Champion T5 Owner's Operating And Maintenance Manual

Procedure chair

Advertisement

Quick Links

T5-X
OWNER'S OPERATING AND
MAINTENANCE MANUAL
T5 and T5-X PROCEDURE CHAIR
1-800-237-3377
www.championchair.com
customerservice@championchair.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the T5 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Champion T5

  • Page 1 T5-X OWNER’S OPERATING AND MAINTENANCE MANUAL T5 and T5-X PROCEDURE CHAIR 1-800-237-3377 www.championchair.com customerservice@championchair.com...
  • Page 2: Important Notes About This Manual

    Champion assumes no responsibility for damage or injury caused by improper assembly, installation, use, or maintenance of these products. No part of this manual may be duplicated in any form without the prior consent of Champion Manufacturing, Inc. Unauthorized duplication/distribution of these materials may result in civil prosecution to the maximum extent allowed by law.
  • Page 3: Table Of Contents

    Manual…............................Intended Use Statement................................. Transportation, Storage, Handling & Disposal..........................Preparation: Before You Begin..............................Specifications....................................... Power Cable Requirements................................Safety Precautions.................................... Product Labeling Diagrams (T5 Shown)............................Pinch Points ......................................Operating Instructions..................................Quick Release Back Section................................ Caster Brake Operation................................Side Rail Operation................................... Adjusting Side Rail Angle................................
  • Page 4: Intended Use Statement

    This device is battery powered and employs a pendant control to operate a series of electric motors for patient positioning. The pendant control is NOT INTENDED FOR PATIENT USE. A surgeon and/or clinician must always rely on his or her own clinical knowledge when deciding whether to use a T5 Procedure- Chair when treating a patient.
  • Page 5: Specifications

    Power consumption (standby) max. 0.5 W (depending on input voltage) Power consumption (charging) max. 30 W (depending on input voltage) DUTY CYCLE 10%, Max / 2 min, MIN / 18 min IP RATING All T5 Procedure chairs are rated IPX0 1-800-237-3377 www.championchair.com customerservice@championchair.com manual...
  • Page 6: Power Cable Requirements

    T5 & T5-X POWER/CABLE REQUIREMENTS Only use the power/charging cord that came with your product. SAFETY PRECAUTIONS WARNING IMPORTANT: PLEASE READ WARNINGS CHAIR OPERATION BY QUALIFIED, TRAINED MEDICAL PERSONNEL ONLY The chair is intended to be operated only by qualified, trained medical staff. Operation of chair by unauthorized, untrained and/or lay people must be avoided.
  • Page 7 If mounting accessories to back surgical rails, ensure accessory is properly installed and securely engaged prior to transporting patient or chair and prior to use. Only equipment approved by Champion is to be mounted on surgical rails. Champion Manufacturing, Inc. is not responsible for damage and assumes no liability caused by the use of unapproved equipment or accessories.
  • Page 8 For T5, chair must be at least 24 inches from nearest wall or obstruction to allow for full range of activation. Leave chair in lowest position whenever possible. This practice will decrease potential injury during an unsupervised patient egress from chair.
  • Page 9: Important: Please Read

    CAUTION: Use of accessories, transducers, and cables other than those specified or provided by Champion could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation. The AC power cable supplied with your chair is 3,000 mm long (+/- 300 mm). The pendant cable is 820 mm long coiled and 7,400 mm long uncoiled.
  • Page 10 T5 & T5-X TEST SPECIFICATIONS FOR ENCLOSURE PORT IMMUNITY TO RF WIRELESS COMMUNICATION EQUIPMENT Test Level Test Frequency (MHz) Band (MHz) Service Modulation (V/m) 380 to 390 TETRA 400 Pulse Modulation, 18 Hz FM ± 5 kHz deviation 1 kHz sine...
  • Page 11: Product Labeling Diagrams

    Every three (3) months, inspect all labels and ensure they are legible and not tattered, torn or missing. Refer to the label locations in figure below. If labels need replacing, contact Champion Customer Service at 1-800-237-3377. Back rest Quick Release...
  • Page 12: Pinch Points

    T5 & T5-X Item No. Part No. Description TMM-2544-10 T5 Base Cover Label TMM-2448-10 Label Base Cover TMM-2355-10 Surgical Support Locked Label (X-Models Only) TMM-2354-10 Surgical Support Notice Label (X-Models Only) TMM-754-10 Max Weight Label 500 lb TMM-684-10 Cover Clearance Caution Label (X-Models Only)
  • Page 13: Operating Instructions

    T5 & T5-X OPERATING INSTRUCTIONS QUICK RELEASE BACK SECTION FIG. 1 1. LOCATE RELEASE LEVER Back section quick release (red) lever is located under seat on patient’s right side. (Fig. 1) FIG. 2 2. ACTIVATE RELEASE LEVER To activate back section quick release, pull red lever out towards the arm rail.
  • Page 14: Caster Brake Operation

    T5 & T5-X CASTER BRAKE OPERATION BRAKE MODE FIG. 1 Activate braking system by pressing down on red end of either brake pedal located at base of chair. (Fig. 1) NOTE: This mode prevents all four casters from swiveling and all wheels from spinning.
  • Page 15: Side Rail Operation

    T5 & T5-X SIDE RAIL OPERATION FIG. 1 LOWERING SIDE RAIL Grasp top of side rail and push inward slightly, while pulling out red rail release tab. Once released, lower rail. (Fig. 1) STOWING SIDE RAIL To stow side rail, lower rail into “down” position below seat (or FIG.
  • Page 16: Pendant (Controller)

    T5 & T5-X IMPORTANT FIG. 3 FIG. 4 • You do not need to hold the bottom nut while tightening the top nut. • When adjusting, ensure bottom jam nut is contacting seat frame. • Ball joint head angle does not matter.
  • Page 17 T5 & T5-X PENDANT (CONTROLLER) BACK SECTION ADJUSTMENT First row of pendant buttons adjusts angle of back section. Press left button to raise chair back. Press right button to lower chair back. LEG SECTION ADJUSTMENT Second row of pendant buttons adjusts angle of leg section.
  • Page 18 T5 & T5-X PENDANT (CONTROLLER) INDICATOR LIGHTS MAINTENANCE REQUIRED BATTERY STATUS (see page 19) AC POWER CONNECTED LOCKOUT INDICATOR When illuminated, this function has been manually locked by the user, or has been automatically locked to prevent damage to the unit.
  • Page 19: Control Box Indicator Lights

    T5 & T5-X CONTROL BOX INDICATOR LIGHTS CONTROL BOX INDICATOR LIGHTS 1-800-237-3377 www.championchair.com customerservice@championchair.com manual Document No.: TMM-2572-13 - Revision: REV A...
  • Page 20: Fault Indicator Leds

    T5 & T5-X CONTROL BOX INDICATOR LIGHTS FAULT INDICATOR LEDs 1-800-237-3377 www.championchair.com customerservice@championchair.com manual Document No.: TMM-2572-13 - Revision: REV A...
  • Page 21: Patient Ingress/Egress

    T5 & T5-X PATIENT INGRESS/EGRESS WARNING PATIENT SHOULD NEVER BE PERMITTED TO ENTER / EXIT FROM ENDS OF PROCEDURE CHAIR WHEN IN AN UPRIGHT, PARTIALLY, OR TOTALLY RECLINED POSITION. EXCESSIVE WEIGHT ON ENDS COULD CAUSE CHAIR TO TILT, RESULTING IN POSSIBLE PATIENT INJURY.
  • Page 22: Push Bar Operation

    T5 & T5-X PUSH BAR OPERATION LOWERING PUSH BAR Grasp push bar handle while pulling out red release knob. Lower push bar into “down” position. RAISING PUSH BAR Lift push bar handle until it is in “up” position. NOTE: The red release knob will engage (lock) automatically.
  • Page 23: Batteries

    T5 & T5-X BATTERIES Your procedure chair comes with one rechargeable Lithium-Ion battery pack. The pack may be charged on the chair, or by using the optional tabletop charger (TMA301-15). To charge the battery on the chair, plug the AC power cord into the bottom of the control box. The other end of the cord may be plugged into any standard 110-220 VAC outlet.
  • Page 24: Options

    T5 & T5-X OPTIONS “W” OPTION: WIDE WIDTH For “W” option, patient surface width is 28”. Standard width is 24”. 24” standard 28” “W” option LOW HEIGHT OPTION Low “L” option reduces all heights (high and low) by 1.5” NOTE: To prevent chair damage, ensure ground clearance is at least 1.75”...
  • Page 25: Q" Option: Scale

    T5 & T5-X “Q” OPTION: SCALE TO MEASURE PATIENT’S WEIGHT WITH SCALE: Ensure chair base is level. Ensure chair is not contacting nearby objects. Activate chair’s caster brakes. To tare scale, press and hold the “ZERO” button for approximately 3 seconds until “ZERO” appears on the display, release button and the display will read 0.0.
  • Page 26: Support Leg Operation

    T5 & T5-X CHANGING PENDANT BATTERY Follow these instructions if pendant does not wake up when “WEIGH” button is pressed. Remove battery cover on back of pendant by sliding it downwards. Take out battery and replace it with a new 9V Industrial battery.
  • Page 27: Cleaning Instructions

    T5 & T5-X SUPPORT LEG OPERATION (CONTINUED) ADJUSTING CHAIR POSITION While the support leg is locked, the seat back and tilt functions cannot be adjusted, but the seat height and leg position are can still be adjusted. NOTE: Ensure the locking lever is completely stowed to return the chair to normal function.
  • Page 28: General Care And Cleaning

    Use of Facility approved detergents and disinfectants must comply with the instructions provided by the manufacturer(s) of those products. GENERAL CARE & CLEANING Proper care is essential in ensuring the durability and reliability of Champion upholstery. In general, all products should Cleaned Disinfected (per facility policy)
  • Page 29: Pad Removal Instructions

    T5 & T5-X While deflecting the tab, slide the cover outward until it disengages. Do the same for the other side of the cover. PAD REMOVAL When removing cushions, pull laterally. Do not pull from top to bottom. PREVENTIVE MAINTENANCE...
  • Page 30: Battery Safety & Charging Information

    WARNING Use of Batteries that are NOT factory approved will void any and all warranties. Replace only with Champion part number TMA300-15. SUGGESTED PROTOCOL FOR CHARGING YOUR BATTERY: Benchtop charger (CHL50): Ensure the cord for your battery charger is fully inserted into the battery charger inlet.
  • Page 31 Information contained in this document is based on the latest product data available at the time of printing. Photography, artwork, text and specifications are sub- ject to change. Photographs may show optional items. Champion Manufacturing Inc. reserves the right to make changes to products, materials, options, and/or specification at any time without notice or obligation.