CONTENTS CONTENTS STANDARD ACCESSORIES ..................1 OPTIONS (OPTIONAL PRODUCTS) ................2 SAFETY PRECAUTIONS (BE SURE TO READ THIS BEFORE USE) ............... 3 LIMITED LIFESPAN PARTS AND PREVENTIVE MAINTENANCE ............. 9 EQUIPMENT RENEWAL ....................9 NOTATIONS USED IN THE INSTRUCTION MANUAL ............. 10 PRECAUTIONS ON CE MARKING .................
Page 3
CONTENTS 2-5. EXTERNAL DEVICES CONNECTIONS ............. 40 Connection with USB-80/70 Series Units ............. 40 2-6. CONNECTION WITH A PC ................. 42 RS-232C Connection ................... 42 USB Connection................... 44 Ethernet Connection ..................48 2-7. SD CARD INSERTION ..................51 2-8. USB MEMORY DEVICE INSERTION ..............53 3.
Page 4
CONTENTS Delta V and Delta T Settings ................ 91 5-2. CHANNEL CONDITION SETTINGS ..............92 Measurement ON/OFF ................. 94 Calibration Coefficient Settings ..............95 Unit Settings ....................96 Reference Resistance Value Settings ............97 Scanning Speed Settings ................98 Channel Mode Settings ................100 Decimal Point Position Settings ..............
STANDARD ACCESSORIES STANDARD ACCESSORIES The following standard accessories are enclosed with the UCAM-80A. After opening the package, confirm that all standard accessories are present. UCAM-80A Data Logger Name Quantity "READ BEFORE USING THE PRODUCT" document "Terms and Conditions of Product Warranty" document...
SAFETY PRECAUTIONS (BE SURE TO READ THIS BEFORE USE) SAFETY PRECAUTIONS (BE SURE TO READ THIS BEFORE USE) This device has been designed as described in "10-1. UCAM-80A PRODUCT SPECIFICATIONS." Do not use in environments outside of specifications. Doing so could cause failure.
Page 11
・To prevent an electric shock always connect the protective ground wire to GND. ・To prevent fire, always replace with a specified rating fuse. DANGER ・This device uses an internal backup battery. Devices will be collected by KYOWA for battery replacement. ・Do not apply heat or throw into fire.
Page 12
SAFETY PRECAUTIONS (BE SURE TO READ THIS BEFORE USE) WARNING ● Power supply • To prevent fire, be sure to check that the voltage of the power supply that you are using matches the power supply voltage specifications of this device, prior to connecting the power supply. ●...
Page 13
SAFETY PRECAUTIONS (BE SURE TO READ THIS BEFORE USE) CAUTION ● Use within the operating temperature range (0°C to 50°C). • Use outside of the operating temperature range may worsen performance or result in failure. If the device must be used in direct sunlight or in the cold, ensure that a sunshade is installed or heating measures are implemented.
Page 14
SAFETY PRECAUTIONS (BE SURE TO READ THIS BEFORE USE) CAUTION ● Use indoors. • Do not place this device outdoors, so that it can be used normally and safely. ● After turning the power OFF, be sure to wait at least 5 seconds before turning the power back ON again.
Page 15
SAFETY PRECAUTIONS (BE SURE TO READ THIS BEFORE USE) CAUTION ● Do not apply impact or force to the front screen LCD. • Doing so may cause damage. When packing this instrument in a box for transportation, do not face the front side (screen LCD) to the ground, and place the instrument flat or with the handle facing up.
In order for the product to be used reliably for a long time, parts must be inspected and replaced through periodic maintenance. Refer to "7-1. MAINTENANCE DETAILS" for guidelines on when to replace parts. Contact KYOWA sales personnel or your nearest dealer for information on preventive maintenance and replacing limited lifespan parts.
NOTATIONS USED IN THE INSTRUCTION MANUAL Notations used in the instruction manual ● "This device" refers to the UCAM-80A Data Logger. ● "Tap" refers to lightly pressing the device screen with your finger. ● Screen names displayed on the device screen are enclosed with rectangle frames.
PRECAUTIONS ON CE MARKING PRECAUTIONS ON CE MARKING Product has been confirmed that it complies with EN61326-1:2013 (EMC test requirements for equipment intended to be used in an industrial electromagnetic environment, Class A Group 1) under the following conditions. When using this product in EU countries, be sure to meet the following conditions.
1. PRODUCT OUTLINE 1-1. PRODUCT OUTLINE The UCAM-80A Data Logger ("this device") is a highly accurate multi-input and multichannel static strain measuring instrument. It uses a touch panel for intuitive operation, allowing for measurements to be made efficiently. Strain gages, strain gage...
PRODUCT OUTLINE 1-3. CONTROLS AND INDICATORS Front (4) (5) (6) (10) (11) <Printer lid closed> (12) (13) (14) (15) (16) (17) <Printer lid open> (1) Printer lid This covers the printer. In order to replace recording paper, insert an SD card, or connect a USB memory device, the lid must first be opened.
Page 22
PRODUCT OUTLINE (3) Recording paper output Recording paper is output here when printing. When cutting recording paper, grab either the left or right side of the recording paper, pull it upward, and then cut it. (4) feed key Used to feed printer paper. (5) printer key Used to select whether to print measurement values.
PRODUCT OUTLINE Back (3) (4) (11) (7) (8) (10) (1) External scanner connector Used to connect USB-80/70 Series External Scanner. (2) Expansion panel Covers the expansion connector. The expansion connector is currently disabled. Do not remove the expansion panel. (3) USB 2.0 connector USB interface connector.
PRODUCT OUTLINE Grab the arrow and remove the scanner cover. To remove the scanner cover, grab the rear side and pull it in the direction indicated with the arrow.
Page 25
PRODUCT OUTLINE SLOT1 SLOT2 SLOT3 channel channel channel 0 to 9 10 to 19 20 to 29 * The illustration above shows three USM-82A units installed. (1) Scanner cover Used to cover the scanning units. (2) Input terminal block Input terminals (screw soldered terminal block) for measurement. The screw diameter for the terminal block is M3, and the recommended tightening torque is 0.49 Nm.
PRODUCT OUTLINE Scanning Unit Top USM-81A, USM-83A (1)-1 (1)-2 (1) Input terminals Input terminals (screw soldered terminal block) for measurement. (1)-1 Soldering lug Used to solder a cable. (1)-2 Screw with metal plate The screw diameter for the terminal block is M3, and the recommended tightening torque is 0.49 Nm.
Page 27
PRODUCT OUTLINE USM-82A (1)-1 (1)-2 (1) Input terminals Input terminals (screw soldered terminal block) for measurement. (1)-1 Soldering lug Used to solder a cable. (1)-2 Screw with metal plate The screw diameter for the terminal block is M3, and the recommended tightening torque is 0.49 Nm.
Page 28
PRODUCT OUTLINE CAUTION ● Do not connect KYOWA "FV-1A" or "FV-2A" NDIS connector to BNC conversion adapters. ● Do not connect to the input terminal and the NDIS connector on the same channel at the same time.
PRODUCT CONNECTIONS 2. PRODUCT CONNECTIONS 2-1. GROUND CONNECTION Be sure to connect the GND terminal to ground. In addition to helping prevent noise, this can also help prevent electric shock caused by current leaking from the power supply noise filter. Use green plastic-coated electrical wire (0.75 mm or more).
PRODUCT CONNECTIONS Measured Object Is Insulated from Ground (1) If not using an external scanner, connect the measured object and the GND terminal of this device, and then ground the GND terminal of this device. (2) If using an external scanner, connect the measured object and the GND terminal of USB-80/70 Series External Scanner.
PRODUCT CONNECTIONS Measured Object Has Potential to Ground (1) Ground the GND terminal on either this device or on USB-80/70 Series External Scanner. CAUTION If the measured object has potential to ground, ensure that the potential of the object is not input into this device. Doing otherwise may worsen performance or cause failure.
PRODUCT CONNECTIONS Connecting with External Device Where DCOM and GND Are Not Connected (1) Connect the GND terminals on this device and on the external device. (2) Ground the GND terminal on either this device or on the external device. Connecting with External Device Where DCOM and GND (or Chassis) Are Connected (1) Connect the GND terminals on this device and on the external device.
PRODUCT CONNECTIONS 2-2. LIGHTNING SURGE PROTECTION GROUND CONNECTION When implementing lightning surge protection using this device and an external scanner, be sure to connect the lightning surge protection ground to the E terminal and GND terminal on each input. Prepare a separate Class A ground for use as the lightning surge protection ground. The effectiveness of lightning surge protection will depend on the quality of grounding and connective wiring.
Sensor Lightning Surge Protection Ground Connection (1) KYOWA strain gage transducers Connect the JB-100D KYOWA Lightning Arrester Kit near (approx. 1 m) each transducer, and connect a separate Class A ground for lightning surge protection. (Refer to the JB-100D Instruction Manual for details.)
In this case, use an insulation transformer (optional product: UPT-300BR) or commercially available Noisecuttrans to implement anti-noise measures. KYOWA Insulation Transformer (model UPT-300BR) ● Input: 100 / 120 / 220 / 240 V ● Output: 100 V, capacity 300 VA...
Page 36
PRODUCT CONNECTIONS CAUTION ● If using 100 VAC generated from an uninterruptible power supply (UPS) or battery as the power supply for this device, ensure that AC is output as a sine wave. Some UPS products output AC as a square wave instead of a sine wave. Square wave AC may cause a malfunction.
Page 38
(due to gage lead wire temperature effect, initial value discrepancy, etc.). ● If measuring in an environment with a high amount of noise, use KYOWA low- noise lead wires (L-13/14/15/16).
Page 40
PRODUCT CONNECTIONS CAUTION ● If using a strain gage with a gage resistance of 120 Ω or lower with BV 5 V (applicable channel modes: 1-gage 120 Ω_5 V, 1-gage 120 Ω D_5 V,1-gage 120 Ω A_5 V, 2-gage A-D_5 V, 2-gage A-A_5 V, 4-gage HR), and performing strain measurement for a material with poor heat dissipation (such as plastic or wood) or using a strain gage with a short gage length, strain gage self-heating may have an effect on measurement.
Page 41
PRODUCT CONNECTIONS Bridge Circuit Connection Method : Strain gage : Built-in fixed resistor Measurement Channel Mode A: Active gage A: Active gage Method D: Dummy gage D: Dummy gage E: Bridge exciter e: Bridge output 4-gage method (Full bridge system) (Opposite side White active method)
Page 42
PRODUCT CONNECTIONS 4-gage method (Full bridge system) Full bridge method White Civil engineering Black transducer with 4-gage T Green temperature Shield measuring function Bridge resistance 350 Ω ■ (*1): If using channel mode 4-gage HR, 4-gage I (120 Ω), or 4-gage I (350 Ω)_HR, this cannot be used with an external scanner.
PRODUCT CONNECTIONS Potentiometer Sensor Connection Measurement Connection Method Channel Mode Method Potentiometer sensor Potentiometer Potentiometer resistance 1 kΩ to 10 kΩ Shield Voltage/Current Output Transducer Connection Measurement Connection Method Channel Mode Method Voltage Voltage Voltage 500 mV measurement source Voltage 50 V Shield Current measurement...
Page 44
● Do not input voltage or current outside of the specified range. Doing so may cause electric shock, reduced performance, or failure. ● When connecting input, make sure that the positive and negative sides are connected correctly. ● Do not connect KYOWA "FV-1A" or "FV-2A" NDIS connector to BNC conversion adapters.
PRODUCT CONNECTIONS Thermocouple Connection Measurement Connection Method Channel Mode Method Thermocouple K Thermocouple Thermocouple T temperature Thermocouple E measurement Thermocouple J Thermocouple R Shield CAUTION ● When measuring, be sure to cover the terminals with the terminal cover and ensure that the terminal temperature reaches room temperature. ●...
PRODUCT CONNECTIONS 2-5. EXTERNAL DEVICES CONNECTIONS Connection with USB-80/70 Series Units (1) Connect "external scanner connector" and the "TO INDICATOR connector" on the USB-80/70 Series unit using the connection cable included with the scanner. (2) If connecting a second scanner, connect the "TO NEXT SCANNER connector" on the first scanner and the "TO INDICATOR connector"...
Page 48
PRODUCT CONNECTIONS If using the following type of connection with USB-80/70 Series, attach an optional AC power supply unit (UPS-80A/70B) to all scanners. ● One USB-80/70 unit, with a cable length of 240 m or more ● Two USB-80/70 units, with a total cable length exceeding 120 m ●...
PRODUCT CONNECTIONS 2-6. CONNECTION WITH A PC There are three interfaces for connecting to a PC: RS-232C, USB, and Ethernet. RS-232C Connection ● Connection with PC Connect this device to the PC using an RS-232C cable. Use the following cable to connect to a PC. Compatible cable (D-sub 9P, female-female) Recommended: Sanwa Supply RS-232C Crossover Cable (3 m) Model: KRS-403XF3K2, etc.
Page 50
PRODUCT CONNECTIONS ● USB-RS232C converter Recommended: I-O DATA USB-RSAQ7R, etc. (Device side: D-sub 9P male, PC side: USB Type B) * A USB driver will need to be installed separately on the PC. Refer to the manufacturer's USB-RS232C converter website for information on compatible OS and how to install the USB driver.
PRODUCT CONNECTIONS USB Connection ● Connection with PC Connect this device to a PC using a USB cable (Type A - Type B). USB cable (Type B - Type A) ● USB driver setup In order to control the device over USB, a USB driver will need to be setup separately on the PC.
Page 52
PRODUCT CONNECTIONS (1) Select CDM212364_Setup.exe on the included CD-ROM, right-click, and then select "Run as administrator (A)." (2) If the User Account Control pop-up window is displayed, click Yes (Y) to continue. (3) Click Extract.
Page 53
PRODUCT CONNECTIONS (4) Click Next >. (5) The license agreement is displayed. To agree to the license agreement, select "I accept this agreement" and then click Next > to start installation. Selecting "I don’t accept this agreement" will prevent installation from continuing.
Page 54
PRODUCT CONNECTIONS (6) Installation is complete. Click Finish to finish installation. If no USB driver is installed on the PC, connecting this device to the PC will cause the PC to automatically search for the required driver on the network and then install it.
PRODUCT CONNECTIONS Ethernet Connection ● Connection with PC Connect this device to the PC using LAN cable (Category 5 or higher). LAN cable (Category 5) CAUTION ● In order to connect to a network, an IP address and other settings must first be configured.
Page 56
PRODUCT CONNECTIONS The procedure for configuring network settings on the PC is provided below. In order to connect to this device using an Ethernet connection, the IP address on the PC will need to be manually set to the same network as this device. In the "Run"...
Page 57
PRODUCT CONNECTIONS Enter an IP address and subnet mask so that the PC is in the same network as this device, and then press OK. CAUTION Communication is not possible if the IP address of this device is the same as that of the PC.
PRODUCT CONNECTIONS 2-7. SD CARD INSERTION This device is capable of saving measurement values and measurement conditions to an SD card. To use an SD card, insert it as shown in the photo below. Notch PROHIBITED SD cards must be inserted correctly. Inserting an SD card incorrectly may damage the product or SD card.
Page 59
● If the SD card is not recognized, start this device with the SD card removed. Insert the SD card after a while after the Menu screen is displayed. ● The following SD card has been tested by KYOWA. Manufacturer : HAGIWARA Solutions...
● If the USB memory device is not recognized, start this device with the USB memory device removed. Insert the USB memory device after a while after the Menu screen is displayed. ● The following USB memory device has been tested by KYOWA. Manufacturer : Green house Model...
TURNING THE POWER ON 3. TURNING THE POWER ON Turn the POWER switch on the rear of this device ON to start the software. The following Menu screen will be displayed once it has finished starting. Number Name Function Displays the measurement status of this device. •...
Page 62
TURNING THE POWER ON CAUTION ● The device will be ready to use approximately 1 minute after turning the power ON. However, internal parts will not be thermally stable immediately after receiving power. In order to reliably operate the device, allow it to warm up for approximately 30 to 40 minutes.
BASIC OPERATING METHOD 4. BASIC OPERATING METHOD 4-1. HARDWARE KEYS The keys on the outside of the screen are called the "hardware keys." Key Upper-Right LED Number Name Function Starts/stops real-time ON: Scanning scan key measurement OFF: Scanning stopped ON: Monitoring monitor key Starts/stops monitor OFF: Monitor stopped...
BASIC OPERATING METHOD Scan Key Used to start/stop real-time measurement. Refer to "5-6-1. Real-Time Measurement" for details. Monitor Key Used to start/stop the monitor. Refer to "5-6-3. Monitor" for details. Program Run/Stop Key Press and hold for at least 2 seconds to start/stop automatic measurement. Refer to "5- 6-2.
BASIC OPERATING METHOD Auto Save Key Press and hold for at least 2 seconds to select whether to save measurement values to a measurement data file. Printer Key Press and hold for at least 2 seconds to select whether to print measurement values. Feed Key Used to feed printer paper.
BASIC OPERATING METHOD 4-2. SCREEN This device can be operated by tapping the screen. The basic methods of operating the screen are described below. Moving between Screens/Pages Number Name Function Back button Returns to the previous screen. Moves to the next screen lower in the Arrow button hierarchy.
BASIC OPERATING METHOD Selecting Items Number Name Function Radio button Used to select one of multiple items. Check box Used to select one or several of multiple items. Select range button Used to select a range of items. Select all button Used to select all items at once.
BASIC OPERATING METHOD Entering Numbers Number Name Function Used to enter a number. Entry field Tap to begin entering information. Number buttons Used to enter numbers. Move cursor buttons Moves the entry field cursor left/right. +/- button Used to select positive/negative numbers. Decimal point button Used to enter a decimal point.
BASIC OPERATING METHOD Entering Text Number Name Function Text buttons Used to enter text characters. Switch text button Switches the text buttons. Switch case button Switches between uppercase and lowercase.
BASIC OPERATING METHOD Selecting Channels Number Name Function Tap to display a keyboard for selecting a channel. Select channel button The keyboard can be used to enter a channel number and move to that channel number. Previous channel button Moves to the previous channel number. Next channel button Moves to the next channel number.
BASIC OPERATING METHOD Confirming Measurement Values Number Name Function Tap to display a screen for selecting a group. Filter button A group can be selected to display only certain channels. Refer to "5-1-5. Group Settings" for details on groups. Shrink button Increases the number of channels displayed.
BASIC OPERATING METHOD Menu Screen Once the software has finished starting, the first screen that is displayed is the Menu screen. Tap a button on the Menu screen to access the corresponding screen. Number Name Function Channel Moves to the screen used to configure channels. Measure Moves to the screen used to configure measurement.
MEASUREMENT PROCEDURE 5. MEASUREMENT PROCEDURE The procedure from connecting the device to measuring data is shown below. Follow this procedure to start working. Refer to "6. FUNCTIONS" for an explanation of other functions. Install software (only If this device will be controlled from a PC, first install when using this device the control software on the PC.
Page 74
MEASUREMENT PROCEDURE Configure channel settings. Set channel conditions Refer to "5-2. CHANNEL CONDITION SETTINGS." Configure measurement value settings. Configure Refer to "5-3. MEASUREMENT VALUE SETTINGS." measurement values Configure saving Configure settings for saving measurement values. Refer to "5-4. SAVING SETTINGS." Configure automatic If automatic measurement will be used, configure measurement...
Page 75
MEASUREMENT PROCEDURE If this device will be controlled from a PC, use the Collect measurement data files control software to collect measurement data files. If this device will not be controlled from a PC, use a USB memory device or SD card to collect measurement data files.
MEASUREMENT PROCEDURE 5-1. MEASUREMENT CONDITIONS SETTINGS Measurement settings are called "measurement conditions." This section describes how to set measurement conditions. 1. Tap Measure on the Menu screen. 2. The Measure screen is displayed.
Page 77
MEASUREMENT PROCEDURE The Measure screen is used to set measurement conditions, such as the range of channels used. The following functions are available on the Measure screen. (10) Reference Page Number Name Function Internal Scanner Scanning unit settings 5-1-1 External Scanner External scanner settings 5-1-2 Cold junction...
Page 78
MEASUREMENT PROCEDURE *1) No need to be executed if the stroke change function is not used. *2) No need to be configured if the External I/O Unit rainfall input function is not used. If an External I/O Unit is not connected, Rainfall cannot be tapped. *3) No need to be configured if the External I/O Unit alarm function is not used.
MEASUREMENT PROCEDURE Scanning Unit Settings Devices connected to the top of this device are called "scanning units." The relationship between scanning units and channel numbers is shown below. SLOT1 SLOT2 SLOT3 channel channel channel 0 to 9 10 to 19 20 to 29 Channel 0 Channel 1...
Page 80
MEASUREMENT PROCEDURE 1. Tap Internal Scanner. 2. If using a scanning unit, tap ON. If not using a scanning unit, tap OFF. 3. If using a scanning unit, tap Meas. Channel. 4. After entering the range of channels to use, tap the Enter button.
Page 81
MEASUREMENT PROCEDURE For example, if using channels 0 through 4 on SLOT1 (indicated below in the yellow box), the first channel number would be 0 and the last channel number would be 4. Enter a channel range from 0000 to 0004. Range of channels to use...
MEASUREMENT PROCEDURE External Scanner Settings Devices connected to the external scanner connectors on the rear of this device are called "external scanners." The channel number of a USB-80/70 Series External Scanner is determined through a combination of the number specified using the USB-80/70 Series CHANNEL switch (00 to 99), and the channel number (0 to 9).
Page 83
MEASUREMENT PROCEDURE 1. Tap External Scanner. 2. If using an external scanner, tap ON. If not using an external scanner, tap OFF. 3. If using an external scanner, tap Meas. Channel. 4. After entering the range of channels to use, tap Enter.
Page 84
MEASUREMENT PROCEDURE 5. CHANNEL switch settings when using USB-80A/70B To prevent any overlap with the range of channels set for Internal Scanner, configure the USB-80A/70B CHANNEL switches. For example, if the range of channels set for Internal Scanner is 0000 to 0029, configure the USB-80A/70B CHANNEL switches as follows.
Page 85
MEASUREMENT PROCEDURE 6. CHANNEL switch settings when using USB-70A To prevent any overlap with the range of channels set for Internal Scanner, configure the USB-70A CHANNEL switches. For example, if the range of channels set for Internal Scanner is 0000 to 0029, configure the USB-70A CHANNEL switches as follows.
Page 86
MEASUREMENT PROCEDURE 7. MODE SET switch setting when using USB-70A If using the screen of this device to set the channel mode of USB-70A, set the MODE SET switch to REMOTE. If using the USB-70A MODE switch to set the channel mode of USB-70A, set the MODE SET switch to SCANNER.
Page 87
MEASUREMENT PROCEDURE 8. MODE switch setting when using USB-70A If the USB-70A MODE SET switch is set to SCANNER, the USB-70A MODE switch can be set to select the USB-70A channel mode. The relationship between the MODE switch number and channel mode is shown toward the top of USB-70A.
Page 88
MEASUREMENT PROCEDURE 9. Enter the range of channels to use for USB-70A/70B/80A. For example, if using channels 30 through 34 in the first slot (indicated in the yellow box below), the first channel number would be 30 and the last channel number would be 34.
MEASUREMENT PROCEDURE Cold Junction Compensation Settings If temperature measurement will be performed using a thermocouple, configure the cold junction compensation method. There are two cold junction compensation methods: (1) Internal Perform cold junction compensation inside this device. (2) External Perform cold junction compensation outside this device, take the output after compensation, and then input it into this device.
MEASUREMENT PROCEDURE Measurement Timing Settings Set whether to synchronize with the power frequency when performing measurement. There are three types of measurement timing: (1) Power synchronized Measure in synchronization with the power frequency. When the measurement timing is set to power synchronized, this device will automatically detect the power frequency.
Page 91
MEASUREMENT PROCEDURE 1. Tap Meas. Timing. 2. Tap the measurement timing to set. If the measurement timing is set to power synchronized but this device cannot detect the power frequency, the Meas. Timing screen will appear as shown below. If the power frequency could not be detected, set the measurement timing to power non- synchronized.
MEASUREMENT PROCEDURE Group Settings Groups are used to display only certain channels during measurement. Channels to display in a group can be registered ahead of time, in order to display only certain channels from among multiple channels. The procedure for registering channels to a group is described below. 1.
Page 93
MEASUREMENT PROCEDURE 4. Tap the check boxes of the channels to register to the group. 5. Tap the Back button. 6. The channels are added to the group. There are 20 groups. Up to 50 channels may be registered to a single group.
Page 94
MEASUREMENT PROCEDURE The procedure for using a group to display only certain channels is described below. The procedure when monitoring is shown here as an example. 1. After moving to the Menu screen, press the monitor key. 2. After moving to the Monitor screen, tap CH.
Page 95
MEASUREMENT PROCEDURE 5. Only the channels registered to the group are displayed.
Page 96
MEASUREMENT PROCEDURE The procedure for deleting a channel from a group is described below. 1. Tap Group. 2. Tap the group from which to delete a channel. 3. Tap the channel to delete. 4. Tap the Option button.
Page 97
MEASUREMENT PROCEDURE 5. Tap Delete. 6. The channel is deleted.
MEASUREMENT PROCEDURE Delta V and Delta T Settings The Delta V function is used to subtract the initial value from the measurement value on a channel with the channel mode set to voltage as the measured object, when the measurement value setting is "measured value." The Delta T function is used to subtract the initial value from the measurement value on a channel with the channel mode set to temperature as the measured object, when the measurement value setting is "measured value."...
MEASUREMENT PROCEDURE 5-2. CHANNEL CONDITION SETTINGS Channel settings are called "channel conditions." This section describes how to set channel conditions. 1. Tap Channel on the Menu screen. 2. Tap the arrow button of the channel to set. 3. The Condition screen is displayed.
Page 100
MEASUREMENT PROCEDURE The Condition screen is used to set the calibration coefficient and other channel conditions. Channel conditions are set for each channel. The following functions are available on the Condition screen. (10) (11) (12) Reference Number Name Function Page 5-2-1 Measurement Measurement ON/OFF...
MEASUREMENT PROCEDURE Measurement ON/OFF Turn Measurement ON for any channels used for measurement. 1. Tap Measurement. 2. To turn Measurement ON, tap ON. To turn Measurement OFF, tap OFF. Real-Time Measurement, Measurement Automatic Measurement, During Monitoring Measure Do not measure (measurement values will not be saved) Refer to "5-6-1.
MEASUREMENT PROCEDURE Calibration Coefficient Settings The calibration coefficient is used to convert measurement values into physical quantities. Set the calibration coefficient in order to convert measurement values into physical quantities. 1. Tap Calibr. Const.. 2. Enter the calibration coefficient and then tap Enter.
MEASUREMENT PROCEDURE Unit Settings If calibration coefficient calculation is enabled, set the unit for displaying physical quantities. 1. Tap Unit. 2. Tap the unit to set. The unit can also be set by tapping one of the units shown in Log. ...
MEASUREMENT PROCEDURE Reference Resistance Value Settings The reference resistance value is the resistance at 0°C for a platinum resistance thermometer or civil engineering transducer with temperature measuring function. If using a platinum resistance thermometer or civil engineering transducer with temperature measuring function, set the reference resistance value.
MEASUREMENT PROCEDURE Scanning Speed Settings This device performs measurement while switching channels. Measuring while switching channels is called "scanning." The time required for measurement on one channel is called the "scanning speed." Scanning Description Speed Measurement is performed at a scanning speed of 20 ms per channel.
Page 106
MEASUREMENT PROCEDURE Some scanning speeds may not be compatible depending on the channel mode. For example, a channel set to a channel mode of 4-gage HR will not scan faster than 0.28 sec. The following table summarizes channel modes and incompatible scanning speeds. Incompatible Scanning Channel Mode Speeds...
MEASUREMENT PROCEDURE Channel Mode Settings Modes compatible with each sensor connected to this device or an external scanner must be set for each channel. These modes are called "channel modes". Measured Channel Mode Function Object 1G 120 Ω 1-gage (120 Ω, BV2V) Strain 1G 240 Ω...
Page 108
MEASUREMENT PROCEDURE Measured Channel Mode Function Object 1G 120 Ω_5 V 1-gage (120 Ω, BV5V) Strain 1G 240 Ω_5 V 1-gage (240 Ω, BV5V) Strain 1G 350 Ω_5 V 1-gage (350 Ω, BV5V) Strain 1G 120 Ω D_5 V 1-gage (120 Ω, dummy, BV5V) Strain 1G 120 Ω...
Page 109
MEASUREMENT PROCEDURE 1. Tap Mode. 2. Tap the channel mode to set. You can also tap a channel mode in Log to set the channel mode. Log shows the most recently set channel modes.
MEASUREMENT PROCEDURE Decimal Point Position Settings If calibration coefficient calculation is enabled, set the decimal point position when displaying measurement values as physical quantities. Decimal Point Description Position ###### No decimal point #####.# Display 1 digit after decimal point Display 2 digits after decimal point ####.## ###.### Display 3 digits after decimal point...
MEASUREMENT PROCEDURE Offset Value Settings The offset value is a variable applied to measurement values during calibration coefficient calculation. If using calibration coefficient calculation, set the offset value. 1. Tap Offset. 2. Enter the offset value and then tap Enter. Calibration coefficient calculation must be turned ON in order to apply the offset value to measurement values.
MEASUREMENT PROCEDURE Initial Value Settings The initial value is the variable used when the measurement value setting is "measured value." Follow the procedure below to set the initial value manually. 1. Tap Initial. 2. Enter the initial value and then tap Enter.
MEASUREMENT PROCEDURE Multiple Channel Condition Settings The procedure for setting channel conditions for multiple channels is shown below. 1. Tap Channel on the Menu screen. 2. Tap Sel.. 3. Tap the entry field on the left side. 4. Tap either the Backspace button or the Clear all button, and then enter the first channel number to configure.
Page 114
MEASUREMENT PROCEDURE 5. Tap the entry field on the right side. 6. Tap either the Backspace button or the Clear all button, and then enter the last channel number to configure. 7. Tap Enter. 8. The check boxes for the selected range of channels are selected.
Page 115
MEASUREMENT PROCEDURE 9. Tap the arrow button for any of the channels in the selected range. 10. Tap the arrow button for the channel condition to set for all selected channels. In this example, Measurement will be set. 11. Tap ON or OFF. 12.
Page 116
MEASUREMENT PROCEDURE 13. The display returns to the Condition screen. Any channel conditions indicated with ― here will not be set. 14. Tap the Back button. 15. Tap Set to set the channel condition for all selected channels. To cancel, tap Cancel.
MEASUREMENT PROCEDURE 5-3. MEASUREMENT VALUE SETTINGS This section describes how to set measurement values. There are two types of measurement values: (1) Original value Raw sensor output is displayed as the measurement value. Use this if the initial unbalance has already been compensated on the sensor, or to confirm raw sensor output.
MEASUREMENT PROCEDURE Original Value or Measured Value Selection 1. Tap Function on the Menu screen. 2. The Function screen is displayed. The Function screen is used to select either original value or measured value. To use the original value, select the Original radio button. To use the measured value, select the Measure radio button.
MEASUREMENT PROCEDURE Initial Value Measurement If the measurement value setting is "measured value," perform initial value measurement. This device stores the measurement value obtained during initial value measurement as the initial value. 1. Tap the Measure arrow button. 2. Tap Initial. 3.
Page 120
MEASUREMENT PROCEDURE 4. Tap the Meas. button to begin initial value measurement. CAUTION Performing initial value measurement will overwrite the previous initial value. Do not perform initial value measurement on a channel unless you want to change the initial value. 5.
Page 121
MEASUREMENT PROCEDURE The following procedure can also be used to directly change the initial value. 1. Tap the arrow button for the channel to change the initial value. 2. Enter the initial value, and then tap Enter. 3. The initial value is changed. CAUTION Changing the initial value will overwrite the previous initial value.
MEASUREMENT PROCEDURE Calibration Coefficient Calculation Settings Calibration coefficient calculation is used to convert measurement values into physical quantities. If calibration coefficient calculation is enabled, this device will perform the following calculations on measurement values. (1) If the measurement value setting is "original value" Physical quantity = Measurement value ×...
Page 123
MEASUREMENT PROCEDURE 1. If using "original value," tap the Original arrow button. If using "measured value," tap the Measure arrow button. 2. To turn calibration coefficient calculation ON, tap ON. To turn calibration coefficient calculation OFF, tap OFF.
MEASUREMENT PROCEDURE 5-4. SAVING SETTINGS This section describes how to configure saving settings. 1. Tap Storage on the Menu screen. 2. The Storage Setting screen is displayed.
Page 125
MEASUREMENT PROCEDURE The Storage Setting screen is used to configure settings such as file names and storage locations. The following functions are available on the Storage Setting screen. Reference Number Name Function Page 5-4-1 File Name Set the name for measurement data files Storage Set the file storage location 5-4-2...
MEASUREMENT PROCEDURE Saved File Name Settings Set the name for measurement data files created when saving measurement values. 1. Tap File Name. 2. Enter the file name. The file name may be up to 32 characters long. Enter a file name so that the total number of characters will be 32 or less when the file name format is included.
Page 127
MEASUREMENT PROCEDURE 4. Tap the file name format to set. The yymmdd portion will be replaced with the year (yy), month (mm), and day (dd) when saving the file. The hhmmss portion will be replaced with the hour (hh), minutes (mm), and seconds (ss) when saving the file.
MEASUREMENT PROCEDURE Storage Location Settings Set the file storage location. 1. Tap Storage. 2. Tap the location to save files. Storage Description Location • Save files to the internal memory of this device. Internal • On the right is shown the amount of internal memory used, followed by the total capacity, and then the remaining amount (in parentheses).
MEASUREMENT PROCEDURE Saving Method Settings Set the saving method for measurement data files. 1. Tap Method. 2. The current saving method is displayed. Saving Description Method Measurement values are saved appended within a single Addition measurement data file.
Page 130
MEASUREMENT PROCEDURE This device creates measurement data files in the following situations. (1) If the file name format is "yymmdd_XXX.extension" A measurement data file is created the first time that measurement values are saved. Measurement values are then appended whenever real-time measurement or automatic measurement is performed.
MEASUREMENT PROCEDURE Saving Format Settings Set the saving format for measurement data files. 1. Tap Format. 2. The current saving format is displayed. Saving Description Format ASCII Save measurement values in ASCII format.
MEASUREMENT PROCEDURE 5-5. AUTOMATIC MEASUREMENT SETTINGS The automatic measurement function is used to automatically perform measurements under preset conditions. There are three types of automatic measurement: (1) Interval measurement The time to start measurement and the time interval are set in advance. Once the measurement start time is reached, automatic measurement begins.
Page 133
MEASUREMENT PROCEDURE As an example, an interval measurement operation using the following settings is shown below. Setting Item Setting Details Measurement start time 2022/01/01 12:00:00 Time interval 1 hour Repeat times Step 1 Operation once Go to next step repetition is complete Time interval 2 hours Repeat times...
Page 134
MEASUREMENT PROCEDURE (2) Trigger measurement The channel to monitor ("trigger channel") is set in advance. Once the "trigger channel measurement value" exceeds the threshold, measurement starts automatically. Scanning is then performed whenever the "trigger channel measurement value" exceeds the threshold. Trigger measurement settings are shown below. ...
Page 135
MEASUREMENT PROCEDURE The trigger mode is set to "relative value," the trigger value is set to "100," and the reference value is set to "0" in the following example. Example of relative value operation CAUTION If the trigger mode is set to "relative value," do not set the trigger value to "0."...
Page 136
MEASUREMENT PROCEDURE Step Trigger measurement is composed of at least one step. Set the trigger value, repeat times, and operation once repetition is complete, for each step. Steps are represented as numbers (from 1 to 15), so up to 15 steps can be created.
Page 137
MEASUREMENT PROCEDURE As an example, a trigger measurement operation using the following settings is shown below. Setting Item Setting Details Trigger mode Absolute value Trigger value Repeat times Step 1 Operation once Go to next step repetition is complete Trigger value Repeat times Step 2 Operation once...
Page 138
MEASUREMENT PROCEDURE (3) Trigger interval measurement The trigger channel and time interval are set in advance. When the "trigger channel measurement value" exceeds the threshold, measurement starts automatically. Scanning will then be performed at the set time intervals. Trigger channel Set the channel number of the trigger channel.
Page 139
MEASUREMENT PROCEDURE As an example, a trigger interval measurement operation using the following settings is shown below. Setting Item Setting Details Reference value Time interval 1 hour Repeat times Step 1 Operation once repetition Go to next step is complete Time interval 2 hours Repeat times...
Page 140
MEASUREMENT PROCEDURE Automatic measurement is configured on the Program screen. 1. Tap Program on the Menu screen. 2. The Program screen is displayed.
Page 141
MEASUREMENT PROCEDURE The Program screen is used to create, configure, assign, and delete programs. Programs are collections of automatic measurement settings. Configuring a program allows you to start automatic measurement simply by pressing the program run/stop key. The following functions are available on the Program screen. Reference Number Name...
MEASUREMENT PROCEDURE New Program Creation The procedure for creating a new program is shown below 1. Tap New. 2. A new program is created. Up to 15 programs may be created.
MEASUREMENT PROCEDURE Programs Settings The procedure for configuring a new program is shown below 1. Tap the arrow button of the program to configure. 2. The Program settings screen is displayed.
Page 144
MEASUREMENT PROCEDURE The Program settings screen is used to configure a program. The items displayed on the Program settings screen will vary depending on the automatic measurement mode setting. The following functions are available on the Program settings screen. Automatic measurement mode set to "interval" Number Name Function...
Page 145
MEASUREMENT PROCEDURE Automatic measurement mode set to "trigger" Number Name Function Name Change the program name. Set the automatic measurement mode. Any of the three following modes can be set. Mode • Interval • Trigger • Trigger interval Set the trigger channel. Trigger Also, set the trigger mode (absolute value or relative value).
Page 146
MEASUREMENT PROCEDURE Automatic measurement mode set to "trigger interval" Number Name Function Name Change the program name. Set the automatic measurement mode. Any of the three following modes can be set. Mode • Interval • Trigger • Trigger interval Set the trigger channel. Trigger Also, set the reference value to begin automatic measurement.
Page 147
MEASUREMENT PROCEDURE (1) Procedure for setting interval measurement 1. Tap Name. 2. Enter a program name, and then tap Enter. 3. Tap Mode. 4. Tap Interval.
Page 148
MEASUREMENT PROCEDURE 5. Tap the Back button. 6. Tap Start Condition. 7. To start interval measurement from a given time, tap the Start Time radio button. In order to press the program run/stop key and start interval measurement, tap the program run key radio button.
Page 149
MEASUREMENT PROCEDURE 9. Tap the Back button. 10. Tap Step. 11. Tap STEP01. 12. Tap Time Interval.
Page 150
MEASUREMENT PROCEDURE 13. Enter the time interval to perform scanning, and then tap Enter. CAUTION Set a time interval with some leeway for the "time for one scan." An example of calculating the "time for one scan" is shown below. •...
Page 151
MEASUREMENT PROCEDURE 14. Tap Repeat. 15. To indicate that interval measurement should continue until the program run/stop key is pressed, tap Infinite. To specify the count, tap the Sel. Times radio button. 16. To specify the count, tap the Sel. Times arrow button, and then enter the count.
Page 152
MEASUREMENT PROCEDURE 18. Tap Action 19. Tap the operation to set. 20. To set the specified step, select the To Selected Step arrow button, and then enter the step number. 21. Tap the Back button.
Page 153
MEASUREMENT PROCEDURE 22. Tap the Back button. 23. If setting step 02 onward, tap the arrow button of the step to set, and then set the step.
Page 154
MEASUREMENT PROCEDURE (2) Procedure for setting trigger measurement 1. Tap Name. 2. Enter a program name, and then tap Enter. 3. Tap Mode. 4. Tap Trigger.
Page 155
MEASUREMENT PROCEDURE 5. Tap the Back button. 6. Tap Trigger. 7. Tap Channel. 8. Determine the trigger channel and tap the corresponding channel radio button. You can also tap Sel. CH and enter the channel directly.
Page 156
MEASUREMENT PROCEDURE 9. Tap the Back button. 10. Tap Mode. 11. To set the trigger mode to "absolute value," tap the Absolute radio button. To set the trigger mode to "relative value," tap the Relative radio button. 12. To set the trigger mode to "relative value,"...
Page 157
MEASUREMENT PROCEDURE 13. Tap the Back button. 14. Tap the Back button. 15. Tap Step. 16. Tap STEP01.
Page 158
MEASUREMENT PROCEDURE 17. Tap Trigger Value. 18. Enter the trigger value and then tap Enter. 19. Tap Repeat. 20. To indicate that trigger measurement should continue until the program run/stop key is pressed, tap Infinite. To specify the count, tap the Sel. Times radio button.
Page 159
MEASUREMENT PROCEDURE 21. To specify the count, tap the Sel. Times arrow button, and then enter the count. 22. Tap the Back button. 23. Tap Action. 24. Tap the operation to set.
Page 160
MEASUREMENT PROCEDURE 25. To set the specified step, select the To Selected Step arrow button, and then enter the step number. 26. Tap the Back button. 27. Tap the Back button. 28. If setting step 02 onward, tap the arrow button of the step to set, and then set the step.
Page 161
MEASUREMENT PROCEDURE (3) Procedure for setting trigger interval measurement 1. Tap Name. 2. Enter a program name, and then tap Enter. 3. Tap Mode. 4. Tap Trigger Interval.
Page 162
MEASUREMENT PROCEDURE 5. Tap the Back button. 6. Tap Trigger. 7. Tap Channel. 8. Determine the trigger channel and tap the corresponding channel radio button. You can also tap Sel. CH and enter the channel directly.
Page 163
MEASUREMENT PROCEDURE 9. Tap the Back button. 10. Tap Reference. 11. Enter the reference value and then tap Enter. 12. Tap the Back button.
Page 164
MEASUREMENT PROCEDURE 13. Tap Step. 14. Tap STEP01. 15. Tap Time Interval. 16. Enter the time interval to perform scanning, and then tap Enter.
Page 165
MEASUREMENT PROCEDURE CAUTION Set a time interval with some leeway for the "time for one scan." An example of calculating the "time for one scan" is shown below. • Channel range: Use from CH0 to CH2 • CH0 scanning speed: 0.5 sec •...
Page 166
MEASUREMENT PROCEDURE 17. Tap Repeat. 18. To indicate that interval measurement should continue until the program run/stop key is pressed, tap Infinite. To specify the count, tap the Sel. Times radio button. 19. To specify the count, tap the Sel. Times arrow button, and then enter the count.
Page 167
MEASUREMENT PROCEDURE 21. Tap Action. 22. Tap the operation to set. 23. To set the specified step, select the To Selected Steparrow button, and then enter the step number. 24. Tap the Back button.
Page 168
MEASUREMENT PROCEDURE 25. Tap the Back button. 26. If setting step 02 onward, tap the arrow button of the step to set, and then set the step.
MEASUREMENT PROCEDURE Program Assignment The procedure for assigning a program to run when the program run/stop key is pressed is shown below. 1. Tap the check box of the program to assign. 2. Tap Alloc.. 3. On the right of the assigned program, Allocated is displayed.
Page 170
MEASUREMENT PROCEDURE 4. Tap the Back button. 5. Beneath Program on the Menu screen, the name of the assigned program is displayed. 6. While the Menu screen is displayed, pressing and holding the program run/stop key will start automatic measurement according to the assigned program.
MEASUREMENT PROCEDURE Program Deletion The procedure for deleting programs is shown below 1. Tap the check box of the program to delete. 2. Tap Del.. 3. Tap Delete. 4. The program is deleted.
MEASUREMENT PROCEDURE Program Duplication The procedure for duplicating programs is shown below. 1. Tap the check box of the program to duplicate. 2. Tap the Option button. 3. Tap Copy. 4. The program is duplicated.
MEASUREMENT PROCEDURE Program Locking The procedure for locking a program (preventing it from being configured or deleted) is shown below. 1. Tap the check box of the program to lock. 2. Tap the Option button. 3. Tap Lock. Tap Unlock to unlock. 4.
MEASUREMENT PROCEDURE Detailed Preview Display The detailed preview function displays the schedule for performing a scan during a program. The procedure for displaying a detailed preview is described below. 1. To display a detailed preview for a program, tap its arrow button. 2.
Page 175
MEASUREMENT PROCEDURE 4. The detailed preview is displayed. A detailed preview can be displayed only for programs where the automatic measurement mode is "interval."...
MEASUREMENT PROCEDURE 5-6. MEASUREMENT There are three types of measurement: (1) Real-time measurement This function obtains measurement values one time only. When the scan key is pressed, real-time measurement starts. The screen during real-time measurement is shown below. Number Name Function Displays the measurement status of this device.
Page 177
MEASUREMENT PROCEDURE (2) Automatic measurement This function automatically obtains measurement values based on program settings. Press and hold the program run/stop key to start automatic measurement. For details on programs, refer to "5-5. AUTOMATIC MEASUREMENT SETTINGS." The screen during automatic measurement is shown below. Number Name Function...
Page 178
MEASUREMENT PROCEDURE (3) Monitor This function displays measurement values in series. Press the monitor key to start monitoring. Use this to confirm measurement values prior to starting real-time measurement or automatic measurement. The screen during monitoring is shown below. Number Name Function Displays the measurement status of this device.
MEASUREMENT PROCEDURE Real-Time Measurement The procedure for performing real-time measurement is shown below. 1. To save a measurement data file, press and hold the auto save key until the LED to the upper right of the auto save key lights up. CAUTION If the LED to the upper right of the auto save key is not lit up, a measurement data file will not be saved.
Page 180
MEASUREMENT PROCEDURE 4. Real-time measurement starts. To stop real-time measurement, press the scan key again. 5. Once real-time measurement ends, the display will automatically return to the Menu screen. 6. Tap Data. 7. Measurement values obtained during real-time measurement are displayed. Refer to "5-7-1.
MEASUREMENT PROCEDURE Automatic Measurement The procedure for performing automatic measurement is shown below. 1. To print measurement values, press and hold the printer key until the LED to the upper right of the printer key lights up. For details on printing, refer to "6-14.
Page 182
MEASUREMENT PROCEDURE CAUTION During automatic measurement, do not move to any screen other than the Menu screen. Scanning will not be performed on any screen other than the Menu screen. 4. Scanning will automatically be performed once the conditions are met.
Page 183
MEASUREMENT PROCEDURE 7. The File screen is displayed. For details on this screen, refer to "5- 7-2. Measurement Data File Confirmation." During interval measurement, the following information is displayed at the top of the screen. (4) (5) Number Details Displays the time to start measurement. In the example above, measurement will start at 2022/01/01 12:00:00.
Page 184
MEASUREMENT PROCEDURE During trigger measurement with the trigger mode set to "absolute value," the following information is displayed at the top of the screen. (2) (3) Number Details Displays the number of the next step to perform. In the example above, "Step 1" will be performed next. Displays the scan count of the next scan to perform.
Page 185
MEASUREMENT PROCEDURE During trigger measurement with the trigger mode set to "relative value," the following information is displayed at the top of the screen. (3) (4) Number Details Displays the reference value to start measurement. In the example above, measurement will start once the trigger channel measurement value exceeds 100.
Page 186
MEASUREMENT PROCEDURE During trigger interval measurement, the following information is displayed at the top of the screen. (3) (4) Number Details Displays the reference value to start measurement. In the example above, measurement will start once the trigger channel measurement value exceeds 100. Displays the number of the next step to perform.
MEASUREMENT PROCEDURE Monitor The procedure for performing monitoring is shown below. 1. While the Menu screen is displayed, press the monitor key. 2. Monitoring starts. To stop monitoring, press the monitor key again. 3. To increase/decrease the number of channels displayed, tap the Shrink button or Expand button.
Page 188
MEASUREMENT PROCEDURE 5. To display only certain channels, tap the Filter button. 6. Tap a group. 7. Tap the Back button. 8. Only the channels registered to the group are displayed.
MEASUREMENT PROCEDURE 5-7. MEASUREMENT VALUE CONFIRMATION Latest Measurement Values Confirm measurement values obtained through real-time measurement or automatic measurement. 1. Tap Data on the Menu screen. 2. The Data screen is displayed.
Page 190
MEASUREMENT PROCEDURE The Data screen is used to confirm measurement values. The following functions are available on the Data screen. Number Name Function Filter button A group can be selected to display only certain channels. Increases the number of channels displayed. Shrink button The number of channels can be set to either 4, 12, or 24.
MEASUREMENT PROCEDURE Measurement Data File Confirmation Confirm measurement data files saved through real-time measurement or automatic measurement. 1. Tap File on the Menu screen. 2. The File screen is displayed.
Page 192
MEASUREMENT PROCEDURE The File screen is used to confirm files saved to memory. The following functions are available on the File screen. Number Name Function • Displays files saved to internal memory. Internal • On the right is shown the amount of internal memory memory used, followed by the total capacity.
Page 193
MEASUREMENT PROCEDURE 3. Tap the memory location containing the file. 4. Tap the arrow button of the file to confirm. For information on file extensions, refer to the following table. Extension Meaning Measurement data file saved during real-time measurement Measurement data file saved during interval measurement Measurement data file saved during trigger measurement Measurement data file saved during trigger interval measurement Measurement data file saved during initial value measurement...
Page 194
MEASUREMENT PROCEDURE 5. File information is displayed. This screen can be used to confirm measurement values for data items 1 through 101. Information is not displayed for csv files, prm files, and txt files. CAUTION Do not remove the USB memory device or SD card while confirming files. ...
MEASUREMENT PROCEDURE Measurement Data File Operations Measurement data files can be copied, deleted, renamed, or converted to CSV files. (1) Copying a file 1. Tap File on the Menu screen. 2. Tap the memory location containing the file to copy. 3.
Page 196
MEASUREMENT PROCEDURE 4. Tap the Option button. 5. Tap the memory location in which to save the copy. 6. Once copying is complete, tap Complete. CAUTION Wait until copying is finished before removing the USB memory device or SD card. ...
Page 197
MEASUREMENT PROCEDURE (2) Deleting a file 1. Tap File on the Menu screen. 2. Tap the memory location containing the file to delete. 3. Tap the check box of the file to delete. 4. Tap the Option button.
Page 198
MEASUREMENT PROCEDURE 5. Tap Delete File. 6. The file is deleted. CAUTION Wait until deleting is finished before removing the USB memory device or SD card. Wait until deleting is finished before turning the power of this device OFF.
Page 199
MEASUREMENT PROCEDURE (3) Renaming a file 1. Tap File on the Menu screen. 2. Tap the memory location containing the file to change. 3. Tap the check box of the file to change. 4. Tap the Option button.
Page 200
MEASUREMENT PROCEDURE 5. Tap Change File Name. 6. Enter the file name and then tap Enter. 7. Once changing is complete, tap Complete.
Page 201
MEASUREMENT PROCEDURE (4) Converting a file to CSV 1. Tap File on the Menu screen. 2. Tap the memory location containing the file to convert. 3. Tap the check box of the file to convert. 4. Tap the Option button.
Page 202
MEASUREMENT PROCEDURE 5. Tap Convert File to CSV. 6. Once conversion is complete, tap Complete. The CSV file will be created in the same location as the original file. CAUTION Wait until conversion is completed before removing the USB memory device or SD card.
Page 203
MEASUREMENT PROCEDURE (5) Sorting files 1. Tap File on the Menu screen. 2. Tap the memory location containing the file. 3. Tap Date. 4. Select the sort order and condition. Order: Select ascending/descending Date: Sort by file update date Name: Sort by file name Type: Sort by file extension...
Page 204
MEASUREMENT PROCEDURE 5. Tap the Back button to use the selected order and condition for sorting. If a file is copied from one memory location to another memory location, the file update date will change. For example, if a file is copied from internal memory to a USB memory device, sorting by update date in internal memory and sorting by update date in USB memory device may have different results.
Page 205
MEASUREMENT PROCEDURE (6) Displaying only certain extensions 1. Tap File on the Menu screen. 2. Tap the memory location containing the file. 3. Tap Filter. 4. Tap the extension to display.
Page 206
MEASUREMENT PROCEDURE 5. Tap the Back button to use the selected extension to display files.
FUNCTIONS 6. FUNCTIONS 6-1. STROKE CHANGE The stroke change function is used to measure displacement in excess of the rated capacity of a displacement meter. Displacement can be measured even after the rated capacity is exceeded, by performing stroke change each time displacement reaches the rated capacity. Stroke change is performed on a channel basis.
Page 208
FUNCTIONS 3. Select the channel to perform stroke change. 4. Tap Data before Moving. 5. Before moving the displacement meter, tap Measure Data. 6. Tap Complete.
Page 209
FUNCTIONS 7. Tap Data after Moving. 8. After moving the displacement meter, tap Measure Data. 9. Tap Complete. 10. Tap OK. Stroke change is completed. To cancel stroke change, tap Cancel.
FUNCTIONS 6-2.(OPTIONAL) UIO-80A External I/O Unit FUNCTIONS Installing an optional UIO-80A External I/O Unit allows this device to perform rainfall input, alarm output, external start/stop input, and reset input. Procedure for Adding UIO-80A Remove the two screws on the back, and then remove the dummy panel. Install UIO-80A, and then attach the two included screws.
FUNCTIONS Connector Pin Assignment The pin assignment of the external input/output connector on UIO-80A is shown below. Name Description Name Description Common potential for RESET RESET RESET input ISOCOM input Common potential for STOP STOP STOP input ISOCOM input Common potential for START START START input ISOCOM...
● Rainfall pulses will be counted even when the power of this device is OFF. ● If the rainfall count is reset while the power supply is OFF, the internal battery will need to be replaced. Contact KYOWA sales personnel or your nearest dealer.
Page 213
FUNCTIONS In order to use a rainfall meter, rainfall input must be set. 1. Tap Measure on the Menu screen. 2. Tap Rainfall. 3. Tap 0.5 mm or 1.0 mm. 4. Measurement value CH---- for the rainfall meter during measurement is displayed.
Page 214
FUNCTIONS The value and unit displayed for the rainfall meter will vary depending on the measurement value setting. The relationships between rainfall meter display values and measurement value settings are shown below. Calibration Rainfall Measurement Rainfall Meter Display Value and Coefficient Input Value Setting...
FUNCTIONS Alarms The alarm function outputs a signal from the set alarm channel once the measurement value exceeds the upper limit value or lower limit value. Upper limit value Lower limit value Important points when using alarms The circuit diagram for alarm output is shown below. Alarm output is photocoupler insulated open collector output.
Page 216
FUNCTIONS Alarm hysteresis Hysteresis can be set at the upper/lower limit level of the alarm. Hysteresis can be set from 000.00 to 999.99 for the upper/lower limit value. Upper limit alarm ON Upper limit value (Hysteresis) Upper limit alarm OFF Lower limit alarm OFF (Hysteresis) Lower limit value...
Page 217
FUNCTIONS If using an External I/O Unit alarm, configure the alarm. 1. Tap Measure on the Menu screen. 2. Tap Alarm. 3. Tap the alarm to configure. 4. Tap ON.
Page 218
FUNCTIONS 5. Configure the alarm channel, upper limit, lower limit, and hysteresis.
FUNCTIONS External Start/Stop Operations While the Menu screen is shown on the screen of this device, real-time measurement can be started by turning the START signal ON (shorted). Measurement can be stopped by turning the STOP signal ON (shorted) during measurement. BUSY output can be checked to confirm whether this device is currently measuring.
FUNCTIONS Reset Operation Turning the RESET signal ON (shorted) and then OFF (open) will restart this device. The circuit diagram for reset input is shown below. RESET (No-voltage contact input) (Internal circuit) ● This device will not start while RESET is ON (shorted). Confirm that the RESET signal turns OFF (open) when the power supply is turned ON.
When using an insulation resistance tester, be sure to remove the gage from the terminal block of this device. Contact the sensor manufacturer for information on how to accurately measure the insulation resistance of the sensor. Object Insulation is reduced. UCAM-80A Insulation is reduced.
Page 222
FUNCTIONS (4) Mode detection Detects the bridge configuration for strain measurement and displays the applicable mode. (5) Terminal temperature check Displays the measurement value of the reference junction temperature used for temperature measurement by a thermocouple. When measuring temperature, you will need to measure the temperature of the input terminal block serving as the reference junction, in order to obtain the measurement value (thermocouple output from 0°C).
Page 223
FUNCTIONS The following table summarizes check function limitations. ○ indicates checking is possible. ▲ indicates checking is possible only for certain channel modes. × indicates checking is not possible. Scanning Unit External Scanner General Civil Civil General Use Engineering Engineering Check USM-81A USB-80A-10...
Page 224
FUNCTIONS 1. Tap Channel on the Menu screen. 2. Tap the radio button of the channel to check. 3. Tap the Option button. 4. Tap Check.
Page 225
FUNCTIONS 5. Tap the check function to run. Tapping an i button will display a description of that function.
FUNCTIONS Disconnection/Burnout Check 1. Tap Disconnection. 2. Once the check has completed, the check result will be displayed. Check Meaning Result Normal Disconnection or burnout detected Channel mode that cannot be checked...
FUNCTIONS Input/Output Resistance Check 1. Tap Resistance. 2. Once the check has completed, the input/output resistance values are displayed. Input indicates an input resistance value. Output indicates an output resistance value. - indicates a channel mode that cannot be checked.
FUNCTIONS Insulation Check 1. Tap Insulation. 2. Once the check has completed, the check result will be displayed. Check Result Meaning High Insulation resistance value is higher than 100.0 MΩ Insulation resistance value is lower than 1.0 MΩ Channel mode that cannot be checked Other than above Insulation resistance value is displayed in MΩ...
FUNCTIONS Mode Detection 1. Tap Mode Detection. 2. Once the check has completed, the detection result is displayed. The upper row displays the mode detection result. The lower row displays the currently set channel mode in parentheses. Detection Result Meaning 1-gage 120 Ω...
FUNCTIONS Terminal Temperature Check 1. Tap Terminal Temperature. 2. Once the check has completed, the terminal temperature is displayed. The terminal temperature is checked for each scanning unit. Therefore, the check result channel number will be the first channel of the scanning unit.
FUNCTIONS 6-4. AUTOMATIC MODE SETTING The automatic mode setting is a function that detects the channel mode for the connected sensor, and sets the mode automatically for each channel. CAUTION The automatic mode setting cannot be used for external scanner channels. 1.
Page 234
FUNCTIONS 4. Tap Auto Mode Setting. 5. Tap Execute. 6. Once setting is complete, the setting results are displayed. Setting Result Meaning Unable to detect channel mode. The channel mode will not be changed for any item with "-" indicated as the setting result. Manual configuration is required.
Page 235
FUNCTIONS The following channel modes can be set using automatic mode setting. 1-gage connection: 1-gage 120 Ω, 1-gage 240 Ω, 1-gage 350 Ω 2-gage connection: 2-gage A-A 4-gage connection: 4-gage Civil engineering transducer with temperature measuring function: 4-gage T However, if the same type of channel mode has already been set, the channel mode prior to automatic mode setting will be retained.
Page 236
FUNCTIONS Channel Mode Channel Mode Set Channel Mode Prior to Detected by This Automatic Mode Setting Device 4-gage 4-gage 4-gage 4-gage A-A 4-gage 4-gage A-A 4-gage HR 4-gage 4-gage HR 4-gage I (120 Ω) 4-gage 4-gage I (120 Ω) 4-gage I (350 Ω) 4-gage 4-gage I (350 Ω) 4-gage I (350 Ω)_HR...
FUNCTIONS 6-5. SETTING CHANNEL CONDITIONS FROM TEDS INFORMATION The channel mode, calibration coefficient, decimal point position, and unit can be set automatically from information read from a Transducer Electronic Data Sheet (TEDS) sensor. CAUTION It is not possible to read all information from a TEDS sensor. ...
Page 238
FUNCTIONS 4. Tap Set Channel Conditions from TEDS. 5. Tap Execute. 6. Once setting is complete, the setting results are displayed. Result Meaning Information was successfully read from the TEDS sensor. Information could not be read from the TEDS sensor. Manual configuration is required.
FUNCTIONS 6-6. CONDITION FILES Condition File Saving The settings of this device can be saved as a condition file. 1. Tap File on the Menu screen. 2. Tap the memory location in which to save the condition file. 3. Tap the Option button.
Page 240
FUNCTIONS 4. Tap Save File as Condition File. 5. Enter the name of the condition file, and then tap Enter. 6. Once saving is complete, tap Complete.
FUNCTIONS Condition File Loading This device can be configured by loading an existing condition file. 1. Tap File on the Menu screen. 2. Tap the memory location containing the condition file. 3. Tap the check box of the condition file to load. 4.
Page 242
5. Tap Read Condition File. 6. Once loading is complete, tap Complete. CAUTION UCAM-80A cannot load a condition file saved on UCAM-60C. Loading a condition file will not change the following settings. • Date and time • Language • IP address, subnet mask, default gateway...
FUNCTIONS 6-7. DATE AND TIME SETTINGS Set the date and time of this device. 1. Tap Setting on the Menu screen. 2. Tap Time. 3. Enter the date and time, and then tap Enter. 4. The date and time are set.
Page 244
FUNCTIONS CAUTION The date and time of this device are displayed in Japan Standard Time. The date and time can be set up to December 31, 2037 23:59:59. Do not use this device after January 19, 2038.
FUNCTIONS 6-8. VOLUME SETTINGS Set the volume. 1. Tap Setting on the Menu screen. 2. Tap Volume. 3. Tapping OFF for Sound will mute this device. Tapping ON for Sound will enable sound from this device. 4. Tapping + will increase the volume. Tapping - will decrease the volume.
FUNCTIONS 6-9. LANGUAGE SWITCHING The language can be switched between Japanese and English. 1. Tap Setting on the Menu screen. 2. Tap Language. 3. Tap either Japanese or English. 4. Tap Restart.
Page 247
FUNCTIONS 5. After this device restarts, the language will be switched. Tapping Restart will restart this device. The language will be switched once it has restarted.
FUNCTIONS 6-10. INTERFACE SETTINGS Configure settings related to the communication interface between this device and the 1. Tap Setting on the Menu screen. 2. Tap Interface. 3. The Interface screen is displayed.
Page 249
FUNCTIONS The Interface screen is used to configure the communication interface. The following functions are available on the Interface screen. Type: RS-232C or USB Type: Ethernet Reference Number Name Function Page 6-10-1 Format Output format settings Type Communication interface selection 6-10-2 Baud Rate Communication speed settings...
FUNCTIONS Output Format Set the output format for measurement values sent from this device to the PC, when sending/receiving control commands. Refer to the Control Command Instruction Manual for details on control commands. 1. Tap Format. 2. Tap Header. 3. Tap Header ON or Header OFF.
Page 251
FUNCTIONS 4. Tap the Back button. 5. Tap Data. 6. Tap All Data or Channel & Data. Data Description All Data Output data for all items. Channel & Data Output only channel numbers and measurement values.
FUNCTIONS Baud Rate Set the communication speed when using RS-232C. 1. Tap Baud Rate. 2. Tap the communication speed to use. The baud rate cannot be set for Ethernet or USB.
FUNCTIONS Termination Set the termination code for when using RS-232C. 1. Tap Termination. 2. Tap the termination code to use. Refer to the Control Command Instruction Manual for details on termination codes. Termination cannot be set for Ethernet or USB.
FUNCTIONS Timeout Set the timeout period (the time that this device will wait to receive a control command from the PC) for when using RS-232C or USB. 1. Tap Timeout. 2. Set the timeout period to either 5 seconds or 50 seconds.
FUNCTIONS IP Address, Subnet Mask, Default Gateway Set the IP address, subnet mask, or default gateway. 1. Tap the item to set. 2. Enter the value to set, and then tap Enter.
FUNCTIONS 6-11. MAINTENANCE SETTING The maintenance setting is used for performing maintenance. 1. Tap Setting on the Menu screen. 2. Tap Maintenance. 3. Tap the maintenance setting to use. Maintenance Description ACOM-GND Connect ACOM and GND inside this device. Connection This should normally be set to Connect ACOM-GND.
FUNCTIONS 6-12. SCREEN BRIGHTNESS SETTINGS Set the brightness of the screen. 1. Tap Setting on the Menu screen. 2. Tap Brightness. 3. Tapping + will increase the brightness. Tapping - will decrease the brightness. The brightness can be set to any of several levels (20/40/60/80/100%).
FUNCTIONS 6-13. CONFIRMATION SCREEN ON/OFF Select whether to display confirmation screens when deleting files, etc. 1. Tap Setting on the Menu screen. 2. Tap Check Screen. 3. Tap either Display ON or Display OFF. Check Description Screen Display ON Display confirmation screens when deleting files, etc. Display OFF Do not display confirmation screens when deleting files, etc.
FUNCTIONS 6-14. PRINTING SETTINGS Set what to print to recording paper. 1. Tap Setting on the Menu screen. 2. Tap Printer. 3. Tap the item to set. 4. Select ON or OFF.
Page 261
FUNCTIONS Item Description Print with space between the "End of printing" and "Next header." Feed Print without any space between the "End of printing" and "Next header." Print the header when printing measurement values, etc. Header Do not print the header when printing measurement values, etc. Print the title when printing measurement values, etc.
Page 262
FUNCTIONS The status number indicates the status of the measurement value. If the status number is any number other than 00, the measurement value will be 999999. Number Status Cause & Solution Normal data The measurement value has been exceeded. →...
Page 263
The A/D converter is faulty. A/D converter fault → Remeasure and check whether this error occurs again. → If this occurs frequently, request service from KYOWA. The channel mode of this device does not match that of Device and USB- USB-80/70 Series.
FUNCTIONS 6-15. CHECK MENU Check various settings, etc. 1. Tap Setting on the Menu screen. 2. Tap Check Menu. 3. The Check Menu screen is displayed.
Page 265
FUNCTIONS The following functions are available on the Check menu screen. Reference Number Name Page 6-15-1 Check Dead Pixel Check Printer 6-15-2 Display Scanning Slot 6-15-3 Display Frequency of Power 6-15-4 Version 6-15-5...
FUNCTIONS Check for Dead Pixels on Screen Display red/green/blue screens to check for dead pixels. 1. Tap Check Dead Pixel. 2. Tap OK. 3. A red screen is displayed. 4. A green screen is displayed.
Page 267
FUNCTIONS 5. A blue screen is displayed. 6. After checking, Tap Complete.
FUNCTIONS Check Print Quality Print certain characters on recording paper to check printer soundness. 1. Tap Check Printer. 2. Tap OK. 3. Printing starts. 4. After checking, Tap Complete.
FUNCTIONS Display Scanning Unit Slots Display any scanning units connected to this device. 1. Tap Display Scanning Slot. 2. Any scanning units are displayed. Optional slot indicates whether an External I/O Unit is present.
FUNCTIONS Display Power Frequency Detection Result Display the power frequency detected when this device started. 1. Tap Display Frequency of Power. 2. The power frequency detection result is displayed. Result Meaning 50 Hz Indicates that 50 Hz was detected. 60 Hz Indicates that 60 Hz was detected.
FUNCTIONS Display Version Information Display version information, etc. for this device. 1. Tap Version. 2. The version information is displayed. Version Description Main Firmware Displays the main firmware version of this device. Sub Firmware Displays the sub firmware version of this device. Hardware Displays the FPGA version of this device.
FUNCTIONS 6-16. COLOR MODE SWITCHING Switch between a dark or light color mode. 1. Tap Setting on the Menu screen. 2. Tap Color Mode. 3. Tap the color mode to set. Color Mode Description Dark Screen colors will have a darker tone. Light Screen colors will have a lighter tone.
FUNCTIONS 6-17. SCREEN CONTENT PRINTING Screens that include an Option button can be printed. 1. Tap the Option button. 2. Tap Print Contents of Current Page to print the page displayed on the previous screen. Tap Print Contents of All Pages. to print all pages from the previous screen.
FUNCTIONS 6-18. SCREEN CONTENT OUTPUT TO EXTERNAL MEMORY DEVICE Screens that include an Option button can be output to an external memory device. Screens will be output as text files. 1. Tap the Option button. 2. Output Contents to USB Memory Tap this to output a text file to the USB memory device connected to this device.
Page 275
FUNCTIONS 3. Once output is complete, tap Complete. 4. The display returns to the Option screen. Text files will be output to the "ucam" in the root directory (highest level directory) of the external memory device. If there is no "ucam" folder, it will be automatically created by this device.
FUNCTIONS 6-19. MEASUREMENT DATA FILE AUTOMATIC COLLECTION USING USB MEMORY DEVICE A USB memory device (containing a certain folder) can be connected to this device, in order to automatically collect measurement data files stored in internal memory. The automatic collection procedure is shown below. CAUTION ...
Page 277
FUNCTIONS (4) The following screen is displayed. Tap [Copy] to start automatic collection. Tap [Cancel] to return to the Menu screen. (5) The following screen is displayed during automatic collection. Tapping [Cancel] will cancel automatic collection. (6) Once automatic collection is complete, the display will return to the Menu screen. (7) Remove the USB memory device.
In order to ensure the operation of this device, it is Device operation recommended to perform periodic inspections. Once per year confirmation Devices will be collected by KYOWA for periodic inspection. Limited Lifespan Part Replacement Item Details Schedule The internal backup battery is used to operate the clock Backup battery function when the power supply is OFF.
MAINTENANCE 7-3. RECORDING PAPER REPLACEMENT Follow the procedure below to replace recording paper in the printer built into this device. (1) Open the printer lid. (2) Pull the printer open lever to open the printer cover. (3) Set the recording paper in place as shown here. (4) Stick the end of the recording paper out from the printer output.
(3) Use a tester or other device to check whether the fuse inserted into the FUSE holder is blown. If it is blown, replace it with a spare fuse. (4) Attach the FUSE holder to the rear. UCAM-80A recommended fuse Model: 217 004 MXP (Littelfuse) Fusing characteristic: Fast-blow type Rated voltage: 250 V, Rated current: 4 A Dimensions: φ5 ×...
4 years if a UIO-80A External I/O Unit is installed. ● Devices will be collected by KYOWA for battery replacement. ● The internal backup battery is not used when the power of this device is ON.
ERROR MESSAGES 8. ERROR MESSAGES A message will be displayed on the screen when an error occurs. These messages are described below. If an error message is displayed, follow the instructions provided. Code Message E001 A value outside the settable range was entered. There is a blank entry field.
Page 283
ERROR MESSAGES Unable to convert to CSV. E017 File data is abnormal. An overheat error has occurred in the printer. E018 Wait a short while and then try printing again.
This table explains what to check and provides possible solutions for issues. If operation continues to be problematic, contact KYOWA sales personnel or your nearest dealer. Note that a repair request may be denied if the device was damaged through use not described in this Instruction Manual or if the device was disassembled or modified by the user.
Page 285
BV voltage is output. the sensor connected, the BV circuit may have failed. Contact KYOWA sales personnel or your nearest dealer. Refer to "6-3-2. Input/Output Resistance If input/output resistance differs Check." Perform an input/output resistance...
Page 286
Contact abnormal. roughly same ambient KYOWA sales personnel or your temperature (this device generates heat, nearest dealer. so a temperate a few °C higher than the ambient temperature will be displayed). Is the thermocouple disconnected? If the result is not OK, the Refer to "6-3-1.
Page 287
BEFORE CONCLUDING THAT THE PRODUCT IS DEFECTIVE Issue Check Solution Thermocouple Was this device sufficiently warmed up? Measurement values will drift if this measurement device is not sufficiently warmed values up. Warm up this device for around abnormal. 30 to 60 minutes. thermocouple attached If the thermocouple is moved by...
Page 288
Is the printer LED flashing? This will flash if the printer is hot. Wait a short while and then try printing again. continues to flash even after waiting a while, contact KYOWA sales personnel or your nearest dealer.
Is the USB cable connected firmly? Try reconnecting the USB cable. Is the interface type set to USB? Set the interface type to USB. Turning the power The built-in battery is no longer working. Contact KYOWA sales personnel causes your nearest dealer time to drift.
Strain gages, strain gage transducers, civil engineering transducers with a thermal sensor, DC voltage-output or DC current-output instruments, potentiometer sensors, thermal sensors(Thermocouples and platinum resistance thermometer bulbs) ・Connectable Scanning Units(mounted on top of the UCAM-80A) USM-81A,USM-82A,USM-83A ・Connectable Scanners USB-80/70 series...
Page 291
350 Ω Transducers with ○ ― ○ a thermal sensor *1. Can not use remote sensing sensor directly. *2. 120 to 1000 Ω in high-resolution mode. *3. Accessory terminal cover not attachable. When mounted on the UCAM-80A, calibration is required.(with charge)
Page 292
PRODUCT SPECIFICATIONS Scanners Scanning Units External Scanners General Civil USM-81A purpose engineering USM-82A USB-80A-10 USM-83A USB-80A-20 USB-80A-30 Measuring Targets USS-61B USB-70A/B-10 USB-70A/B-30 USS-62B USB-70A/B-20 USS-63B Voltage DC voltage-output instruments ○ ○ ○ Current DC current-output instruments ○ ○ ○ K(CA) ○...
Page 293
PRODUCT SPECIFICATIONS (5) Scanning speed Line Frequencies 50 Hz Zone 60 Hz Zone Scanners Scanning Units (standard mode) 50 ms/ channel Scanning Units (High-resolution mode) 280 ms/ channel Scanning Units (High-speed mode) 20 ms/channel 16.7 ms/channel USB-80/70 series 60 ms/channel 58.4 ms/channel (standard mode only) Notes:...
Page 294
PRODUCT SPECIFICATIONS (8) Coefficient Calculation Function Multiplication by calibration coefficient, calibration by TEDS, conversion of measured values to physical quantities, scaling and correction (9) Unit 59 units (10) Automatic Measurement Function ・Trigger Measurement A relative value (certain changing quantity) or an absolute value triggers measurement. ・Interval Measurement Measurement is automatically performed at preset time intervals.
Page 295
PRODUCT SPECIFICATIONS (14) Strain Measurement(High-Resolution Mode) ・Constant voltage excitation Approx. 5 VDC ・Constant current excitation Approx. 14.3 mA(Bridge resistance 350 Ω) ・Scanning speed 0.28 s/channel ・Gage factor 2.00 fixed (Coefficient calculation function enables correction with 2.00/Ks) ・Initial value memory range Same as measuring range ・Measuring range,Resolution and Accuracy Measuring Range...
Page 296
PRODUCT SPECIFICATIONS (16) Voltage Measurement(Standard Mode) ・Scanning speed 50 ms/channel ・Initial value memory range Same as measuring range ・Measuring range,Resolution and Accuracy Range Resolut Input Measuring Range Accuracy Mode Resistance In case of Scanning Units ±(0.05 % of reading+0.003)mV 0 to ±50.000 mV 1 µV In case of USB-80/70 ±(0.07 % of reading+0.004)mV...
Page 297
PRODUCT SPECIFICATIONS (18) Current Measurement(Standard Mode) ・Scanning speed 50 ms/channel ・Initial value memory range Same as measuring range ・Measuring range,Resolution and Accuracy Channel Mode Measuring Range Resolution Accuracy In case of Scanning Units ±(0.05 % of reading+0.01) mA I/50 mA 0 to ±50.00 mA 10 µA In case of USB-80/70...
Page 298
PRODUCT SPECIFICATIONS (20) Temperature Measurement with Thermocouple (Standard Mode) ・Scanning speed 50 ms/channel ・Measuring range,Resolution and Accuracy Accuracy Accuracy Resoluti Type Measuring range External reference junction Internal reference junction compensator compensator In case of Scanning Units ●Ambient temperature 15 to 35 °C ・Above -50°C In case of Scanning Units -200.0 to 1230.0 °C...
Page 299
PRODUCT SPECIFICATIONS (21) Temperature Measurement with Civil Engineering Transducers with a thermal sensor (Standard Mode) ・Scanning speed 50 ms/channel ・Measuring range,Resolution and Accuracy Measuring range Resolution Accuracy In case of Scanning Units Within±0.5 °C -50.0 to 200.0 °C 0.1 ℃ In case of USB-80/70 Within±0.7 °C Notes: 1.
Page 300
PRODUCT SPECIFICATIONS (24) Measurement with Potentiometer sensor(High-speed Mode) ・Scanning speed 20 ms/channel(50 Hz Zone), 16.7 ms/channel(60 Hz one) ・Initial value memory range Same as measuring range ・Sensor power supply Approx. 2 VDC ・Potentiometer resistance 1 kΩ to 10 kΩ ・Measuring range,Resolution and Accuracy Channel Mode Measuring range Resolution...
Page 301
USB2.0(collect measurement data by USB Flash Drive) Notes: 1 Measurement data cannot be saved directly to USB memory. Notes: 2 USB Flash Drives Capacity:32GB or less FileFormat:FAT32 Notes: 3 The following USB memory device has been tested by KYOWA. Manufacturer : Green house...
Page 302
PRODUCT SPECIFICATIONS (36) Setting Maintenance Function ACOM at measurement circuit is switchable between floating and GND connect. (37) Power Supply AC 100 to 240 V 50/60 Hz (38) Current Consumption 0.5 A or less: 100 VAC (With 3 Scanning Units mounted) (39) Degrees of Protection IP20(JIS C 0920/IEC 60529)...
TEDS compatible) USM-83A(For civil engineering measurement, TEDS compatible, with lightning arresters) (2) Type Scanning Unit (3) Number of Mountable Units 3 units can be mounted on UCAM-80A (4) Switching Terminal Semiconductor Relays (5) Channels 10 / Unit (6) Input Terminals...
Page 304
PRODUCT SPECIFICATIONS Standard Accessory “READ BEFORE USING THE PRODUCT” “Terms and Conditions of Product Warranty” NDIS4102(7 pins) connectors caps(USM-82A only) Terminal cover Channel label Short bar ML-1000-3H1 Optional Accessories One-touch terminal block (JT-1A)
Platinum resistance thermometer bulbs Pt100, JPt100 (USB-80A-30) Potentiometer sensors (4) Input Channels 50/unit (5) Connection data loggers UCAM-80A, UCAM-60C M14, UCAM-65C M14 When connecting UCAM-60C or UCAM-65C, an optional scanner interface (USI-67A) is required. (6) Number of Connection Up to 20...
Page 307
PRODUCT SPECIFICATIONS (11) Degrees of Protection IP30(JIS C 0920/IEC 60529) (12) Dimensions 302(W) × 107(H) × 500(D) mm (Excluding protrusions) USB-80A-10: Approx. 7.0 kg (Excluding UPS-80A) (13) Weight USB-80A-20: Approx. 8.0 kg (Excluding UPS-80A) USB-80A-30: Approx. 7.3 kg (Excluding UPS-80A) (14) Compliance Directive 2014/30/EU (EMC) Directive 2014/35/EU (LVD)
TECHNICAL INFORMATION NECESSARY FOR MEASUREMENT 12. TECHNICAL INFORMATION NECESSARY FOR MEASUREMENT 12-1. STRAIN GAGE NONLINEARITY The ideal bridge output for strain measurement is: ε ⋅ ⋅ Output voltage Gage factor ε: Strain value Voltage applied to bridge If one side of a wheatstone bridge is used as the strain gage and only that side changes, there will be nonlinearity generated in the relationship between the strain value and output voltage.
Page 314
TECHNICAL INFORMATION NECESSARY FOR MEASUREMENT The following table lists compensation equations for each channel mode during strain measurement Measurement Channel Mode Compensation External Scanning Method Equation Scanner Unit 1-gage method 1-gage 120 Without With ε (3-wire system) 1-gage 240 compensation compensation −...
TECHNICAL INFORMATION NECESSARY FOR MEASUREMENT 12-2. DUMMY GAGE USAGE The true dummy method is used by KYOWA UCAM Series Data Loggers. This method uses software to effectively eliminate apparent strain caused by the linear expansion coefficient of the measured object (object being tested) when the ambient temperature changes.
Page 316
TECHNICAL INFORMATION NECESSARY FOR MEASUREMENT (2) Common dummy method In this method, the active gages are switched in sequence and bridge excitation voltage is applied only when measurement is performed. On the other hand, bridge excitation voltage is applied to the dummy gage every time an active gage is measured. The impact of self-generated heat therefore will be different between the active and dummy gages, making it difficult to provide complete temperature compensation.
Page 317
TECHNICAL INFORMATION NECESSARY FOR MEASUREMENT * Calculations when setting true dummy For the active gage, compensation is performed using the dummy gage nearest that active gage (that is, with the smaller measurement channel number). Example: 1-gage 120 Ω_D set to CH0 and CH3, 1-gage 120 Ω_A set to CH1/CH2/CH4, and then measurement performed Dummy Active...
TECHNICAL INFORMATION NECESSARY FOR MEASUREMENT 12-3. LEAD WIRE TEMPERATURE EFFECT COMPENSATION METHOD (3-WIRE SYSTEM CONNECTION METHOD) When using a SELCOM gage, if the lead wire from the gage to the bridge circuit is long, the change in the resistance of the lead wire due to temperature will show as apparent strain.
TECHNICAL INFORMATION NECESSARY FOR MEASUREMENT If the same calculation is performed using a parallel vinyl wire with cross section of 0.08 (resistance value 0.239 Ω/m, resistance temperature coefficient 3.93 × 10 /°C), the results are as follows. 10 m: 145 × 10 strain/°C 20 m: 290 ×...
TECHNICAL INFORMATION NECESSARY FOR MEASUREMENT 12-5. COMPENSATION WHEN LEAD WIRE OR CABLE IS LONG If a long lead wire is used for 1-gage method or 2-gage method connections, a bridge circuit will be created where the resistance of the unnecessary lead wire will be introduced Gage resistance value + Lead wire resistance value in series with the strain gage, reducing the apparent gage factor.
Nominal bridge resistance value Input bridge resistance value True strain = Measured strain value × The nominal bridge resistances are as follows: UCAM-80A: 4-gage I (120 Ω) :120 Ω 4-gage I (350 Ω) :350 Ω 4-gage I (350 Ω)_HR :350 Ω...
APPENDIX 13. APPENDIX 13-1. INTERNAL MEMORY DATA AMOUNT The amount of data for a single scan can be calculated using the following simplified formula. Data amount (bytes) = 29 (bytes) + [No. of channels] × 26 (bytes) 26 bytes used per channel Number of channels with measurement ON Header information: 29 bytes used each time Amount of data (bytes) used for a single scan...
External I/O Unit while updating. Failure to follow the above warnings may cause this device to fail. If this device fails, repairing it will require it to be collected by KYOWA for paid service. (1) Prepare the following.
Page 324
APPENDIX (5) Tap the Yes button on the screen to start the update. To cancel the update, tap the No button on the screen, and then remove the USB memory device. This device will restart. It will take approximately 7 minutes to update. (6) Once updating has completed, the following message will be displayed on the screen.
Page 325
APPENDIX Kyowa Electronic Instruments Co., Ltd. 3-5-1, Chofugaoka, Chofu, Tokyo 182-8520 Japan http://www.kyowa-ei.com/...
Need help?
Do you have a question about the UCAM-80A and is the answer not in the manual?
Questions and answers