Before using this equipment you must read and fully understand this user manual. This Aries 2000 bath is intended for assisted bathing of residents in care facilities or in a home care environment. The bath must not be used for any other purpose.
NOTICES The system batteries must be charged WARNING: Risk of electric shock; for 24 hours prior to using the bath. do not allow electrical appliances You may test the bath to ensure it is within 1.5m of the bath. working correctly. To avoid the risk of electric shock, •...
OPERATION INSTRUCTIONS Before you use your bath for the first time FIG. 1 run through these basics checks: Check that the remote is connected Check there is power to the bath Check the handset works and the bath Bath Bath down goes up and down as intended HANDSET...
OPERATION INSTRUCTIONS BATH HEIGHT AND TRANSFER FIG. 3 1 Set the bath at the height appropriate for transfer of the bather (as low as possible for an ambulant transfer ( FIG. 3 ) or raised to allow access for a portable hoist)� 2 Allow/assist the bather to transfer into the bath ( FIG.
OPERATION INSTRUCTIONS SAFE EDGE OBSTRUCTION SENSOR FIG. 6 The underside of the bath is fitted with a obstruction sensor for added safety ( FIG. 6 ) Whilst in use, upward pressure on the underside of the bath will activate the sensor� Immediately stopping motion�...
TRANSFERS The below diagram is for guidance only. Always ensure carers are trained in the appropriate NOTE transfer techniques and equipment that may be used in conjunction with the bath. FIG. 7 A. Ambulant B. Ceiling track hoist C. Mobile/portable hoist A R I E S 2 0 0 0 U S E R M A N U A L...
CLEANING INFORMATION The Aries 2000 bath does BIOCOTE Gainsborough Specialist baths not require any major feature integrated BioCote maintenance, but you antimicrobial technology� should follow these simple instructions. This silver-ion based technology has been • The bath should be wiped...
TROUBLE SHOOTING Some basic checks are detailed below. If in doubt contact service support on Tel: 01527 400024 • Check power Bath Does Not Move • Check fused spur to ensure there is power to the unit� • Ensure the RCD is working and has power to it� To test the RCD push the test button;...
WARRANTY If the bath is installed by a 3rd party we Electrical All electrical components fitted to Gainsborough highly recommend that it is commissioned by Specialist Bathing are CE marked to comply with a Gainsborough Healthcare Group Ltd engineer. current Standards� Our bath products are guaranteed for a period of twelve months from the date of delivery�...
Abacus Healthcare – 01527 400024 environment, its common practices and processes� It is vital that regular assessments of the bather are undertaken to ensure the bath or product remains There is no other warranty expressed or suitable�...
10 & 11 The Oaks Clews Road Redditch Worcestershire B98 7ST Declare that: The Aries 2000 Conforms to the following Directives: • 2006/42/EC Machinery Directive UK MDR 2002 – Class 1 Medical Device • and has been designed and manufactured to the relevant parts of the following standards: ISO 10535:2006 Hoists for the Transfer of Disabled Persons.
KEY TO LABEL SYMBOLS Read instruction manuals and accompanying documents� This product is powered� Warning dangerous voltage� Application part “Type B” according to DIN EN 60601-1� No domestic waste� Separate electrical and electronic components for recycling in accordance with European Directive 2012/19/EU (WEEE)� Protection Class (ie, the product is protected against spraying water)�...