Page 1
Digital Torque Tester INSTRUCTION MANUAL “GENERAL SAFETY RULES” WARNING! Read and understand all instructions. Failure to follow all instructions listed may result in electric shock, fire or serious personal injury. FOLLOW THESE INSTRUCTIONS. Version 1.9 URYU SEISAKU, LTD.
Table of contents Outline................2 1.1. Feature ................2 1.2. Installation place ..............2 1.3. Environmental condition ............2 Safety instructions ..............3 Dimensions ................ 6 Parts names and the functions ............7 4.1. Standard accessories .............. 8 4.2. Specifications ..........................9 How to use ................
Installation place Install and fix UDT-25 as per the following instructions. 1)Within a building where no water no direct rays of the Sun affects UDT-25 because this is not waterproof. 2)Place where UDT-25 does not suffer from corrosives, flammable gas, grinding fluid, oil mist, metal powder, and etc.
●Fix UDT-25 on the place bearing the weight to avoid personal injury due to accidental drop. ●Keep your working place well lighted, clean and tidy to avoid personal injury. ●Make sure of UDT-25 fixing to avoid personal injury due to installation or fixing in the event of earthquake.
DANGER ●Never touch any switches of UDT-25 with wet hands to avoid electric shock. ●Do not get UDT-25 connected with battery charger when you do not recharge UDT-25. ●Use UDT-25 genuine battery charger for recharging the battery. ●Remove battery from UDT-25 if you do not use UDT-25 for long time.
●Stop to use UDT-25 immediately if you feel abnormality. GENERAL WARNING ●Keep other people than operators away from UDT-25. ●Do not use at the place where you can wet UDT-25 because this is not waterproof. Doing so will cause short-circuit inviting fire and electric shock. DISCLAIMER...
⑬SYSTEM RESET button: This switch is in the hole available on the left side of power switch. Press this button switch by point of a ballpoint pen or by the similar, and UDT-25 will initialize the setting to the original status or erace memorized data .
4.1. Standard accessories 3/8-SQUARE POINT BIT HEX BOLT (HARD) TX-C46E-75 M8×20 (919-700-0 DRIVER ADAPTER) (946-928-0) HEXAGON SOCKET BIT 13×100(6.35Hex) SQUARE SOCKET ADAPTER (918-223-0) (830-520-6) Genuine charger for 100V/115V use ( AC adapter with flat blades) (910-950-0) *for 230V use (AC adapter with round HEXAGON BOLT M8×35 blades) (910-951-0) (946-930-0)...
・Do not use UDT-25 while recharging is in process. ・Recharge the battery regularly if you do not use UDT-25 for more than Power 3-6 months. ・Keep the battery and UDT-25 separately when you do not use UDT-25 for a long time. AC100V 50/60Hz Charger...
INDI.OFF TIME (Torque display timer) function and the setting Use this timer when you use UDT-25 under [P-P] or [P-D] setting, and the digital meter will maintain torque display just for the preset time. The original setting is 0.0C maintaining torque display until CLEAR button is pressed.
4.6. BLOW COUNTER UDT-25 displays not only torque but also blow numbers counted in a period of time by SURVEY TIME preliminary set 0.1~9.9 seconds. Display of maximum number is 99. Input of 100 or greater number will restart the display from 0. This function works only under P-P mode. Press SURVEY TIME switch ▲...
Begin this action when UDT-25 is switched off status. Press CLEAR MEM.DATA button, and digital meter will change the display On or OFF. It takes 5 seconds for UDT-25 to come back to measuring mode after On or OFF option was displayed.
RS232C port 6.1. Communication specifications Order No rule Synchronize Start-stop synchronous All/half duplex Half duplex Transmission code ASCII Transmission speed 9600bps Start bit Data bit Stop bit Parity operation None Cable Cross cable for PC. For devices except for PC, use cable whose pin configuration is appropriate to the devices.
Page 15
⑤0DH [CR] (Control code): 1 byte Example: Torque value 13.21, and Blow numbers 16 pieces 1 6 P (3)Output memorized data (start a new line) *Press MEM.DATA, and UDT-25 will make line feed (line insert). ① ② ③ ④ ①18H [CAN] (Control code): 1 byte ②20H Space:...
Attachment Use attachment appropriate to your fastening conditions. Soft joint Hard joint Pulse tools Stall torque tools Use hard joint bolt Use soft joint bolt (hexagon head Use soft joint bolt (hexagon head bolt), (hexagon bolt), bolt), nut, spring holder spacer, nut, spring holder...
Battery replacement ① Open and remove 2 pieces screws available on the right and left side of UDT-25, and back cover plate will be separated d from UDT-25. ② Disconnect battery connector from print circuit board, and remove the battery from the battery retainer of the back cover plate.
Page 19
INSTRUCTION MANUAL for UDT-25 Digital Torque Testers Version 1.7 URYU SEISAKU ,LTD. 1-2-11, Fukae-Minami, Higashinari, Osaka, Japan. : +81-6-6973-9415 TEL. : +81-6-6972-0346 : uryuair@uryu.co.jp E-Mail DISTRIBUTED BY:...
Need help?
Do you have a question about the UDT-25 and is the answer not in the manual?
Questions and answers