Trouble
Cannot calibrate
Reading is always
slowly changing,
won't stabilize.
Reason
Incorrect standard solutions
Bad standard solutions
Contaminated probe
Aged probe
Dried-out probe
Probe is not fully submerged
in the solution
Air bubbles around the
sensor shield
Contaminated probe
Clogged junction
Aged probe
Testing purified water like
tap water, drinking water
and RO water
Testing low ion
concentration solutions e.g.
distilled water and deionized
water
How to fix
When powered on, calibrate pH 7 first, then
pH 4. After pH 4 is calibrated, if you want to
calibrate pH7 again, you need to power off
and restart the tester.
Replace with a new reliable buffer solution.
Clean the electrode, it is best to use a cleaning
solution.
Replace the electrode.
Soak in the soaking solution overnight.
Make sure the electrode is immersed in the
solution at least 3 cm.
Make a quick stir in the solution to remove air
bubbles.
Clean the electrode, it is best to use a cleaning
solution.
Clean the electrode with cleaning solution,
then soak it in 3M KCL storage solution
overnight.
Replace the electrode.
Be patient, wait for 2-5 minutes to fully
stabilize. If still not stabilizing, add in 1ml of
3M KCL solution to 1000ml of test solution.
Contact Apera for a specialized pH electrode
suitable for low ion concentration solutions.

Advertisement

loading
Need help?

Need help?

Do you have a question about the PH60-MS and is the answer not in the manual?

Questions and answers

Summary of Contents for Apera Instruments PH60-MS

  • Page 1 Trouble Reason How to fix When powered on, calibrate pH 7 first, then pH 4. After pH 4 is calibrated, if you want to Incorrect standard solutions calibrate pH7 again, you need to power off and restart the tester. Bad standard solutions Replace with a new reliable buffer solution.
  • Page 2 If you don’t find any visible damage of the electrode and it’s within the 6-month electrode warranty, contact us for warranty Broken probe fulfillment; Display similar readings in any If there is visible damage, replace the solutions or always electrode. display 7.0 pH Instrument defect Contact us for warranty fulfillment...
  • Page 3 Test strips or drop tests’ accuracy is not comparable to pH meters’. Make sure your calibration standards are Calibration standards are not fresh and clean, and made by a legitimate in good condition scientific equipment manufacturer. Your electrode is not suitable Contact us to find the most appropriate for your test sample or electrode for your specific application.