 5 divisions, and turn VAR clockwise to the calibration position. (3) Adjust level to make waveform steady.
  • Page 12: Time Measurements

    AT7016 User Manual 4.2.2 DC VOLTAGE MEASUREMENT STEP: (1) Setup front panel connector to obtain a sweep baseline on the screen. (2) Setup input coupling options as “⊥”. (3) Setup POSITION, let sweep baseline to coincide with horizontal center, define it as the zero reference level.
  • Page 13 AT7016 User Manual Horizontal factor Shown in figure 4-6, distance between A & B is 8 Div. sensitivity is 2 µS/Div, Horizontal factor x 1, then Time measurement is 16µS Fig 4-6 4.3.2 CYCLE & FREQUENCY MEASUREMENTS Shown in Fig 4-6, frequency measurements are made by measuring the time period of one cycle of waveform (T), and calculating the frequency that equals the reciprocal of the time period.
  • Page 14: Y Mode Applications

    AT7016 User Manual Horizontal distance (div) × sensitivity (Time/div) Rise Time = ----------------------------------------------------------- Horizontal factor For the example shown in Fig.4-7, the horizontal distance from 10% to 90% is 2.4 divisions, the sweep TIME/DIV setting is 1µS/DIV, factor x 1. The rise time is calculated as follows: 1µS/DIV ×...
  • Page 15 AT7016 User Manual WARNING DO NOT OPEN THE CASE, HIGH VOLTAGE EXISTED. SHENZHEN ATTEN ELECTRONICS CO.,LTO Building A29,Tanglang Industrial Zone,Xili,Nanshan, Shenzhen,518055,P,R,China Tel:86-755-86021376 Fax:86-755-86021337 E-mail:atten@atten.com.cn...