8058902 | Instructions for Use for Podiatry Drill nova PRO
REV04 | Revision status: 2025-01-10
5800001

Advertisement

loading
Need help?

Need help?

Do you have a question about the nova PRO and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Ruck nova PRO

  • Page 1 Instructions for Use EN | English 5800001 8058902 | Instructions for Use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 2: Dear Customer

    Thank you for your trust Dear Customer, With the nova PRO, you have acquired a podiatry drill that sets standards in terms of functionality and performance. It was resolutely developed "based on best practice for use in the best practices". You will notice the many details that are taken into consideration in this new development, details that reflect over 90 years of experience and expertise.
  • Page 3 You can find the language variants on our websites: hellmut-ruck.de ch.hellmut-ruck.de hellmut-ruck.uk Search term: 5800001 Open Gebrauchsanweisungen | Manuals in the Downloads area. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 4: Table Of Contents

    3.8 Hazards due to incorrect settings or incorrect use ............. 20 3.9 Hazards due to moving parts ....................21 3.10 Hazards due to wetness and humidity ................... 21 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 5 5.10 Functions in the "Device settings" menu ................44 5.11 Functions in the "User profile" menu ..................48 5.12 Functions in the "Help" menu ....................50 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 6 6 Maintenance and warranty .................... 69 Regular maintenance by RUCK service department ............69 6.2 Regular maintenance by the operating personnel .............. 70 6.2.1 Overview ........................... 70 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 7 10.1 General data ........................... 95 10.2 Manufacturer’s declaration – Electromagnetic emissions ..........95 10.3 Electromagnetic interference emission and immunity ............96 10.4 Radio technologies used ......................98 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 8: Information Relevant To These Instructions For Use

    Protection against contact with fingers and against foreign bodies with a diameter of more than 12.5 mm. Protection against water: Protection against vertically dripping water. Protect from moisture and humidity. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 9 Follow the instructions for use. Observe the disposal notice. Manufacturer Manufacturing date Item number Serial number of the device Product identification number (Unique Device Identification) 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 10: Meaning Of Signal Words

    Declaration relating to these instructions for use Hellmut Ruck GmbH reserves all rights to modify and supplement these instructions for use. Hellmut Ruck GmbH has taken great care to keep these instructions free of inaccuracies and omissions. Please contact us if you find any errors or inaccuracies.
  • Page 11: Target Group Of The Instructions For Use

    The responsible person must inform both RUCK as the manufacturer and the competent authority of the Member State in which the operator and/or the treated person is established.
  • Page 12: Product Description

    Cover of the dust filter bag Chuck (interchangeable clamping head + coupling) Handpiece Flexible suction hose Positions 3 to 6 form the "handpiece unit" assembly. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 13: Type Plates In The Cover Of The Dust Filter Bag And On The Display Unit

    Inside of the cover of the dust filter bag T2 Type plate of the handpiece unit (handpiece, flexible suction hose and cover of the dust filter bag) 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 14: Functional Description

    The additional HEPA filter in the rear filter slot collects the smallest dust particles and thus ensures maximum air purity. Dry technology The podiatry drill works with dry technology and is suitable for continuous operation. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 15: Intended Usage

    › for injured skin › for hypersensitivity of the skin › in case of incompatibility with the used rotary instruments (instruments may contain allergenic materials such as nickel) 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 16: Prerequisites And Qualifications Of The Operating Personnel

    Make sure that no part of the device is in direct sunlight. Exposure to direct sunlight can impair the legibility of the display and hinder the suction function and cooling of the device. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 17: Basic Safety Instructions

    There is then a risk of electric shock when touching. In this case, have the device checked by the manufacturer or an authorised service centre before using it again. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 18: Hazards Due To Unauthorised Maintenance Or Unauthorised Conversions

    › Risk of strangulation due to suction hose and mains cable. › Also keep the packaging material away from children. Choking hazard due to parts of the packaging. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 19: Hazards Due To An Unsuitable Installation Site

    › If possible, change the filters before they are completely full. A high filling level reduces the suction power. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 20: Hazards Due To Unsuitable Or Faulty Instruments

    Excessive speed can lead to deformation or breakage of the instrument. This endangers the person being treated and yourself. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 21: Hazards Due To Moving Parts

    Use packaging or a transport box designed for the device for transport in order to protect the device against moisture and mechanical damage. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 22: Electromagnetic Disturbances - Requirements And Tests

    WARNUNG Der Nova Pro ist ein mobiles Kommunikationsgerät, sorgen Sie daher für einen Mindestabstand von 30cm / 12 in zu anderen elektrischen Geräten, um diese nicht zu stören. Bei Personen mit aktiven Implantanten, z.B. Herzschrittmacher, ist der Mindestabstand von 30cm / 12 in ebenfalls einzuhalten.
  • Page 23: Hazards Due To Inadequate Disinfection

    Insufficient disinfection can lead to transmission of disease. Please also observe the guidelines and recommendations of the national competent body for hygiene and disinfection. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 24: Scope Of Delivery And Assembly

    Please retain the packaging (cardboard box, polybag, PU foam parts) and use the transport box to return the device for maintenance work. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 25: Storage Conditions

    The main unit and the handpiece with the flexible suction hose are delivered separately. Handpiece with flexible suction hose Main unit Assemble the components as described on the next two pages. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 26 Pull the plug-in tube off the cover. Guide the motor cable plug through the opening in the cover. Screw the flexible suction hose to the cover. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 27 The plug-in tube has a cut-out for the motor cable. Slide the cover (together with the hose hanging from it) into the main unit. Assembled device 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 28: Operation And Use

    Remove the dust filter bag, see page 63. Make sure that the mains switch is switched off. Make sure that the mains cable is disconnected. Place the main unit on its back. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 29 Remove the display unit and place it in a suitable location. Release the release pin. Inserting the dust filter bag Insert a dust filter bag into the main unit as described on page 64. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 30: Connecting The Display Unit With The Main Unit

    Make sure that no part of the device is in direct sunlight. Exposure to direct sunlight can impair the legibility of the display and hinder the suction function and cooling of the device. Pull the protective foil off the display. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 31: Connecting And Switching On

    Answer the questions and follow the instructions on the display. * Only available with certain device variants. Further information on this functionality can be found on page 56. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 32 As soon as you have inserted an instrument (see page 58), you can start treatment. Information on operating the display can be found in the following section. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 33: Display With Touch Functionality

    PRO favourites view, see page 38. nova PRO basic view John Doe nova PRO favourites view Instrument 1 Instrument 2 Instrument 1 Switch between the main views 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 34: Standby Mode

    Switch off the suction function. Navigation and info bar Start/Stop key for the handpiece motor Favourites key: Switch to the nova PRO favourites view, see page 38 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 35: Setting The Speed

    You can change the direction of rotation before switching on the handpiece motor or when the motor is already running. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 36: Setting The Suction Power

    High suction power keeps the room air clean and reduces the risk of transferring potentially pathogenic particles and abrasive dust. A high suction power also keeps the device cool and increases the service life of the device. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 37: Switching On And Off The Handpiece Motor

    46. The start/stop key then contains the symbol A (for automatic). 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 38: The Nova Pro Favourites View

    Operation and use The nova PRO favourites view In the nova PRO favourites view, you can quickly access saved values for specific instruments on the podiatry drill. One set of saved values is called "Favourite". You can define up to 30 favourites per user profile.
  • Page 39: Moving, Editing Or Deleting A Favourite

    Activates the edit function, see page 40. Instrument Moves the favourite to the right. Confirms the position of the favourite. Editing options in the nova PRO favourites view 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 40: Creating Favourites

    Selection of the direction of rotation Cancel Selection of suction power (in %) Field for text input Selection of the symbol for the instrument head Confirm 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 41: Navigation Elements

    As soon as you switch to a view in which the start/stop key for the handpiece motor is no longer visible, the handpiece motor stops automatically. The automatic switch-off of the motor is for your safety. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 42: Functions In The "Settings" Menu

    Device settings Calls up the Device settings menu, see page 44. Device settings User profile Calls up the User profile menu, see page 48. User profile 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 43 Tap on Search for radio connection and then tap on a displayed device to establish a radio-linked connection. The function is only visible if you are logged in with the "ADMIN" (administrator) profile property. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 44: Functions In The "Device Settings" Menu

    Standby You can also call up standby mode with a gesture: In any view, swipe upwards from the bottom edge of the display. John Doe 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 45: Automatic Standby

    You can also call up disinfection mode with a gesture: In any view, swipe down from the top edge of the display with 2 fingers. John Doe 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 46 Hose cleaning Factory setting: 20 operating hours. You can also switch off the reminder completely. Description of the removal of the suction hose: See page 74. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 47 The function is only visible if you are logged in with the "ADMIN" (administrator) profile property. Reset device Tap the tick to reset the device to factory settings. Tap on the x to cancel the process. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 48: Functions In The "User Profile" Menu

    › Option to establish radio connections, see page 43. › Option to reset all device settings to the factory settings, see page 47. › Option to synchronise the user profiles with the RUCK cloud, see page 47. 8058902 | Instructions for use for Podiatry Drill nova PRO...
  • Page 49 "ADMIN" property. Tap on the tick to delete the user profile. Delete user profile Tap on the x to cancel the process. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 50: Functions In The "Help" Menu

    Shows a link to a website as a QR code. Scan the QR code to receive answers to frequently asked questions. We are constantly updating and adding to this online collection of knowledge. Häufig gestellte Fragen 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 51: Operating A Ruck Podiatry Couch

    Operation and use 5.13 Operating a RUCK podiatry couch If you have a podiatry couch from RUCK that is equipped with a radio function, you can operate the podiatry couch via the display of the podiatry drill. The podiatry couch must be connected to the podiatry drill via a radio connection, see page 43.
  • Page 52 Swivels the left leg rest of the podiatry couch downwards. Only available with certain versions of the podiatry couch. Switch to the favourites view of the podiatry couch, see page 53. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 53: Favourites View Of The Podiatry Couch

    The tile turns white. Keep your finger on the tile until the position is Position 3 Position 3 reached. When the position is reached, the tile remains white. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 54 Position 2 Position 3 Shows the keypad. Moves the favourite to the right. Behandlungsliege Confirms the position of the Editing options in the favourites view favourite. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 55: Creating User Profiles / Switching Between User Profiles

    Switch to the User profile menu if you want to change the properties (e.g. the colour) of the user profile or if you want to delete a user profile, see page 48. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 56: Extended Functionality Through "Nova Connect

    RUCK server. This personal data can only be used by you. The data cannot be viewed by RUCK employees or third parties. The protection of your data is guaranteed. 8058902 | Instructions for use for Podiatry Drill nova PRO...
  • Page 57: Application Areas Of "Nova Connect

    CLOUD If you have activated the "Synchronisation with the cloud" function, the user profile settings are saved in a special area of a RUCK server (cloud). This data includes user names, favourites created, the language of the user interface and the brightness of the display. You can activate or deactivate the function in the Settings >...
  • Page 58: Instruments

    › labelled as a medical device (CE) › straight shank (not bent) › undamaged working part You can find suitable instruments on our websites: hellmut-ruck.de ch.hellmut-ruck.de (Switzerland) hellmut-ruck.uk (United Kingdom) hellmut-ruck.uk 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 59 Tip: To check the diameter and straightness of the shank of an instrument, you can use the pull-off aid. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 60: Inserting An Instrument Into The Handpiece

    Set the appropriate speed for the instrument being used. Always work at the speed specified by the instrument manufacturer. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 61: Ergonomic Holding Of The Handpiece

    5.17.2 Overhand grip When carrying out surface work, gripping the handpiece from above and supporting it with the thumb makes for accurate and relaxed working. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 62: Changing The Dust Filter Bag

    However, the handpiece motor must be switched off. When the device is switched on, the display shows messages during the process. Follow the information on the display. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 63: Removing A Full Dust Filter Bag

    This requires a certain amount of strength. Pull the dust filter bag out of the filter slot. Dispose of the dust filter bag. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 64: Inserting A New Dust Filter Bag

    When the device is switched on, the message “Filter change: dust filter bag” appears after inserting the cover. Answer the question after you have changed the filter. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 65: Changing The Hepa Filter

    Follow the information on the display. Once you have made and confirmed the change, the operating hours counter of the HEPA filter is reset to 0 (zero). 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 66: Changing The Hepa Filter When The Device Is Switched Off Or When The Display Is Inactive

    HEPA filter. This requires a certain amount of strength. Pull the HEPA filter out of the filter slot. Dispose of the HEPA filter. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 67: Inserting A New Hepa Filter

    HEPA filter” appears after inserting the HEPA filter. Answer the question after you have changed the filter. A positive response resets the operating hours counter of the HEPA filter to 0 (zero). 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 68: Integrity Of The Main Unit

    Make sure that the mains switch and the mains plug are easily accessible so that the device can be quickly disconnected from the mains in the event of danger. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 69: Maintenance And Warranty

    In such cases, worn parts will be replaced with original spare parts. We therefore recommend that you have the product serviced by the RUCK service department every 1200 operating hours or every 36 months – whichever comes first.
  • Page 70: Regular Maintenance By The Operating Personnel

    Use the “Disinfection mode” function to clean the display when the device is switched on, see page 45. WARNING Do not use any liquids or sprays for cleaning and disinfecting. Danger of electric shock. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 71: Suitable Agents For Cleaning And Disinfection

    Maintenance and warranty 6.2.2 Suitable agents for cleaning and disinfection Use alcohol-based RUCK surface disinfectant wipes (item no. 2966201, 2967601 or 2970901) for cleaning and disinfecting. Allow sufficient time for cleaning and drying. Exposure times for disinfectants: 1 minute for bacteria...
  • Page 72: Cleaning And Disinfection Of The Chuck, Handpiece Motor And The Inside Of The Handpiece Sleeve

    These elements and the inside of the handpiece sleeve can now be cleaned and disinfected. Handpiece sleeve Chuck Handpiece motor 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 73: Replacing The Interchangeable Clamping Head Of The Handpiece

    Use the supplied removal aid to pull the interchangeable clamping head out of the handpiece body. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 74: Cleaning The Flexible Suction Hose And Motor Cable

    Pull the plug-in tube off the cover. Press the locking tab (1) away from the motor cable plug and slide the plug out of the holder (2). 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 75 Unscrew the cover from the suction hose. 8. Unscrew the handpiece sleeve from the suction hose. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 76: Cleaning And Drying

    High humidity and poor air circulation prolong the drying time. To hang the hose vertically, we recommend using the accessory set for hose cleaning (item no. 5801901). 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 77: Assembly

    Make sure that you insert the plug-in tube correctly into the cover: The plug-in tube has a cut-out for the motor cable. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 78: Replacing The Coupling For The Interchangeable Clamping Head

    Disconnect the device from the mains supply. Danger of electric shock. Unscrew the handpiece sleeve from the suction hose. Unscrew the coupling from the handpiece motor. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 79 Screw the coupling onto the motor until there is no gap between the two parts. Screw the handpiece sleeve back onto the suction hose. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 80: Replacing The Handpiece Motor

    Remove the interchangeable clamping head as described on page 73 to access the bearing bush. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 81: Warranty

    Within the scope of the warranty, HELLMUT RUCK GmbH will, at no charge, either repair or replace the defective products, at RUCK’s discretion. This requires carriage-paid shipment of the defective product together with the purchase receipt.
  • Page 82: Filters And Spare Parts

    We offer suitable filters and spare parts to maintain or restore the function of the device. Only use original parts from RUCK. The use of other parts may impair the functionality and properties of the device. Among other things, there may be an increased emission of electromagnetic radiation or reduced interference immunity.
  • Page 83 HEPA filter cover, item no. 4959501 Handpiece motor incl. cable and plug, item no. 4960101 Handpiece unit, complete set, item no. 4977501 Flex suction hose, 145 cm, item no. 4959401 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 84 Mains cable, type G plug, (for the item no. 4960301 United Kingdom), item no. 4960401 Coupling for interchangeable clamping Handpiece sleeve for nova PRO, head, item no. 4959301 item no.  4960501 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 85: Accessories

    Handpiece holder for ROOM classic for nova PRO/SE, item no. 5802502 Handpiece holder for wall for third-party cabinet for nova PRO/SE, item no. 5802504 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 86 5801901 2 pcs, item no. 5802301 Cutter pliers, chrome plated, item no. 2126401) Click on the link or scan the QR code to see an instructional video. qr.de/Fraeserzange 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 87 Accessories Instrument tray for nova PRO for Instrument tray for nova PRO for right-handers, item no. 5802601 left-handers, item no. 5802602 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 88: Disposal

    Protect our environment by disposing of waste responsibly. Observe the regulations and laws that apply in your state, country and region. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 89: Messages

    The message appears if there is no HEPA filter in the device. 22003 Insert a new HEPA filter, see page 65. You can only continue working once a HEPA filter has been inserted. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 90 The message appears if the dust filter bag is full. 50001 Remove the dust filter bag and insert a new dust filter bag, see page 62. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 91 Replace the HEPA filter as soon as possible. The HEPA filter must be replaced at the latest when it has reached 400 operating hours. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 92 The message is purely informative. You can continue working without any restrictions. During maintenance by RUCK service department, the operating hours are reset and the message is cancelled. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 93 The message appears when the suction turbine has become too hot. 32404 Wait until the suction turbine has cooled down to 52 °C. If the error occurs again: Contact the RUCK service department. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 94: Device Error

    The message remains until a connection to the cloud is re-established. As soon as the connection is re-established, the settings are synchronised and the message disappears. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 95: Technical Specifications

    Radioélectriques) [Special International Committee on Radio Interference] 11 Harmonic components according to IEC 61000-3-2 Class A Voltage fluctuations / flicker according to IEC 61000-3-3 Fulfilled 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 96: Electromagnetic Interference Emission And Immunity

    13.56 MHz ; 7.5 A/m; PM: 50 kHz = mains AC voltage before implementation of the test level. PM = Pulse modulation CW = Continuous wave 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 97 As an alternative to frequency modulation, the carrier can be modulated with a square wave signal with a duty cycle of 50 % at 18 Hz. Although it does not represent the actual modulation, it would be the most unfavourable case. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 98: Radio Technologies Used

    If you remove these other devices or switch them off, communication can be re-established. 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 99 Notes 8058902 | Instructions for use for Podiatry Drill nova PRO REV04 | Revision status: 2025-01-10...
  • Page 100 HELLMUT RUCK GmbH Daimlerstr. 23 DE–75305 Neuenbürg, Germany fon +49 7082 944 20 fax +49 7082 944 22 22 contact@hellmut-ruck.uk hellmut-ruck.uk HELLMUT RUCK LTD RUCK Swiss GmbH Raceside Barn, Kirksanton Birseckstrasse 38 CH-4144 Arlesheim, Basel Millom, GB-LA18 4NW...