Advertisement

Quick Links

YZ15T/YZ25T
Pump Heads Operation Manual
Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 ,3
mrc@mrclab.com
MRC. 2.25

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the YZ15T and is the answer not in the manual?

Questions and answers

Summary of Contents for MRC YZ15T

  • Page 1 YZ15T/YZ25T Pump Heads Operation Manual Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 ,3 mrc@mrclab.com MRC. 2.25...
  • Page 2: Safety Precautions

    Safety Precautions & Important information: Be sure to read the manual carefully before operation! This icon warns: Keep fingers away from moving parts. This icon warns: Be careful. � Ai!\ This icon warns: Be careful, the surface is hot. & This icon warns: Be careful, electric shock hazard.
  • Page 3 Warning: Before using, please confirm that the transmitted liquid will not react chemically with the tube or pump head, otherwise it will damage the tube or pump head. If unsure, please consult our engineers. Tubes are vulnerable parts, please pay attention to regl!lar inspections. Our company is not responsible for the losses caused by tube damage, especially the leakage of toxic, harmful, and precious liquids! &...
  • Page 4: Table Of Contents

    Contents Description -------------------- --- ------- 2 Pump Head Structure Component Name and Function Usage Method Install YZ15TNZ25T Pump Head Tube Selection Install the Tube forYZ15TNZ25T Pump Head Take Out the Tube - -- ---- ---- - -- ---- - 4 Precautions for Using Malfunction and Maintenance Warranty and After-sales...
  • Page 5: Description

    The PPS material pump casing is resistant to organic solvents and other corrosive liquids, making it sturdy and durable. Pump Head Structure Suitable tubes for each model: • YZ15T tube specifications: 13#, 14#, 19#, 16#,25#, 17#, 18# • YZ25T tube specifications: 15#,24# ---·2 -------1...
  • Page 6: Usage Method

    Usage Method ♦ Install YZ15T/YZ25T Pump Head • Put one side of the pump head flat shaft towards to drive, push on the groove of coupling in front of drive. Make the pump head in suitable position through the two positions column of pump head support, put the pump head cling to the drive.
  • Page 7: Tube Selection

    ♦ Tube Selection • The tubes suitable for YZ15T/YZ25T series pump heads a- r e divided into domestic tubes and imported tubes.The size accuracy of the wall thickness and inner diameter of the tube can affect the flow accuracy.
  • Page 8: Precautions For Using

    movable tube clamps will be lifted with the upper pressure block at the same time, and the tube will be taken out. ♦ Precautions for Using • The tube flow will fluctuate with the extension of service time. Please pay attention to timely adjust the speed of peristaltic pump (for flow type peristaltic pump, please recalibrate the flow) •...
  • Page 9: Malfunction Solutions

    Suitable Tube and Flow Rate of YZ15T-3 Pump Head (unit: mUmin) � u b e peed 30rpm 16.2 28.2 60.4 94.1 141.1 60rpm 15.6 31.7 116.2 54.9 174.7 255.6 100 rpm 26.4 53.9 94.1 313.7 470.2 300 rpm 21.1 79.1 161.6...
  • Page 10: Technical Parameters

    Technical Parameters YZ15T 3 rollers pump head Suitable Drive BT series, WT series 13#, 14#, 19#, 16#,25#, 17#, 18# (Wall Thickness Suitable Tube Size 1.6mm, ID 0.8-7.9mm) Silicone, pharmed, A-60-G, A-60-F, F5500A,2275, Suitable Tube Type 2475,3603, Viton Speed Range Maximum 600 rpm...