Sensor interfacing with a SENTRO X Controller TX6310 SENTRO X Controller with integrated audio/ visual (A/V) alarm Local area A/V alarm, with high brightness LEDs and loud sounder. SENTRO X Controller with an A/V alarm fitted trolex.com/sentro-x TX6310-UM-EN · P5600.1606-A...
Page 4
fitted above a gas permeable Gas inlet cover membrane, o ering a high level of ingress protection in severe operating environments. The complete assembly is available as a quick-change service replacement after exposure to extreme levels of contamination. trolex.com/sentro-x TX6310-UM-EN · P5600.1606-A...
Plug-in calibrated gas sensing gBlocs for a range of oxygen, toxic and flammable gas • Optional remote gas sensor configuration with interconnecting flexible armoured conduit with 4-core cable Dimensions With AV alarm fitted 2 x M8 fixing holes 2 x M6 fixing holes trolex.com/sentro-x TX6310-UM-EN · P5600.1606-A...
The Remote Sensor can be equipped with any standard gBloc in the range and uses the identical gas inlet cover with the same stainless-steel grille and moisture resistant membrane. Option Option Flex conduit/cable TX6320.00 Not fitted 10 m trolex.com/sentro-x TX6310-UM-EN · P5600.1606-A...
1 x threaded M16 cable entry (for Remote Sensor) Connections TX6320 Remote Sensor: • M16 cable entry provided with stainless steel cable gland (for connection to SENTRO X Controller) Graphic OLED Information display 16 V to 30 V DC PoE Supply voltage trolex.com/sentro-x TX6310-UM-EN · P5600.1606-A...
Ethernet card (10 Mbs) Minimum 16 V +40 mA gBlocs (Gas) Electrochemical Minimum 16 V +25 mA Pellistor Minimum 16 V +45 mA Infrared Minimum 16 V +65 mA Note: Current consumption details are based on minimum input voltage. trolex.com/sentro-x TX6310-UM-EN · P5600.1606-A...
Page 9
SENTRO X Controller with an A/V alarm fitted 86 dB peak at 1 m, sweeping frequency 3 kHz to 4 kHz Audio warning Colour red, multi-point flashing high-brightness LED Visual warning IP65 Protection trolex.com/sentro-x TX6310-UM-EN · P5600.1606-A...
Controller to be powered and exchange data with wide area remote monitoring systems and SCADA networks. In this case, only 1 Relay card can be fitted. Note: The ethernet card can be configured for PoE or powered from a local source. Ethernet card trolex.com/sentro-x TX6310-UM-EN · P5600.1606-A...
This way, the earliest possible warning of a gas release can be given to initiate shutdown functions, alarm functions or safe evacuation of the area. trolex.com/sentro-x TX6310-UM-EN · P5600.1606-A...
A range of test gases from 0.5 L/min source Test gas Flow regulators are available from Trolex on request A test gas hood is available from Trolex for application of test gas on to Test gas hood the SENTRO X gas sensor...
Page 13
Replace the gas inlet cover • Tighten the 4 x M4 screws to 0.7 Nm Note: Connect directly to the quick-connect terminals or remove the inner module for even greater ease of connection. Wire into Remote Sensor terminals trolex.com/sentro-x TX6310-UM-EN · P5600.1606-A...
Page 14
flexible armoured conduit/cable. Wire into rBloc SENTRO X Controller Cable connections wired to Remote Sensor 0.5 mm 4-core, flexible data cable Cable Up to 10 m (where specified) Length TX6310 SENTRO X Controller TX6320 Remote Sensor Wire connections trolex.com/sentro-x TX6310-UM-EN · P5600.1606-A...
Page 15
Note: SENTRO X is configured to always de-energise relay coils when power is absent as a fail-safe condition. SENTRO X · Power o SENTRO X · Power on Setpoint not activated Setpoint activated Coil de-energised Coil energised Coil de-energised trolex.com/sentro-x TX6310-UM-EN · P5600.1606-A...
Page 16
PoE Ethernet card An ethernet card can be fitted to specification and will occupy Slot B only. • Address: Static or dynamic assignment ~10 Mbps to 100 Mbps operation • • Max distance: ~100 m Ethernet card trolex.com/sentro-x TX6310-UM-EN · P5600.1606-A...
Activate OVER the setpoint • Activate UNDER the setpoint • Hysteresis Customised settings Operating parameters can be customised as required within main menu > ‘Functions’. Any changes will be permanently retained by the SENTRO X Controller. trolex.com/sentro-x TX6310-UM-EN · P5600.1606-A...
A spare gas identity label will be supplied with a change-of-duty gBloc to replace the original gas identity label located on the gas inlet cover. We will also supply a spare gas ID label with all gBlocs. trolex.com/sentro-x TX6310-UM-EN · P5600.1606-A...
Power on 1. Splash screen 2. Insert sensing bloc Note: Timeout fault will occur if no sensing bloc is inserted within 5 minutes of operation. 3. Sensor warmup 4. Gas value screen SENTRO X is operating. trolex.com/sentro-x TX6310-UM-EN · P5600.1606-A...
Passive review of firmware and settings information. Step through the information pages using the keypad. Note: No settings or outputs will be a ected. The screen will return to main display if the device times out. trolex.com/sentro-x TX6310-UM-EN · P5600.1606-A...
Note: SENTRO X will automatically return to the main gas sensing screen if no commands or a period of inactivity takes place within 3 minutes. This time limit is extended to 10 minutes during maintenance routines, where the test will be cancelled. trolex.com/sentro-x TX6310-UM-EN · P5600.1606-A...
3. Wait for the gas reading to stabilise 4. Use the ‘Adjust’ option if necessary 5. Save the adjusted value Note: All gas sources have error tolerances, this should be taken into account when performing a Span test. trolex.com/sentro-x TX6310-UM-EN · P5600.1606-A...
1. Navigate to the ‘Main Menu’ > ‘Maintenance’ > ‘Linearity Test’ 2. Follow the on screen instructions 3. Ensure that the gas reading has stabilised before confirming the test 4. A summary table will be presented at the end of the test for recording trolex.com/sentro-x TX6310-UM-EN · P5600.1606-A...
3. Ensure that the values on the multimeter match the outputted value indicated on the SENTRO X screen. 5.8 Remove bloc Safe removal of a gBloc is required to avoid the possibility of data corruption. ‘Main Menu’ > ‘Maintenance’ > ‘Remove’ trolex.com/sentro-x TX6310-UM-EN · P5600.1606-A...
4. Set the trigger mode for OVER or UNDER the setpoint 5. Latching or auto-reset mode 6. Any setpoint can be assigned to operate a WARNING or an ALARM 7. Any relay can be ASSIGNED to be controlled by a selected Setpoint trolex.com/sentro-x TX6310-UM-EN · P5600.1606-A...
The SENTRO X is programmed with the following default A/V assignment settings, when fitted. Sounder Beacon Warning Alarm Over-range (including ‘Pellistor Protect’) Fault Note: All setpoints have a unique MODBUS status flag. All alarm generators can be assigned to an alarm severity. ‘Warning’ or ‘Alarm’. trolex.com/sentro-x TX6310-UM-EN · P5600.1606-A...
Default values are set in accordance with EH40 standards. Ensure local guidelines and legislation are consulted where applicable. If STEL or TWA function is enabled, ensure that all accumulated data is reset to Zero before the commencement of a working period. trolex.com/sentro-x TX6310-UM-EN · P5600.1606-A...
0.4 V (indicating 0%) Maintenance / warmup state 3.1 mA / 0.31 V Maintenance warmup • When ‘Drift mute’ is disabled, the analogue output is proportional to Fault 3 mA / 0.3 V the gas reading Fault trolex.com/sentro-x TX6310-UM-EN · P5600.1606-A...
Page 31
‘IP address’, ‘Subnet mask’ and ‘Gateway’ address settings Set the static IP address. This should be defined in your network router IP address Set the Subnet mask Subnet mask Used only by the DHCP server Gateway trolex.com/sentro-x TX6310-UM-EN · P5600.1606-A...
SENTRO X Controller can be updated in the field with the latest firmware as required using a programming Bloc (pBloc) where required. Note: Trolex will notify users and Distributors of the need to pBloc firmware upgrade infield devices. 7.4 Passcode protection SENTRO X main menu is passcode protected by default.
Any previously remembered maintenance function gas values Factory programmed information such as serial numbers will be retained. The password is not reset to default. ‘Reboot’ Perform a software reboot of the internal micro-controller. This does not power cycle the SENTRO X. trolex.com/sentro-x TX6310-UM-EN · P5600.1606-A...
Repair The product shall only be serviced and repaired by Trolex Ltd or a local Trolex service agent approved by Trolex Ltd in order to maintain the safety integrity of the product. Refer to the following standard for additional guidance: IEC/EN 60079-19 trolex.com/sentro-x...
Under-range Gas concentration is below -10% of full scale Diagnostic messages can be cleared via the ‘Clear’ prompt following resolution of the Fault condition. If the fault condition remains, please contact Trolex for further information. trolex.com/sentro-x TX6310-UM-EN · P5600.1606-A...
Gas sensors have a known Zero and Span movement related to time, level of exposure to gas and the nature of the operating environment. Trolex recommend that the sensors are tested with suitable test gas at regular intervals. This should always be carried out in compliance with local standards and legislations. Following a Zero test and trolex.com/sentro-x...
• Methane (CH ) test gas must be combined with balance air if being used to calibrate catalytic pellistor-based devices • Oxygen (O ) test gas can only be supplied at a maximum concentration of 23% trolex.com/sentro-x TX6310-UM-EN · P5600.1606-A...
At the end of the product’s life, do not dispose of any electronic sensor, component or instrument in the domestic waste. Contact Trolex or the Distributor for disposal instructions. trolex.com/sentro-x...
Trolex shall not be responsible or liable for misuse of the information contained herein. When instruments are used for applications with technical safety requirements, the relevant instructions must be followed.
Page 42
For more information about Trolex, please contact us at: For more information about Trolex, please contact us at: Enquiries Website...
Need help?
Do you have a question about the SENTRO X TX6310 and is the answer not in the manual?
Questions and answers