Page 1
EGL-T16 Quadruple Oil-less Air Compressor 2023 USER MANUAL/Installation Instructions 1953 W. Gulf To Lake Hwy, Lecanto, FL 34461 Phone: 352-527-1975 FAX: 352-513-4801 www.Dentalvacuums.com sierradentalequipment@outlook.com...
Page 2
Troubleshooting Cont. Maintenance Log Congratulations! On the Purchase of your Eagle-T16 oil-less dental compressor. Our Eagle-T16 oil-less dental compressors provide clean dry air to each chair flawlessly, requiring less maintenance than traditional compressors. Now that’s peace of mind you can stand behind! We are certain you will love your new...
CUSTOMER SUPPORT “We are happy to help you with any questions or concerns you may have about your equipment” Contact Us at 352-527-1975 We're here to help! sierradentalequipment@outlook.com DentalVacuums.com Look for us… on social media! Sierra Dental Equipment Inc. 1953 W. Gulf to Lake Hwy Lecanto, FL 34461...
WARRANTY POLICY Sierra Dental Equipment Limited Warranty Warranty for the EAGLE-T16 oil-less compressor is limited to the original purchaser of the unit. Sierra Dental Equipment is only obligated under this warranty to the repair or replacement of defective parts or materials The defect(s) must be reported to Sierra Dental Equipment within the valid warranty period.
Page 5
SAFETY Use of the Eagle-T16 air compressor not in compliance with the instructions specified in this manual may result in permanent failure and voided warranty of the unit. WARNING To prevent fire or electrical shock, do not expose this unit to rain or moisture, excessive heat or improper ventilation.
Page 6
SAFETY CONTINUED IMPORTANT - This equipment: Has hot, electrical, and moving parts; May start or stop automatically, depending on the tank’s pressure or the tripping of electrical protection elements; May cause mechanical or electrical interferences in nearby equipment; Must not be operated in places where non-authorized people, children, or animals may have access;...
SYSTEM LAYOUT ¾ HP Compressor 24v Remote Low voltage control box heads (x4) wires Capacitors (x4) Desiccant Wall-mounted Dryer (x2) Heat Exchanger Low voltage reset Tank Power cord Purge solenoid valve Drain Air muffler & Auto purge Purge shut Coalescing filter off valve Purge silencer Solenoid Coil...
SPECIFICATIONS SITE REQUIREMENTS Voltage 230 Volts Circuit breaker rating 15 Amp Minimum wire size 12 Gauge Minimum trunk line size ½” 100° F Max ambient temp. 50° F Minimum ambient temp. This equipment must be installed in a well-ventilated mechanical room to prevent overheating and failure not covered by the warranty.
INSTALLATION YOU MUST DO A PRESSURE TEST AFTER THE COMPRESSOR IS INSTALLED TO VALIDATE WARRANTY! (See pg. 12) 1. Location: Install the compressor in a covered, well ventilated area, free of dust, gases, toxic gases, humidity, or any other kind of pollutant. The maximum ambient temperature recommended during operation should be no greater than 95 degrees a.
Page 11
INSTALLATION 5. The electrical input wires must be in accordance to the motor power, voltage and distance from electrical energy source. 6. Before installing the product, verify that the product voltage is compatible to the local electrical power supply (220-240) – (208) require the 24v transformer to be rewired.
Once all the leaks are addressed. Repeat steps 1-4. DO NOT INSTALL this compressor on a leaking air system, THIS WILL VOID YOUR WARRANTY. Your EAGLE-T16 Compressor is ready to operate. This compressor is preset from the factory to automatically start at 80 psi and shut off at 100 psi.
PREVENTATIVE MAINTENANCE In order to guarantee the compressor’s perfect operation and good warranty status which will ultimately extend the service life, please follow these recommendations below: 1. Monthly a. Check the operation of the pressure switch (80-100psi) b. Check the tank for condensation using the drain petcock (The tank’s pressure must not be greater than 10 psi) If moisture is present you need to replace the filter &...
TROUBLESHOOTING Motor(s) will not start -Check supply voltage with a multimeter. (220- 240 is acceptable) Buck-boost transformer will be necessary if your voltage is above or below the acceptable voltage. -Damaged motor or start capacitor. -Automatic Thermal protector tripped . (Motor will re-start as soon as temperature decreases) If this happens, contact technical support.
Page 15
TROUBLESHOOTING The compressor unit overheats -High ambient temperature (MAX 100 degrees Fahrenheit) Improve installation conditions (climate control, ventilation,) -Inadequate electric power supply. -Check supply voltage with a multimeter. (220- 240 is acceptable) Buck-boost transformer will be necessary if your voltage is above or below the acceptable voltage....
MAINTENANCE LOG Date of Working Ambient Additional Technician Services Provided Service Hours Temp. Remarks Signature...
Page 17
Once all the leaks are addressed. Repeat steps 1-4. DO NOT INSTALL this compressor on a leaking air system, THIS WILL VOID YOUR WARRANTY. Your EAGLE-T16 Compressor is ready to operate. This compressor is preset from the factory to automatically start at 80 psi and shut off at 100 psi.