ENG INSTRUCTIONS FOR USE 1 Symbols 1.1 Description of symbols used Description of symbols used S S ymbol ymbol Description Description S S ymbol ymbol Description Description Manufacturer. Catalogue number. Consult instructions for use or consult CE Marking with number of the electronic notified body.
Bien-Air Dental DMX3 drive motor. c c onditions onditions As defined in the Bien-Air Dental brushless Description: micromotor IFUs. The DMX3 electronic board* is dedicated to drive 2.7 P P atient c atient contr ontra-indic a-indications ations up to three Bien-Air Dental brushless and sensorless micromotors.
EMC compliance to IEC 60601-1-2 can only be claimed when the device is used with observed: Bien-Air Dental compatible motors and hoses. WARNING • During integration, only use a medical power • Ensure to follow the installation procedure...
4 Electromagnetic Compatibility (EMC) 4.1 EMC Caution EMC Caution CAUTION • Since compliance with the international standard IEC 60601-1-2 does not guarantee immunity against 5G worldwide (due to the different frequency bands used locally), avoid the presence of devices equipped with 5G broadband cellular networks in the clinical environment or ensure that the network functionality of these devices is disabled during the clinical procedure.
4.1.1 Electromagnetic compatibility – emissions & immunity The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment. Guidance and manufacturer’s declaration – Electromagnetic emissions: Emissions test Emissions test...
Guidance and manufacturer’s declaration – Electromagnetic immunity: IEC 60601 test IEC 60601 test Electr Electromagnetic en omagnetic envir vironment onment Immunity test Immunity test Compliance lev Compliance level el level el - guidance - guidance Electrostatic ±8kV contact ±8kV contact Floors should be wood, concrete or ceramic discharge (ESD) tile.
Page 10
IEC 60601 test IEC 60601 test Electr Electromagnetic en omagnetic envir vironment - onment - Immunity test Immunity test Compliance lev Compliance level el level el guidance guidance Conducted 3 VRMS 3 VRMS Field strengths from fixed RF transmitters, disturbances 0,15 MHz –...
5 Electrostatic precautions The device contains ESD sensitive elements, appropriate handling precautions must be observed. CAUTION The device uses semiconductors that can be damaged by electrostatic discharge (ESD). When handling, care must be taken so that the device is not damaged. Damage due to inappropriate handling is not covered by the warranty.
They may not give rise to any claim. Note : Note : The original language of those instructions for use is English. Note : Note : For any further information, please contact Bien-Air Dental SA at the address given on the back cover.
6.2 Main functions Main functions • Controls up to three motors. • Pneumatic control. • Electric controlled by analog inputs or digital interface (RS-232). The system's variable parameters are as follows: • Speed range: 100 to 40'000 rpm. • Progressive or ON/OFF mode speed adjustment. •...
2 light connections. The relays are switched simultaneously and controlled by the MUX Control input (24 Vdc). The DMX3 can drive up to three motors with 2 dual motor switch. To connect the dual motor switch, please consult the wiring diagram.
6.4 T T echnic echnical data al data T T echnic echnical data al data Dimensions 102 x 58 x 27 mm Weight approx. 58 g Voltage 32 Vdc ±10% (min. 28.8 Vdc, max. 35.2 Vdc) Nominal power 60 W (MCX and MX2) 120 W (MX-i) 6.5 Classific Classification ation...
Bien-Air Dental; • Must ensure that the device is used only for the purpose for which it is intended, must protect themselves, their patients and third parties from any danger, and must avoid contamination through the use of the product.
For more information or if you have any questions about the integration, wiring configuration or programming of the electronic board system, please contact your Bien-Air Dental representative (addresses below). • Only use accessories and cables specified by Bien-Air Dental SA. FIG. Connectors specifications...
7.2 Wiring diagr Wiring diagrams The wiring diagrams below show a standard configuration of the system in the four operating modes. The connections depend on the integration in the unit and the desired functions. 7.2.1 Mode 0 (Electrical mode from 100 rpm to 40'000 rpm) The motor's target speed is attained using a potentiometer (10 kΩ) or an electric pedal.
7.2.2 Mode 1: Pneumatic mode from 100 rpm to 40'000 rpm FIG. The motor's target speed is attained using an air-driven pedal connected to the pressure sensor (0 to 3 bar). A 10 kΩ potentiometer is required in order to vary the luminous intensity. FIG.
7.2.3 Mode 2: Pneumatic mode with electric limitation FIG. Pneumatic mode (mode 1) with maximum speed limitation. Identical to mode 1 but with the maximum speed limited by a potentiometer (10 kΩ). A 10 kΩ potentiometer is required in order to vary the luminous intensity.
7.2.4 Mode 3: Serial mode (RS-232) FIG. The DMX3 is controlled using an RS-232 communication protocol. The RS-232 communication protocol is available by request from Bien-Air Dental SA. FIG. 6 . 6...
(see table below). The system installation depends on the chosen operating mode. For more information and technical support, please contact your Bien-Air Dental dealer. Dip S...
7.4 F F aul ault list and instal t list and installed pr led protection otection FIG. 10 . 10 The electronic board includes 3 diagnostic LEDs (see FIG. 1 point 7=Diagnostic LEDs). Power ON: The green LED lights up when the board is powered on Diagnostic: The red LED flashes (1-7 times) when a fault occurs (see fault list) RS232 Communication:...
Page 26
Embedded Protection: Temperature: The electronic board temperature is continuously controlled by the software. Power supply: The electronic control system is protected against over-and undervoltage. Motor and light: The motor output (phases) is protected against short circuits. The light output is protected against short circuits. Interruption of one, two or three phases is detected by the system, and the motor either does not start or stops.
No maintenance can be performed on the device. 8.2 Servicing Servicing Never dismantle the device. For all servicing and repairs, it is recommended that you contact your usual supplier or Bien-Air Dental directly. 9 Transport - Storage & Disposal 9.1 T T r r ansport and stor...
The warranty period is: • 12 months from the date of invoicing. In the event of justified claim, Bien-Air Dental or its authorised representative will fulfil the company’s obligations under this guarantee by repairing or replacing the product free of charge.