Vestfrostsolutions.com Vestfrostsolutions.com WARNING WARNING: Keep ventilation openings in the appliance’s cabinet or in the built-in structure clear of obstruction. WARNING: Do not use other mechanical devices or other means to accel- erate the defrosting process than those recommended by the manufacturer.
Page 5
Vestfrost Solutions Vestfrost Solutions WARNING: Turn of power before access remote alarm connection. CAUTION: Possibility of electric shock. WARNING: Crushing of hands and risk of hand injuries CAUTION: Cold surface avoid contact. Temperature may cause frostbite or injury. Wear protective...
Page 6
Vestfrostsolutions.com Vestfrostsolutions.com WARNING: The appliance must be connected to power minimum 12 hours before using it for storage of medicine. ● Always keep the keys in a separate place and out of reach of children. ● Children should be supervised to ensure that they do not play with the appliance. ●...
ULTF-51i ULTF-91i We recommend that you read this user manual before using the unit for the first time. Vestfrost Solutions does not guarantee safe operation if the unit is used for anything other than its intended use. Contents of the user manual can be subject to change without notice.
Vestfrostsolutions.com Vestfrostsolutions.com Electrical connection Wiring and connections in power supply systems must been all applicable (local and national) electrical codes. Consult these codes lengths and sizes prior to cabinet installation. This device complies with relevant EU directives including Low Voltage Directive 2014/35/EU and Electromagnetic Compatibility Directive 2014/30/EU The socket should be freely accessible.
Vestfrost Solutions Vestfrost Solutions Intended use Vestfrost biomedical freezers are designed to provide reliable temperature stability and precision and ensure an ideal cold temperature storage for sensitive items is achieved. The products are designed for the following operating ranges: Models...
Vestfrostsolutions.com SPARE PART LIST ULTF-7i ULTF-7i Position Item number Item name 0061 6530123 Filter drier, 0087 0-6038175 Base plate fittings, complete 0126 8-036510026 Compressor - NLV8.0CN, Complete, 0153 7535234 Mains lead/Power cord UK plug with angled C13 connector 0153 7535254 Power cord EU H05VV-F 3G1 C13 jack 90°...
Vestfrostsolutions.com ULTF-51i ULTF-51i Position Item number Item name 0061 6530123 Filter drier, 0087 0-6038175 Base plate fittings, complete 0126 8-036510176 Compressor -NLV12.6CN 0153 7535234 Mains lead/Power cord UK plug with angled C13 connector 0153 7535254 Power cord EU H05VV-F 3G1 C13 jack 90° black 2500mm EU-plug 0153 7535400 Mains lead...
Vestfrostsolutions.com ULTF-91i ULTF-91i Position Item number Item name 0061 6530123 Filter drier, 0087 0-6038175 Base plate fittings, complete 0126 8-036510176 Compressor -NLV12.6CN 0153 7535234 Mains lead/Power cord UK plug with angled C13 connector 0153 7535254 Power cord EU H05VV-F 3G1 C13 jack 90° black 2500mm EU-plug 0153 7535400 Mains lead...
Vestfrostsolutions.com GET ACCES TO THE ELECTRICAL BOX ULTF -7i Electrical box Use a screwdriver to loosen the 2 screws Open the door before you remove the front plate to access engine compartment...
Page 17
Vestfrost Solutions Electrical box can easily be accessed from front. Use a screwdriver to loosen the electrical...
Vestfrostsolutions.com Vestfrostsolutions.com ULTF -51i/91i Inverter Electrical box Use a screwdriver to loosen the 2 screws Open the door before you remove the front plate to access engine compartment...
Page 19
Vestfrost Solutions Vestfrost Solutions Electrical box can easily be accessed from front. Use a screwdriver to loosen the electrical...
Vestfrostsolutions.com Get access to the inverter To take out the Invert loosen the screws Inverter can easily be accessed from back on ULTF -7i To take out the Invert loosen the screws Inverter can easily be accessed from front on ULTF -51i. To take out the Invert loosen the screws Inverter can easily be accessed from back on ULTF - 91i.
Vestfrostsolutions.com Vestfrostsolutions.com GET ACCES TO THE COMPRESSOR ULTF -7i Use a screwdriver to loosen the screws on the back cover Access to compressor and inverter (7i) Inverter Dry filet compressor...
Vestfrostsolutions.com Vestfrost Solutions Remote alarm function Fig. Internal controller with built-in alarm relay Your device is equipped with a remote alarm connection to send alarm signals to a building management system, a monitoring unit, visual or acoustical warning device etc.
Vestfrostsolutions.com Vestfrost Solutions MAINTENANCE Frequency TASK Description Quar- terly nual- Years need- The compressor com- 1. Unplug unit from power supply partment and the con- 2. Remove the lower panel. denser must be kept 3. Use a vacuum cleaner to free of dust and dirt.
Vestfrostsolutions.com Vestfrost Solutions GENERAL INFORMATION Warranty, spare parts and service Warranty disclaimer Faults and damage caused directly or indirectly by incorrect operation, misuse, insufficient main- tenance, incorrect building, installation or mains connection. Fire, accident, lightening, voltage variation or other electrical interference, including defective fuses or faults in mains installations.
Vestfrostsolutions.com DISPOSAL Information for Users on Collection and Disposal of Old Equipment and used Batteries These symbols on the products, packaging, and/or accompanying documents mean that used electrical and electronic products and batteries should not be mixed with general household waste. For proper treatment, recovery and recycling of old products and used batteries, please take them to applicable collection points, in accordance with your nation- al legislation and the Directives 2012/19/EU and 2006/66/EC.
Need help?
Do you have a question about the ULTF-7i and is the answer not in the manual?
Questions and answers