Page 1
Type III RO Water Purification System LHWP-B11 www.labtron.com | info@labtron.com...
Page 2
Type III RO Water Purification System LHWP-B11 Index Sr.no Title Page no Safety Measures Introduction Features Specifications Applications Instrument Introduction Installation Operations Replacement...
Page 3
Type III RO Water Purification System LHWP-B11 1. Safety Measures This manual is designed to assist users in properly using and maintaining the equipment to ensure it remains in optimal working condition and prolongs its service life. Kindly read the instructions carefully before using the equipment, approaching it with a scientific attitude.
Page 4
Type III RO Water Purification System LHWP-B11 2. Introduction Type III RO Water Purification System LHWP-B11 is a revolutionary, modular design which helps in elimination of large amount of salts and other minerals from water using 2 stages of RO. It provides Type III level water which is required for chemical and physical inspection or for washing lab glassware’s.
Dimensions ( L x W x H ) 340 x 550 x 530 mm Weight 35 kg 5. Applications Our Type III RO Water Purification System LHWP-B11 is best suited for mid-sized research facilities, it offers a dependable water supply for daily lab activities and sample preparation.
Type III RO Water Purification System LHWP-B11 7. Installation Working Conditions of Water Purification System 1) Voltage: AC220V/50Hz, with reliable ground connection 2) Max power: 100W 3) Tap water temperature: 5-40℃, TDS: 200ppm 4) Tap water pressure: 0.15-0.3Mpa (if above 0.3 Mpa, Then the user needs to take decompression measures.
Page 8
Type III RO Water Purification System LHWP-B11 7.2.1 Installation steps Open the package and confirm the goods as shown in Figure 3, Figure 4, and Figure 5. Figure-3 Host (frame) Figure-4 Pure water tank Figure-5 1/4” Kit...
Page 9
Type III RO Water Purification System LHWP-B11 7.2.2 Installation and operation method 1) Open the tee joint, connect one end to the tap water pipe, and connect the other end to the original tap water outlet or block it with an end cap.
Page 10
Type III RO Water Purification System LHWP-B11 Figure-8 6) Kindly contact the signal wire (above the pure water tank), to the signal socket (rear of the main case) Figure-9 7) After installation, the machine and the site shall be clean and tidy.
Page 11
Type III RO Water Purification System LHWP-B11 8. Operations The operation panel of the machine Figure-10 8.2 Standard procedure for operation maintenance of the machine 8.2.1 Functional specifications of buttons 1) Start: Start/Stop the machine. 2) RO FLUSH/SELECT: Manual RO membrane flush can be started and stopped manually.
Page 12
Type III RO Water Purification System LHWP-B11 At setting the status, the following buttons can be reused: Name of buttons Functions Notes In standby mode, press the button for 3 seconds to PUREWATER/SET enter the Set After choosing the pressure control mode on the circuit board, select the high and low-pressure control points, High-Hxx, L-Lxx.
Page 13
Type III RO Water Purification System LHWP-B11 Manual recycling Timing intake ultrapure water Flicker and display the remaining time Ultrapure Random intake for ultrapure water...
Page 14
Type III RO Water Purification System LHWP-B11 9. Replacement Consumables Replacement Procedure Turn off the water inlet valve first and turn off the power when the low-pressure switch alarm is triggered, which can reduce the residual water in the filter and make it convenient to replace the consumables.
Page 15
Type III RO Water Purification System LHWP-B11 Figure-11 Figure-12...
Page 16
Type III RO Water Purification System LHWP-B11 Replacement and judging method of consumables Replacement Items Quantity Judgment Method Period Replace when the colour of the filter 5μmPP filter 1 piece 2-3 Monthly cartridge becomes deep yellow. If the tap water bleaching powder is...
Need help?
Do you have a question about the LHWP-B11 and is the answer not in the manual?
Questions and answers