Clave 8+ Instruction Manual Serial numbers starting with ‘Y’ (3 Gen) Steam Sterilizer...
Page 2
Thank you for choosing our steam sterilizer. Prior to operating this device, please read the instruction manual carefully and follow all installation instructions. Regular maintenance: If this symbol appears on the screen when the device is powered on or if it shows E88 on the sterilization report, please contact your dealer or local service company for a maintenance service.
Rev 1.0 Contents 1 Scope of manual General 1 Intended use 1 General safety instructions 1 Standards and directives 2 Symbols Description of the sterilizer 2 Sterilizer views 2 Control panel 4 Technical specification Installation 4 Packing content 5 General conditions 5 Power supply Connection Setup 5 Location requirements...
Instructions Manual Rev. 1.0 1 General 1.1 Scope of Manual This manual contains information concerning the installation, operation and maintenance of the CLAVE8+ steam sterilizer. To ensure proper performance of the sterilizer, the instructions given in this manual should be thoroughly understood and followed.
Instructions Manual 1.5 Symbols For safe operation, please pay close attention to the alert symbols below which can be found in the sterilizer or throughout this manual. This symbol represents an electrical caution - ground protection. Hot Surface This symbol represents a warning of a potential hot surface. Important safety information.
Instructions Manual Rev. 1.0 2.2 Control panel Program temperature Program Holding time Pressure Distilled water / Fill if flashing. Used water tank full / Drain if flashing Printer status Door Open Door Closed Door Locked Time Current internal temperature Up button Down button Start/Stop button Menu button...
Instructions Manual 2.3 Technical specifications Item Clave8+ Chamber(mm) Φ(6.7 x 12.2”) 170 x 320mm Rated Voltage 110-130 VAC,60Hz Circuit breaker F16A/400V Nominal power 1600VA Sterilization 121°C /134°C temperature Capacity of the distilled Approx 2.5 L (Water at level Max) water tank Approx 0.5 L (Water at level Min) Operation temperature 5°C-40°C...
Instructions Manual Rev. 1.0 3. Installation 3.1 General conditions Position the device on a solid surface with a minimum weight capacity of 60 kgs. The sterilizer should be placed on a level worktable. Leave at least 10 cm between the device rear part and the wall. The clearance required to open the door is 40 cm. ...
Instructions Manual 4 Setup Open the door to remove all of the inner contents for unpacking. Connect the power cord to the rear of the unit, and to an outlet of the appropriate voltage. Turn on the main power switch on the right side. After switching on, the machine turns on the display and shows the door position, water level, working program, date, time, etc.
Instructions Manual Rev. 1.0 4.4 Basic Set Press the ↖ button to access the main menu. From the main menu, select "Basic Set”. Pr ogr am D ate : 11 -07 -20 19 Ba sic Se t T ime : 12: 05 :35 Re por t L ang uag e:...
Page 11
Instructions Manual 4.7 Parameter The “Parameter” menu allows you to set the following options: Holding time • • Dry time Select (with ↑↓ buttons) “Parameter” from the menu by pressing ↖ button. Select the program by pressing ↑↓ then press ↖ button. Select the parameter by pressing ↖...
Page 12
Instructions Manual Rev. 1.0 4.11 Water Quality (optional) If your sterilizer is equipped with a water quality sensor and you want to deactivate it, select (with ↑↓ buttons) “Water quality” from the menu by pressing ↖ button. Adjust it by pressing ↑↓. Press button to save and exit.
Instructions Manual 5. Operation 5.1 Prerequisites Switch On. Check the status of the icons in the screen, . (Refer to section 2. Description of the sterilizer) ** THE DOOR MUST BE OPEN PRIOR TO TURNING ON THE POWER TO ALLOW THE UNIT TO CALIBRATE ** W rap pe d I nit ia liz e..
Instructions Manual Rev. 1.0 Caution: Always use the tray handle to load or unload the tray into the sterilizer. Failure to do so can result in burns. 5.6 Manual interruption of the cycle To interrupt a cycle prematurely, hold for 3 seconds. If the cycle is manually interrupted after it reaches the drying phase, the items inside the sterilizer may be considered sterile and since the cycle has been interrupted during the drying phase the materials and instruments inside the chamber may be wet.
Page 15
Instructions Manual Example report: ================================ Program: Wrapped Temperature: 134 C Pressure: 204 kPa Drying Time: 2.0 MIN Holding Time: 04.0 MIN -------------------------------- Time Temp. Pressure HH:MM:SS Start 11:38:02 028.2 002.5 11:50:46 115.9 081.0 11:51:17 107.2 020.6 11:51:29 111.0 058.8 11:51:52 107.3 020.1 11:52:26...
Page 16
Instructions Manual Rev. 1.0 5.10 Label printer (Optional) Report 5.9.1 Connect the label printer and switch on. Label 5.9.2 Select “Label”, press ↖ button. About device 5.9.3 Select the records with ↑↓ buttons. Setup 5.9.4 Press ↖ button to set the quantity. 5.9.5 Change the quantity of labels that you want to print with ↑↓...
Instructions Manual 6.Maintenance To ensure proper operation and maximum steam sterilizer life, carefully follow all recommendations for periodic maintenance. One of the most important steps you can take to prevent problems with your sterilizer is to use only distilled water. Frequency Number of cycles Maintenance operation...
Instructions Manual Rev. 1.0 6.4 Door adjustment Under normal circumstances, the chamber door does not require adjustments. However, if the seal fails (resulting in steam leaking from the front of the chamber), you may need to adjust the door tension. Open the door.
Instructions Manual 7 Troubleshooting If an error is displayed, the unit will refuse to run a cycle until it is cleared. If an error persists or returns, contact your dealer. Most error codes can be cleared by taking the following steps: Open door (assuming pressure has reached zero, and error isn't about the door) Power cycle the unit (off, wait 5 seconds, on) If error still present, press and hold the ↖...
Instructions Manual Rev. 1.0 8 Transportation and storage Switch off the sterilizer, unplug power cord, and let the machine cool down. Drain the distilled water tank and the used water tank. Conditions for transport and storage Temperature: -20°C ~ +50°C •...
Certificate of Warranty All new Clave16 or Clave 23 or Class B autoclaves installed by a Flight Dental Systems authorized dealer for a period of two (2) full year from the time of purchase. The warranty covers defects in parts, workmanship and materials for two (2) year except door gaskets and filters which are wear and tear items.
Instructions Manual Contact Information Thank you for taking the time to review the Flight Dental Unit’s Instruction Manual. Your feedback or comments regarding the document is welcomed. For any comments and concerns please mail, e-mail or phone us at: Sales and Marketing Joseph Hui 1-866-799-0517 905-799-0517...