INSTRUCTIONS FOR USE Refraction and diagnostic unit HS-810 Version: IFU HS-810 V05 Date: 28.03.2023 WARNING! Read the instructions for use carefully before you use this product. It contains important information about safety of users and patients. HS DOMS GmbH Tel. + 49 3671 5450 0 E-mail: info_dmd@haag-streit.com...
With your decision to purchase an HS DOMS examination unit, you have given us a high level of trust. We thank you for that. You have chosen a mature product and modern design. The HS-810 is available in a not height-adjustable and a height-adjustable version. The unit is available as right-hand and left- hand unit.
Safety instructions Operation HS DOMS patient chair ..........9 Image 3 Safety instructions Operation patient chair with calf support ........9 Image 4 HS-810 unit (right version, motor height adjustable) ..........10 Image 5 Switch on front base unit ................... 12 Image 6 System switch on wall junction box ................
Refraction and diagnostic unit HS-810 Symbol explanations Follow the instructions for use Disconnect from the mains before opening Warning! Be sure to follow the instructions electrical device of protection class I, Protective earth CE — Certificate of Conformity Manufacturer Year of manufacture...
Page 7
Refraction and diagnostic unit HS-810 Do not sit, step on or lean on the device Table 1 Symbol Explanation Page 5...
Refraction and diagnostic unit HS-810 Environmental conditions and transport Before using the examination unit for the first time, please read the instructions for use. The examination unit may only be operated in accordance with the instructions for use. 5.1 Environmental conditions...
Installation and dismantling as well as repair, maintenance and repair work may only be carried out by Haag-Streit or by authorised technical service of the Haag-Streit sales partners. If a third- party device is connected, this must be done in accordance with the standard EN 60601-1. Only original HS spare parts may be used.
Refraction and diagnostic unit HS-810 Safety instructions The examination unit must be set up by Haag-Streit customer service or by authorized specialists. The mains connection cable may only be connected to a correctly and permanently installed power connec- tion. It must be installed in accordance with the applicable standards and guidelines.
Refraction and diagnostic unit HS-810 Safety instructions patient chair Protection against collision with the patient’s thigh When operating the electromotive lifting column, avoid colliding with the patient's thigh. Maximum chair load Pay attention to the maximum chair load. Do not place a patient with a weight of more than 125 kg (see technical data).
HS-810 Variants, construction and components 8.1 Variants HS-810 Standard or HS-810 motorised height adjustable The standard version of the HS-810 is not height-adjustable and therefore equipped with a fixed column. The height-adjustable version of the HS-810 has an adjustable column and associated control buttons (table UP and table DOWN).
Refraction and diagnostic unit HS-810 8.4 Description of the components Column (1) To mount a chart projector or monitor, Workplace lamp, LED spot (reading light), phoropter arm Phoropter arm (2) manually movable optional adapters for different phoropters (Option: cable chain for cables of automatic phoropter)
When commissioning, first turn on the system switch on the wall junction box (Image 6). Thus, the unit HS-810 and the WJB have mains voltage. To activate the unit, turn on the main switch on the body (Image 5). After switching on the main switch (lights green), the unit is ready for operation.
10.2 Unit with wall junction box (optional) The examination unit HS-810 is optionally equipped with wall junction box. When decommissioning with wall junction box, the main switch on the body (Image 7) must first be switched off. After that, the main switch no longer lights.
Refraction and diagnostic unit HS-810 11 Operation of the components 11.1 instruments table Bring the instruments table into working position 1 by swiveling it. The instruments table is locked in place by turning the table locking lever (Image 9) clockwise.
Refraction and diagnostic unit HS-810 The instruments table is brought by a swiveling movement from the resting position into working position 1. By sliding movement, you switch from working position 1 to working position 2. In reverse sequence, put the device table back into rest position. The table instruments are automatically turned on and off. If a 230V device is connected to an IEC connector, it always remains “on”...
Refraction and diagnostic unit HS-810 11.6 Led workplace lamp (optional) Dimming Image 13 Button light head 11.6.1 Switching the workplace lamp on and off You can switch the workplace lamp on or off with the switch on the unit. 11.6.2 Dimming the workplace lamp This function allows you to change the brightness of the lamp.
Refraction and diagnostic unit HS-810 11.7 Led Reading lamp (optional) Switch the LED reading lamp “ON/OFF” with the switch on the unit. Led Spot — Reading lamp Image 14 11.8 Dimming room light/automatic room light (optional) The room light is switched via the switch on the button “ON/OFF”.
11.9 Safety shut-off bar HS-810 has a safety shut-off bar under the instruments table. If the device table is in front of the patient, accidental clamping the patient’s thighs (due to the upward movement of the patient’s chair) is pre- vented.
During cleaning/wipe disinfection, no liquid may enter the examination unit or the control buttons. Before you start, turn off the HS-810. The unit must then be disconnected from the mains. Before you turn on the unit again, make sure that it is completely dried. The specified disinfection products must be used in accordance with the manufacturer’s instructions for...
Refraction and diagnostic unit HS-810 13 Electrical fuses Important note! Do not open the HS-810 unit or wall junction box. The change of fuses and bulbs may only be performed by the technical customer service of Haag-Streit/Haag-Streit Distributors or authorised technical personnel.
Refraction and diagnostic unit HS-810 16 Electromagnetic compatibility Observe the EMC information Medical electrical devices and systems are subject to specific measures regarding EMC and must be installed in accordance with the EMC information contained in this accompa- nying document. Portable radio frequency devices such as mobile phones, or other RF phone accessories, including antennas, can affect the function of medical devices.
Page 26
HS DOMS GmbH Langenschader Straße 67 D-07318 Saalfeld, Germany Phone: + 49 3671 54 50 0 info_dmd@haag-streit.com www.hs-doms.com Design and form are subject to change without prior notice.