Endress+Hauser Proline Promass S 500 Operating Instructions Manual
Endress+Hauser Proline Promass S 500 Operating Instructions Manual

Endress+Hauser Proline Promass S 500 Operating Instructions Manual

Coriolis flowmeter profinet over ethernet-apl
Hide thumbs Also See for Proline Promass S 500:
Table of Contents

Advertisement

Quick Links

BA02130D/06/EN/02.24-00
71674748
2024-11-01
Valid as of version
01.00.zz (Device firmware)
advanced physical layer
Products
Operating Instructions
Proline Promass S 500
Coriolis flowmeter
PROFINET over Ethernet-APL
Solutions
Services

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Proline Promass S 500 and is the answer not in the manual?

Questions and answers

Summary of Contents for Endress+Hauser Proline Promass S 500

  • Page 1 Products Solutions Services BA02130D/06/EN/02.24-00 71674748 2024-11-01 Valid as of version 01.00.zz (Device firmware) Operating Instructions Proline Promass S 500 Coriolis flowmeter PROFINET over Ethernet-APL advanced physical layer...
  • Page 2 • The manufacturer reserves the right to modify technical data without prior notice. Your Endress+Hauser sales organization will supply you with current information and updates to this manual. Endress+Hauser...
  • Page 3: Table Of Contents

    Proline Promass S 500 PROFINET over Ethernet-APL Table of contents Table of contents About this document ....6 Installation ..... . . 21 Document function .
  • Page 4 Table of contents Proline Promass S 500 PROFINET over Ethernet-APL Structure and function of the operating 10.5.2 Displaying the communication menu ......
  • Page 5 13.2 Measuring and test equipment ... 271 13.3 Endress+Hauser services ....271 Repair ......272 14.1 General notes .
  • Page 6: About This Document

    About this document Proline Promass S 500 PROFINET over Ethernet-APL About this document Document function These Operating Instructions contain all the information required in the various life cycle phases of the device: from product identification, incoming acceptance and storage, to installation, connection, operation and commissioning, through to troubleshooting, maintenance and disposal.
  • Page 7: Tool Symbols

    Proline Promass S 500 PROFINET over Ethernet-APL About this document 1.2.4 Tool symbols Symbol Meaning Torx screwdriver Phillips head screwdriver Open-ended wrench 1.2.5 Symbols for certain types of information Symbol Meaning Permitted Procedures, processes or actions that are permitted. Preferred Procedures, processes or actions that are preferred.
  • Page 8: Documentation

    • Device Viewer (www.endress.com/deviceviewer): Enter the serial number from the nameplate • Endress+Hauser Operations app: Enter serial number from nameplate or scan matrix code on nameplate. The following documentation may be available depending on the device version ordered: Document type...
  • Page 9: Safety Instructions

    Proline Promass S 500 PROFINET over Ethernet-APL Safety instructions Safety instructions Requirements for the personnel The personnel for installation, commissioning, diagnostics and maintenance must fulfill the following requirements: ‣ Trained, qualified specialists must have a relevant qualification for this specific function and task.
  • Page 10: Workplace Safety

    Verification for borderline cases: ‣ For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance in verifying the corrosion resistance of fluid-wetted materials, but does not accept any warranty or liability as minute changes in the temperature, concentration or level of contamination in the process can alter the corrosion resistance properties.
  • Page 11: Device-Specific It Security

    Proline Promass S 500 PROFINET over Ethernet-APL Safety instructions Device-specific IT security The device offers a range of specific functions to support protective measures on the operator' s side. These functions can be configured by the user and guarantee greater in- operation safety if used correctly.
  • Page 12: Infrastructure Mode

    Safety instructions Proline Promass S 500 PROFINET over Ethernet-APL WLAN passphrase: Operation as WLAN access point A connection between an operating unit (e.g. notebook or tablet) and the device via the WLAN interface (→  88), which can be ordered as an optional extra, is protected by the network key.
  • Page 13: Product Description

    Proline Promass S 500 PROFINET over Ethernet-APL Product description Product description The measuring system consists of a transmitter and a sensor. The transmitter and sensor are mounted in physically separate locations. They are interconnected by connecting cables. Product design Two versions of the transmitter are available.
  • Page 14 Product description Proline Promass S 500 PROFINET over Ethernet-APL 3.1.2 Proline 500 Signal transmission: analog Order code for "Integrated ISEM electronics", option B "Transmitter" For use in applications required to meet special requirements due to ambient or operating conditions. As the electronics are located in the transmitter, the device is ideal in the event of: •...
  • Page 15: Incoming Acceptance And Product

    • Enter the serial numbers from the nameplates in the Device Viewer (www.endress.com/deviceviewer): all the information about the device is displayed. • Enter the serial numbers from the nameplates into the Endress+Hauser Operations app or scan the DataMatrix code on the nameplate with the Endress+Hauser Operations app: all the information about the device is displayed.
  • Page 16: Transmitter Nameplate

    Incoming acceptance and product identification Proline Promass S 500 PROFINET over Ethernet-APL 4.2.1 Transmitter nameplate Proline 500 – digital Order code: Ser. no.: Ext. ord. cd.: Date: A0029194  3 Example of a transmitter nameplate Name of the transmitter Manufacturer address/certificate holder...
  • Page 17 Proline Promass S 500 PROFINET over Ethernet-APL Incoming acceptance and product identification Proline 500 3 4 5 Order code: Ser. no.: Ext. ord. cd.: Date: A0029192  4 Example of a transmitter nameplate Manufacturer address/certificate holder Name of the transmitter Order code Serial number (Ser.
  • Page 18: Sensor Nameplate

    Incoming acceptance and product identification Proline Promass S 500 PROFINET over Ethernet-APL 4.2.2 Sensor nameplate Order code: Ser. no.: Ext. ord. cd.: Date: A0029199  5 Example of a sensor nameplate Name of the sensor Manufacturer address/certificate holder Order code Serial number (Ser.
  • Page 19: Symbols On The Device

    Proline Promass S 500 PROFINET over Ethernet-APL Incoming acceptance and product identification 4.2.3 Symbols on the device Symbol Meaning WARNING! This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious or fatal injury. Please consult the documentation for the measuring instrument to discover the type of potential danger and measures to avoid it.
  • Page 20: Storage And Transport

    Storage and transport Proline Promass S 500 PROFINET over Ethernet-APL Storage and transport Storage conditions Observe the following notes for storage: ‣ Store in the original packaging to ensure protection from shock. ‣ Do not remove protective covers or protective caps installed on process connections.
  • Page 21: Measuring Devices With Lifting Lugs

    Proline Promass S 500 PROFINET over Ethernet-APL Installation 5.2.2 Measuring devices with lifting lugs CAUTION Special transportation instructions for devices with lifting lugs ‣ Only use the lifting lugs fitted on the device or flanges to transport the device. ‣...
  • Page 22 Installation Proline Promass S 500 PROFINET over Ethernet-APL Installation in down pipes However, the following installation suggestion allows for installation in an open vertical pipeline. Pipe restrictions or the use of an orifice with a smaller cross-section than the nominal diameter prevent the sensor running empty while measurement is in progress.
  • Page 23: Inlet And Outlet Runs

    Proline Promass S 500 PROFINET over Ethernet-APL Installation Orientation Recommendation Horizontal orientation, transmitter at bottom Exception: →  7,  23 A0015590 Horizontal orientation, transmitter at side A0015592 This orientation is recommended to ensure self-draining. Applications with low process temperatures may reduce the ambient temperature. To maintain the minimum ambient temperature for the transmitter, this orientation is recommended.
  • Page 24: Environmental And Process Requirements

    Installation Proline Promass S 500 PROFINET over Ethernet-APL 6.1.2 Environmental and process requirements Ambient temperature range Measuring device • –40 to +60 °C (–40 to +140 °F) • Order code for "Test, certificate", option JP: –50 to +60 °C (–58 to +140 °F) Readability of the local –20 to +60 °C (–4 to +140 °F)
  • Page 25: Special Installation Instructions

    Proline Promass S 500 PROFINET over Ethernet-APL Installation A0034391  8 Thermal insulation with exposed extended neck Heating NOTICE Electronics can overheat due to elevated ambient temperature! ‣ Observe maximum permitted ambient temperature for the transmitter. ‣ Depending on the medium temperature, take the device orientation requirements into account.
  • Page 26: Hygienic Compatibility

    Installation Proline Promass S 500 PROFINET over Ethernet-APL Hygienic compatibility When installing in hygienic applications, please refer to the information in the "Certificates and approvals/hygienic compatibility" section →  303 Securing with mounting clamp in the case of hygiene connections It is not necessary to provide additional support for the sensor for operational performance purposes.
  • Page 27: Weather Protection Cover

    Proline Promass S 500 PROFINET over Ethernet-APL Installation Verification and adjustment cannot be carried out if the following process conditions are present: • Gas pockets Ensure that the system has been sufficiently flushed with the medium. Repeat flushing can help to eliminate gas pockets •...
  • Page 28: Installing The Measuring Instrument

    Installation Proline Promass S 500 PROFINET over Ethernet-APL Installing the measuring instrument 6.2.1 Required tools For transmitter For mounting on a post: • Proline 500 – digital transmitter • Open-ended wrench AF 10 • Torx screwdriver TX 25 • Proline 500 transmitter...
  • Page 29: Mounting The Transmitter Housing: Proline 500 - Digital

    Proline Promass S 500 PROFINET over Ethernet-APL Installation 6.2.4 Mounting the transmitter housing: Proline 500 – digital CAUTION Ambient temperature too high! Danger of electronics overheating and housing deformation. ‣ Do not exceed the permitted maximum ambient temperature. ‣ If operating outdoors: Avoid direct sunlight and exposure to weathering, particularly in warm climatic regions.
  • Page 30: Mounting The Transmitter Housing: Proline 500

    Installation Proline Promass S 500 PROFINET over Ethernet-APL 17 (0.67) 5.8 (0.23) 5.8 (0.23) 149 (5.85) A0029054  12 Engineering unit mm (in) Depends on order code for "Transmitter housing" Order code for "Transmitter housing" • Option A, aluminum, coated: L = 14 mm (0.55 in) •...
  • Page 31: Turning The Transmitter Housing: Proline 500

    Proline Promass S 500 PROFINET over Ethernet-APL Installation 18 (0.71) 10 (0.39) 20 (0.79) 100 (3.94) A0029068  13 Engineering unit mm (in) 1. Drill the holes. 2. Insert wall plugs into the drilled holes. 3. Screw in the fixing screws slightly.
  • Page 32: Turning The Display Module: Proline 500

    Installation Proline Promass S 500 PROFINET over Ethernet-APL 4 mm 7 Nm (5.2 lbf ) A0043150  15 Ex housing 1. Loosen the fixing screws. 2. Turn the housing to the desired position. 3. Tighten the securing screws. 6.2.7 Turning the display module: Proline 500 The display module can be turned to optimize display readability and operability.
  • Page 33: Post-Installation Check

    Proline Promass S 500 PROFINET over Ethernet-APL Installation Post-installation check Is the device undamaged (visual inspection)?  Does the measuring instrument correspond to the measuring point specifications? For example: • Process temperature →  294  • Pressure (refer to the "Pressure-temperature ratings" section of the "Technical Information"...
  • Page 34: Electrical Connection

    Electrical connection Proline Promass S 500 PROFINET over Ethernet-APL Electrical connection WARNING Live parts! Incorrect work performed on the electrical connections can result in an electric shock. ‣ Set up a disconnecting device (switch or power-circuit breaker) to easily disconnect the device from the supply voltage.
  • Page 35 Proline Promass S 500 PROFINET over Ethernet-APL Electrical connection Pulse /frequency /switch output Standard installation cable is sufficient. Relay output Standard installation cable is sufficient. Current input 4 to 20 mA Standard installation cable is sufficient. Status input Standard installation cable is sufficient.
  • Page 36 Electrical connection Proline Promass S 500 PROFINET over Ethernet-APL A: Connecting cable between sensor and transmitter: Proline 500 – digital Standard cable A standard cable with the following specifications can be used as the connecting cable. Design 4 cores (2 pairs); uninsulated stranded CU wires; pair-stranded with common...
  • Page 37 Proline Promass S 500 PROFINET over Ethernet-APL Electrical connection Cross-section Cable length [max.] Termination 2 x 2 x 0.50 mm 50 m (150 ft) 2 x 2 x 0.50 mm (AWG 20) (AWG 20) BN WT YE GN • +, – = 0.5 mm •...
  • Page 38: Terminal Assignment

    Electrical connection Proline Promass S 500 PROFINET over Ethernet-APL C: Connecting cable between sensor and transmitter: Proline 500 Design 6 × 0.38 mm PVC cable with individual shielded cores and common copper shield Conductor resistance ≤ 50 Ω/km (0.015 Ω/ft) Capacitance: core/shield ≤...
  • Page 39: Device Plug Pin Assignment

    Proline Promass S 500 PROFINET over Ethernet-APL Electrical connection 7.2.5 Device plug pin assignment Assignment Coding Plug/socket APL signal - Socket APL signal + Cable shield Not used Metal plug Cable shield housing If a cable shield is used 7.2.6...
  • Page 40: Preparing The Measuring Device

    Electrical connection Proline Promass S 500 PROFINET over Ethernet-APL A0047536  16 Connection example for PROFINET with Ethernet-APL Cable shield Measuring device Local grounding Potential equalization Trunk or TCP Field switch 7.2.7 Preparing the measuring device Carry out the steps in the following order: 1.
  • Page 41: Connecting The Measuring Instrument: Proline 500 - Digital

    Proline Promass S 500 PROFINET over Ethernet-APL Electrical connection Connecting the measuring instrument: Proline 500 - digital NOTICE An incorrect connection compromises electrical safety! ‣ Only properly trained specialist staff may perform electrical connection work. ‣ Observe applicable federal/national installation codes and regulations.
  • Page 42 Electrical connection Proline Promass S 500 PROFINET over Ethernet-APL Connecting the sensor connection housing via terminals For the device version with the order code for "Sensor connection housing": Option A "Aluminum coated" 10 (0.4) 3 mm TX 20 TX 20...
  • Page 43 Proline Promass S 500 PROFINET over Ethernet-APL Electrical connection Connecting the sensor connection housing via terminals For the device version with the order code for "Sensor connection housing": Option B "Stainless" 10 (0.4) 8 mm 22 mm 24 mm A0029613 1.
  • Page 44 Electrical connection Proline Promass S 500 PROFINET over Ethernet-APL Connecting the sensor connection housing via the connector For the device version with the order code for "Sensor connection housing": Option C "Ultra-compact hygienic, stainless" A0029615 1. Connect the protective ground.
  • Page 45 Proline Promass S 500 PROFINET over Ethernet-APL Electrical connection Connecting the connecting cable to the transmitter TX 20 TX 20 10 (0.4) 22 mm 24 mm A0029597 1. Loosen the 4 fixing screws on the housing cover. 2. Open the housing cover.
  • Page 46: Connecting The Transmitter

    Electrical connection Proline Promass S 500 PROFINET over Ethernet-APL 7.3.2 Connecting the transmitter A0028200 Terminal connection for supply voltage Terminal connection for signal transmission, input/output Terminal connection for signal transmission, input/output Terminal connection for connecting cable between sensor and transmitter Terminal connection for signal transmission, input/output;...
  • Page 47 Proline Promass S 500 PROFINET over Ethernet-APL Electrical connection 7. Plug in the RJ45 connector. 8. Firmly tighten the cable glands.  This concludes the connection process. Connecting the supply voltage and additional inputs/outputs TX 20 TX 20 10 (0.4)
  • Page 48 Electrical connection Proline Promass S 500 PROFINET over Ethernet-APL 3 (0.12) A0029598  17 Engineering unit mm (in) 1. Use a flat-blade screwdriver to press down on the slot between the two terminal holes. 2. Remove the cable end from the terminal.
  • Page 49: Integrating The Transmitter Into A Network

    Proline Promass S 500 PROFINET over Ethernet-APL Electrical connection 7.3.3 Integrating the transmitter into a network This section only presents the basic options for integrating the device into a network. For information on the procedure to follow to connect the transmitter correctly →  41.
  • Page 50: Connecting The Measuring Instrument: Proline 500

    Electrical connection Proline Promass S 500 PROFINET over Ethernet-APL Connecting the measuring instrument: Proline 500 NOTICE An incorrect connection compromises electrical safety! ‣ Only properly trained specialist staff may perform electrical connection work. ‣ Observe applicable federal/national installation codes and regulations.
  • Page 51 Proline Promass S 500 PROFINET over Ethernet-APL Electrical connection Connecting the sensor connection housing via terminals For the device version with the order code for "Housing": Option A "Aluminum coated" 10 (0.4) 3 mm 22 mm 24 mm A0029612 1. Loosen the securing clamp of the housing cover.
  • Page 52 Electrical connection Proline Promass S 500 PROFINET over Ethernet-APL Connecting the sensor connection housing via terminals For the device version with the order code for "Housing": Option B "Stainless" 10 (0.4) 8 mm 22 mm 24 mm A0029613 1. Release the securing screw of the housing cover.
  • Page 53 Proline Promass S 500 PROFINET over Ethernet-APL Electrical connection Attaching the connecting cable to the transmitter 10 (0.4) 3 mm 22 mm 24 mm A0029592 1. Loosen the securing clamp of the connection compartment cover. 2. Unscrew the connection compartment cover.
  • Page 54: Connecting The Transmitter

    Electrical connection Proline Promass S 500 PROFINET over Ethernet-APL 7.4.2 Connecting the transmitter A0026781 Terminal connection for supply voltage Terminal connection for signal transmission, input/output Terminal connection for signal transmission, input/output or terminal connection for network connection via service interface (CDI-RJ45)
  • Page 55 Proline Promass S 500 PROFINET over Ethernet-APL Electrical connection 10 (0.4) mm (in) – A0051111 7. Push the cable through the cable entry. To ensure tight sealing, do not remove the sealing ring from the cable entry. 8. Strip the cable and cable ends and connect to terminals 26-27. In the case of stranded cables, also fit ferrules.
  • Page 56 Electrical connection Proline Promass S 500 PROFINET over Ethernet-APL 5. Firmly tighten the cable glands.  This concludes the cable connection process. 6. Close the terminal cover. 7. Fit the display module holder in the electronics compartment. 8. Screw on the connection compartment cover.
  • Page 57: Integrating The Transmitter Into A Network

    Proline Promass S 500 PROFINET over Ethernet-APL Electrical connection 7.4.3 Integrating the transmitter into a network This section only presents the basic options for integrating the device into a network. For information on the procedure to follow to connect the transmitter correctly →  50.
  • Page 58: Special Connection Instructions

    Electrical connection Proline Promass S 500 PROFINET over Ethernet-APL Special connection instructions 7.6.1 Connection examples PROFINET with Ethernet-APL A0047536  19 Connection example for PROFINET with Ethernet-APL Cable shield Measuring device Local grounding Potential equalization Trunk or TCP Field switch Current output 4-20 mA 4...20 mA...
  • Page 59: Pulse/Frequency Output

    Proline Promass S 500 PROFINET over Ethernet-APL Electrical connection 4...20 mA A0028759  21 Connection example for 4-20 mA current output (passive) Automation system with current input (e.g. PLC) Active barrier for power supply (e.g. RN221N) Analog display unit: observe maximum load...
  • Page 60: Relay Output

    Electrical connection Proline Promass S 500 PROFINET over Ethernet-APL Relay output A0028760  24 Connection example for relay output (passive) Automation system with relay input (e.g. PLC) Power supply Transmitter: observe input values →  283 Current input A0028915  25...
  • Page 61: Hardware Settings

    Proline Promass S 500 PROFINET over Ethernet-APL Electrical connection Hardware settings 7.7.1 Setting the device name A measuring point can be quickly identified within a plant on the basis of the tag name. The factory-assigned device name can be changed using the DIP switches or the automation system.
  • Page 62 Electrical connection Proline Promass S 500 PROFINET over Ethernet-APL 1 2 3 4 5 6 7 8 Name of Station A0034497 1. Loosen the 4 fixing screws on the housing cover. 2. Open the housing cover. 3. Fold open the terminal cover.
  • Page 63: Activating The Default Ip Address

    Proline Promass S 500 PROFINET over Ethernet-APL Electrical connection 5. Reconnect the device to the power supply.  The configured device address is used once the device is restarted. Setting the device name via the automation system DIP switches 1-8 must all be set to OFF (factory setting) or all be set to ON to be able to set the device name via the automation system.
  • Page 64: Ensuring The Degree Of Protection

    Electrical connection Proline Promass S 500 PROFINET over Ethernet-APL ‣ Disconnect the device from the power supply. A0034499 1. Depending on the housing version, loosen the securing clamp or fixing screw of the housing cover. 2. Depending on the housing version, unscrew or open the housing cover and disconnect the local display from the main electronics module where necessary .
  • Page 65: Endress+Hauser

    Proline Promass S 500 PROFINET over Ethernet-APL Electrical connection Are the installed cables strain-relieved and securely routed?  Are all cable glands installed, securely tightened and leak-tight? Cable run with "water trap"  →  64? Is the terminal assignment correct ? ...
  • Page 66: Operation Options

    Operation options Proline Promass S 500 PROFINET over Ethernet-APL Operation options Overview of operation options A0046226 Local operation via display module Computer with web browser or with operating tool (e.g FieldCare, DeviceCare, SIMATIC PDM) Field Xpert SMT70 Mobile handheld terminal Automation system (e.g.
  • Page 67: Structure And Function Of The Operating Menu

    Proline Promass S 500 PROFINET over Ethernet-APL Operation options Structure and function of the operating menu 8.2.1 Structure of the operating menu For an overview of the operating menu for experts: see the "Description of Device Parameters" document supplied with the device →  306...
  • Page 68: Operating Philosophy

    Operation options Proline Promass S 500 PROFINET over Ethernet-APL 8.2.2 Operating philosophy The individual parts of the operating menu are assigned to certain user roles (e.g. operator, maintenance etc.). Each user role contains typical tasks within the device life cycle.
  • Page 69: Access To Operating Menu Via Local Display

    Proline Promass S 500 PROFINET over Ethernet-APL Operation options Menu/parameter User role and tasks Content/meaning Expert Function- Tasks that require detailed knowledge of Contains all of the device parameters and allows direct access to these by oriented the function of the device: means of an access code.
  • Page 70: Display Area

    Operation options Proline Promass S 500 PROFINET over Ethernet-APL Display area In the display area, each measured value is prefaced by certain symbol types for further description: Measured variables Symbol Meaning Mass flow • Volume flow • Corrected volume flow •...
  • Page 71: Navigation View

    Proline Promass S 500 PROFINET over Ethernet-APL Operation options 8.3.2 Navigation view In the submenu In the wizard /../Operation 0091-1 /../Select medium Access stat.disp Select medium Operator Liquid Locking status Display A0013993-EN A0013995-EN Navigation view Navigation path to current position...
  • Page 72: Locking Procedure

    Operation options Proline Promass S 500 PROFINET over Ethernet-APL Setup Is displayed: • In the menu next to the "Setup" selection • At the left in the navigation path in the Setup menu Diagnosis Is displayed: • In the menu next to the "Diagnostics" selection •...
  • Page 73: Editing View

    Proline Promass S 500 PROFINET over Ethernet-APL Operation options 8.3.3 Editing view Numeric editor +0.000 Xx – A0034250  28 For entering values in parameters (e.g. limit values) Entry display area Input screen Confirm, delete or reject entry Operating elements...
  • Page 74 Operation options Proline Promass S 500 PROFINET over Ethernet-APL Operating key Meaning Enter key • Pressing the key briefly confirms your selection. • Pressing the key for 2 s confirms your entry. Escape key combination (press keys simultaneously) Close the editing view without accepting a change.
  • Page 75: Operating Elements

    Proline Promass S 500 PROFINET over Ethernet-APL Operation options 8.3.4 Operating elements Operating key Meaning Minus key In menu, submenu Moves the selection bar upwards in a picklist In wizards Goes to previous parameter In the text and numeric editor Move the entry position to the left.
  • Page 76 Operation options Proline Promass S 500 PROFINET over Ethernet-APL Calling up and closing the context menu The user is in the operational display. 1. Press the  and  keys for longer than 3 seconds.  The context menu opens.
  • Page 77: Navigating And Selecting From List

    Proline Promass S 500 PROFINET over Ethernet-APL Operation options 8.3.6 Navigating and selecting from list Different operating elements are used to navigate through the operating menu. The navigation path is displayed on the left in the header. Icons are displayed in front of the individual menus.
  • Page 78: Calling Up Help Text

    Operation options Proline Promass S 500 PROFINET over Ethernet-APL The direct access code consists of a 5-digit number (at maximum) and the channel number, which identifies the channel of a process variable: e.g. 00914-2. In the navigation view, this appears on the right-hand side in the header of the selected parameter.
  • Page 79: User Roles And Related Access Authorization

    Proline Promass S 500 PROFINET over Ethernet-APL Operation options Ent. access code Invalid or out of range input value Min:0 Max:9999 A0014049-EN For a description of the editing view - consisting of the text editor and numeric editor - with symbols →  73, for a description of the operating elements →  75 8.3.10...
  • Page 80: Enabling And Disabling The Keypad Lock

    Operation options Proline Promass S 500 PROFINET over Ethernet-APL 2. Enter the access code.  The -symbol in front of the parameters disappears; all previously write- protected parameters are now re-enabled. 8.3.12 Enabling and disabling the keypad lock The keypad lock makes it possible to block access to the entire operating menu via local operation.
  • Page 81: Requirements

    Proline Promass S 500 PROFINET over Ethernet-APL Operation options 8.4.2 Requirements Computer hardware Hardware Interface CDI-RJ45 WLAN Interface The computer must have a RJ45 The operating unit must have a WLAN interface. interface. Connection Standard Ethernet cable Connection via Wireless LAN.
  • Page 82: Connecting The Device

    Operation options Proline Promass S 500 PROFINET over Ethernet-APL Measuring device: Via CDI-RJ45 service interface Device CDI-RJ45 service interface Measuring device The measuring device has an RJ45 interface. Web server Web server must be enabled; factory setting: ON  For information on enabling the Web server →  86...
  • Page 83 Proline Promass S 500 PROFINET over Ethernet-APL Operation options 3. Connect the computer to the RJ45 plug via the standard Ethernet cable →  88. 4. If a 2nd network card is not used, close all the applications on the notebook.
  • Page 84: Logging On

    Operation options Proline Promass S 500 PROFINET over Ethernet-APL 2. Enter the IP address of the web server in the address line of the web browser: 192.168.1.212  The login page appears. 2 3 4 Device name: Device tag: Signal Status:...
  • Page 85: User Interface

    Proline Promass S 500 PROFINET over Ethernet-APL Operation options 8.4.5 User interface A0029418 Function row Local display language Navigation area Header The following information appears in the header: • Device name • Device tag • Device status with status signal →  195 •...
  • Page 86: Disabling The Web Server

    Operation options Proline Promass S 500 PROFINET over Ethernet-APL Working area Depending on the selected function and the related submenus, various actions can be performed in this area: • Configuring parameters • Reading measured values • Calling up help text •...
  • Page 87: Access To The Operating Menu Via The Operating Tool

    Proline Promass S 500 PROFINET over Ethernet-APL Operation options 3. If no longer needed: Reset the modified properties of the Internet protocol (TCP/IP) →  82. If communication with the web server was established via the default IP address 192.168.1.212, DIP switch no. 10 must be reset (from ON → OFF). Afterwards, the IP address of the device is active again for network communication.
  • Page 88: Service Interface

    Operation options Proline Promass S 500 PROFINET over Ethernet-APL Service interface Via service interface (CDI-RJ45) A point-to-point connection can be established to configure the device onsite. With the housing open, the connection is established directly via the service interface (CDI-RJ45) of the device.
  • Page 89 Proline Promass S 500 PROFINET over Ethernet-APL Operation options A0034569 Transmitter with integrated WLAN antenna Transmitter with external WLAN antenna LED lit constantly: WLAN reception is enabled on measuring device LED flashing: WLAN connection established between operating unit and measuring device Computer with WLAN interface and web browser (e.g.
  • Page 90: Fieldcare

    FieldCare Function range FDT-based (Field Device Technology) plant asset management tool from Endress+Hauser. It can configure all smart field units in a system and helps you manage them. By using the status information, it is also a simple but effective way of checking their status and condition.
  • Page 91: Devicecare

    Proline Promass S 500 PROFINET over Ethernet-APL Operation options 2. In the network: Add a device.  The Add device window opens. 3. Select the CDI Communication TCP/IP option from the list and press OK to confirm. 4. Right-click CDI Communication TCP/IP and select the Add device option in the context menu that opens.
  • Page 92: Simatic Pdm

    Operation options Proline Promass S 500 PROFINET over Ethernet-APL The fastest way to configure Endress+Hauser field devices is with the dedicated "DeviceCare" tool. Together with the device type managers (DTMs) it presents a convenient, comprehensive solution. Innovation brochure IN01047S Source for device description files →  93 8.5.4...
  • Page 93: System Integration

    Proline Promass S 500 PROFINET over Ethernet-APL System integration System integration Overview of device description files 9.1.1 Current version data for the device Firmware version 01.00.zz • On the title page of the manual • On the transmitter nameplate • Firmware version Diagnostics →...
  • Page 94: File Name Of The Manufacturer-Specific Device Master File (Gsd)

    System integration Proline Promass S 500 PROFINET over Ethernet-APL 9.2.1 File name of the manufacturer-specific device master file (GSD) Example of the name of a device master file: GSDML-V2.43-EH-PROMASS_300_500_APL_yyyymmdd.xml GSDML Description language V2.43 Version of the PROFINET specification Endress+Hauser PROMASS...
  • Page 95: Cyclic Data Transmission

    Proline Promass S 500 PROFINET over Ethernet-APL System integration Cyclic data transmission 9.3.1 Overview of the modules The following graphic shows which modules are available to the device for cyclic data transfer. Cyclic data transfer is performed with an automation system.
  • Page 96: Description Of The Modules

    System integration Proline Promass S 500 PROFINET over Ethernet-APL 9.3.2 Description of the modules The data structure is described from the perspective of the automation system: • Input data: Are sent from the measuring device to the automation system. • Output data: Are sent from the automation system to the measuring device.
  • Page 97 Proline Promass S 500 PROFINET over Ethernet-APL System integration Slot Sub-slot Input variables Temperature • Mass flow • Volume flow • Corrected volume flow • Density • Reference density • Temperature • Electronics temperature • Oscillation frequency • Frequency fluctuation •...
  • Page 98 System integration Proline Promass S 500 PROFINET over Ethernet-APL Data structure Output data of Analog Output Byte 1 Byte 2 Byte 3 Byte 4 Byte 5 Measured value: floating point number (IEEE 754) Status Status coding →  105 Application-specific Input module Transmit compensation values from the measuring device to the automation system.
  • Page 99 Proline Promass S 500 PROFINET over Ethernet-APL System integration Slot Sub-slot Device function Status (meaning) Verification carried out successfully. Verification was not performed. Reserved Selection: device function, binary input, slot 81 Slot Sub-slot Device function Status (meaning) Partially filled pipe detection •...
  • Page 100 System integration Proline Promass S 500 PROFINET over Ethernet-APL The Mass Totalizer Control module cyclically transmits a selected totalizer value, including the status, from the measuring device to the automation system. The totalizer value is depicted in the first four bytes in the form of a floating point number as per the IEEE 754 standard.
  • Page 101 Proline Promass S 500 PROFINET over Ethernet-APL System integration Selection: input variable Slot Sub-slot Input variable • Mass flow • Volume flow • Corrected volume flow • Target mass flow • Carrier mass flow • Target volume flow • Carrier volume flow •...
  • Page 102 System integration Proline Promass S 500 PROFINET over Ethernet-APL Selection: input variable Slot Sub-slot Input variable • Mass flow • Volume flow • Corrected volume flow • Target mass flow • Carrier mass flow • Target volume flow • Carrier volume flow •...
  • Page 103: Binary Output Module

    Proline Promass S 500 PROFINET over Ethernet-APL System integration IEEE 754 standard. The fifth byte contains standardized status information pertaining to the compensation value. Assigned compensation values The selection is made via: Expert → Sensor → External compensation Slot Sub-slot...
  • Page 104: Data Structure

    System integration Proline Promass S 500 PROFINET over Ethernet-APL Selection: device function, binary output, slot 210 Slot Sub-slot Device function Status (meaning) Start the verification. A change of status from 0 to 1 starts the Heartbeat Verification Reserved Reserved Reserved...
  • Page 105: Status Coding

    Proline Promass S 500 PROFINET over Ethernet-APL System integration Data structure Concentration output data Byte 1 Control variable Liquid type Enum code Sucrose in water Glucose in water Fructose in water Invert sugar in water Corn syrup HFCS42 Corn syrup HFCS55...
  • Page 106: Factory Setting

    System integration Proline Promass S 500 PROFINET over Ethernet-APL Status Coding (hex) Meaning UNCERTAIN - 0x68 to 0x6B Signs of wear and tear have been detected on the Maintenance demanded measuring instrument. Short-term maintenance is needed to ensure that the measuring instrument remains operational.
  • Page 107: System Redundancy S2

    Proline Promass S 500 PROFINET over Ethernet-APL System integration System redundancy S2 A redundant layout with two automation systems is necessary for processes that are in continuous operation. If one system fails the second system guarantees continued, uninterrupted operation. The measuring device supports S2 system redundancy and can communicate with both automation systems simultaneously.
  • Page 108: Commissioning

    Commissioning Proline Promass S 500 PROFINET over Ethernet-APL Commissioning 10.1 Post-mounting and post-connection check Before commissioning the device: ‣ Make sure that the post-installation and post-connection checks have been performed successfully. • Checklist for "Post-installation" check→  33 • Checklist for "Post-connection" check→  64 10.2...
  • Page 109: Configuring The Measuring Instrument

    Proline Promass S 500 PROFINET over Ethernet-APL Commissioning 10.5 Configuring the measuring instrument The Setup menu with its guided wizards contains all the parameters needed for standard operation. X X X X X X X 20.50 Main menu Display language English Display/operat.
  • Page 110: Defining The Tag Name

    Commissioning Proline Promass S 500 PROFINET over Ethernet-APL ‣ Relay output 1 to n →  133 ‣ Display →  136 ‣ →  140 Low flow cut off ‣ Partially filled pipe detection →  141 ‣ Advanced setup →...
  • Page 111: Parameter Overview With Brief Description

    Proline Promass S 500 PROFINET over Ethernet-APL Commissioning "APL port" submenu Navigation "Setup" menu → Communication → APL port ‣ APL port →  111 IP address (7263) Subnet mask (7265) →  111 Default gateway (7264) →  111 MAC address (7262) →...
  • Page 112: Setting The System Units

    Commissioning Proline Promass S 500 PROFINET over Ethernet-APL Parameter overview with brief description Parameter Description User entry / User interface Factory setting IP address Enter the IP address of the measuring 4 octet: 0 to 255 (in the 192.168.1.212 device. particular octet) Subnet mask Displays the subnet mask.
  • Page 113 Proline Promass S 500 PROFINET over Ethernet-APL Commissioning Volume flow unit →  113 Volume unit →  113 →  113 Corrected volume flow unit Corrected volume unit →  113 Density unit →  113 Reference density unit →...
  • Page 114 Commissioning Proline Promass S 500 PROFINET over Ethernet-APL Parameter Description Selection Factory setting Reference density unit Select reference density unit. Unit choose list Country-specific • kg/Nl • lb/Sft³ Density 2 unit Select second density unit. Unit choose list Country-specific: • kg/l •...
  • Page 115: Selecting And Setting The Medium

    Proline Promass S 500 PROFINET over Ethernet-APL Commissioning 10.5.4 Selecting and setting the medium The Select medium wizard submenu contains parameters that must be configured in order to select and set the medium. Navigation "Setup" menu → Medium selection ‣...
  • Page 116: Configuration Of The Analog Inputs

    Commissioning Proline Promass S 500 PROFINET over Ethernet-APL 10.5.5 Configuration of the Analog Inputs The Analog inputs submenu guides the user systematically to the individual Analog input 1 to n submenu. From here you get to the parameters of the individual analog input.
  • Page 117 Proline Promass S 500 PROFINET over Ethernet-APL Commissioning Parameter overview with brief description Parameter Description User interface / User entry Factory setting Parent class 0 to 255 Endress+Hauser...
  • Page 118 Commissioning Proline Promass S 500 PROFINET over Ethernet-APL Parameter Description User interface / User entry Factory setting Assign process variable Select a process variable. • Mass flow Mass flow • Volume flow • Density • Temperature • Carrier pipe temperature • Electronics temperature •...
  • Page 119: Displaying The I/O Configuration

    Proline Promass S 500 PROFINET over Ethernet-APL Commissioning Parameter Description User interface / User entry Factory setting Damping Enter time constant for input damping (PT1 Positive floating-point number 1.0 s element). Damping reduces the effect of fluctuations in the measured value on the output signal.
  • Page 120: Configuring The Current Input

    Commissioning Proline Promass S 500 PROFINET over Ethernet-APL 10.5.7 Configuring the current input The "Current input" wizard guides the user systematically through all the parameters that have to be set for configuring the current input. Navigation "Setup" menu → Current input ‣...
  • Page 121: Configuring The Status Input

    Proline Promass S 500 PROFINET over Ethernet-APL Commissioning Parameter Prerequisite Description User interface /  Factory setting Selection / User entry Failure mode – Define input behavior in alarm • Alarm Alarm condition. • Last valid value • Defined value Failure value In the Failure mode...
  • Page 122: Configuring The Current Output

    Commissioning Proline Promass S 500 PROFINET over Ethernet-APL Parameter Description Selection / User interface /  Factory setting User entry Active level Define input signal level at which the • High High assigned function is triggered. • Low Response time status input Define the minimum amount of time the...
  • Page 123 Proline Promass S 500 PROFINET over Ethernet-APL Commissioning Parameter Prerequisite Description User interface /  Factory setting Selection / User entry Process variable current output – Select the process variable for • Off Mass flow the current output. • Mass flow • Volume flow •...
  • Page 124 Commissioning Proline Promass S 500 PROFINET over Ethernet-APL Parameter Prerequisite Description User interface /  Factory setting Selection / User entry Current range output – Select current range for • 4...20 mA NE Depends on country: process value output and (3.8...20.5 mA) • 4...20 mA NE upper/lower level for alarm •...
  • Page 125: 10.5.10 Configuring The Pulse/Frequency/Switch Output

    Proline Promass S 500 PROFINET over Ethernet-APL Commissioning 10.5.10 Configuring the pulse/frequency/switch output The Pulse/frequency/switch output wizard guides you systematically through all the parameters that can be set for configuring the selected output type. Navigation "Setup" menu → Advanced setup → Pulse/frequency/switch output ‣...
  • Page 126 Commissioning Proline Promass S 500 PROFINET over Ethernet-APL Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting interface / User entry Operating mode – Define the output as a pulse, • Pulse Pulse frequency or switch output. • Frequency • Switch Terminal number –...
  • Page 127 Proline Promass S 500 PROFINET over Ethernet-APL Commissioning Parameter Prerequisite Description Selection / User Factory setting interface / User entry Failure mode The Pulse option is selected in Define output behavior in • Actual value No pulses the Operating mode alarm condition. • No pulses parameter (→...
  • Page 128 Commissioning Proline Promass S 500 PROFINET over Ethernet-APL Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting interface / User entry Operating mode – Define the output as a pulse, • Pulse Pulse frequency or switch output. • Frequency • Switch Terminal number –...
  • Page 129 Proline Promass S 500 PROFINET over Ethernet-APL Commissioning Parameter Prerequisite Description Selection / User Factory setting interface / User entry Assign frequency output The Frequency option is Select process variable for • Off selected in Operating mode frequency output. • Mass flow parameter (→  125).
  • Page 130 Commissioning Proline Promass S 500 PROFINET over Ethernet-APL Parameter Prerequisite Description Selection / User Factory setting interface / User entry Maximum frequency value The Frequency option is Enter maximum frequency. 0.0 to 10 000.0 Hz 10 000.0 Hz selected in the Operating mode parameter (→  125)
  • Page 131 Proline Promass S 500 PROFINET over Ethernet-APL Commissioning Configuring the switch output Navigation "Setup" menu → Pulse/frequency/switch output ‣ Pulse/frequency/switch output 1 to n Operating mode →  131 Terminal number →  131 Signal mode →  131 Switch output function →...
  • Page 132 Commissioning Proline Promass S 500 PROFINET over Ethernet-APL Parameter Prerequisite Description Selection / User Factory setting interface / User entry Switch output function The Switch option is selected Select function for switch • Off in the Operating mode output. • On parameter. • Diagnostic behavior •...
  • Page 133: 10.5.11 Configuring The Relay Output

    Proline Promass S 500 PROFINET over Ethernet-APL Commissioning Parameter Prerequisite Description Selection / User Factory setting interface / User entry Switch-on value • The Switch option is Enter measured value for the Signed floating-point Depends on country: selected in the Operating switch-on point. number •...
  • Page 134 Commissioning Proline Promass S 500 PROFINET over Ethernet-APL Switch-off value →  135 Switch-off delay →  135 →  135 Switch-on value Switch-on delay →  135 Failure mode →  135 Switch state →  136 →  136...
  • Page 135 Proline Promass S 500 PROFINET over Ethernet-APL Commissioning Parameter Prerequisite Description User interface /  Factory setting Selection / User entry Assign limit The Limit option is selected in Select process variable for limit • Mass flow Mass flow Relay output function function. • Volume flow parameter.
  • Page 136: 10.5.12 Configuring The Local Display

    Commissioning Proline Promass S 500 PROFINET over Ethernet-APL Parameter Prerequisite Description User interface /  Factory setting Selection / User entry Switch state – Shows the current relay switch • Open – status. • Closed Powerless relay status – Select quiescent state for relay. • Open Open •...
  • Page 137 Proline Promass S 500 PROFINET over Ethernet-APL Commissioning Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Format display A local display is provided. Select how measured values • 1 value, max. size 1 value, max. size are shown on the display.
  • Page 138 Commissioning Proline Promass S 500 PROFINET over Ethernet-APL Parameter Prerequisite Description Selection / User Factory setting entry Value 1 display A local display is provided. Select the measured value that • Mass flow Mass flow is shown on the local display. • Volume flow •...
  • Page 139 Proline Promass S 500 PROFINET over Ethernet-APL Commissioning Parameter Prerequisite Description Selection / User Factory setting entry Value 2 display A local display is provided. Select the measured value that For the picklist, see None is shown on the local display. Value 1 display parameter (→...
  • Page 140: 10.5.13 Configuring The Low Flow Cut Off

    Commissioning Proline Promass S 500 PROFINET over Ethernet-APL 10.5.13 Configuring the low flow cut off The Low flow cut off wizard systematically guides the user through all the parameters that must be set to configure low flow cut off. Navigation "Setup"...
  • Page 141: 10.5.14 Configuring Partially Filled Pipe Detection

    Proline Promass S 500 PROFINET over Ethernet-APL Commissioning 10.5.14 Configuring partially filled pipe detection The Partial filled pipe detection wizard guides you systematically through all parameters that have to be set for configuring the monitoring of the pipe filling. Navigation "Setup"...
  • Page 142: Advanced Settings

    Commissioning Proline Promass S 500 PROFINET over Ethernet-APL 10.6 Advanced settings The Advanced setup submenu with its submenus contains parameters for specific settings. Navigation to the "Advanced setup" submenu X X X X X X X 20.50 0104-1 Main menu...
  • Page 143: Using The Parameter To Enter The Access Code

    Proline Promass S 500 PROFINET over Ethernet-APL Commissioning ‣ Totalizer 1 to n →  151 ‣ Display →  153 ‣ →  159 WLAN settings ‣ Viscosity →  161 ‣ Concentration →  161 ‣ Petroleum →  161 ‣...
  • Page 144 Commissioning Proline Promass S 500 PROFINET over Ethernet-APL "Corrected volume flow calculation" submenu Navigation "Setup" menu → Advanced setup → Calculated values → Corrected volume flow calculation ‣ Corrected volume flow calculation →  144 Select reference density (1812) External reference density (6198) →...
  • Page 145: Carrying Out A Sensor Adjustment

    Proline Promass S 500 PROFINET over Ethernet-APL Commissioning Parameter Prerequisite Description Selection / User Factory setting interface / User entry Linear expansion coefficient The Calculated reference Enter linear, medium-specific Signed floating-point 0.0 1/K density option is selected in expansion coefficient for number the Corrected volume flow...
  • Page 146 Commissioning Proline Promass S 500 PROFINET over Ethernet-APL Performing density adjustment Note the following before performing the adjustment: • A density adjustment only makes sense if there is little variation in the operating conditions and the density adjustment is performed under the operating conditions.
  • Page 147 Proline Promass S 500 PROFINET over Ethernet-APL Commissioning 4. Select the Measure density 1 option and confirm.  In the Execute density adjustment parameter the following options are now available: Measure density 2 Restore original 5. Select the Measure density 2 option and confirm.
  • Page 148 Commissioning Proline Promass S 500 PROFINET over Ethernet-APL Parameter Prerequisite Description Selection / User Factory setting entry / User interface Density setpoint 2 In the Density adjustment Enter density for the second The entry depends on 1 kg/l mode parameter, the 2 point reference media.
  • Page 149 Proline Promass S 500 PROFINET over Ethernet-APL Commissioning Zero point verification The zero point can be verified with the Zero verification wizard. Navigation "Setup" menu → Advanced setup → Sensor adjustment → Zero verification ‣ Zero verification Process conditions →  149 →...
  • Page 150 Commissioning Proline Promass S 500 PROFINET over Ethernet-APL Parameter Description Selection / User interface Factory setting Root cause Shows the diagnostic and remedy. • Zero point too high. Ensure – no-flow. • Zero point is unstable. Ensure no-flow. • Fluctuation high. Avoid 2- phase medium.
  • Page 151: Configuring The Totalizer

    Proline Promass S 500 PROFINET over Ethernet-APL Commissioning Parameter overview with brief description Parameter Description Selection / User interface Factory setting Process conditions Ensure process conditions as follows. • Tubes are completely filled – • Process operational pressure applied • No-flow conditions (closed valves) •...
  • Page 152 Commissioning Proline Promass S 500 PROFINET over Ethernet-APL Totalizer 1 to n control (11101–1 to n) →  152 Totalizer 1 to n failure behavior →  152 (11103–1 to n) Parameter overview with brief description Parameter Description Selection Factory setting Assign process variable 1 to n Select process variable for totalizer.
  • Page 153: Carrying Out Additional Display Configurations

    Proline Promass S 500 PROFINET over Ethernet-APL Commissioning 10.6.5 Carrying out additional display configurations In the Display submenu you can set all the parameters associated with the configuration of the local display. Navigation "Setup" menu → Advanced setup → Display ‣...
  • Page 154 Commissioning Proline Promass S 500 PROFINET over Ethernet-APL 0% bargraph value 7 →  158 100% bargraph value 7 →  158 →  158 Decimal places 7 Value 8 display →  158 Decimal places 8 →  158 Display language →...
  • Page 155 Proline Promass S 500 PROFINET over Ethernet-APL Commissioning Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Format display A local display is provided. Select how measured values • 1 value, max. size 1 value, max. size are shown on the display.
  • Page 156 Commissioning Proline Promass S 500 PROFINET over Ethernet-APL Parameter Prerequisite Description Selection / User Factory setting entry Value 1 display A local display is provided. Select the measured value that • Mass flow Mass flow is shown on the local display. • Volume flow •...
  • Page 157 Proline Promass S 500 PROFINET over Ethernet-APL Commissioning Parameter Prerequisite Description Selection / User Factory setting entry Decimal places 1 A measured value is specified Select the number of decimal • x x.xx in the Value 1 display places for the display value.
  • Page 158 Commissioning Proline Promass S 500 PROFINET over Ethernet-APL Parameter Prerequisite Description Selection / User Factory setting entry Decimal places 5 A measured value is specified Select the number of decimal • x x.xx in the Value 5 display places for the display value.
  • Page 159: Wlan Configuration

    Proline Promass S 500 PROFINET over Ethernet-APL Commissioning Parameter Prerequisite Description Selection / User Factory setting entry Display language A local display is provided. Set display language. • English English • Deutsch (alternatively, the • Français ordered language is • Español preset in the device) •...
  • Page 160 Commissioning Proline Promass S 500 PROFINET over Ethernet-APL SSID name →  160 Network security →  160 →  160 Security identification User name →  160 WLAN password →  160 WLAN IP address →  161 →  161...
  • Page 161: Viscosity Application Package

    Proline Promass S 500 PROFINET over Ethernet-APL Commissioning Parameter Prerequisite Description Selection / User Factory setting entry / User interface WLAN IP address – Enter IP address of the WLAN 4 octet: 0 to 255 (in 192.168.1.212 interface of the device. the particular octet) WLAN MAC address –...
  • Page 162: Heartbeat Technology Application Package

    Commissioning Proline Promass S 500 PROFINET over Ethernet-APL 10.6.10 Heartbeat Technology application package For detailed information on the parameter descriptions of the application packages, see the Special Documentation for the device. →  307 Navigation "Setup" menu → Advanced setup → Heartbeat setup 10.6.11 Configuration management...
  • Page 163: Function Scope Of The "Configuration Management" Parameter

    Proline Promass S 500 PROFINET over Ethernet-APL Commissioning Function scope of the "Configuration management" parameter Options Description Cancel No action is executed and the user exits the parameter. Execute backup A backup copy of the current device configuration is saved from the HistoROM backup to the memory of the device.
  • Page 164: Simulation

    Commissioning Proline Promass S 500 PROFINET over Ethernet-APL Parameter overview with brief description Parameter Description User entry Define access code Restrict write-access to parameters to protect the configuration Max. 16-digit character string comprising of the device against unintentional changes. numbers, letters and special characters Confirm access code Confirm the entered access code.
  • Page 165 Proline Promass S 500 PROFINET over Ethernet-APL Commissioning or closed-control loops). The simulation can be performed without a real measurement (no flow of medium through the device). Navigation "Diagnostics" menu → Simulation ‣ Simulation Assign simulation process variable →  166 Process variable value →...
  • Page 166 Commissioning Proline Promass S 500 PROFINET over Ethernet-APL Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Assign simulation process variable – Select a process variable for • Off the simulation process that is • Mass flow activated.
  • Page 167: Protecting Settings From Unauthorized Access

    Proline Promass S 500 PROFINET over Ethernet-APL Commissioning Parameter Prerequisite Description Selection / User Factory setting entry Device alarm simulation – Switch the device alarm on and • Off off. • On Diagnostic event category – Select a diagnostic event • Sensor Process category.
  • Page 168 Commissioning Proline Promass S 500 PROFINET over Ethernet-APL 3. Enter the access code again in the Confirm access code parameter (→  164) to confirm.  The  symbol appears in front of all write-protected parameters. • Disabling parameter write protection via access code →  79.
  • Page 169: Write Protection Via Write Protection Switch

    Commissioning Via Web browser, FieldCare, DeviceCare (via CDI-RJ45 service interface), fieldbus You can only obtain a reset code from your local Endress+Hauser service organization. The code must be calculated explicitly for every device. 1. Note down the serial number of the device.
  • Page 170 Commissioning Proline Promass S 500 PROFINET over Ethernet-APL 4. Enable or disable write protection: Setting the write protection (WP) switch on the main electronics module to the ON position enables hardware write protection/setting to OFF (factory setting) disables hardware write protection.
  • Page 171: Operation

    Proline Promass S 500 PROFINET over Ethernet-APL Operation Operation 11.1 Reading off the device locking status Device active write protection: Locking status parameter Operation → Locking status Function scope of the "Locking status" parameter Options Description None The access authorization displayed in the Access status parameter applies →...
  • Page 172: Measured Variables" Submenu

    Operation Proline Promass S 500 PROFINET over Ethernet-APL 11.4.1 "Measured variables" submenu The Measured variables submenu contains all the parameters needed to display the current measured values for each process variable. Navigation "Diagnostics" menu → Measured values → Measured variables ‣...
  • Page 173 Proline Promass S 500 PROFINET over Ethernet-APL Operation Parameter Prerequisite Description User interface Corrected volume flow – Displays the corrected volume flow that Signed floating-point is currently calculated. number Dependency The unit is taken from: Corrected volume flow unit parameter (→...
  • Page 174: Totalizer

    Operation Proline Promass S 500 PROFINET over Ethernet-APL Parameter Prerequisite Description User interface Carrier corrected volume flow With the following conditions: Displays the corrected volume flow Signed floating-point • Order code for "Application package", currently measured for the carrier fluid.
  • Page 175: Input Values" Submenu

    Proline Promass S 500 PROFINET over Ethernet-APL Operation Parameter overview with brief description Parameter Description Selection / User interface Factory setting Assign process variable 1 to n Select process variable for totalizer. • Mass flow Mass flow • Volume flow • Corrected volume flow •...
  • Page 176: Output Values

    Operation Proline Promass S 500 PROFINET over Ethernet-APL Navigation "Diagnostics" menu → Measured values → Input values → Current input 1 to n ‣ Current input 1 to n Measured values 1 to n →  176 Measured current 1 to n →...
  • Page 177 Proline Promass S 500 PROFINET over Ethernet-APL Operation ‣ Pulse/frequency/switch output →  177 1 to n ‣ Relay output 1 to n →  178 Output values of current output The Value current output submenu contains all the parameters needed to display the current measured values for every current output.
  • Page 178: Adapting The Measuring Device To The Process Conditions

    Operation Proline Promass S 500 PROFINET over Ethernet-APL Parameter overview with brief description Parameter Prerequisite Description User interface Output frequency In the Operating mode parameter, the Displays the value currently measured 0.0 to 12 500.0 Hz Frequency option is selected.
  • Page 179: Function Scope Of "Control Totalizer" Parameter

    Proline Promass S 500 PROFINET over Ethernet-APL Operation Navigation "Operation" menu → Totalizer handling ‣ Totalizer handling Totalizer 1 to n control (11101–1 to n) →  179 Preset value 1 to n (11108–1 to n) →  179 →  179...
  • Page 180: Displaying The Measured Value History

    Operation Proline Promass S 500 PROFINET over Ethernet-APL 11.7 Displaying the measured value history The Extended HistoROM application package must be enabled in the device (order option) for the Data logging submenu to appear. This contains all the parameters for the measured value history.
  • Page 181 Proline Promass S 500 PROFINET over Ethernet-APL Operation Data logging status →  183 →  183 Entire logging duration Endress+Hauser...
  • Page 182 Operation Proline Promass S 500 PROFINET over Ethernet-APL Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry / User interface Assign channel 1 The Extended HistoROM Assign process variable to • Off application package is logging channel. • Mass flow available.
  • Page 183: Gas Fraction Handler

    Proline Promass S 500 PROFINET over Ethernet-APL Operation Parameter Prerequisite Description Selection / User Factory setting entry / User interface Assign channel 2 The Extended HistoROM Assign a process variable to For the picklist, see application package is logging channel. Assign channel 1 available.
  • Page 184: Measurement Mode" Submenu

    Operation Proline Promass S 500 PROFINET over Ethernet-APL In the case of two-phase media, the Gas Fraction Handler stabilizes the output values and enables better readability for operators and easier interpretation by the distributed control system. The level of smoothing is adjusted according to the severity of the disturbances introduced by the second phase.
  • Page 185 Proline Promass S 500 PROFINET over Ethernet-APL Operation Parameter overview with brief description Parameter Prerequisite Description User interface / User Factory setting entry Inhomogeneous medium index – Shows the degree of Signed floating-point – inhomogeneity of the medium. number Cut off inhomogeneous wet gas –...
  • Page 186: Diagnostics And Troubleshooting

    Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting 12.1 General troubleshooting For local display Error Possible causes Remedial action Local display is dark, but signal output is within The cable of the display module is not plugged in...
  • Page 187 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting For access Fault Possible causes Remedial action Write access to parameters is not possible. Hardware write protection is enabled. Set the write protection switch on the main electronics module to the OFF position →...
  • Page 188: Diagnostic Information Via Light Emitting Diodes

    Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL For system integration Error Possible causes Remedy The PROFINET device name is not A device name containing one or Specify a correct device name displayed correctly and contains more underscores has been (without underscores) via the coding.
  • Page 189 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Color Meaning Flashing red A diagnostic event with "Warning" diagnostic behavior has occurred. A diagnostic event with "Alarm" diagnostic behavior has occurred. Flashing red/green The device restarts/self-test. Flashing/network status Green Cyclic data exchange is active.
  • Page 190: Sensor Connection Housing

    Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Color Meaning A diagnostic event with "Alarm" diagnostic behavior has occurred. Flashing red/green The device restarts/self-test. Flashing/network status Green Cyclic data exchange is active. Flashing green Following request from automation system:...
  • Page 191 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Color Meaning Supply voltage Green Supply voltage is ok. Supply voltage is off or too low. Endress+Hauser...
  • Page 192: Diagnostic Information On Local Display

    Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL 12.3 Diagnostic information on local display 12.3.1 Diagnostic message Faults detected by the self-monitoring system of the measuring device are displayed as a diagnostic message in alternation with the operational display.
  • Page 193: Diagnostic Information

    Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic behavior Symbol Meaning Alarm • Measurement is interrupted. • Signal outputs and totalizers assume the defined alarm condition. • A diagnostic message is generated. Warning • Measurement is resumed.
  • Page 194: Calling Up Remedial Measures

    Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL 12.3.2 Calling up remedial measures X X X X X X X X X X X X X X 20.50 S801 Supply voltage Menu Diagnostic list Diagnostics 1 S801 Supply voltage...
  • Page 195: Calling Up Remedy Information

    Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting A0031056 Status area with status signal Diagnostic information Remedial measures with service ID In addition, diagnostic events which have occurred can be shown in the Diagnostics menu: • Via parameter →  264 •...
  • Page 196: Calling Up Remedy Information

    Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Xxxxxx/…/…/ Device name: Xxxxxxx Mass flow: kg/h 12.34 Device tag: Xxxxxxx Volume flow: 12.34 m /h ³ Status signal: Function check (C) Xxxxxx Diagnostics 1: C485 Simu... Remedy information: Deactivate...
  • Page 197: Adapting The Diagnostic Information

    Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting 12.6 Adapting the diagnostic information 12.6.1 Adapting the diagnostic behavior Each item of diagnostic information is assigned a specific diagnostic behavior at the factory. The user can change this assignment for specific diagnostic information in the Diagnostic behavior submenu.
  • Page 198: Overview Of Diagnostic Information

    Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Supported status information Status Coding (hex) BAD - Maintenance alarm 0x24 to 0x27 BAD - Process related 0x28 to 0x2B BAD - Function check 0x3C to 0x3F UNCERTAIN - Initial value...
  • Page 199: Diagnostic Of Sensor

    Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting 12.7.1 Diagnostic of sensor Diagnostic information Remedy instructions Short text 002 Sensor unknown 1. Check if the correct sensor is mounted 2. Check if the 2-D matrix code on the sensor is undamaged...
  • Page 200 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 022 Temperature sensor defective 1. If available: Check connection cable between sensor and transmitter 2. Check or replace sensor electronic module (ISEM) Measured variable status 3.
  • Page 201 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 046 Sensor limit exceeded 1. Check process conditions 2. Check sensor Measured variable status [from the factory] Quality Good Quality substatus Coding (hex) 0x80 to 0x83...
  • Page 202 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 062 Sensor connection faulty 1. If available: Check connection cable between sensor and transmitter 2. Check or replace sensor electronic module (ISEM) Measured variable status 3.
  • Page 203 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 063 Exciter current faulty 1. If available: Check connection cable between sensor and transmitter 2. Check or replace sensor electronic module (ISEM) Measured variable status 3.
  • Page 204 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 082 Data storage inconsistent Check module connections Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83 Status signal Diagnostic behavior Alarm Influenced measured variables •...
  • Page 205 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 083 Memory content inconsistent 1. Restart device 2. Restore S-DAT data Measured variable status 3. Replace S-DAT Quality Good Quality substatus Coding (hex) 0x80 to 0x83...
  • Page 206 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 119 Sensor initialization active Sensor initialization in progress, please wait Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83 Status signal...
  • Page 207 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 140 Sensor signal asymmetrical 1. If available: Check connection cable between sensor and transmitter 2. Check or replace sensor electronic module (ISEM) Measured variable status [from the factory] 3.
  • Page 208 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 141 Zero adjustment failed 1. Check process conditions 2. Repeat commissioning procedure Measured variable status 3. Check sensor Quality Good Quality substatus Coding (hex)
  • Page 209 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 142 Sensor index coil asymmetry too high Check sensor Measured variable status [from the factory] Quality Good Quality substatus Coding (hex) 0x80 to 0x83...
  • Page 210 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 144 Measurement error too high 1. Check process conditions 2. Check or change sensor Measured variable status [from the factory] Quality Good Quality substatus...
  • Page 211: Diagnostic Of Electronic

    Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting 12.7.2 Diagnostic of electronic Diagnostic information Remedy instructions Short text 201 Electronics faulty 1. Restart device 2. Replace electronics Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83...
  • Page 212 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 242 Firmware incompatible 1. Check firmware version 2. Flash or replace electronic module Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83...
  • Page 213 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 252 Module incompatible 1. Check electronic modules 2. Check if correct modules are available (e.g. NEx, Ex) Measured variable status 3. Replace electronic modules...
  • Page 214 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 262 Module connection interrupted 1. Check or replace connection cable between sensor electronic module (ISEM) and main electronics Measured variable status 2. Check or replace ISEM or main electronics...
  • Page 215 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 270 Main electronics defective 1. Restart device 2. Replace main electronic module Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83...
  • Page 216 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 271 Main electronics faulty 1. Restart device 2. Replace main electronic module Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83...
  • Page 217 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 272 Main electronics faulty Restart device Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83 Status signal Diagnostic behavior Alarm Influenced measured variables •...
  • Page 218 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 273 Main electronics defective 1. Pay attention to display emergency operation 2. Replace main electronics Measured variable status Quality Good Quality substatus Coding (hex)
  • Page 219 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 275 I/O module defective Change I/O module Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83 Status signal Diagnostic behavior Alarm Influenced measured variables •...
  • Page 220 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 276 I/O module faulty 1. Restart device 2. Change I/O module Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83 Status signal...
  • Page 221 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 283 Memory content inconsistent Restart device Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83 Status signal Diagnostic behavior Alarm Influenced measured variables •...
  • Page 222 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 302 Device verification active Device verification active, please wait. Measured variable status [from the factory] Quality Good Quality substatus Function check Coding (hex) 0xBC to 0xBF...
  • Page 223 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 304 Device verification failed 1. Check verification report 2. Repeat commissioning procedure Measured variable status [from the factory] 3. Check sensor Quality Good Quality substatus...
  • Page 224 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 311 Sensor electronics (ISEM) faulty Maintenance required! Do not reset device Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83 Status signal...
  • Page 225 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 330 Flash file invalid 1. Update firmware of device 2. Restart device Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83...
  • Page 226 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 331 Firmware update failed 1. Update firmware of device 2. Restart device Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83...
  • Page 227 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 332 Writing in HistoROM backup failed 1. Replace user interface board 2. Ex d/XP: replace transmitter Measured variable status Quality Good Quality substatus Coding (hex)
  • Page 228 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 361 I/O module 1 to n faulty 1. Restart device 2. Check electronic modules Measured variable status 3. Change I/O module or main electronics...
  • Page 229 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 369 Matrix code scanner defective Replace matrix code scanner Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83 Status signal Diagnostic behavior...
  • Page 230 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 371 Temperature sensor defective Contact service Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83 Status signal Diagnostic behavior Warning Influenced measured variables •...
  • Page 231 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 372 Sensor electronics (ISEM) faulty 1. Restart device 2. Check if failure recurs Measured variable status 3. Replace sensor electronic module (ISEM) Quality Good...
  • Page 232 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 373 Sensor electronics (ISEM) faulty Transfer data or reset device Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83 Status signal...
  • Page 233 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 374 Sensor electronics (ISEM) faulty 1. Restart device 2. Check if failure recurs Measured variable status [from the factory] 3. Replace sensor electronic module (ISEM)
  • Page 234 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 375 I/O- 1 to n communication failed 1. Restart device 2. Check if failure recurs Measured variable status 3. Replace module rack inclusive electronic modules...
  • Page 235 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 378 Supply voltage ISEM faulty 1. If available: Check connection cable between sensor and transmitter 2. Replace main electronic module Measured variable status 3. Replace sensor electronic module (ISEM)
  • Page 236 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 382 Data storage 1. Insert T-DAT 2. Replace T-DAT Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83 Status signal Diagnostic behavior...
  • Page 237 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 383 Memory content Reset device Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83 Status signal Diagnostic behavior Alarm Influenced measured variables •...
  • Page 238 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 387 HistoROM data faulty Contact service organization Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83 Status signal Diagnostic behavior Alarm Influenced measured variables •...
  • Page 239: Diagnostic Of Configuration

    Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting 12.7.3 Diagnostic of configuration Diagnostic information Remedy instructions Short text 410 Data transfer failed 1. Retry data transfer 2. Check connection Measured variable status Quality Good Quality substatus Coding (hex)
  • Page 240 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 412 Processing download Download active, please wait Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83 Status signal Diagnostic behavior Warning Influenced measured variables •...
  • Page 241 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 437 Configuration incompatible 1. Update firmware 2. Execute factory reset Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83 Status signal...
  • Page 242 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 438 Dataset different 1. Check dataset file 2. Check device parameterization Measured variable status 3. Download new device parameterization Quality Good Quality substatus Coding (hex)
  • Page 243 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 442 Frequency output 1 saturated 1. Check frequency output settings 2. Check process Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83...
  • Page 244 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 453 Flow override active Deactivate flow override Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83 Status signal Diagnostic behavior Warning Influenced measured variables •...
  • Page 245 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 484 Failure mode simulation active Deactivate simulation Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83 Status signal Diagnostic behavior Alarm Influenced measured variables •...
  • Page 246 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 485 Process variable simulation active Deactivate simulation Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83 Status signal Diagnostic behavior Warning Influenced measured variables •...
  • Page 247 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 491 Current output 1 to n simulation active Deactivate simulation Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83 Status signal...
  • Page 248 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 494 Switch output 1 to n simulation active Deactivate simulation switch output Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83...
  • Page 249 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 520 I/O 1 to n hardware configuration invalid 1. Check I/O hardware configuration 2. Replace wrong I/O module Measured variable status 3. Plug the module of double pulse output on correct slot...
  • Page 250: Diagnostic Of Process

    Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 537 Configuration 1. Check IP addresses in network 2. Change IP address Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83...
  • Page 251 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 830 Ambient temperature too high Reduce ambient temp. around the sensor housing Measured variable status [from the factory] Quality Good Quality substatus Coding (hex)
  • Page 252 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 831 Ambient temperature too low Increase ambient temp. around the sensor housing Measured variable status [from the factory] Quality Good Quality substatus Coding (hex)
  • Page 253 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 832 Electronics temperature too high Reduce ambient temperature Measured variable status [from the factory] Quality Good Quality substatus Coding (hex) 0x80 to 0x83 Status signal...
  • Page 254 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 833 Electronics temperature too low Increase ambient temperature Measured variable status [from the factory] Quality Good Quality substatus Coding (hex) 0x80 to 0x83 Status signal...
  • Page 255 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 834 Process temperature too high Reduce process temperature Measured variable status [from the factory] Quality Good Quality substatus Coding (hex) 0x80 to 0x83 Status signal...
  • Page 256 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 835 Process temperature too low Increase process temperature Measured variable status [from the factory] Quality Good Quality substatus Coding (hex) 0x80 to 0x83 Status signal...
  • Page 257 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 842 Process value below limit 1. Decrease process value 2. Check application Measured variable status [from the factory] 3. Check sensor Quality Good Quality substatus...
  • Page 258 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 862 Partly filled pipe 1. Check for gas in process 2. Adjust detection limits Measured variable status [from the factory] Quality Good Quality substatus...
  • Page 259 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 910 Tubes not oscillating 1. If available: Check connection cable between sensor and transmitter 2. Check or replace sensor electronic module (ISEM) Measured variable status 3.
  • Page 260 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 913 Medium unsuitable 1. Check process conditions 2. Check electronic modules or sensor Measured variable status [from the factory] Quality Good Quality substatus Coding (hex)
  • Page 261 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 915 Viscosity ouf of specification 1. Avoid 2-phase flow 2. Increase system pressure Measured variable status [from the factory] 3. Verify viscosity and density are within range 4.
  • Page 262 Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 942 API/ASTM density out of specification 1. Check process density with selected API/ASTM commodity group 2. Check API/ASTM-related parameters Measured variable status [from the factory]...
  • Page 263 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 944 Monitoring failed Check process conditions for Heartbeat Monitoring Measured variable status [from the factory] Quality Good Quality substatus Coding (hex) 0x80 to 0x83...
  • Page 264: Pending Diagnostic Events

    Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Short text 984 Condensation risk 1. Decrease ambient temperature 2. Increase medium temperature Measured variable status [from the factory] Quality Good Quality substatus Coding (hex) 0x80 to 0x83...
  • Page 265: Diagnostics List

    Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Operating time from restart →  265 Operating time →  265 Parameter overview with brief description Parameter Prerequisite Description User interface Actual diagnostics A diagnostic event has occurred. Shows the current occured diagnostic...
  • Page 266: 12.10.2 Filtering The Event Logbook

    Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Navigation path Diagnostics menu → Event logbook submenu → Events list / ../Eventlist I1091 Config. change I1157 Mem.err. ev.list 0d01h19m10s F311 Electr. failure A0014008-EN  41 Using the example of the local display •...
  • Page 267 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Info number Info name I1089 Power on I1090 Configuration reset I1091 Configuration changed I1092 HistoROM backup deleted I1111 Density adjust failure I11280 ZeroPT verified and adjustm. recommended I11281 ZeroPT verif. and adjust. not recommend.
  • Page 268: 12.11 Resetting The Measuring Device

    Diagnostics and troubleshooting Proline Promass S 500 PROFINET over Ethernet-APL Info number Info name I1626 Write protection deactivated I1627 Web server: login successful I1628 Display: login successful I1629 CDI: login successful I1631 Web server access changed I1632 Display: login failed...
  • Page 269 Proline Promass S 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Device name →  269 Manufacturer →  269 →  269 Order code Extended order code 1 →  269 Extended order code 2 →  269 Extended order code 3 →...
  • Page 270: 12.13 Firmware History

    "Manufacturer' s information" document. The manufacturer' s information is available: • In the Download Area of the Endress+Hauser web site: www.endress.com → Downloads • Specify the following details: • Product root: e.g. 8S5B The product root is the first part of the order code: see the nameplate on the device.
  • Page 271: Maintenance

    Observe the inside diameter of the measuring tube and process connection. 13.2 Measuring and test equipment Endress+Hauser offers a variety of measuring and testing equipment, such as Netilion or device tests. Your Endress+Hauser Sales Center can provide detailed information on the services.
  • Page 272: Repair

    • The measuring devices have a modular design. • Spare parts are grouped into logical kits with the associated Installation Instructions. • Repairs are carried out by Endress+Hauser Service or by appropriately trained customers. • Certified devices can only be converted to other certified devices by Endress+Hauser Service or at the factory.
  • Page 273: Disposal

    Proline Promass S 500 PROFINET over Ethernet-APL Repair 14.5 Disposal If required by the Directive 2012/19/EU on waste electrical and electronic equipment (WEEE), the product is marked with the depicted symbol in order to minimize the disposal of WEEE as unsorted municipal waste. Do not dispose of products bearing this marking as unsorted municipal waste.
  • Page 274: Accessories

    Various accessories, which can be ordered with the device or subsequently from Endress +Hauser, are available for the device. Detailed information on the order code in question is available from your local Endress+Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com.
  • Page 275: For The Sensor

    Is used to stabilize the temperature of the fluids in the sensor. Water, water vapor and other non-corrosive liquids are permitted for use as fluids.  If using oil as a heating medium, please consult with Endress+Hauser. Use the order code with the product root DK8003. ...
  • Page 276: Service-Specific Accessories

    Via the Internet: https://portal.endress.com/webapp/applicator Netilion lloT ecosystem: Unlock knowledge With the Netilion IIoT ecosystem,Endress+Hauser allows you to optimize your plant performance, digitize workflows, share knowledge, and enhance collaboration. Drawing upon decades of experience in process automation, Endress+Hauser offers the process industry an IIoT ecosystem designed to effortlessly extract insights from data.
  • Page 277: System Components

    Proline Promass S 500 PROFINET over Ethernet-APL Accessories 15.4 System components Accessories Description Memograph M graphic The Memograph M graphic data manager provides information on all the relevant data manager measured variables. Measured values are recorded correctly, limit values are monitored and measuring points analyzed.
  • Page 278: Technical Data

    Technical data Proline Promass S 500 PROFINET over Ethernet-APL Technical data 16.1 Application The measuring device is intended only for the flow measurement of liquids. Depending on the version ordered, the measuring device can also measure potentially explosive, flammable, poisonous and oxidizing media.
  • Page 279: Input

    To increase the measurement accuracy of certain measured variables, the automation system can continuously write various measured values to the measuring instrument: • Operating pressure to increase measurement accuracy (Endress+Hauser recommends the use of a pressure measuring device for absolute pressure, e.g. Cerabar M or Cerabar •...
  • Page 280 Technical data Proline Promass S 500 PROFINET over Ethernet-APL Current input 0/4 to 20 mA Current input 0/4 to 20 mA (active/passive) Current span • 4 to 20 mA (active) • 0/4 to 20 mA (passive) Resolution 1 µA Voltage drop Typically: 0.6 to 2 V for 3.6 to 22 mA (passive)
  • Page 281: Output

    Proline Promass S 500 PROFINET over Ethernet-APL Technical data 16.4 Output Output signal PROFINET with Ethernet-APL Device use Device connection to an APL field switch The device may only be operated according to the following APL port classifications: • If used in hazardous areas: SLAA or SLAC •...
  • Page 282 Technical data Proline Promass S 500 PROFINET over Ethernet-APL Damping Configurable: 0 to 999.9 s Assignable measured • Mass flow variables • Volume flow • Corrected volume flow • Density • Reference density • Temperature • Electronics temperature • Oscillation frequency 0 •...
  • Page 283 Proline Promass S 500 PROFINET over Ethernet-APL Technical data Assignable measured • Mass flow variables • Volume flow • Corrected volume flow • Density • Reference density • Temperature • Electronics temperature • Oscillation frequency 0 • Oscillation damping 0 •...
  • Page 284 Technical data Proline Promass S 500 PROFINET over Ethernet-APL Maximum switching • DC 30 V, 0.1 A capacity (passive) • AC 30 V, 0.5 A Assignable functions • Disable • On • Diagnostic behavior • Limit • Mass flow • Volume flow •...
  • Page 285: Local Display

    Proline Promass S 500 PROFINET over Ethernet-APL Technical data Pulse/frequency/switch output Pulse output Fault mode Choose from: • Actual value • No pulses Frequency output Fault mode Choose from: • Actual value • 0 Hz • Definable value between: 2 to 12 500 Hz...
  • Page 286: Light Emitting Diodes (Led)

    Technical data Proline Promass S 500 PROFINET over Ethernet-APL Light emitting diodes (LED) Status information Status indicated by various light emitting diodes The following information is displayed depending on the device version: • Supply voltage active • Data transmission active •...
  • Page 287: Power Supply

    Proline Promass S 500 PROFINET over Ethernet-APL Technical data Supported functions • Identification & Maintenance, simple device identifier via: • Control system • Nameplate • Measured value status The process variables are communicated with a measured value status • Blinking feature via the local display for simple device identification and assignment •...
  • Page 288: Performance Characteristics

    Technical data Proline Promass S 500 PROFINET over Ethernet-APL Electrical connection • →  41 • →  50 Potential equalization →  57 Terminals Spring-loaded terminals: Suitable for strands and strands with ferrules. Conductor cross-section 0.2 to 2.5 mm (24 to 12 AWG).
  • Page 289: Zero Point Stability

    Proline Promass S 500 PROFINET over Ethernet-APL Technical data Density (liquids) Under reference conditions Standard density calibration Wide-range 2) 3) Density specification [g/cm³] [g/cm³] [g/cm³] ±0.0005 ±0.01 ±0.002 Valid over the entire temperature and density range Valid range for special density calibration: 0 to 2 g/cm³, +10 to +80 °C (+50 to +176 °F) order code for "Application package", option EE "Special density"...
  • Page 290 Technical data Proline Promass S 500 PROFINET over Ethernet-APL Accuracy of outputs The outputs have the following base accuracy specifications. Current output Accuracy ±5 µA Pulse/frequency output o.r. = of reading Accuracy Max. ±50 ppm o.r. (over the entire ambient temperature range) Repeatability o.r.
  • Page 291 Proline Promass S 500 PROFINET over Ethernet-APL Technical data Wide-range density specification (special density calibration) If the process temperature is outside the valid range (→  288) the measurement error is ±0.0001 g/cm /°C (±0.00005 g/cm /°F) [kg/m ] [°C] [°F]...
  • Page 292: Mounting

    Technical data Proline Promass S 500 PROFINET over Ethernet-APL Calculation of the maximum measured error as a function of the flow rate Flow rate Maximum measured error in % o.r. ZeroPoint ⋅ ± BaseAccu ³ BaseAccu A0021339 A0021332 ZeroPoint ZeroPoint ⋅...
  • Page 293 The device is suitable for use outdoors and indoors with a relative humidity of 4 to 95 %. Operating height According to EN 61010-1 • ≤ 2 000 m (6 562 ft) • > 2 000 m (6 562 ft) with additional overvoltage protection (e.g. Endress+Hauser HAW Series) Degree of protection Transmitter •...
  • Page 294: Process

    Technical data Proline Promass S 500 PROFINET over Ethernet-APL Internal cleaning • CIP cleaning • SIP cleaning • Cleaning with pigs Options Oil- and grease-free version for wetted parts, without declaration Order code for "Service", option HA Mechanical load Transmitter housing and sensor connection housing: •...
  • Page 295: 16.10 Mechanical Construction

    Proline Promass S 500 PROFINET over Ethernet-APL Technical data testing. The corresponding type test declaration can be ordered with the device (order code for "Additional approval", option LN "Sensor housing burst pressure, type test"). Sensor housing burst pressure [mm] [in]...
  • Page 296: Transmitter Housing

    Technical data Proline Promass S 500 PROFINET over Ethernet-APL Weight in SI units Weight [kg] [mm] Weight in US units Weight [lbs] [in] ½ 1½ Materials Transmitter housing Housing of Proline 500 – digital transmitter Order code for "Transmitter housing": •...
  • Page 297 Proline Promass S 500 PROFINET over Ethernet-APL Technical data Cable entries/cable glands Cable entries and adapters Material Cable gland M20 × 1.5 Plastic • Adapter for cable entry with female thread G ½" Nickel-plated brass • Adapter for cable entry with female thread NPT ½"...
  • Page 298: 16.11 User Interface

    Technical data Proline Promass S 500 PROFINET over Ethernet-APL External WLAN antenna • Antenna: ASA plastic (acrylonitrile styrene acrylate) and nickel-plated brass • Adapter: Stainless steel and nickel-plated brass • Cable: Polyethylene • Plug: Nickel-plated brass • Angle bracket: Stainless steel Process connections •...
  • Page 299: Display Elements

    Proline Promass S 500 PROFINET over Ethernet-APL Technical data Onsite operation Via display module Features: • Order code for "Display; operation", option F "4-line, illuminated, graphic display; touch control"" • Order code for "Display; operation", option G "4-line, illuminated, graphic display;...
  • Page 300 Technical data Proline Promass S 500 PROFINET over Ethernet-APL Supported operating Operating unit Interface Additional information tools FieldCare SFE500 Notebook, PC or tablet • CDI-RJ45 service →  276 with Microsoft Windows interface system • WLAN interface • Fieldbus protocol...
  • Page 301: Historom Data Management

    Proline Promass S 500 PROFINET over Ethernet-APL Technical data HistoROM The measuring device features HistoROM data management. HistoROM data management data management comprises both the storage and import/export of key device and process data, making operation and servicing far more reliable, secure and efficient.
  • Page 302: 16.12 Certificates And Approvals

    The device meets the legal requirements of the applicable EU Directives. These are listed in the corresponding EU Declaration of Conformity along with the standards applied. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. UKCA marking The device meets the legal requirements of the applicable UK regulations (Statutory Instruments).
  • Page 303 Directive a) PED/G1/x (x = category) or b) PESR/G1/x (x = category) on the sensor nameplate, Endress+Hauser confirms compliance with the "Essential Safety Requirements" a) specified in Annex I of the Pressure Equipment Directive 2014/68/EU or b) Schedule 2 of Statutory Instruments 2016 No. 1105.
  • Page 304: 16.13 Application Packages

    The application packages can be ordered with the device or subsequently from Endress+Hauser. Detailed information on the order code in question is available from your local Endress+Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com.
  • Page 305: Heartbeat Verification

    Proline Promass S 500 PROFINET over Ethernet-APL Technical data Diagnostic functionality Order code for "Application package", option EA "Extended HistoROM" Comprises extended functions concerning the event log and the activation of the measured value memory. Event log: Memory volume is extended from 20 message entries (standard version) to up to 100 entries.
  • Page 306: 16.14 Accessories

    For an overview of the scope of the associated Technical Documentation, refer to the following: • Device Viewer (www.endress.com/deviceviewer): Enter the serial number from the nameplate • Endress+Hauser Operations app: Enter serial number from nameplate or scan matrix code on nameplate. Standard documentation Brief operating instructions...
  • Page 307: Special Documentation

    Proline Promass S 500 PROFINET over Ethernet-APL Technical data Contents Documentation code Measuring device cCSAus IS XA01475D cCSAus Ex i XA01509D cCSAus Ex nA XA01510D INMETRO Ex i XA01476D INMETRO Ex ec XA01477D NEPSI Ex i XA01478D NEPSI Ex nA...
  • Page 308: Index

    Index Proline Promass S 500 PROFINET over Ethernet-APL Index 0 … 9 Connecting the measuring instrument Proline 500 ......50 3-A approval .
  • Page 309 Proline Promass S 500 PROFINET over Ethernet-APL Index Overview ......198 FieldCare .
  • Page 310 Index Proline Promass S 500 PROFINET over Ethernet-APL Local display ......299 Mounting dimensions Navigation view .
  • Page 311 Proline Promass S 500 PROFINET over Ethernet-APL Index Corrected volume flow calculation (Submenu) . . . 144 Process connections ......298 Current input .
  • Page 312 Index Proline Promass S 500 PROFINET over Ethernet-APL Relay output ......133 Petroleum ......161 Resetting the device .
  • Page 313 Proline Promass S 500 PROFINET over Ethernet-APL Index Tool tip see Help text Totalizer Assign process variable ....174 Configuring ......151 Totalizer Control module .
  • Page 314 *71674748* 71674748 www.addresses.endress.com...