The manufacturer of the above named equipment is: Proton Products International Limited 10 Aylesbury End Beaconsfield Bucks HP9 1LW ENGLAND Proton Products is an ISO9001:2015 registered company. The declaration is signed by: ……………………… Paul Sives InteliTHERM SRTG03-5 Instruction Manual, issue 2b Page 3 of 21...
NTRODUCTION The Proton Products InteliTHERM SRTG03-5 spectrum radiation temperature gauge is non- contact temperature measuring instrument designed for use on continuous cable production lines for bare metal wire temperature measurement. It monitors temperature by scanning the wire surface to accurately detect wire temperature ranging from 80-400° C.
Gauge OK, Upper tolerance exceeded, User configurable Lower tolerance exceeded. Upper or function lower tolerance exceeded. 1x relay output Maximum voltage 24 Vdc Isolated contact rating Maximum current Communication interfaces RS-232, CAN-bus InteliTHERM SRTG03-5 Instruction Manual, issue 2b Page 5 of 21...
SRTG Gauge head. This version is suitable for single wires up to a temperature of 200C (392F). Proton part number 00055MC030 InteliTHERM SRTG03-5 Instruction Manual, issue 2b Page 7 of 21...
3. Attach the CDI4 display unit to the SRTG using two fixing plates with four screws. SRTG NSTALLING THE GUIDE ROLLER TO THE 1. Attach the guide roller to the SRTG with two fixing screws as shown below. InteliTHERM SRTG03-5 Instruction Manual, issue 2b Page 8 of 21...
SRTG bottom view (with Guide Kit installed) SRTG back view Air wipe / cooling water connector specification Connector type G1/8 BSPP JOINT Pipe Dia. =6mm Compressed Air Quality ISO 8573.1 Solid Particle Class I Class 1 InteliTHERM SRTG03-5 Instruction Manual, issue 2b Page 9 of 21...
It will power up as soon as the CAN cable is connected. Power indications LED status Indication Continuous red Gauge is powered on Power Extinguished Gauge is powered off InteliTHERM SRTG03-5 Instruction Manual, issue 2b Page 10 of 21...
In this case users may clean the window or contact the Proton service personnel. The LED display will flash when this window is dirty. InteliTHERM SRTG03-5 Instruction Manual, issue 2b Page 11 of 21...
1. Press Menu button on the main page of the preheater display unit. 2. Press Comms button on the function menu page. 3. Set the PROTON TG CAN ID then press the “OK” button. InteliTHERM SRTG03-5 Instruction Manual, issue 2b...
NTERFACE SETTINGS 1. Press Interface button on the function menu page. 2. Set the Temp control source as PROTON TG on the Interface page 3. InteliTHERM SRTG03-5 Instruction Manual, issue 2b Page 13 of 21...
3. Select “Auto” on the Control mode then set the Temperature gain, Control range and Distance (TG to PH) values then press >> button to enter the TG page 2. InteliTHERM SRTG03-5 Instruction Manual, issue 2b Page 14 of 21...
Page 15
4. Set the Control I, Control P, Temperature threshold control, Average number and Reference emissivity values then click >> button to enter the TG page 3. 5. Set the materials emissivity values then click Home button to return to the home page. InteliTHERM SRTG03-5 Instruction Manual, issue 2b Page 15 of 21...
Set the emissivity of the measured wire. Press the button to return to the home page. Press the button to access the function menu page. Press the button to advance to the next page. InteliTHERM SRTG03-5 Instruction Manual, issue 2b Page 16 of 21...
AOP zero internal analogue-to-digital converter; it may be adjusted if the SRTG displays a non-zero measured temperature. Relay option Select the relay output from options: Gauge OK; Tol+Tol-; Tol+; Tol-. InteliTHERM SRTG03-5 Instruction Manual, issue 2b Page 17 of 21...
This CAN-bus interface is fitted as standard. It operates independently of the other communications interfaces and may be accessed at the same time as them. The CAN-bus interface uses a proprietary Proton Products protocol. It is exclusively used to communicate between the unit and other Proton Products modules. The unit automatically detects connection to other modules and configures the bus appropriately;...
300pF per metre, the maximum recommended baud rates are as follows: Cable length range / m Maximum Baud rate / s 115200 38400 19200 9600 4800 InteliTHERM SRTG03-5 Instruction Manual, issue 2b Page 19 of 21...
If the relay contact is used to switch inductive loads, then suppressors must be fitted to clamp voltage spikes. Specification Minimum Typical Maximum Units DC voltage between contacts Contact-to-earth voltage Current Ω On resistance (at a current of 10mA) InteliTHERM SRTG03-5 Instruction Manual, issue 2b Page 20 of 21...
Proton Products is not responsible for consequential or incidental damage related to the provision or use of the information contained in this manual. • The information contained in this manual is the property of Proton Products and may not be circulated or distributed to third parties. •...