Fluke 105B User Manual page 149

Scopemeter
Table of Contents

Advertisement

8 - 6
6.
Use
to highlight 10:1, and press
select.
7.
Use
to select CALIBRATE PROBE on A, and
press
to calibrate.
The test tool outputs a voltage on the GENERATOR OUT
banana jack to the connected probe.
8.
Adjust the trimmer screw in the probe housing of
INPUT A to give an optimum square wave within the
equal markers on the display (see Figure 8-3).
Figure 8-3. Adjusting the Probe
9.
Press
to continue.
The test tool automatically calibrates itself to the probe.
A message appears on the display indicating that
calibration is successfully completed.
10. Press
to close the procedure.
to
INPUT B Probe Calibration
1.
Connect the INPUT B BNC input to the GENERATOR
OUT red banana jack. See Figure 8-4. Use the grey
scope probe and the grey 4-mm banana adapter.
Figure 8-4. INPUT B Probe Calibration Setup
2.
Repeat steps 2 trough 10 of the INPUT A calibration.
92B/96B/99B/105B
Users Manual