Flight Dental Systems Clave B Instruction Manual

Flight Dental Systems Clave B Instruction Manual

Third generation system, s/n starting with a

Advertisement

Class B Steam Sterilizer
Third Generation System, s/n starting with A
Instruction Manual
STE-23-D

Advertisement

Table of Contents
loading

Summary of Contents for Flight Dental Systems Clave B

  • Page 1 Class B Steam Sterilizer Third Generation System, s/n starting with A Instruction Manual STE-23-D...
  • Page 2 Thank you for choosing our steam sterilizer. Prior to operating this instrument, please read the instruction manual carefully and follow all installation instructions. IMPORTANT NOTICE If you can’t open the door, please unlock the door according to the instructions “How to open the door in case of power outage”...
  • Page 3: Table Of Contents

    Contents 1. General 4 Scope of manual 4 Intended use 4 General safety instructions 4 Standards and directives 5 Symbols 2. Description of the sterilizer 5 Sterilizer views 6 Control panel 6 Technical specification 6 Packing content 3. Installation 7 General conditions 7 Power supply connection 7 Location requirements 4.
  • Page 4: General

    1 General Scope of Manual This manual contains information concerning the installation, operation and maintenance of the steam sterilizer. To ensure proper performance of the sterilizer, the instructions given in this manual should be thoroughly understood and followed. Keep the manual near the sterilizer in an accessible location for future reference. Intended Use The steam sterilizer described in this manual is intended for the sterilization in all medical, dental, beauty, vet and tattoo fields of the following types of instrument loads: solid, porous, hollow loads...
  • Page 5: Symbols

    Symbols For safe operation, please pay close attention to the alert symbols below which can be found on the pay close attention to the alert symbols below which can be found on the pay close attention to the alert symbols below which can be found on the sterilizer and throughout this manual.
  • Page 6: Control Panel

    Current internal temperature Up button Down button Enter button Menu button Technical Specifications Model Clave B Chamber (m, diameter/depth)(inches) φ247 x 450( 9.724” x 17.716”) Overall dimensions (mm,W*H*D) 490 x 455 x 690 Net Weight (kg) Nominal power (VA) 2000 Electrical supply 220-230V;50/60 Hz;10A...
  • Page 7: Installation

    3 Installation General conditions  Position the device on a solid surface with a minimum weight capacity of 60 kgs.  The sterilizer should be placed on a level worktable.  Leave at least 10 cm between the device rear part and the wall. The clearance required to open the door is 40 cm.
  • Page 8: Setup

    4 Setup Connect the power cord to an outlet of the appropriate voltage. Open the door to remove all of the inner contents for unpacking. Turn on the main power switch on the right side. After switching on, the machine turns on the LCD and shows the door position, water level, working program, date, time, etc.
  • Page 9: Basic Set

    Basic Set From the main menu, select "Basic Set”. Progra m Date: 11-07-2019 Basic Set Time: 1 2:05:35 Report Langu age: ENG Label Counter: 00000 The “Basic Set” menu permits to set the following options: *Date *Time *Language Select the “Basic Set” from the main menu by pressing button.
  • Page 10 Parameter The “Parameter” menu permits setting the following options: *Holding time *Dry time Select “Parameter” from the menu by pressing button. Select the program by pressing then press button. Select the parameter by pressing button. Adjust it by pressing Press button to save and exit.
  • Page 11: Operation

    Expiry date (Optional) To modify the expiration date of the labels, select “Expiry date” from the menu by pressing button. Adjust it by pressing Press button to save and exit. Parameter Unit Month: Preheat Expiry date Water Quality (optional) If your sterilizer is equipped with a water quality sensor and you want to deactivate it, select “Water quality”...
  • Page 12: Select The Program

    Load Open the door then placed the trays inside the chamber by the tray handle. After the instruments are loaded, you may close the door. Select the program Enter to the main menu by pressing button, Select “Program”. Select the program by pressing then press to confirm program, in the screen will appear the information of selected program as the temperature and sterilization time (holding time), also the date, time, current pressure and...
  • Page 13: Manually Interruption Of The Cycle

    Manual interruption of the cycle To interrupt a cycle prematurely, hold for 3 seconds. If the cycle is manually interrupted after it reaches the drying phase, the items inside the sterilizer may be considered sterile and considering that the cycle has been interrupted during the drying phase the materials and instruments inside the chamber may be wet.
  • Page 14: Data

    If the temperature difference between the maximum temperature and the minimum is more than 3°C, it will show void. That means the result of the test is a fail. You will need to run the vacuum test again after the chamber has cooled down.
  • Page 15 Sample of a printer report When reading the printed data records, refer to the diagram below: ================================ ======================= Program: WRAPPED Program:Vacuum test Temperature: 134C Tp:1℃ Pressure: 206.0 kPa P1:-75.0kPa Drying Time: 08Min P2:-74.0kPa Holding Time: 4.0Min rate of pressure rise:0.10 -------------------------------- Start Time:08:22 Time...
  • Page 16: Maintenance

    Print labels Select “Label” from the main menu and press button to enter the menu. Select the cycle number by pressing button. Choose the labels quantity by pressing then press button to print. Press button to exit. Program 00012 Ba sic Set 00011 QT: 01 Report...
  • Page 17: Door Adjustment

    Door adjustment Under normal circumstances, the chamber door does not require adjustments. However, if the seal fails (resulting in steam leaking from the front of the chamber), you may need to adjust it. Open the door. Insert the spanner tool in the gap beneath the plastic cover. Use the spanner to grip the adjusting nut.
  • Page 18: Troubleshooting

    7 Troubleshooting Code Description Proposed solution Steam generator temperature Power off & run a new cycle sensor error. Contact your supplier if error persists. Power off & run a new cycle Inner temperature sensor error. Contact your supplier if error persists. Temperature sensor of the Carefully ensure that the chamber wall is chamber wall error.
  • Page 19: Transportation And Storage

    8 Transportation and storage Switch off the sterilizer before transportation or storage. Pull out the plug. Let the machine cool down. Drain the distilled water tank and the used water tank. Conditions for transport and storage Temperature: -20°C ~ +50°C Relative humidity: ≤...
  • Page 20: Appendix 1

    Appendix 1 Water properties / Characteristics Description Feed water Condensate Evaporate residue ≤ 10mg/ I ≤ 1.0mg/kg Silicon oxide sio2 ≤ 1mg/ I ≤ 1.0mg/kg Iron ≤ 0.2mg/ I ≤ 0.1mg/kg Cadmium ≤ 0.005mg/ I ≤ 0.05mg/kg Lead ≤ 0.05mg/ I ≤...
  • Page 21: Diagrams Of The Sterilization Programs

    Appendix 2 Diagrams of the sterilization programs Programs Type (STE-23-D) SOLID 15-35 Unwrapped solid material 5.00 1.50 (Unwrapped) 31-40 30-50 Unwrapped solid material 5.00 1.50 Single-wrapped solid or WRAPPED 48-60 hollow 4.00 1.20 material Unwrapped porous material 1.25 0.40 36-55 Single-wrapped porous 1.10 0.30...
  • Page 22: Wiring Diagram

    Appendix 3 Wiring Diagram TP1: Steam generator temperature sensor TP2: Inner temperature sensor TP3: Temperature sensor of chamber wall V1: Vacuum pump valve V2: Air filter valve V3: Pump valve V4: Water release valve V5: Vacuum pump start valve V6: Vacuum pump fan...
  • Page 23: Hydraulic Diagram

    Appendix 4 Hydraulic Diagram V1: Vacuum pump valve (Normally closed) V2: Air filter valve (Normally open) V3: Pump valve (Normally closed) V4: Water release valve (Normally open) V5: Vacuum pump start valve (Normally closed)
  • Page 24 Certificate of Warranty All new Clave16 or Clave 23 or Class B autoclaves installed by a Flight Dental Systems authorized dealer for a period of two (2) full year from the time of purchase. The warranty covers defects in parts, workmanship and materials for two (2) year except door gaskets and filters which are wear and tear items.
  • Page 25 Contact Information Thank you for taking the time to review the Flight Dental Unit’s Instruction Manual. Your feedback or comments regarding the document is welcomed. For any comments and concerns please mail, e-mail or phone us at: Sales and Marketing Joseph Hui 1-866-799-0517 905-799-0517...