For AC power supply AC 100 ~ 240 V UCAM-60C-DC For DC power supply DC 10 ~ 16 V STANDARD ACCESSORIES The following accessories are enclosed with the UCAM-60C. Check that there are no missing accessories after opening the package. • Instruction Manual (CD-R) 1 set •...
BEFORE USE Be sure to comply with the following safety precautions during use. Kyowa Electronic Instruments Co., Ltd. shall not be liable for any damage suffered as a result of using the product not in compliance with these safety instructions.
Page 5
WARNING ●Fuse To prevent fire, use only fuses with the specified ratings (current, voltage, type) in this device. Do not use fuses that are not specified, and do not short-circuit the fuse holder. Pull out the power supply cable first before replacing the fuse. ●External connections To prevent electric shock, ground this device properly first before connecting any external devices such as PCs and the measured object.
Page 6
• T o prevent malfunction and accidents due to erroneous handling and installation, and to allow the UCAM-60C for a safe and well performance, understand the handling method and condition by reading the following instructions. CAUTION Do not turn off the power supply when reading/writing a data file and do not take out the USB memory.
1. Controls and Indicators 1.1 Front Printer Display USB Port Keyboard 1.2 Detailed Description of Keyboard START/STOP key: Cursor keys ( ∧ , ∨ , < , > ): Press this key to start or stop Moves the cursor up, down, left, right. FUNC....
Page 8
1.3 Back “GND” terminal: Housing ground terminal “POWER” switch: Power supply switch “LINE IN” Power supply input socket “Fuse” holder “ETHERNET” connector Option slot 1: Option slot 2: “RS-232C” connector plug: plug: Ethernet interface Scanner interface External input/output unit RS-232C interface connector USI-67A is inserted here.
Page 9
Ground the UCAM-60C to the earth, and then connect the “GND” terminal of the UCAM-60C to the measured object. *When using an external scanner, ground the external scanner to the earth, and then connect the external scanner to the measured object.
Page 10
Be careful not to mix up the polarity then. MEMO For the UCAM-60C-DC, the (-) terminal of the DC power supply is grounded to the housing. Therefore, when a measurement device whose (+) terminal is grounded to the housing is connected, or when the (+) terminal of the power supply is connected to or in contact with the housing, this will result in a failure or a drop in the performance.
Page 11
Line F and G need to be separated. Shield Connection during voltage input Note Do not connect the FV-1A and FV-2A sold by KYOWA to the UCAM-60. 1-gage 3-wire-system strain gage’s single line has a red stripe or other marking.
Page 12
4. about the main screen The main screen to be displayed in a ready state can be switched with the cursor keys < , >. The following information appears in each main screen, indicating the current setting status of the UCAM-60C, respectively.
Page 13
5. Tree of UCAM-60C Setting Menu The items to configure by pressing the key on the main unit form a tree structure as shown below. Press the ESC key to return to the previous menu. MEAS CH MEAS Condition Environment...
Page 14
6. Simplified Procedures for Operating UCAM-60C An explanation on how to connect the strain gauge and strain gauge-type transducer, set the measurement conditions and condition of each channel, and perform measurement is given here. Connect the sensor, and Turn the POWER switch at the back to turn ON the power supply [ON].
Page 15
On the “0. MEAS Condition” menu,” select: Repeat Times Setting the repeat count Times 0:MEAS Condition (01~99=Times) (00=Infinite) 1:Repeat Times ・A number from 1 - 99 can be set for the (1) Enter the repeat count with the numeric repeat count. keys (0 - 9).
Page 16
Press the SET key to CAL Coef display the main menu. Setting calibration coefficient 1.2345E+2 *Coefficient to convert the From the main menu, obtain the target 1.0000E+0 measurement value quantitatively menu in the following flow: 1.0000E+0 1:CH Condition This does not need to be set Move the cursor to the input area with the when the coefficient is not used cursor keys.
Page 17
Setting the file name SAVE of the measurement Specifying the file name → data File Name Press the SET key and then, press the test001 This does not need to be set SAVE key. (abcdefghij) when the data is only printed 0123456789 and not saved.
Page 18
Press the FUNC. key and FUNC. 0:MEAS Value Measuring the initial values select “2: Initial Value” on 1:Original Value the display. 2:Initial Value 3:EASY MEAS Press the START START/STOP key to /STOP start the initial measurement. Initial Value When the initial value measurement is Start:START/STOP taken with the “SAVE”...
Page 19
7.Automatically Collecting Measurement Data by Inserting the USB memory In the UCAM-60C, it is possible to collect the data in the built-in memory automatically by inserting a USB memory containing the specified folder. (1) Preparation in a Windows PC Create the folder「_ucam_backup_sddata」in theUSB memory (2) Insert the USB memory into UCAM-60C Insert the USB memory prepared earlier into the USB port of the UCAM-60C.
If troubles are not solved despite countermeasures, contact Kyowa or our representatives. Repairs may be rejected if the product is damaged as a result of not following the instructions given in this instruction manual, or if the product is disassembled or modified by the customer.
Page 21
⇒If the bridge excitation voltage is incorrect, it means that the bridge excitation circuit is faulty. Contact Kyowa or our representatives. ・See INSTRUCTION MANUAL’s "9-3 Input & Output Resistances" to conduct an input &...
Page 22
However, by enabling the deltaV or deltaT function, it becomes possible to measure with the measure value where coefficient calculation was performed. ・Is a FV-1A and FV-2A sold by KYOWA used? Can not use the FV-1A and FV-2A to the UCAM-60. measured value ・See INSTRUCTION MANUAL’s "9-5 Terminal Temperature"...
Page 24
Kyowa Electronic Instruments Co., Ltd. 3-5-1, Chofugaoka, Chofu, Tokyo 182-8520 Japan http://www.kyowa-ei.com/...
Need help?
Do you have a question about the UCAM-60C and is the answer not in the manual?
Questions and answers