The ULTRAFORMER MPT comes with a three-conductor AC power cable and plug. Use a
properly grounded outlet and always plug the ULTRAFORMER MPT directly into the
outlet.
Never remove the ground conductor via any AC adaptor plugs or extension cables.
Disconnect the AC Power cable from the outlet by pulling on the plug not the cord.
Do not touch the AC Power cable with wet hands.
Turn off the AC Power switch and disconnect the AC Power cable from the outlet prior
to cleaning the Main Body.
Do not remove the covers on the Main Body or Hand-piece. The ULTRAFORMER MPT
contains no user serviceable components. If the System requires service, contact local
distributors or Classys Inc.
No modification of this equipment is allowed.
The ULTRAFORMER MPT should not be used near flammable gases or anesthetics. Fire
or explosions may occur. The ULTRAFORMER MPT is not AP or APG rated.
Avoid restricting ventilation under or behind the Main Body of the system. Maintain an
open space of at least 50cm around the Main Body. If ventilation holes are obstructed,
the System may overheat.
The Hand-piece is rated as a Type B patient applied part. It may provide a connection
between the patient and protective earth. This may present a hazard if the patient
becomes connected to other equipment with excessive electrical current leakage.
Use the Cartridge provided by Classys Inc. only. Operating with any other Cartridges may
cause patient injury or malfunction of the device.
Do not suddenly remove the Hand-piece from the Main Body. It may malfunction or
cause damage if separated by force.
Please refer to section "3-3-3" for warnings of Electromagnetic Compatibility and
Immunity".
ULTRAFORMER MPT
Operation Manual