Fluke SigmaPace 1000 Safety Sheet page 2

External pacemaker analyzer
Hide thumbs Also See for SigmaPace 1000:

Advertisement

Defibrillators and
Transcutaneous Pacemakers
 Caution
To avoid possible damage the
Analyzer, do not discharge
defibrillator pulses into the
instrument.
 Warning
To avoid injury to a patient, do not
connect the Analyzer to a patient or
equipment connected to a patient.
The Analyzer is intended for
equipment evaluation only and
should never be used in diagnostics,
treatment or in any other capacity
where the Analyzer would come in
contact with a patient.
Ventilation
 Caution
To avoid heat buildup damage to the
Analyzer, do not block the ventilation
slots during use or while the
instrument is charging.
Avoiding Damage
 Caution
To avoid damage to the Analyzer or
adverse affects on its performance,
do not expose the system to
temperature extremes. Ambient
temperatures should remain between
18 and 40 °C. System performance
may be adversely affected if
temperatures fluctuate above or
below this range.
Cleaning
 Caution
To avoid damage to the Analyzer or
adverse affects on its performance,
clean it only by gently wiping down
with a clean, lint-free cloth dampened
with a mild detergent solution. Do
not immerse the unit.
Service and Calibration
 Caution
To avoid damage to the produt or
adverse affects on its performance,
allow only qualified technical
personnel to service the Analyzer.
SigmaPace 1000
Safety Sheet
2