  • Page 1 Clave 8+ Instruction Manual Serial numbers starting with ‘Y’ (3 Gen) Steam Sterilizer...
  • Page 2 Thank you for choosing our steam sterilizer. Prior to operating this device, please read the instruction manual carefully and follow all installation instructions. Regular maintenance: If this symbol appears on the screen when the device is powered on or if it shows E88 on the sterilization report, please contact your dealer or local service company for a maintenance service.
  • Page 3: Table Of Contents

    Rev 1.0 Contents 1 Scope of manual General 1 Intended use 1 General safety instructions 1 Standards and directives 2 Symbols Description of the sterilizer 2 Sterilizer views 2 Control panel 4 Technical specification Installation 4 Packing content 5 General conditions 5 Power supply Connection Setup 5 Location requirements...
  • Page 4: General

    Instructions Manual Rev. 1.0 1 General 1.1 Scope of Manual This manual contains information concerning the installation, operation and maintenance of the CLAVE8+ steam sterilizer. To ensure proper performance of the sterilizer, the instructions given in this manual should be thoroughly understood and followed.
  • Page 5: Description Of The Sterilizer

    Instructions Manual 1.5 Symbols For safe operation, please pay close attention to the alert symbols below which can be found in the sterilizer or throughout this manual. This symbol represents an electrical caution - ground protection. Hot Surface This symbol represents a warning of a potential hot surface. Important safety information.
  • Page 6: Control Panel

    Instructions Manual Rev. 1.0 2.2 Control panel Program temperature Program Holding time Pressure Distilled water / Fill if flashing. Used water tank full / Drain if flashing Printer status Door Open Door Closed Door Locked Time Current internal temperature Up button Down button Start/Stop button Menu button...
  • Page 7: Technical Specification

    Instructions Manual 2.3 Technical specifications Item Clave8+ Chamber(mm) Φ(6.7 x 12.2”) 170 x 320mm Rated Voltage 110-130 VAC,60Hz Circuit breaker F16A/400V Nominal power 1600VA Sterilization 121°C /134°C temperature Capacity of the distilled Approx 2.5 L (Water at level Max) water tank Approx 0.5 L (Water at level Min) Operation temperature 5°C-40°C...
  • Page 8: Installation

    Instructions Manual Rev. 1.0 3. Installation 3.1 General conditions Position the device on a solid surface with a minimum weight capacity of 60 kgs.  The sterilizer should be placed on a level worktable.  Leave at least 10 cm between the device rear part and the wall. The clearance required to open the door is 40 cm. ...
  • Page 9: Fill The Distilled Water Tank

    Instructions Manual 4 Setup Open the door to remove all of the inner contents for unpacking. Connect the power cord to the rear of the unit, and to an outlet of the appropriate voltage. Turn on the main power switch on the right side. After switching on, the machine turns on the display and shows the door position, water level, working program, date, time, etc.
  • Page 10: Basic Set

    Instructions Manual Rev. 1.0 4.4 Basic Set Press the ↖ button to access the main menu. From the main menu, select "Basic Set”. Pr ogr am D ate : 11 -07 -20 19 Ba sic Se t T ime : 12: 05 :35 Re por t L ang uag e:...
  • Page 11 Instructions Manual 4.7 Parameter The “Parameter” menu allows you to set the following options: Holding time • • Dry time Select (with ↑↓ buttons) “Parameter” from the menu by pressing ↖ button. Select the program by pressing ↑↓ then press ↖ button. Select the parameter by pressing ↖...
  • Page 12 Instructions Manual Rev. 1.0 4.11 Water Quality (optional) If your sterilizer is equipped with a water quality sensor and you want to deactivate it, select (with ↑↓ buttons) “Water quality” from the menu by pressing ↖ button. Adjust it by pressing ↑↓. Press button to save and exit.
  • Page 13: Operation

    Instructions Manual 5. Operation 5.1 Prerequisites Switch On. Check the status of the icons in the screen, . (Refer to section 2. Description of the sterilizer) ** THE DOOR MUST BE OPEN PRIOR TO TURNING ON THE POWER TO ALLOW THE UNIT TO CALIBRATE ** W rap pe d I nit ia liz e..
  • Page 14: Manually Interruption Of The Cycle

    Instructions Manual Rev. 1.0 Caution: Always use the tray handle to load or unload the tray into the sterilizer. Failure to do so can result in burns. 5.6 Manual interruption of the cycle To interrupt a cycle prematurely, hold for 3 seconds. If the cycle is manually interrupted after it reaches the drying phase, the items inside the sterilizer may be considered sterile and since the cycle has been interrupted during the drying phase the materials and instruments inside the chamber may be wet.
  • Page 15 Instructions Manual Example report: ================================ Program: Wrapped Temperature: 134 C Pressure: 204 kPa Drying Time: 2.0 MIN Holding Time: 04.0 MIN -------------------------------- Time Temp. Pressure HH:MM:SS Start 11:38:02 028.2 002.5 11:50:46 115.9 081.0 11:51:17 107.2 020.6 11:51:29 111.0 058.8 11:51:52 107.3 020.1 11:52:26...
  • Page 16 Instructions Manual Rev. 1.0 5.10 Label printer (Optional) Report 5.9.1 Connect the label printer and switch on. Label 5.9.2 Select “Label”, press ↖ button. About device 5.9.3 Select the records with ↑↓ buttons. Setup 5.9.4 Press ↖ button to set the quantity. 5.9.5 Change the quantity of labels that you want to print with ↑↓...
  • Page 17: Maintenance

    Instructions Manual 6.Maintenance To ensure proper operation and maximum steam sterilizer life, carefully follow all recommendations for periodic maintenance. One of the most important steps you can take to prevent problems with your sterilizer is to use only distilled water. Frequency Number of cycles Maintenance operation...
  • Page 18: Door Adjustment

    Instructions Manual Rev. 1.0 6.4 Door adjustment Under normal circumstances, the chamber door does not require adjustments. However, if the seal fails (resulting in steam leaking from the front of the chamber), you may need to adjust the door tension. Open the door.
  • Page 19: Troubleshooting

    Instructions Manual 7 Troubleshooting If an error is displayed, the unit will refuse to run a cycle until it is cleared. If an error persists or returns, contact your dealer. Most error codes can be cleared by taking the following steps: Open door (assuming pressure has reached zero, and error isn't about the door) Power cycle the unit (off, wait 5 seconds, on) If error still present, press and hold the ↖...
  • Page 20: Transportation And Storage

    Instructions Manual Rev. 1.0 8 Transportation and storage Switch off the sterilizer, unplug power cord, and let the machine cool down. Drain the distilled water tank and the used water tank. Conditions for transport and storage Temperature: -20°C ~ +50°C •...
  • Page 21: Appendix 1

    Instructions Manual Appendix 1 Water properties / Characteristics Description Feed water Condensate Evaporate residue ≤ 10mg/ I ≤ 1.0mg/kg Silicon oxide sio2 ≤ 1mg/ I ≤ 1.0mg/kg Iron ≤ 0.2mg/ I ≤ 0.1mg/kg Cadmium ≤ 0.005mg/ I ≤ 0.05mg/kg Lead ≤...
  • Page 22: Diagrams Of The Sterilization Programs

    Instructions Manual Rev. 1.0 Appendix 2 Diagrams of the sterilization programs Programs (CLAVE 8+ Type STE-8-K) 10-20 FLASH Unwrapped solid material 2.00 0.60 20-35 30-50 Unwrapped solid material 2.00 0.60 Single-wrapped solid or 48-60 hollow 1.50 0.50 WRAPPED material Unwrapped porous material 0.50 0.15 Single-wrapped porous...
  • Page 23 Instructions Manual...
  • Page 24: Wiring Diagram

    Instructions Manual Rev. 1.0 Appendix 3 Wiring Diagram TP1: Steam generator temperature sensor TP2: Inner temperature sensor TP3: Temperature sensor of chamber wall V1: Air release valve (normally open) V4: Water release valve (normally closed)
  • Page 25: Hydraulic Diagram

    Instructions Manual Appendix 4 Hydraulic Diagram...
  • Page 26: Certificate Of Warranty

    Certificate of Warranty All new Clave16 or Clave 23 or Class B autoclaves installed by a Flight Dental Systems authorized dealer for a period of two (2) full year from the time of purchase. The warranty covers defects in parts, workmanship and materials for two (2) year except door gaskets and filters which are wear and tear items.
  • Page 27: Contact Information

    Instructions Manual Contact Information Thank you for taking the time to review the Flight Dental Unit’s Instruction Manual. Your feedback or comments regarding the document is welcomed. For any comments and concerns please mail, e-mail or phone us at: Sales and Marketing Joseph Hui 1-866-799-0517 905-799-0517...