Solo/Solo+ ® OPERATING HANDBOOK This document is not contractual and the equipment specification may be modified at any time without prior notice. sales@gds-technologies.co.uk +44 (0)1 13 286 0166 www.gds–technologies.co.uk GDS TECHNOLOGIES LTD FUSION POINT ASH LANE GARFORTH LEEDS LS25 2GA...
Contents System Overview ...............................3 Sample Line Identification ............................4 Translator Control / Flow Monitor ...........................5 System Components ..............................6 GasVac Solo – Data ............................7–10 ® 15/30J Flammable Sensor – Adjustments (C1883) ..................1 1–12 15/30J Toxic/Oxygen Sensor – Adjustments (C1882) ................13–14 Gas Sensor – Sample Point Locations (G973) .....................15 Sample And Test Gas Line Installation (C1892) ....................16 Maintenance ................................17 Notes:...
System Overview The unit has a unique system controller which enables the three main functions to be adjusted according to site conditions and sampling requirements. SAMPLE TIME (minutes) DRAIN CYCLES See Fig.1 Is the period were the sensors are exposed to the sample gas. Switches / Cycles Following an initial 45 second sample draw period, sensor 1 = 8 cycles...
GAS VAC SOLO Sample line and gas reading Sample Line Identification – Solo+ (4-20mA signal outputs) Line Identification 4 – 20mA Signal Output. Sensor output 5 translator board Fig.3 (2mA Increments) 20 mA Line Number Selected sample time period Gas Signal mA Fig.4 20 mA Example:...
System Components 1-5 Gas sensor transmitters (1-4 Solo+) Fig.6 Sensor cell flowblock Pump Unit Translator PCB (PCB 364) Sensor Sample pump unit – page 5 Auto drain unit (peristaltic pump) 10 Solonoid valve / purge Sensor 1 1 Condensate trap 12 Power supply 13 Field terminals 14 1-8 sample line valves (Solo+)
Page 7
GasVac Solo+ ® MULTI LINE GAS SAMPLE UNIT T E C H N O L O G I E S Specification Sheet C1856Av.4 • Low cost, high reliability • Monitor up to 4 gas types •1~8 sample lines • Switch selectable, sample, purge and drain times •...
Page 8
GasVac Solo+ ® MULTI LINE GAS SAMPLE UNIT T E C H N O L O G I E S Specification Sheet C1856Av.4 Weight 25kg Selectable purge/sample times – minutes SOLO+ 1, 2, 4, 8, 16, 32, 64, 128 Selectable drain cycles Internal or remote 1, 2, 4, 8 CAN status indicator...
Page 9
GasVac Solo ® SINGLE LINE GAS SAMPLE UNIT T E C H N O L O G I E S Specification Sheet C1814 Issue Av.4 • Low cost, high reliability • Monitor up to 5 gas types • Single line sampling •...
GasVac Solo ® SINGLE LINE GAS SAMPLE UNIT T E C H N O L O G I E S Specification Sheet C1814 Issue Av.4 MPS Module Network Maximum number –10 All weather – IP66 Orientation – any Module size – 100mm (L) x 32mm (D) Specials –...
Page 11
Storage – 2,880 readings Requires USB A to USB micro B cable Sensor Voltage DISPLAY CONTRAST PROG SENSOR GDS Terminal (GDS part No. 160-511) VOLTAGE Connection (Micro USB B) PC or laptop (dedicated) Zero Pot (4mA) GDS Terminal ZERO (download from GDS website)
Page 12
9. Connect PC to sensor via GDS terminal using USB-A to USB Micro-B readings to display, up to 2880 cable (GDS part No. 160-51 1) at 4800 baud connected to J1 1. The % = Clear gas log Set all 2880 log readings to 0.00...
Connection CONTRAST (Micro USB B) ZERO Requires USB A to USB micro B cable Zero Pot GAIN (4mA) (GDS part No. 160-511) TP10 PC or laptop (dedicated) GDS Terminal Gas Status LED SENSOR GAIN (download from GDS website) CELL Gain Pot (20mA)
Page 14
LED D9 is Green. head. Pressing (R) on the PC causes a reset to occur. Gas type with 7. Connect PC to sensor via GDS terminal using USB-A to USB address and serial number are then output to the PC together with Micro-B cable at 4800 baud connected to J1 1.
For inaccessible the formation of dangerous layers occurs at ground high-level locations, GDS pump aspirated sensors level. These gases tend to behave like water and will located at chest height, having fixed sample lines to the run down gradients and pool at the lowest point.
It is expected that the sensor be gas checked at the controller GDS Sample Systems are designed to monitor for sample line by allowing test gas to be drawn directly from a gas bag until a blockages but are not able to detect any change in the porosity maximum gas value read-out is obtained.
2. Disconnect sensor tube at point (x) and inject cylinder gas at a flow rate of approximately 1liter/min, 3. For calibration adjustments see 4~20mA adjustment Toxic/Oxygen page 8 and Flammable page 10. sales@gds-technologies.co.uk +44 (0)1 13 286 0166 www.gds–technologies.co.uk GDS TECHNOLOGIES LTD FUSION POINT ASH LANE GARFORTH LEEDS...