Page 2
01458 Ottendorf-Okrilla Germany Legal notice This work is protected by copyright. Endress+Hauser SICK GmbH+Co KG retains the rights conferred by it. Reproduction of this work or parts thereof is only permitted within the limits of the statutory provisions of copyright law.
Sources of danger ................... 8 System warranty....................9 RoHS Directive ....................9 Safety conventions..................9 Warning signs....................10 Mandatory signs....................11 Product description .................13 Features of the POWERCEMS100..............13 Method of operation ..................13 Functional principle ..................13 Internal functional units................14 External functional units................15 Measuring components ................15 Interfaces ......................15 Description of system components..............16...
Page 4
Connecting to the customer plant ............... 24 Commissioning ..................25 Commissioning of the humidifier containers (option)......... 25 Before switching on ..................25 Switching on the POWERCEMS100 analysis system ........26 Commissioning of sample gas cooler ............26 Commissioning of analyzers ................ 26 Commissioning of NO converter (option) ...........
Page 5
1.43 Perform Leak Test..................56 Disposal.....................57 Specifications...................59 13.1 EU Declaration of Conformity ...............59 13.2 Dimensions....................59 13.3 Technical data....................60 Endress+Hauser O P E R A T I N G I N S T R U C T I O N S 8030341/1AOJ/V4-0/2023-07...
Page 6
CONTENT Endress+Hauser O P E R A T I N G I N S T R U C T I O N S 8030341/1AOJ/V4-0/2023-07...
Follow all local laws, regulations, technical rules and company policies applicable at the installation location. 8030341/1AOJ/V4-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 Subject to change without notice...
All units and of measurement in this document are metric units. Subject to change without notice. Illustrations may differ from the actual design. O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 8030341/1AOJ/V4-0/2023-07| Endress+Hauser...
Use suitable undamaged lifting equipment. ▸ Wear personal protective equipment (safety helmet, safety gloves). O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 8030341/1AOJ/V4-0/2023-07| Endress+Hauser Subject to change without notice...
Indicates a situation that can lead to material damage if ignored. 8030341/1AOJ/V4-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 Subject to change without notice...
Warning of dangers to environment, nature and organic life Hazard warning on the control cabinet door O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 8030341/1AOJ/V4-0/2023-07| Endress+Hauser...
Wear foot protection Switch off prior to maintenance or repair 8030341/1AOJ/V4-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 Subject to change without notice...
Page 14
FOR YOUR SAFETY O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 8030341/1AOJ/V4-0/2023-07| Endress+Hauser Subject to change without notice...
Should a malfunction occur, the device switches automatically to “Stand-By”. ● Functional principle Fig. 1: Functional principle PowerCEMS100 (example) 8030341/1AOJ/V4-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 Subject to change without notice...
Span gas / test gas feed unit: ● – Solenoid valves O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 8030341/1AOJ/V4-0/2023-07| Endress+Hauser Subject to change without notice...
– 115V, 50Hz – 115V, 60Hz UPS connection / supply (option) ● 8030341/1AOJ/V4-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 Subject to change without notice...
Optional RJ45 socket - BCU Ethernet connection Note The layout can vary specific to the system. O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 8030341/1AOJ/V4-0/2023-07| Endress+Hauser Subject to change without notice...
Refer to the Technical System Documentation (EPLAN) for a detailed parts list of the system components. 8030341/1AOJ/V4-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 Subject to change without notice...
PRODUCT DESCRIPTION Gas flow plan Fig. 4: PowerCEMS100 gas flow plan O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 8030341/1AOJ/V4-0/2023-07| Endress+Hauser Subject to change without notice...
Filter elements and bubbling frits must be installed and/or filled before commission- ing. 8030341/1AOJ/V4-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 Subject to change without notice...
Page 22
TRANSPORT O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 8030341/1AOJ/V4-0/2023-07| Endress+Hauser Subject to change without notice...
Normally, the converter fitted in the POWERCEMS100 analysis system is delivered with the catalyzer cartridge already fitted and ready for operation. 8030341/1AOJ/V4-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 Subject to change without notice...
Sample gas line inlet Zero gas/span gas Gas outlet Roof vent opening O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 8030341/1AOJ/V4-0/2023-07| Endress+Hauser Subject to change without notice...
Page 25
The pressure at the sample gas outlet should not be increased significantly. 8030341/1AOJ/V4-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 Subject to change without notice...
– Voltage supply – Signal lines – Integration of all external components O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 8030341/1AOJ/V4-0/2023-07| Endress+Hauser Subject to change without notice...
▸ Catalyzer cartridge inserted in NO converter. 8030341/1AOJ/V4-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 Subject to change without notice...
Refer to the details on clearing malfunctions in this system description as well as in the Analyzer Operating Instructions. O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 8030341/1AOJ/V4-0/2023-07| Endress+Hauser...
Serious injuries of the respiratory system or death. ▸ Wear suitable respiratory equipment. ▸ Ensure adequate ventilation. O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 8030341/1AOJ/V4-0/2023-07| Endress+Hauser Subject to change without notice...
2 Fill, if necessary 3 Replace the water trap when condensate is present 8030341/1AOJ/V4-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 Subject to change without notice...
Page 32
Refer to the Operating Instructions for the individual system components for fur- ther details and information. O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 8030341/1AOJ/V4-0/2023-07| Endress+Hauser...
Function buttons F1, F2, F3, F4 Current menu branch Current menu Status bar Menu number 8030341/1AOJ/V4-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 Subject to change without notice...
Refer to “Supplementary Operating Instructions BCU” for further, detailed information on operation via the BCU operator panel. O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 8030341/1AOJ/V4-0/2023-07| Endress+Hauser...
CAN baud rate Time | Temperature System Serial number | Tag No. 8030341/1AOJ/V4-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 Subject to change without notice...
Page 36
“Supplementary Operating Instructions BCU” for detailed information on the BCU operat- ing functions. O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 8030341/1AOJ/V4-0/2023-07| Endress+Hauser Subject to change without notice...
The specific adjustment cycles depend on the application and are not the responsibility of SICK AG 8030341/1AOJ/V4-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 Subject to change without notice...
When the adjustment procedure is aborted, it has to be acknowledged by pressing the key switch. Only then can a new adjustment procedure be started. O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 8030341/1AOJ/V4-0/2023-07| Endress+Hauser...
After a manual adjustment abort, the adjustment values are discarded. ▸ Maintenance request is not activated. 8030341/1AOJ/V4-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 Subject to change without notice...
Page 40
ADJUSTING O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 8030341/1AOJ/V4-0/2023-07| Endress+Hauser Subject to change without notice...
Refer to the Operating Instructions of the relevant analyzer for details on operation and menu functions. 8030341/1AOJ/V4-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 Subject to change without notice...
To avoid condensation, purge the sample gas paths of the POWERCEMS100 analysis sys- tem with a dry, neutral gas before storage. O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 8030341/1AOJ/V4-0/2023-07| Endress+Hauser...
▸ Wear personal protective equipment (gloves, respiratory protection) 8030341/1AOJ/V4-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 Subject to change without notice...
– Connections of the POWERCEMS100 – Connections of the probe O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 8030341/1AOJ/V4-0/2023-07| Endress+Hauser Subject to change without notice...
Caution: Serious damage to environment, nature and health by toxic condensate. 8030341/1AOJ/V4-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 Subject to change without notice...
Page 46
▸ Clean or replace lines / hoses as necessary. O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 8030341/1AOJ/V4-0/2023-07| Endress+Hauser Subject to change without notice...
1 pc. / 1 Y. Fi56NK3 Front fitted filter Diaphragm fine filter (visible contamination) 8030341/1AOJ/V4-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 Subject to change without notice...
Page 48
1 pc. / 1 Y. Sales unit=1 pc. PV-30-A (5317088) O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 8030341/1AOJ/V4-0/2023-07| Endress+Hauser Subject to change without notice...
Page 49
5 310 689 Sample gas cooler ECM 1 pc. / 1 Y. 8030341/1AOJ/V4-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 Subject to change without notice...
Page 50
1 pc. / 3 Y. in housing, flow rate approx. 250 l/h O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 8030341/1AOJ/V4-0/2023-07| Endress+Hauser Subject to change without notice...
Page 51
Stainless steel wool filling, T=660 °C, with spare O-ring set. 8030341/1AOJ/V4-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 Subject to change without notice...
5 306 625 Gas sampling probe 2 pcs. / 1 Y. SP180-H O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 8030341/1AOJ/V4-0/2023-07| Endress+Hauser Subject to change without notice...
Disposal according to local and legal environment regulations and guidelines. 8030341/1AOJ/V4-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 Subject to change without notice...
/ too low Temperature sensor PT100 Replace temperature sensor. defective. O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 8030341/1AOJ/V4-0/2023-07| Endress+Hauser Subject to change without notice...
Replace faulty or corroded grounding. Inadequate power supply. Check. Faulty interfaces Check. 8030341/1AOJ/V4-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 Subject to change without notice...
● – Monitoring the pump function is controlled via the gas analyzer. O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 8030341/1AOJ/V4-0/2023-07| Endress+Hauser Subject to change without notice...
To protect the Testo 324, all lines must be free of oil, dust and moisture. 8030341/1AOJ/V4-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 Subject to change without notice...
– Measurement result is displayed. 3. Document and save the measurement result in the test log. O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 8030341/1AOJ/V4-0/2023-07| Endress+Hauser...
Air conditioning: ● Refrigerant in the air conditioner 8030341/1AOJ/V4-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 Subject to change without notice...
Page 60
DISPOSAL O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 8030341/1AOJ/V4-0/2023-07| Endress+Hauser Subject to change without notice...
Fig. 10: PowerCEMS100 outer dimensions Legend 800 mm 2100 mm 600 mm 8030341/1AOJ/V4-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 Subject to change without notice...
RS-422 (option) Interfaces RS-485 (option) Ethernet TCP/IP Modbus TCP Bus protocol O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 8030341/1AOJ/V4-0/2023-07| Endress+Hauser Subject to change without notice...
Page 63
Automatic check and adjustment with test gases converter Cooling unit Options Heating Bypass pump 8030341/1AOJ/V4-0/2023-07| Endress+Hauser O P E R A T I N G I N S T R U C T I O N S | POWERCEMS100 Subject to change without notice...