RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Technical
Service
Manual
Part Number: 4117104
Rev: G
Date: 17 April 2003
© 2003 Draeger Medical, Inc.
Fabius GS®
Anesthesia System

Advertisement

loading
Need help?

Need help?

Do you have a question about the Fabius GS and is the answer not in the manual?

Questions and answers

Summary of Contents for Draeger Fabius GS

  • Page 1 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS Technical Service Manual Part Number: 4117104 Rev: G Date: 17 April 2003 © 2003 Draeger Medical, Inc. Fabius GS® Anesthesia System...
  • Page 2 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS...
  • Page 3 FUNCTION DESCRIPTION ..........2-1 General Information about the Fabius GS ......2-1 Fabius GS Function Diagram .
  • Page 4 PMS PROCEDURE, FABIUS GS ........
  • Page 5 Updating Send Image (utility) file........8-4 INSTALLING FABIUS GS SOFTWARE FROM A PC ....8-4 Boot Strap Download Procedure .
  • Page 6 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS ANESTHESIA SYSTEM DISPLAY AUXILIARY OXYGEN FLOWMETER CONTROL PANEL VENTILATOR VAPORIZER BREATHING SYSTEM CYLINDER PRESSURE GAUGES PIPELINE O2 FLUSH PRESSURE GAUGES FLOW SCAVENGER CONTROLS TROLLEY TOTAL FRESH GAS...
  • Page 7 After any component is replaced, verify that the unit is operating properly by running the appropriate diagnostic procedure. The PMS PROCEDURE in Section 7 must also be performed after any component is replaced. The general arrangement of the Fabius GS anesthesia system is shown on the opposite page.
  • Page 8 • Technical Specifications Copyright Copyright© 2003 by Draeger Medical, Inc. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or stored in a retrieval system in any form or by any means, electronic or mechanical, including photocopying and recording, without written permission of Draeger Medical, Inc.
  • Page 9 Monitoring, Electrical and Gas connections are illustrated on following pages. AUXILIARY DISPLAY OXYGEN FLOWMETER CONTROL PANEL VENTILATOR VAPORIZER BREATHING CYLINDER SYSTEM PRESSURE GAUGES PIPELINE PRESSURE GAUGES O2 FLUSH FLOW CONTROLS TROLLEY TOTAL FRESH GAS FLOWMETER SV00543 Figure 2-1. Front View of Fabius GS Anesthesia System...
  • Page 10 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS FUNCTION DESCRIPTION (continued) SV00457 SERIAL COMMUNICATION PORT HOSE CONNECTION FOR PEEP VALVE HOSE CONNECTION FOR APL BYPASS VALVE O2 SENSOR CORD CONNECTION BREATHING PRESSURE PILOT LINE CONNECTION...
  • Page 11 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS FUNCTION DESCRIPTION (continued) SV00505 PIPELINE HOSE CONNECTIONS O2 CYLINDER YOKE N2O OR AIR CYLINDER YOKE O2 CYLINDER YOKE Figure 2-3. Back View Showing Gas Pipeline and Cylinder Yokes...
  • Page 12 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS FUNCTION DESCRIPTION (continued) N2O CYLINDER CONNECTION O2 CYLINDER CONNECTION PIPELINE HOSE CONNECTIONS SV00551 Figure 2-4. Back View Showing Gas Pipeline and Cylinder Hose Connections (machines without cylinder yokes)
  • Page 13 VALVE VALVE AUXILIARY VALVE LUNG SAFETY VALVE VENTILATOR P VENT BREATHING (AIR ENTRAINMENT) ABSORBER PEEP EXPIRATORY CONTROL VALVE VALVE APL VALVE ELECTRONIC SPONT PEEP VALVE BYPASS VALVE PISTON PUMP SV00499 OUTLET Figure 2-5. Function diagram of Fabius GS Rev. B...
  • Page 14 FUNCTION DESCRIPTION (continued) Battery Backup Fabius GS backup power is provided by two series-connected 12 V rechargeable batteries. These batteries remain on charge as long as the machine is plugged into an active AC outlet. Should power to the main controller PCB assembly fail...
  • Page 15 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS FUNCTION DESCRIPTION (continued) Fabius GS Piping Diagram SV00454 Figure 2-7. Fabius GS Piping Diagram Rev. B...
  • Page 16 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS FUNCTION DESCRIPTION (continued) Function Description of Gas Box The supply gases flow through the filters and non-return valves in the gas inlet assembly. Pipeline supply pressures are indicated on gauges located on the flowmeter assembly.
  • Page 17 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS FUNCTION DESCRIPTION (continued) If the O flush button is pressed, oxygen is delivered to the fresh gas outlet. The fresh-gas ejector prevents the fresh gas from flowing back into the anesthetic vaporizer.
  • Page 18 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS FUNCTION DESCRIPTION (continued) SORC (Sensitive Oxygen Ratio Controller) The SORC is a control element that functions like an N O shut-off device and ensures a vital O concentration in the fresh gas.
  • Page 19 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS FUNCTION DESCRIPTION (continued) The flow control valves are used to adjust the O and N O flows. Restrictors located at the outlets of the SORC generate back-pressures. These back-pressures exert a force on the control diaphragms of the SORC.
  • Page 20 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS FUNCTION DESCRIPTION (continued) Compact Breathing System, Cosy II The Cosy II compact breathing system allows three modes of patient ventilation: manual and spontaneous breathing, volume controlled, and pressure controlled.
  • Page 21 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS FUNCTION DESCRIPTION (continued) SV00493 Figure 2-12. Compact breathing system, CosyII PEEP/Pmax valve Expiratory valve Flow sensor (Spirolog) Expiratory port MAN/SPONT-Volume Control selector APL Bypass valve...
  • Page 22 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS FUNCTION DESCRIPTION (continued) 2.7.1 Function Description: Manual Ventilation Manual Ventilation: General During manual ventilation, the APL valve is set to the "MAN" position. The patient system safety valve is activated. The piston of the ventilator is in the upper end position in order to reduce the dead space volume of the ventilator.
  • Page 23 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS FUNCTION DESCRIPTION (continued) Manual Ventilation: Expiration During expiration, the inspiratory valve remains closed and thus prevents the expiratory gas from flowing back into the inspiratory branch.
  • Page 24 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS FUNCTION DESCRIPTION (continued) 2.7.2 Function Description: Spontaneous Breathing Spontaneous Breathing: General A prerequisite for spontaneous breathing is that the patient is supplied with a sufficient amount of fresh gas. The APL valve selector must be set to the "SPONT"...
  • Page 25 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS FUNCTION DESCRIPTION (continued) Spontaneous Breathing: Expiration During expiration, the inspiratory valve remains closed thus preventing the expiratory gas from flowing back into the inspiratory branch.
  • Page 26 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS FUNCTION DESCRIPTION (continued) 2.7.3 Function Description: Volume Mode Ventilation Volume Control Mode: General A prerequisite for volume control is that the patient is supplied with a sufficient amount of fresh gas.
  • Page 27 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS FUNCTION DESCRIPTION (continued) Volume Control Mode: Inspiration During inspiration, the PEEP Pmax valve remains closed. The control pressure present at the PEEP Pmax valve varies with the set pressure limit (Pmax).
  • Page 28 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS FUNCTION DESCRIPTION (continued) Volume Control Mode: Expiration During expiration, the inspiratory valve remains closed thus preventing rebreathing into the inspiratory branch. The expiratory gas from the lung flows through the expiratory hose, the flow...
  • Page 29 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS FUNCTION DESCRIPTION (continued) 2.7.4 Cosy II Absorber The absorber canister is filled with fresh soda lime. The soda lime scrubs from the respiratory expired gas.
  • Page 30 The ventilator is located in a swing-out compartment at the left side of the Fabius GS. A hose terminal is provided on the left side of the compartment for connection to the breathing system. Fresh gas is delivered to the patient by a piston that is driven by a motor and ball-screw arrangement.
  • Page 31 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS FUNCTION DESCRIPTION (continued) NEGATIVE PRESSURE RELIEF VALVE VENT CHAMBER PRESSURE HIGH PRESSURE SENSOR LINE VALVE VACUUM LINE TO PNEUMATIC ASSEMBLY UPPER DIAPHRAGM CYLINDER PISTON LOWER...
  • Page 32 Negative Pressure Relief Valve The negative pressure relief valve allows the patient to spontaneously breathe ambient air should the medical gas supply and/or Fabius GS fail. The opening pressure of this valve is indicated in the table below. threaded ring...
  • Page 33 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS FUNCTION DESCRIPTION (continued) Pneumatic System The pneumatic assembly provides pressure for the PEEP valve control, and also provides vacuum for the ventilator bag-type rolling seals and the APL bypass valve control.
  • Page 34 2.9.1 PEEP Pmax Valve Control When the Fabius GS is operating in the automatic mode, the pump on the pneumatic assembly is running, and the electronic PEEP valve receives a signal from the main control PCB. The amount of current...
  • Page 35 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS FUNCTION DESCRIPTION (continued) 2.10 Electrical Block Diagram GAS INLET ASM FLOW METER BEZEL ASM FRESH LOW O2 PRESS DISPLAY PNEUMATIC ASM SERIAL PORT PEEP INTERFACE...
  • Page 36 FABIUS GS FUNCTION DESCRIPTION (continued) 2.11 Function Description: Control PCB The control PCB provides the following functions in the Fabius GS: 2.11.1 Flowmeter Receives flow rate data from the N2O, Air, and O2 flow sensors and provides fresh gas display information for these gases.
  • Page 37 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS FUNCTION DESCRIPTION (continued) FLASH XCVR BUFFER LATCH FRESH GAS DISPLAY FRESH GAS FLOW SENSORS DISPLAY VIDEO PANEL CONTROL O2 SENSOR MICROCONTROLLER SPKR SENSOR LEDs TABLE LAMP VACUUM KEYPAD &...
  • Page 38 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS FUNCTION DESCRIPTION (continued) 2.12 Control Panel Assembly The control panel assembly consists of a 320 x 240 pixel graphical display, a table lamp with six LEDs, a membrane keypad, rotary encoder and speaker.
  • Page 39 Pressure Gas Selection Freshgas Alarm PEEP Ambient Pressure Figure 2-26. Fabius GS Control Panel (Main Service Screen Illustrated) Item Function Selects volume controlled ventilation mode. Refer to Operator’s Manual Selects pressure controlled ventilation mode. Refer to Operator’s Manual Reserved for future use Controls table lamp: Off/On Places ventilator in Man Spont mode.
  • Page 40 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS FUNCTION DESCRIPTION (continued) 2.13 FiO2 Measurement The O sensor measures the fraction of inspired O (FiO ) in the respiratory gas. The O sensor contains an alkaline electrolyte, a lead anode, two gold cathodes, and a Teflon membrane.
  • Page 41 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS FUNCTION DESCRIPTION (continued) 2.14 Respiratory Flow Measurement The flow sensor functions according to the constant temperature hot-wire anemometer principle. Respiratory gas flows past a thin platinum wire. This platinum wire (A) is located in a measuring tube and is electrically heated.
  • Page 42 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS FUNCTION DESCRIPTION (continued) 2.15 Gas Flow Rate Measurement The gas flow sensors operate on the principle of specific heat for individual gases. In each sensor, as the gas flows through a heated chamber the gas molecules carry away a certain amount of heat relative to the specific heat index for that gas.
  • Page 43 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS FUNCTION DESCRIPTION (continued) 2.16 Vaporizer Refer to separate technical documentation of the anesthetic vaporizer. 2-35...
  • Page 44 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS...
  • Page 45 Power Supply and Voltage Distribution In the Fabius GS, the power supply delivers 28 VDC to the main controller unit PCB assembly. Some of the other voltages derived on this assembly can be measured at the connectors listed in Table 3-1. Fuse data is listed in Table 3-2.
  • Page 46 O2 SENSOR PUMP SPIROLOG VOL SENSOR MAIN CONTROLLER PCB ASSEMBLY FRONT MOTOR PANEL DISPLAY TACH SENSOR KEYPAD INTERFACE HOME SENSOR VENTILATOR TABLE LAMP BOOST ON/OFF CAPS ROTARY ENCODER BATT POWER MAINS SUPPLY PWR ENTRY FIGURE 3-1. Fabius GS Block Diagram (Electrical)
  • Page 47 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS TROUBLESHOOTING GUIDE (continued) J18 SPIROLOG SENSOR J17 O2 SENSOR J2 FRONT PANEL DISPLAY J9 KEYPAD INTERFACE J15 SERIAL PORT INTERFACE VENT CHAMBER PRESSURE SENSOR BREATHING...
  • Page 48 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS TROUBLESHOOTING GUIDE (continued) FABIUS GS SV00454 FIGURE 3-3. Fabius GS Piping Diagram Rev. B...
  • Page 49 TO UPPER PORT OF LS1 ON PCB ASM VENT CHAMBER PRESSURE PNEUMATIC ASSEMBLY VENTILATOR ELECTRONIC PEEP VALVE PUMP RESERVOIR APL BYPASS VALVE PEEP VALVE INTERFACE BREATHING CONTROL SIGNAL PANEL SYSTEM FROM J7 ON PCB ASM FIGURE 3-4. Fabius GS Pneumatic Control Diagram...
  • Page 50 Troubleshooting Guides Table 3-3 lists common failure modes and symptoms (excluding simultaneous multiple faults) for the monitoring and display devices in the Fabius GS. Each failure mode or symptom is keyed to a troubleshooting guide flow chart at the back of this section to assist the service representative in locating the cause of a problem.
  • Page 51 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS TROUBLESHOOTING GUIDE (continued) GUIDE 1: Loss of O Monitor START IS O2 MONITOR WORKING PROPERLY? CONNECT SENSOR CORD TO INTERFACE IS SENSOR CORD CONNECTED TO...
  • Page 52 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS TROUBLESHOOTING GUIDE (continued) FABIUS GS GUIDE 2: Loss of Breathing Pressure Monitor START BREATHING PRESSURE MONITOR WORKING PROPERLY? CHECK PILOT LINE CONNECTION AT BOTH ENDS PILOT LINE...
  • Page 53 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS TROUBLESHOOTING GUIDE (continued) GUIDE 3: Loss of Respiratory Volume Monitor START RESPIRATORY VOLUME MONITOR WORKING PROPERLY? CONNECT SENSOR CORD TO FLOW SENSOR AND CALIBRATE IS SENSOR...
  • Page 54 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS TROUBLESHOOTING GUIDE (continued) FABIUS GS GUIDE 4: No Audio Alarms START AUDIO ALARMS WORKING? CHECK CABLE FROM J1 ON FRONT PANEL KEYPAD INTERFACE (MONITOR BEZEL) TO J9...
  • Page 55 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS TROUBLESHOOTING GUIDE (continued) GUIDE 5: Serial Port Communication Failure START DOES HOST UNIT COMMUNICATE WITH PERIPHERAL DEVICE? CONNECT CABLE TO INTERFACE PERIPHERAL CABLE CONTACT LOCAL...
  • Page 56 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS TROUBLESHOOTING GUIDE (continued) FABIUS GS GUIDE 6: No O Supply Pressure Alarms START LOW O2 SUPPLY PRESSURE ALARM FUNCTIONAL? VERIFY NO O2 PRESSURE IS IN THE SYSTEM CHECK THE FOLLOWING ITEMS: LOW O2 SUPPLY PRESS.
  • Page 57 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS TROUBLESHOOTING GUIDE (continued) GUIDE 7: Main Display Blank Upon System Power-up START IS DISPLAY FUNCTIONAL UPON SYSTEM POWER-UP? CHECK POWER SUPPLY VOLTAGE, REPLACE POWER SUPPLY IF OUT...
  • Page 58 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS TROUBLESHOOTING GUIDE (continued) FABIUS GS GUIDE 8: Keypad Inoperative START KEYPAD WORKING PROPERLY? CHECK CABLE FROM J1 ON FRONT PANEL KEYPAD INTERFACE (MONITOR BEZEL) TO J9 ON CONTROLLER PCB...
  • Page 59 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS TROUBLESHOOTING GUIDE (continued) GUIDE 9: Ventilator Inoperative START IS VENTILATOR WORKING PROPERLY? VERIFY GASKET, VENTILATOR COVER, EXTERNAL HOSE CONNECTIONS AND BREATHING SYSTEM ARE SECURE IS VENTILATOR...
  • Page 60 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS TROUBLESHOOTING GUIDE (continued) FABIUS GS GUIDE 10: Fresh Gas Display Blank or Not Working Properly START ARE ALL FRESH GAS DISPLAYS WORKING PROPERLY? CHECK CABLE FROM ARE ALL...
  • Page 61 DIAGNOSTICS DIAGNOSTICS The Fabius GS diagnostic system monitors and records the status of its internal hardware when the machine is turned on. The status of each test is displayed on the Power-up screen as shown in Figure 4-1. This screen is displayed for several seconds before proceeding to the Standby screen.
  • Page 62 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) Standby 13:24 Fabius GS will activate Sleep Mode in 3 min 49 sec To start Pressure operation Control press one Volume of the keys...
  • Page 63 System Service Screen The System Service Screen shown in Figure 4-3 displays system service data. Press the rotary knob to go to the Main Service screen. System Service Screen 13:32 Fabius GS SW 1.20 Serial Number 12345678 Last Service Date 7-5-2001...
  • Page 64 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) Main Service Screen The Main Service Screen shown in Figure 4-4 lists the service functions available under each of the displayed categories. Turn the rotary knob to move the scroll box to the desired category, then press the knob to confirm.
  • Page 65 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) Service Log When the Service Log category is selected, the screen shown in Figure 4-5 will appear. Turn the rotary knob to scroll down to select the desired function, then press the knob to confirm.
  • Page 66 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) 4.3.1 Service Log Entries When Entries is selected, the screen shown in Figure 4-6 appears. Use the rotary knob to scroll through the list of service log entries. The service log holds up to 256 entries.
  • Page 67 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) Fabius GS Service Log Entries: Communication Errors Code Description Possible Cause C001 Handler Busy Bad processor, external device incorrect operation C002 Incoming Overflow...
  • Page 68 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) Diagnostic Errors Code Description Possible Cause F001 Diag Watchdog Bad hardware F002 Diag System Ram System RAM fails on pwr up, bad RAM chip, bad CPU, bad trace, etc.
  • Page 69 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) System Errors Code Description Possible Cause NVF2 Sec NV File Inv Secondary NVRAM file invalid NVC0 Non-Matching CRCS CRCs for primary and secondary NVRAM did not match Rev.
  • Page 70 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) Service Log Events Code Description Possible Cause S006 Service Log Clear Log cleared by service tech, SW upgrade SS90 test message Factory test - no failure...
  • Page 71 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) Ventilator Errors Code Description Possible Cause V001 VENT PEEP NOT CAL Auto ventilation was attempted without valid PEEP calibration V002 ZERO POS INVALID...
  • Page 72 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) 4.3.2 Real Time Values When Real Time Values is selected, the typical screen shown in Figure 4- 7 appears. Service 14:33 Real Time Values...
  • Page 73 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) Preventive Maintenance When the Preventive Maintenance category is selected, the screen shown in Figure 4-8 will appear. Turn the rotary knob to select Activate, then press the knob to confirm. The next preventive maintenance due date can then be set using the screen described on the next page.
  • Page 74 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) 4.4.1 Activate Preventive Maintenance Date With Activate selected, the screen will appear as shown in Figure 4-9. Turn the rotary knob to position the box under Mth and press the knob to highlight the box.
  • Page 75 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) General When the General category is selected, the screen shown in Figure 4-10 will appear. Turn the rotary knob to scroll down to select the desired function, then press the knob to confirm.
  • Page 76 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) 4.5.1 Serial Number When Serial Number is selected, the screen will appear as shown in Figure 4-11. Turn the rotary knob to position the box on the first digit of the serial number, then press the knob to highlight the number.
  • Page 77 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) 4.5.2 Reset Hours Run When Reset Hours Run is selected, the screen will appear as shown in Figure 4-12. Turn the rotary knob to select Reset Hours Run, and press the knob to highlight the selection.
  • Page 78 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) 4.5.3 Reset Last Service Date When Reset Last Service Date is selected, the screen will appear as shown in Figure 4-13. Turn the rotary knob to select Reset Last Service Date, and press the knob to highlight the selection.
  • Page 79 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) 4.5.4 Select Market Kit When Select Market Kit is selected, the screen will appear as shown in Figure 4-13A. Turn the rotary knob to scroll to Select Market Kit, and press the knob to highlight the selection.
  • Page 80 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) This page intentionally left blank Rev. B 4-16B...
  • Page 81 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) Calibration When the Calibration category is selected, the screen shown in Figure 4-14 will appear. Turn the rotary knob to scroll down to select the desired function, then press the knob to confirm.
  • Page 82 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) 4.6.1 Fresh Gas Flow Calibration When Fresh Gas Flow is selected, the screen shown in Figure 4-15 appears. Follow the on-screen instructions to perform the calibration. The Store Zero and Exit keys are located just below their legends at the bottom of the screen.
  • Page 83 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) 4.6.2 Pressure Calibration When Pressure is selected, the screen shown in Figure 4-16 appears. Follow the on-screen instructions to perform the calibration. The Store Zero and Exit keys are located just below their legends at the bottom of the screen.
  • Page 84 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) 4.6.3 O2 Zero Calibration When O2 is selected, the screen shown in Figure 4-17 appears. Follow the on-screen instructions to perform the calibration. The Store Zero and Exit keys are located just below their legends at the bottom of the screen.
  • Page 85 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) 4.6.4 PEEP Valve Calibration When PEEP is selected, the screen shown in Figure 4-18 appears. Caution: Use of the test hose (P/N 9995112, Corrugated Tube, 22mm x 12 in.
  • Page 86 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) Configure When the Configure category is selected, the screen shown in Figure 4-19 will appear. Turn the rotary knob to scroll down to select the desired function, then press the knob to confirm.
  • Page 87 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) 4.7.1 Flowmeter When Flowmeter is selected, the screen shown in one of the following illustrations will appear. Turn the knob to scroll down to the desired selection, and press the knob to enter that function.
  • Page 88 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) 4.7.1.1 O2 Position (Virtual Flowtubes) When O2 Position for virtual flowtubes is selected, the screen will appear as shown in Figure 4-21. Turn the rotary knob to select O2 Position, and press the knob to highlight the choices.
  • Page 89 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) 4.7.1.2 Gas Selection When Gas Selection is chosen, the screen will appear as shown in Figure 4-22. Turn the rotary knob to choose Gas Selection, and press the knob to highlight the choices.
  • Page 90 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) 4.7.2 O2 Whistle When O2 Whistle is chosen, the screen will appear as shown in Figure 4- 23. Turn the rotary knob to select Change State, and press the knob to highlight the choices.
  • Page 91 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) 4.7.3 Low Fresh Gas Alarm When Fresh Gas Alarm is chosen, the screen will appear as shown in Figure 4-24. Turn the rotary knob to select Change State, and press the knob to highlight the choices.
  • Page 92 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) 4.7.4 Ambient Pressure When Ambient Pressure is chosen, the screen will appear as shown in one of the following illustrations. Turn the rotary knob to select Adjust, and press the knob.
  • Page 93 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) This page intentionally left blank Rev. B 4-27B...
  • Page 94 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) Serial Port When the Serial Port category is selected, the screen shown in Figure 4-25 will appear. Turn the rotary knob to select Parameters, then press the knob to confirm. The serial port communication parameters can then be set using the screen described on the next page.
  • Page 95 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS DIAGNOSTICS (continued) 4.8.1 Serial Port Parameters With Parameters selected, the screen will appear as shown in Figure 4- 26. Turn the rotary knob to select the parameter you want to set, and press the knob to highlight that parameter.
  • Page 96 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS...
  • Page 97 (TSR). The following are the only procedures authorized by DrägerService to be performed in the field. All other service procedures shall be referred to Draeger Medical, Inc., DrägerService. NOTE: The PMS PROCEDURE given in Section 7 must be performed after...
  • Page 98 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS REPLACEMENT PROCEDURES (continued) FABIUS GS Cylinder Yokes and Regulators Replacement of a yoke, check valve, or cylinder regulator requires that the yoke be removed from the anesthesia machine. Figure 5-1 shows the cylinder yokes mounting arrangement.
  • Page 99 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS REPLACEMENT PROCEDURES (continued) BOTTOM VIEW SV00472 N2O OR AIR REGULATOR YOKE MOUNTING O2 REGULATOR SCREWS AND LOCK WASHERS (2X PER YOKE) CHECK VALVE (TYP) O2 YOKE...
  • Page 100 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS REPLACEMENT PROCEDURES (continued) FABIUS GS 5.1.12 Reattach the yoke to the machine with the hardware that screws and lock washer that were previously removed. 5.1.13 If a new cylinder is being installed, remove the old sealing washer from the gas inlet of the yoke and install a new washer.
  • Page 101 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS REPLACEMENT PROCEDURES (continued) Cylinder Pressure Gauges Each pressure gauge is secured by two 6-32 kep nuts As shown in Figure 5-2. Access to gauge mounting nuts and tubing connections is at the underside of the table top at the back of the machine.
  • Page 102 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS REPLACEMENT PROCEDURES (continued) FABIUS GS 5.2.14.2 For any gas, the pressure should not drop more than 50 psi in two minutes. 5.2.15 Re-install the cylinder on the yoke.
  • Page 103 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS REPLACEMENT PROCEDURES (continued) Auxiliary Oxygen Flow Meter The auxiliary oxygen flowmeter is attached to the outside of the ventilator door on the left side of the machine, by a stud and nut arrangement - accessible by opening the door.
  • Page 104 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS REPLACEMENT PROCEDURES (continued) FABIUS GS Figure 5-3. Auxiliary O Flowmeter...
  • Page 105 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS REPLACEMENT PROCEDURES (continued) Vaporizers The vaporizer mounting is shown in Figure 5-4. Before removing a vaporizer from the machine, it must be completely drained and dried in accordance with the procedure given below.
  • Page 106 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS REPLACEMENT PROCEDURES (continued) FABIUS GS SV00518 LOCKING LEVER SELECTOR LEVER O-RINGS VAPORIZER MOUNTING AND EXCLUSION SYSTEM Figure 5-4. Vaporizer Installation 5-10...
  • Page 107 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS REPLACEMENT PROCEDURES (continued) NOTE: Should a 19.x series vaporizer containing anesthetic agent be accidentally tilted more than 45 degrees, it must be drained and flushed in accordance with instructions given in the manual supplied with the vaporizer.
  • Page 108 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS REPLACEMENT PROCEDURES (continued) FABIUS GS O2 Flush Valve The O2 flush valve is accessible from the underside of the table top. Valve removal requires disconnecting the plastic tubing and removal of an L-fitting.
  • Page 109 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS REPLACEMENT PROCEDURES (continued) UNDERSIDE VIEW OF TABLE SV00474 STRAIGHT FITTING L-FITTING O2 FLUSH VALVE RETAINING NUT Figure 5-5. O Flush Valve 5-13...
  • Page 110 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS REPLACEMENT PROCEDURES (continued) FABIUS GS 5.5.12 Disconnect the fresh gas hose from the breathing system. Connect a test gauge and B.P. bulb to the fresh gas hose via Fresh Gas Test Connector P/N 4117361, and perform the following test: 5.5.12.1...
  • Page 111 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS REPLACEMENT PROCEDURES (continued) Caster The casters are threaded into the frame rails on the trolley. A typical arrangement is shown in Figure 5-6. Caster replacement requires that the machine be tilted to provide enough clearance for the caster stem to be withdrawn from the frame rail.
  • Page 112 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS REPLACEMENT PROCEDURES (continued) FABIUS GS SV00507 THREADED STEM Figure 5-6. Caster Replacement 5-16...
  • Page 113 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS REPLACEMENT PROCEDURES (continued) Ventilator The ventilator and door assembly are retained in the monitor housing by a hinge rod. Figures 5-7 and 5-7A show the hinge arrangement and the disconnects needed to remove the ventilator from the machine.
  • Page 114 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS REPLACEMENT PROCEDURES (continued) FABIUS GS E-RING UPPER HINGE BUSHING VENTILATOR ASSEMBLY HINGE ROD LOWER HINGE BUSHING SPACERS J11 TACH SENSOR J12 HOME SENSOR DETAIL OF CONTROLLER PCB WITH VENTILATOR...
  • Page 115 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS REPLACEMENT PROCEDURES (continued) TOP VIEW OF HINGE DETAILS VENTILATOR ASSEMBLY HINGE UPPER HINGE BUSHING VENT PRESSURE HOSE DOOR CONNECTION HIGH PRESSURE VALVE NEGATIVE PRESSURE LOWER...
  • Page 116 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS REPLACEMENT PROCEDURES (continued) FABIUS GS Ventilator Servicing: Upper Diaphragm (patient side) 5.7.11 Release the three clamps holding the patient assembly to the upper cylinder; lift the patient assembly up and out.
  • Page 117 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS REPLACEMENT PROCEDURES (continued) Lower Diaphragm: (motor & case assembly) 5.7.15 Remove the bolt securing the band clamp holding the upper and lower cylinders together. Open the band clamp and move it down onto the lower cylinder.
  • Page 118 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS REPLACEMENT PROCEDURES (continued) FABIUS GS O-RING LOWER DIAPHRAGM SV00544 Figure 5-7D. Lower Diaphragm Installation 5.7.21 Reinstall the O-ring in the groove on the outside of the diaphragm.
  • Page 119 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS REPLACEMENT PROCEDURES (continued) SV00514 MARK PISTON SCREW (3X) RETAINING RING TOP VIEW OF PISTON PISTON LOWER DIAPHRAGM LOWER CYLINDER MOTOR/BALLSCREW ASSEMBLY O-RING GROUND PLATE SCREW (2X)
  • Page 120 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS REPLACEMENT PROCEDURES (continued) FABIUS GS 5.7.26 While holding the cylinders together, move the band clamp into position and reinstall its bolt to join the cylinders. Make sure the band clamp is positioned as shown in Figure 5-7A, or the ventilator door may not close properly.
  • Page 121 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS REPLACEMENT PROCEDURES (continued) Tach Sensor Disc and Encoder NOTE: Upper and lower cylinders must be separated, and piston must be removed. 5.7.37 At the bottom of the lower cylinder, remove the ground plate, and remove the center cap.
  • Page 122 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS REPLACEMENT PROCEDURES (continued) FABIUS GS Ventilator Reassembly: 5.7.43 Position the ventilator against the door; slide the upper and lower hinge bushings in place, and reinstall Screw A (Figure 5-7A) to attach the ventilator to the door.
  • Page 123 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS REPLACEMENT PROCEDURES (continued) Patient Assembly: In case of repair, replace the high pressure valve, the negative pressure relief valve, or the complete patient system. See Figure 5-7A.
  • Page 124 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS REPLACEMENT PROCEDURES (continued) FABIUS GS 5.7.66 Carefully reinstall the patient system, making sure the diaphragm seats properly. 5.7.67 Secure the three clamps holding the patient assembly to the upper cylinder.
  • Page 125 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS REPLACEMENT PROCEDURES (continued) Gas Inlet Assembly (including O2 supply pressure switch) The gas inlet assembly is attached to the inside of the back panel on the gas box assembly as shown in Figure 5-8.
  • Page 126 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS REPLACEMENT PROCEDURES (continued) FABIUS GS GAS INLET ASSEMBLY WIRE HARNESS O2 SUPPLY PRESSURE ALARM SWITCH GAS INLET ASSEMBLY MOUNTING SCREWS BACK PANEL SV00473 Figure 5-8. Gas Inlet Assembly 5-30 Rev.
  • Page 127 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS REPLACEMENT PROCEDURES (continued) Battery The two 12 V rechargeable batteries are located in the vent & monitor assembly behind the front bezel. The batteries and their connections are accessible through the ventilator compartment as shown in Figure 5-9.
  • Page 128 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS REPLACEMENT PROCEDURES (continued) FABIUS GS TOP CUT AWAY TO SHOW BATTERY ARRANGEMENT BATTERIES BATTERY RETAINER BRACKET RETAINER SCREW BLACK WIRE CABLE CLIP YELLOW WIRES RED WIRE SV0052 VENTILATOR COMPARTMENT Figure 5-9.
  • Page 129 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS REPLACEMENT PROCEDURES (continued) 5.10 Power Supply The power supply is attached to the inside of the controller assembly back panel. Figure 5-10 shows the mounting and connection arrangement for the power supply.
  • Page 130 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS REPLACEMENT PROCEDURES (continued) FABIUS GS WIRE CLIPS GROUND WIRE CONNECTION INPUT CABLE (J1) POWER SUPPLY OUTPUT CABLE (J3) BACK PANEL OF CONTROLLER ASSEMBLY POWER SUPPLY MOUNTING SCREWS Figure 5-10.
  • Page 131 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS REPLACEMENT PROCEDURES (continued) 5.11 Control PCB Assembly The control PCB assembly is accessible by removing the back cover of the controller housing. Figure 5-11 shows the mounting arrangement and electrical connections to this assembly.
  • Page 132 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS REPLACEMENT PROCEDURES (continued) FABIUS GS O2 Whistle Language Date Format Time Format O2 Position Alarm Sounds Pres. Units Disabled English M/D/Y 12 Hour Right Drager cmH2O UK/Ireland...
  • Page 133 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS REPLACEMENT PROCEDURES (continued) 5.12 Pneumatic (PEEP Control) Assembly The Pneumatic assembly is accessible by removing the back cover of the controller housing. Figure 5-12 shows the mounting, tubing and electrical connections for this assembly.
  • Page 134 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS REPLACEMENT PROCEDURES (continued) FABIUS GS SV00459 CONTROLLER PCB LS1 SENSOR CONNECTION PUMP CABLE CONNECTION CAPTIVE MOUNTING SCREW REAR MOUNTING PEEP VALVE SCREW AND CONNECTION CABLE CLAMP PEEP HOSE...
  • Page 135 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS REPLACEMENT PROCEDURES (continued) 5.13 Flow Meter Bezel Assembly Following is the procedure for removing and replacing the flow meter bezel assembly. See Figure 5-13. This assembly comprises the fresh gas flow meter, flow control valves, flow sensors and filter, pipeline pressure gauges, SORC, and the fresh gas display PCB.
  • Page 136 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS REPLACEMENT PROCEDURES (continued) FABIUS GS SV00506 FLOWMETER BEZEL ASSEMBLY RIBBON CABLE FLOWSENSOR CONNECTION (TYPICAL) FLOWMETER BEZEL ASSEMBLY RETAINER SCREWS BACK VIEW OF MACHINE (PARTIAL) Figure 5-13. Flowmeter Bezel Assembly 5-40 Rev.
  • Page 137 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS REPLACEMENT PROCEDURES (continued) 5.14 Fresh Gas Flow Meter Access to the fresh gas flow meter requires removal of the flow meter bezel assembly as described in a previous paragraph. Figure 5-14 shows the tubing and mounting arrangement for the fresh gas flow tube.
  • Page 138 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS REPLACEMENT PROCEDURES (continued) FABIUS GS BEZEL ASSEMBLY SV00478 TUBE TO VAPOR FLOW MOUNT TUBE TUBE FROM FLOW SENSOR MANIFOLD FLOW TUBE MOUNTING BRACKET (2X) FLOW TUBE MOUNTING NUT LUER FITTING AND COLLAR (2X) Figure 5-14.
  • Page 139 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS REPLACEMENT PROCEDURES (continued) 5.15 Flow Control Valves Flow control valves can be replaced from the front panel while the flow meter bezel assembly is in place. Figure 5-15 shows the hardware arrangement for a typical flow control valve.
  • Page 140 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS REPLACEMENT PROCEDURES (continued) FABIUS GS COVER SPRING WASHER FLAT WASHER KNOB CHUCK COLLET SPACER FLOW CONTROL VALVE FLOW METER BEZEL ASSEMBLY VALVE MANIFOLD SV00485 Figure 5-15. Flow Control Valves...
  • Page 141 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS REPLACEMENT PROCEDURES (continued) 5.16 Fresh Gas Flow Sensors and Filter Assembly Access to the flow sensors and filter assembly requires removal of the flow meter bezel assembly as described in a previous paragraph. Figure 5-16 shows the tubing and mounting arrangement for this sub-assembly.
  • Page 142 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS REPLACEMENT PROCEDURES (continued) FABIUS GS FLOW SENSOR BRACKET SAFETY VALVE FLOW SENSOR (3X) FILTER MANIFOLD SORC BYPASS VALVE FLOWMETER BEZEL ASSEMBLY SV00484 Figure 5-16. Fresh Gas Flow Sensors and Filter Assembly...
  • Page 143 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS REPLACEMENT PROCEDURES (continued) 5.17 Pipeline Pressure Gauges Access to the pipeline pressure gauges requires removal of the flow meter bezel assembly as described in a previous paragraph. Figure 5-17 shows the tubing and mounting arrangement for the gauges.
  • Page 144 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS REPLACEMENT PROCEDURES (continued) FABIUS GS FLOW METER SV00481 BEZEL ASSEMBLY TUBES FROM GAS INLET ASSEMBLY STRAIGHT FITTINGS (3X) RETAINER PLATE PIPELINE PRESSURE GAUGE (3X) SCREW (4X) GAUGE RETAINER PLATE GAUGE MOUNTING NUT Figure 5-17.
  • Page 145 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS REPLACEMENT PROCEDURES (continued) 5.18 SORC (Sensitive Oxygen Ratio Controller) Access to the SORC requires removal of the flow meter bezel assembly as described in a previous paragraph. Figure 5-18 shows the tubing and mounting arrangement for the SORC.
  • Page 146 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS REPLACEMENT PROCEDURES (continued) FABIUS GS Figure 5-18. SORC 5-50...
  • Page 147 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS REPLACEMENT PROCEDURES (continued) 5.19 Fresh Gas Display PCB Access to the fresh gas display PCB requires removal of the flow meter bezel assembly as described in a previous paragraph. Figure 5-19 shows the cable connections and PCB mounting details.
  • Page 148 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS REPLACEMENT PROCEDURES (continued) FABIUS GS FLOWMETER SV00483 BEZEL ASSEMBLY PCB MOUNTING SCREWS AND LOCKWASHERS (4X) RIBBON CABLE FROM CONTROLLER (J1) INSULATOR FRESH GAS DISPLAY PCB CABLE FROM O2 SUPPLY PRESSURE SWITCH (J3) Figure 5-19.
  • Page 149 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS REPLACEMENT PROCEDURES (continued) 5.20 Monitor Bezel Assembly The monitor bezel assembly comprises the keypad, flat panel display, rotary switch, and LED lamp PCB assembly. Access to the bezel assembly mounting screws is through the ventilator compartment as shown in Figure 5-20.
  • Page 150 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS REPLACEMENT PROCEDURES (continued) FABIUS GS SV00463 SEE TEXT BEZEL SCREWS GROUND WIRE KEYPAD CONNECTOR LOCKING TABS DISPLAY PANEL CONNECTOR Figure 5-20. Monitor Bezel Assembly 5-54...
  • Page 151 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS REPLACEMENT PROCEDURES (continued) 5.21 Spirolog Sensor and Cable 5.21.1 Set machine power switch to OFF and disconnect AC power from the machine. 5.21.2 Remove the expiratory hose terminal from the breathing system, and slide the Spirolog sensor out of the breathing system.
  • Page 152 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS...
  • Page 153 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS ADJUSTMENT AND CALIBRATION PROCEDURES ADJUSTMENT AND CALIBRATION PROCEDURES Equipment Required: • Test Gauge for setting cylinder pressure regulator, DMI Part No. 4114807 • Test gauge (0 - 100 mbar) for setting SORC control pressure...
  • Page 154 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS ADJUSTMENT AND CALIBRATION PROCEDURES (continued) FABIUS GS Cylinder Pressure Regulator Adjustment 6.1.1 Set the System Power switch to OFF. NOTE: Minimum cylinder pressures for this adjustment shall be: N2O: 700 psi;...
  • Page 155 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS ADJUSTMENT AND CALIBRATION PROCEDURES (continued) SV00465 O2 REGULATOR REGULATOR ACORN NUT O2 PIPELINE HOSE TEST GAUGE Figure 6-1. Cylinder Pressure Regulator Adjustment: O2 Connection Illustrated...
  • Page 156 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS ADJUSTMENT AND CALIBRATION PROCEDURES (continued) FABIUS GS NOTE: Cylinder pressure compensation for the N2O regulator is not required. Cylinder Pressure Compensated Regulator (psi) output tolerances (-4/+2) 2000...
  • Page 157 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS ADJUSTMENT AND CALIBRATION PROCEDURES (continued) Gas Inlet Regulator Output Adjustment O2 Inlet Regulator: 6.2.1 Bleed all cylinder and pipeline pressures. Disconnect all cylinders and pipeline hoses.
  • Page 158 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS ADJUSTMENT AND CALIBRATION PROCEDURES (continued) FABIUS GS 6.2.4 Reconnect the O2 pipeline supply hose to the O2 pipeline inlet connector and pressurize the O2 supply. 6.2.5 Set the O2 flow to 4 L/min.
  • Page 159 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS ADJUSTMENT AND CALIBRATION PROCEDURES (continued) Oxygen Supply Pressure Alarm Switch Adjustment 6.3.1 Disconnect all pipeline hoses and set the System Power switch to ON. 6.3.2 Close all cylinder valves except the O valve.
  • Page 160 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS ADJUSTMENT AND CALIBRATION PROCEDURES (continued) FABIUS GS SIDE VIEW SV00504 OF GAS INLET ASSEMBLY SUPPLY SWITCH ADJUSTMENT SCREW O2 PIPELINE HOSE TEST GAUGE Figure 6-3. Oxygen Supply Pressure Alarm Switch Adjustment...
  • Page 161 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS ADJUSTMENT AND CALIBRATION PROCEDURES (continued) Sensitive Oxygen Ratio Controller (SORC) Adjustment 6.4.1 Set the System Power switch to OFF. 6.4.2 Remove the two screws at the back of the machine securing the flow meter bezel assembly.
  • Page 162 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS ADJUSTMENT AND CALIBRATION PROCEDURES (continued) FABIUS GS SV00476 Figure 6-4. Sensitive Oxygen Ratio Controller Adjustment Rev. A 6-10...
  • Page 163 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS ADJUSTMENT AND CALIBRATION PROCEDURES (continued) 6.4.14 The N2O flow should be 3 ±0.3 mL/min, resulting in an O2 concentration of 25 ± 3 vol. % O2 at the fresh gas outlet.
  • Page 164 ‘Calibrate O2 Sensor’ softkey shall illuminate. Following calibration the screen returns to normal. NOTE: If an “O2 Sensor Calibration Failed” message appears, refer to the oxygen monitoring section of the Fabius GS Operator’s Instruction Manual for further information. 6.5.3.8 Reinstall the O2 sensor assembly in the inspiratory valve dome.
  • Page 165 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS ADJUSTMENT AND CALIBRATION PROCEDURES (continued) OXYGEN SENSOR HOUSING SENSOR CAPSULE SENSOR HOUSING COVER INSPIRATORY VALVE DOME SV00492 Service 9:34 O2 Calibration Current cell A: Current cell B: Zero calibration procedure: - Remove O2 cell from housing.
  • Page 166 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS ADJUSTMENT AND CALIBRATION PROCEDURES (continued) FABIUS GS Pressure Calibration 6.6.1 Set the main power switch to ON. 6.6.2 Access the main service screen (ref. Section 4). 6.6.3 Scroll to Calibration and select Pressure to bring up the pressure calibration screen.
  • Page 167 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS ADJUSTMENT AND CALIBRATION PROCEDURES (continued) Fresh Gas Flow Calibration 6.7.1 Set the main power switch to ON. 6.7.2 Access the main service screen (ref. Section 4).
  • Page 168 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS ADJUSTMENT AND CALIBRATION PROCEDURES (continued) FABIUS GS PEEP Valve Calibration NOTE: Before performing the PEEP valve calibration, a valid pressure calibration and system leak test must be performed first. Otherwise an ‘Inv pres.
  • Page 169 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS ADJUSTMENT AND CALIBRATION PROCEDURES (continued) INSPIRATORY VALVE TERMINAL EXPIRATORY SV00494 VALVE TERMINAL Service 10:06 PEEP Calibration Calibration: PASS PEEP calibration procedure: - Attach breathing hose from expiratory valve to inspiratory valve and remove breathing bag.
  • Page 170 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS ADJUSTMENT AND CALIBRATION PROCEDURES (continued) FABIUS GS Vacuum Adjustment The vacuum adjustment is located on the pneumatic assembly and is accessible by removing the back cover of the controller housing.
  • Page 171 Fabius GS Part Replacement Schedule ........
  • Page 172 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS Test Equipment Required: -- Multi-Meter (Fluke or equivalent) -- Electrical Safety Analyzer (Biotek 501 Pro or equivalent) -- Test Pressure Gauge, P/N 4114807 or equivalent...
  • Page 173 Annually: Fabius GS 1 year kit 4117360-001 4199912 Filter, Pressure & Pneumatic asm 8402868 Diaphragm, Patient 2600650 O-ring, Vapor 4115864 Every 3 Years: Fabius GS 3 year kit 4117360-002 4199911 Diaphragm, Piston 2600651 O-ring, for Diaphragm G60337 Battery, 12V Rechargeable 4114229...
  • Page 174 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS 4113119 FRESH GAS LEAK TEST DEVICE CONNECT TO TEST PRESSURE GAUGE TEST CONNECTOR 4117361 SV00517...
  • Page 175 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) 4115128 SIEMENS TEST LUNG SVOOO25...
  • Page 176 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS S010159 BREATHING SYSTEM LEAK TEST DEVICE 4104389 TEST TERMINAL ADAPTER SV00515...
  • Page 177 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) SV00559 S000081 FLOW METER TEST STAND 0-250cc/MIN...
  • Page 178 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS SV00552 Liter 2212300 vol/min MINUTE VOLUMETER TEST VOLUME METER 4115087 33mm x 22mm ADAPTER...
  • Page 179 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) M33972 TEST CAP SV00516...
  • Page 180 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS 4114807 PRESSURE TEST ASSEMBLY , WITH ADAPTERS SV00376 4114830-002 4114830-001 4114830-004 4114830-003 4114830-006 4114830-005 7-10...
  • Page 181 Also included are sample PMC labels marked to show several levels of certifications. An excerpt from Draeger Medical, Inc.’s Anesthesia System Risk Analysis and Risk Reduction is included, and also a sample of an Executive Summary to be furnished to the hospital’s Risk Manager or Chief of Anesthesia.
  • Page 182 RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS The letters A, B, C, D and the Roman Numerals I, II are used as codes in the individual matrix for each model of anesthesia machine. The letters A, B, C, and D are used in descending order to indicate the certification level of the equipment.
  • Page 183 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) A Narkomed AMIII or Narkomed II could have recommendations 6 apply, and failure code 46.2. 6 - Oxygen ratio monitor instead of oxygen ratio monitor controller. Code D II.
  • Page 184 PMS PROCEDURE (continued) FABIUS GS PM Certification Procedure for Fabius GS Anesthesia System 1. Use the PM Certification form P/N 4117715 for the Fabius GS Anesthesia System. 2. Completely fill in the header information. 3. Perform the vapor concentration test on all Dräger vapor vaporizers every six months in accordance with SP00073 at a six month maximum interval.
  • Page 185 10. Have the customer sign each PM Certification form or the Executive Summary, and review the equipment conditions and recommendations with the customer. 11. Return the top copy to Draeger Medical, Inc. Service Department, keep middle copy for service organization records, give bottom copy to customer.
  • Page 186 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS ELECTRICAL SAFETY ( ü ü ü ü ) 7.1.1 Protective Ground Continuity 7.1.1.1 Turn the System Power switch to STANDBY. Plug the unit into the safety analyzer, and plug the power cord of the safety analyzer into an AC receptacle.
  • Page 187 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) ( ü ü ü ü ) 7.1.3 Convenience Outlet and Auxiliary Outlet Strip (if applicable) NOTE: This test will check the convenience outlets and the auxiliary strip outlets for fault conditions such as open ground (>31 Ω...
  • Page 188 Main Power Main Power Pass Pass Battery Battery Pass Pass Fabius GS SW 1.03 CRC 6DBA Fabius GS SW XX ( ü ü ü ü ) 7.2.4 Record the machine software version on the header of the checklist form. 7-18...
  • Page 189 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) ( ü ü ü ü ) BATTERY CIRCUIT 7.3.1 Press the Man Spont key, and press the rotary dial to confirm. 7.3.2 Unplug the AC power cord and verify the “Power Fail”...
  • Page 190 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS ( ü ü ü ü ) CONFIGURATION 7.4.1 Press tje Standby key, then press the rotary key to confirm. 7.4.2 Press the (ok) key, and press the rotary dial to confirm.
  • Page 191 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) 7.6.5 Press the Exit key. Pressure: 7.6.6 Scroll to Pressure and confirm. ( ü ü ü ü ) 7.6.7 Follow the on-screen instructions and perform the Zero calibration.
  • Page 192 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS O2 Whistle: ( ü ü ü ü ) 7.7.5 Press to confirm, scroll to O2 whistle. Verify O2 whistle selection (enabled or disabled) is in accordance with local requirements or customer demand.
  • Page 193 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) Elevation Range in Barometric Pressure Elevation Range in Feet Meters Setting -200 - 199 -60 - 60 1013 200 - 599 61 - 182...
  • Page 194 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS Scavenger, AGS Scavenger Cleaning 7.8.1 Remove all scavenger hoses one at a time and drain all accumulated moisture. Inspect all scavenger hoses for deterioration and replace any worn hoses.
  • Page 195 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) BREATHING SYSTEM ( ü ü ü ü ) 7.9.1 Breathing System Inspection ( ü ü ü ü ) 7.9.1.1 Record the serial number (located on the side of the COSY) on the PMS form.
  • Page 196 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS 7.9.1.16 Examine all pneumatic hoses connecting from the interface panel to the breathing system. Are the hoses kink free and in good condition? ___(Y) 7.9.1.17 Inspect the breathing circuit and manual breathing bag.
  • Page 197 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) ( ü ü ü ü ) 7.9.2.11 If applicable, turn on the right mounted vaporizer to the first graduated marking; repeat Steps 7.9.2.5 thru 7.9.2.10 ___ (>40 cm H 7.9.2.12 Remove the test equipment from the fresh gas hose.
  • Page 198 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS 7.9.3 Breathing System Leak NOTE: Use the appropriate column in the following table - according to the software version of the machine you are testing: Software ≥...
  • Page 199 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) SET APL VALVE TO MAN, 70 cmH2O SV00549a INSTALL TEST CAP M33972 ON VENTILATOR HOSE TERMINAL (UNDERSIDE) TEST TERMINAL 4104389 12 IN. TEST HOSE...
  • Page 200 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS 7.9.4 APL Valve Verification 7.9.4.1 Connect a digital manometer between the inspiratory and expiratory ports of the breathing system (using two test terminals and modified 4113119 fresh gas leak test device).
  • Page 201 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) Lever Style APL Valve Rotary Knob Style APL Valve 7.9.4.3 Verify the fresh gas hose is connected to the breathing system. 7.9.4.4 Verify the APL valve is in the MAN position and set to 10 cmH2O.
  • Page 202 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS 7.9.5 Inhalation and Exhalation Valves Inhalation: 7.9.5.1 Install test terminal on the 22 mm inspiratory connector. 7.9.5.2 Connect 0 - 250 cc flowmeter (S000081) with a test terminal as shown, to the ventilator port of the breathing system.
  • Page 203 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) 7.9.5.3 Remove the freshgas hose from the breathing system and connect it to the freshgas leak connector (4117163). 7.9.5.4 Connect the fresh gas leak connector to the fresh gas leak test device (4113119 modified) as shown.
  • Page 204 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS Exhalation: 7.9.5.8 Occlude ventilator and inspiratory port, each with test cap M33972. 7.9.5.9 Interconnect digital manometer and 0-250 cc test flowmeter to bag terminal as shown.
  • Page 205 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) 7.9.6 Reverse Flow Test (Fresh Gas Decoupling Valve) 7.9.6.1 Verify the breathing bag port, fresh gas connection and expiration port connections are open and not occluded.
  • Page 206 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS 7.9.6.5 Verify the APL valve is in the MAN position and set to 70 cmH2O. 7.9.6.6 Slowly increase O2 flow to maintain 40 cmH2O on the digital pressure meter.
  • Page 207 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) 7.9.7 Leakage Control Port 7.9.7.1 Disconnect the APL and PEEP control hoses from the interface panel at rear of machine. 7.9.7.2 Connect the PEEP control hose to the input of the 0-250 cc test flowmeter.
  • Page 208 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS 7.9.7.3 Short circuit the inspiratory and expiratory ports, attach a test terminal to the breathing bag port and attach a digital manometer to the test terminal.
  • Page 209 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) ( ü ü ü ü ) 7.10 VAPOR INTERLOCK SYSTEM 7.10.1 Set all vapors to zero (0). 7.10.2 Adjust the handwheel on the left vapor to any concentration above zero (0).
  • Page 210 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS 7.11 YOKES & GAUGES ( ü ü ü ü ) 7.11.1 Yokes & Check Valves (if applicable) 7.11.1.1 Turn the System Power switch Off.
  • Page 211 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) ( ü ü ü ü ) 7.11.2 Cylinder Gauges (if applicable) 7.11.2.1 Are the pressure gauges correct for the gases indicated by the flowmeters? ___ (Y) 7.11.2.2 Bleed all pressure from the cylinder circuits using the flow...
  • Page 212 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS 7.12 Gas Inlet Regulator Output 7.12.1 O2 Inlet Regulator 7.12.1.1 Bleed all cylinder and pipeline pressures. Disconnect all cylinders and pipeline hoses.
  • Page 213 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) ( ü ü ü ü ) 7.12.1.5 After the digital manometer stabilizes, what is the regulator output pressure? ___psi (28 - 32 psi) 7.12.1.6 Deplete O2 pressure from the pipeline supply.
  • Page 214 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS 7.12.3 Pipeline Check Valves O2 Pipeline Check Valve: 7.12.3.1 Attach a Swagelock fitting (P/N 4109318) to the inlet of the S000081 test flowmeter.
  • Page 215 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) N2O Pipeline Check Valve (if applicable) 7.12.3.5 Attach the appropriate N2O adapter to the N2O connector on the inlet block, and transfer the end of the hose with the quick disconnect fitting to the N2O adapter.
  • Page 216 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS 7.13 Cylinder Regulator & Pipeline Gauges Minimum cylinder pressure requirements for this test are: N2O: 600 psi; O2, Air: 1000 psi. ( ü ü ü ü ) 7.13.1...
  • Page 217 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) 7.13.1.2 Connect the test fixture hose to the machine’s nitrous oxide pipeline inlet. 7.13.1.3 Does the back panel correctly identify the nitrous oxide inlet? ___(Y) 7.13.1.4 Connect the nitrous oxide pipeline supply hose to the test fixture.
  • Page 218 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS 7.13.3 N2O Pipeline Leak 7.13.3.1 Close the nitrous oxide flow control valve. 7.13.3.2 Release the test device push button. ( ü ü ü ü ) 7.13.3.3 After 30 seconds, what is the pressure loss? ___psi (<5)
  • Page 219 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) ( ü ü ü ü ) 7.13.5 Air Pipeline Gauge Accuracy 7.13.5.1 Close the air cylinder valve and drain all air pressure.
  • Page 220 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS ( ü ü ü ü ) 7.13.7.7 Release the push button. After the pressure decay stabilizes, what is the regulator output pressure? ___psi (based upon...
  • Page 221 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) 7.14 HIGH PRESSURE LEAK 7.14.1 Oxygen High Pressure Leak 7.14.1.1 Verify the System Power switch is OFF. 7.14.1.2 Open the oxygen cylinder valve.
  • Page 222 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS 7.14.3 Air High Pressure Leak 7.14.3.1 Turn the System Power switch to ON. 7.14.3.2 Open one (1) oxygen cylinder valve and the air cylinder valve.
  • Page 223 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) 7.15 OXYGEN SUPPLY FAILURE PROTECTION ( ü ü ü ü ) 7.15.1 Nitrous Oxide (if applicable) 7.15.1.1 Open and close the oxygen cylinder valve.
  • Page 224 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS 7.15.2 Oxygen Supply Pressure Alarm 7.15.2.1 If not already connected, connect the Pressure Test Assembly (P/N 4114807) with O adapters between the central supply hose and the machine.
  • Page 225 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) 7.16 FLOWMETERS ( ü ü ü ü ) 7.16.1 Oxygen Flowmeter 7.16.1.1 Open the O cylinder valve. 7.16.1.2 Is it possible to adjust the flow of oxygen to 10 L/min.? ___ (Y) 7.16.1.3 Close the O...
  • Page 226 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS ( ü ü ü ü ) 7.16.3 Air Flowmeter 7.16.3.1 Connect the Air pipeline supply (if applicable). 7.16.3.2 Is it possible to adjust the flow of air to 10 L/min.? ___ (Y) 7.16.3.3 Close the air flow control valve and disconnect the Air pipeline...
  • Page 227 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) ( ü ü ü ü ) 7.17 OXYGEN MONITOR 7.17.1 Press the Man Spont key and press the rotary knob to confirm.
  • Page 228 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS 7.17.18 Verify that the high alarm limit has a range from 100 to 19%. 7.17.19 Set the high alarm limit to below the correct O2 reading.
  • Page 229 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) 7.18 OXYGEN CONCENTRATIONS ( ü ü ü ü ) 7.18.1 Oxygen + Nitrous Oxide Concentration (if applicable) 7.18.1.1 Verify the oxygen flow is at 4 L/min.
  • Page 230 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS 7.19 SORC 7.19.1 Depress the O FLUSH for 5 seconds. 7.19.2 Set the O2 and N2O flow to 10 L/min. 7.19.3 Set the O2 flow control valve to 0.8 L/min.
  • Page 231 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) ( ü ü ü ü ) 7.20 PRESSURE MONITOR 7.20.1 Verify unit is in Man Spont mode. 7.20.2 Disconnect the breathing pressure sensor line from the absorber.
  • Page 232 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS ( ü ü ü ü ) 7.20.16 At what pressure does the APNEA PRESSURE alarm deactivate? ___ cm H O (7-13) 7.20.17 Adjust the threshold to 18 cm H 7.20.18 Increase the pressure to 20 cm H...
  • Page 233 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) 7.21 VENTILATOR ( ü ü ü ü ) 7.21.1 Manual Ventilation 7.21.1.1 Cycle machine power-using On/Off switch and allow completion of the self-test.
  • Page 234 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS ( ü ü ü ü ) 7.21.4 Ventilator Delivery 7.21.4.1 Set APL to Man position and set to 30 cmH2O. 7.21.4.2 Switch to Volume Control mode and set the fresh gas flow to 3 L/ min.
  • Page 235 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) 7.21.6 Pmax Accuracy 7.21.6.1 Connect a test pressure gauge in line with the pressure port connector of breathing system. 7.21.6.2 Remove test lung and seal off Y-piece.
  • Page 236 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS 7.21.9 Remove ventilator hose from ventilator and connect a digital manometer to the ventilator outlet port. ( ü ü ü ü ) 7.21.10 Verify relief valve opens between 70 and 80 mbar (cmH2O) when...
  • Page 237 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) 7.21.11 Auxiliary Air Valve Test ( ü ü ü ü ) 7.21.11.1 Close all flow control valves. Verify vacuum valve opens at the pressure indicated in the following table, when ventilator is in expiratory phase.
  • Page 238 FRESH GAS LEAK TEST SV00548 DEVICE 4113119 (MODIFIED) 7.21.13 Vacuum Pressure 7.21.13.1 Set Fabius GS to Volume mode. 7.21.13.2 Remove APL bypass hose from breathing system and attach to digital manometer. ( ü ü ü ü ) 7.21.13.3 Verify vacuum is between -150 and -240 cmH2O (mbar) (-110 and -175 mmHg) as shown on digital manometer.
  • Page 239 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) ( ü ü ü ü ) 7.22 Volume Alarms 7.22.1 Set ventilator to Volume Control mode and confirm. 7.22.2 Set: pmax = 40 mbar...
  • Page 240 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS 7.22.17 Fresh Gas Low 7.22.17.1 Select Volume Control mode and confirm. 7.22.17.2 Disconnect the hose attached to the breathing system’s inspiratory port.
  • Page 241 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) ( ü ü ü ü ) 7.23 AUDIO SILENCE 7.23.1 Press the key. 7.23.2 Does the LED on the Silence Alarms key light? Is the audio alarm silenced for 120 sec.?
  • Page 242 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS PMS PROCEDURE (continued) FABIUS GS ( ü ü ü ü ) 7.24 OXYGEN FLUSH VALVE 7.24.1 Press and release the O FLUSH button. 7.24.2 Does the flow of oxygen stop immediately? __ (Y) 7.24.3...
  • Page 243 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS PMS PROCEDURE (continued) 7.25 FINAL TESTS ( ü ü ü ü ) 7.25.1 Operator's Instruction Manual 7.25.1.1 Verify that the availability/location of the machine's Operator's Instruction Manual is in close proximity of the machine.
  • Page 244 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS...
  • Page 245 SOFTWARE UPDATE PROCEDURE This section outlines the software installation procedure, including the equipment needed and its connections. Software updates to the Fabius GS anesthesia system are done through a serial port connection to an external PC using the batch file LOADGS.BAT. REQUIREMENTS NOTE: U.S.
  • Page 246 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS SOFTWARE UPDATE PROCEDURE (continued) FABIUS GS 8.2.2 Minimize all programs and check desktop area for the following icon. 8.2.3 Using Windows Explorer double click on the appropriate drive.
  • Page 247 RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS SOFTWARE UPDATE PROCEDURE (continued) 8.2.7 Choose ‘Next’ to the following screen. 8.2.8 The following screen shall appear. Choose ‘Finish’. Proceed to Section 8.4 ‘Installing Fabius GS Software from a PC’. Rev. F...
  • Page 248 Double click on the ‘copy_si.bat’ file. This updates your current ‘Send Image’ file to the latest version. INSTALLING FABIUS GS SOFTWARE FROM A PC NOTE: The screen illustrations contained in this document are for reference only and therefore may or may not reflect the software version being installed.
  • Page 249 FABIUS GS SOFTWARE UPDATE PROCEDURE (continued) 8.4.3 Press Enter or click OK. Connect the download cable from Com1 on the Figure laptop computer to the serial interface port on the Fabius GS. See REAR VIEW OF FABIUS GS SERIAL INTERFACE PORT SV00524 FIGURE 8-1.
  • Page 250 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS SOFTWARE UPDATE PROCEDURE (continued) FABIUS GS 8.4.4 From the drop-down box, select the CD Drive. 8.4.5 Double click the ‘software image’ folder (fgs_vxxx). The ‘Fabius GS.IMG’ is displayed. Rev. F...
  • Page 251 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS SOFTWARE UPDATE PROCEDURE (continued) 8.4.6 Double-click the ’.IMG’ file or single-click (highlight), then press ‘Enter’. The following screen will be displayed: 8.4.7 Set the machine’s power switch to ON. The download process will start immediately.
  • Page 252 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS SOFTWARE UPDATE PROCEDURE (continued) FABIUS GS In Progress Fabius GS Image Loader Loading programming blocks FIGURE 8-2. Image Loader Screen Complete - Successful Fabius GS Image Loader...
  • Page 253 SOFTWARE UPDATE PROCEDURE (continued) 8.4.9 Set the power switch on the Fabius GS to OFF. Close the PC loader screen and power down the PC. Disconnect the interface cable. 8.4.10 Switch machine power to ON and verify all self-diagnostics indicate a Pass condition.
  • Page 254 Fabius GS boot Strap load function to the original SENDIMG.EXE program that is currently used in existing products. When a Fabius GS does not have a valid bootable image installed, it can not perform an Automatic Image Reload Operation. The same PC and files are used (including SENDIMG.EXE V2).
  • Page 255 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS SOFTWARE UPDATE PROCEDURE (continued) 8.5.3 Remove the back panel and change the JP2 connector on the control PCB from the Boot Flash position to the Boot Strap position 8.5.4...
  • Page 256 8.5.14 Reinstall the back panel of the control unit. 8.5.15 Access the appropriate service screens (see Section 7.5 of the Fabius GS Service Manual) and update the machine serial number, clear service log and reset hours run. The machine serial number is located on the rear of the machine.
  • Page 257 SOFTWARE UPDATE PROCEDURE (continued) 8.5.16 Perform PMS test Steps 7.6, 7.7, 7.17, 7.18, 7.20 and 7.21 in Section 7 of the Fabius GS Service Manual. Verify the correct market kit settings are entered per original machine configuration or customer requirements.
  • Page 258 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS...
  • Page 259 SPARE AND REPLACEMENT PARTS 9.0 Spare and Replacement Parts Part numbers for field-replaceable items on the Fabius GS anesthesia system are listed on the following pages, along with part numbers for related hardware and cables. The item numbers are keyed to the accompanying illustrations to aid in identifying the item and its location.
  • Page 260 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS SPARE AND REPLACEMENT PARTS (continued) FABIUS GS WRITING TRAY ASSEMBLY SV00522 Rev. D...
  • Page 261 Tray Top, Fabius GS ........
  • Page 262 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS SPARE AND REPLACEMENT PARTS (continued) FABIUS GS SV00558 Rev. B...
  • Page 263 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS SPARE AND REPLACEMENT PARTS (continued) ITEM DESCRIPTION PART NUMBER 1. Spirolog flow sensor, 5 pack ......... . . 8403735 2.
  • Page 264 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS SPARE AND REPLACEMENT PARTS (continued) FABIUS GS FRESH GAS DECOUPLING VALVE APL BYPASS VALVE PEEP/PMAX VALVE SV00508 Rev. F...
  • Page 265 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS SPARE AND REPLACEMENT PARTS (continued) ITEM DESCRIPTION PART NUMBER Breathing system assembly w/APL bypass........4116398 Breathing system assembly w/APL bypass, Service Exchange .
  • Page 266 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS SPARE AND REPLACEMENT PARTS (continued) FABIUS GS SV00418 Rev. C...
  • Page 267 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS SPARE AND REPLACEMENT PARTS (continued) ITEM DESCRIPTION PART NUMBER 1. Absorber top ............M33013 2.
  • Page 268 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS SPARE AND REPLACEMENT PARTS (continued) FABIUS GS SV00419 Rev. C 9-10...
  • Page 269 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS SPARE AND REPLACEMENT PARTS (continued) ITEM DESCRIPTION PART NUMBER 1. Canister assembly ........... M29320 Screen insert .
  • Page 270 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS SPARE AND REPLACEMENT PARTS (continued) FABIUS GS SV00547 0.19 M [7 1/2 IN] 0.33 M [13 IN] 0.94 M [37 IN] Rev. F 9-12...
  • Page 271 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS SPARE AND REPLACEMENT PARTS (continued) ITEM DESCRIPTION PART NUMBER 1. Breathing system mount assembly ........4117400 Gauge assembly, patient airway pressure .
  • Page 272 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS SPARE AND REPLACEMENT PARTS (continued) FABIUS GS SV00557 9-14...
  • Page 273 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS SPARE AND REPLACEMENT PARTS (continued) ITEM DESCRIPTION PART NUMBER Mount, auxiliary assembly, fresh gas, left (illustrated) ......4117403 Mount, auxiliary assembly, fresh gas, right .
  • Page 274 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS SPARE AND REPLACEMENT PARTS (continued) FABIUS GS SV00477 9-16...
  • Page 275 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS SPARE AND REPLACEMENT PARTS (continued) ITEM DESCRIPTION PART NUMBER Fresh Gas Flow Meter [part of flow meter bezel assembly 4116543 (3-gas) or 4116542 (2-gas)] 1. Flowmeter tube assembly (incl. brackets and mounting nuts) ....4116950 2.
  • Page 276 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS SPARE AND REPLACEMENT PARTS (continued) FABIUS GS SV00479 Rev. B 9-18...
  • Page 277 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS SPARE AND REPLACEMENT PARTS (continued) ITEM DESCRIPTION PART NUMBER Flow control valves and valve manifold [part of flow meter bezel assembly 4116543 (3-gas) or 4116542 (2-gas)] 1.
  • Page 278 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS SPARE AND REPLACEMENT PARTS (continued) FABIUS GS Rev. B 9-20...
  • Page 279 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS SPARE AND REPLACEMENT PARTS (continued) ITEM DESCRIPTION PART NUMBER Fresh gas flow sensors and filter assembly [part of flow meter bezel assembly 4116543 (3-gas) or 4116542 (2-gas)] 1.
  • Page 280 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS SPARE AND REPLACEMENT PARTS (continued) FABIUS GS SV00480 9-22...
  • Page 281 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS SPARE AND REPLACEMENT PARTS (continued) ITEM DESCRIPTION PART NUMBER Pipeline pressure gauges [part of flow meter bezel assembly 4116543 (3-gas) or 4116542 (2-gas)] 1. Plate, pipeline gauge ..........4116987 2.
  • Page 282 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS SPARE AND REPLACEMENT PARTS (continued) FABIUS GS SORC SV00456 Rev. A 9-24...
  • Page 283 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS SPARE AND REPLACEMENT PARTS (continued) ITEM DESCRIPTION PART NUMBER SORC [part of flow meter bezel assembly 4116543 (3-gas) or 4116542 (2-gas)] 1. Kep nut, M5 (4x)........... HW55013 2.
  • Page 284 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS SPARE AND REPLACEMENT PARTS (continued) FABIUS GS SV00482 9-26...
  • Page 285 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS SPARE AND REPLACEMENT PARTS (continued) ITEM DESCRIPTION PART NUMBER Fresh gas display PCB [part of flow meter bezel assembly 4116543 (3-gas) or 4116542 (2-gas)] 1. Insulator, flowmeter display ......... . .4116965 2.
  • Page 286 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS SPARE AND REPLACEMENT PARTS (continued) FABIUS GS SV00470 Rev. B 9-28...
  • Page 287 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS SPARE AND REPLACEMENT PARTS (continued) ITEM DESCRIPTION PART NUMBER Cylinder Yokes, Regulators, and Gauges: 1. O2 yoke (2x) ............1101620 N2O yoke.
  • Page 288 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS SPARE AND REPLACEMENT PARTS (continued) FABIUS GS SV00462 Rev. B 9-30...
  • Page 289 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS SPARE AND REPLACEMENT PARTS (continued) ITEM DESCRIPTION PART NUMBER Gas Inlet Assembly, 3-connector DISS (includes back panel) ....4116535-004 Alternate part number .
  • Page 290 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS SPARE AND REPLACEMENT PARTS (continued) FABIUS GS SV00521 Rev. D 9-32...
  • Page 291 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS SPARE AND REPLACEMENT PARTS (continued) ITEM DESCRIPTION PART NUMBER 1. PCB assembly, Service Exchange ........SE4116632 Alternate part number .
  • Page 292 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS SPARE AND REPLACEMENT PARTS (continued) FABIUS GS SV00519 9-34...
  • Page 293 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS SPARE AND REPLACEMENT PARTS (continued) ITEM DESCRIPTION PART NUMBER Pneumatic (PEEP Control) Assembly ........4117004 1.
  • Page 294 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS SPARE AND REPLACEMENT PARTS (continued) FABIUS GS SV00461 Rev. B 9-36...
  • Page 295 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS SPARE AND REPLACEMENT PARTS (continued) ITEM DESCRIPTION PART NUMBER Bezel Assembly, Monitor ..........4117019 1.
  • Page 296 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS SPARE AND REPLACEMENT PARTS (continued) FABIUS GS 10 10A SV00512 Rev. F 9-38...
  • Page 297 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS SPARE AND REPLACEMENT PARTS (continued) ITEM DESCRIPTION PART NUMBER Ventilator (w/-8 mbar pressure relief valve) ......4117080-001 Ventilator, Service Exchange (w/-8 mbar pressure relief valve) .
  • Page 298 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS SPARE AND REPLACEMENT PARTS (continued) FABIUS GS Rev. B 9-40...
  • Page 299 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS SPARE AND REPLACEMENT PARTS (continued) ITEM DESCRIPTION PART NUMBER 1. Auxiliary O2 Flow Meter, USA, complete ......4109310-003 Auxiliary O2 Flow Meter, ISO .
  • Page 300 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS SPARE AND REPLACEMENT PARTS (continued) FABIUS GS Rev. D 9-42...
  • Page 301 NOTE: The adapter fitting is needed when ordering a two-vapor mounting system as a replacement part for Fabius GS units with serial numbers < 10685. O-rings, vaporizer mount (4x) ........4115864 One-vapor mounting system .
  • Page 302 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS SPARE AND REPLACEMENT PARTS (continued) FABIUS GS SV00560 9-44...
  • Page 303 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS SPARE AND REPLACEMENT PARTS (continued) ITEM DESCRIPTION PART NUMBER Scavenger, AGS ............M33300 1.
  • Page 304 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS SPARE AND REPLACEMENT PARTS (continued) FABIUS GS SV00598 9-46...
  • Page 305 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS FABIUS GS SPARE AND REPLACEMENT PARTS (continued) ITEM DESCRIPTION PART NUMBER AGS Scavenger w/adj valve ..........4117494 1.
  • Page 306 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS...
  • Page 307 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS...
  • Page 308 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS DrägerService is a division of Draeger Medical, Inc. 3122 Commerce Drive Telford, PA 18969 Tel: (215) 721-5402 (800) 543-5047 Fax: (215) 721-5784 Web: www.draegermedical.com Printed in the U.S.A.
  • Page 309 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS Fabius GS Service Manual Rev. F and G summary of changes Page Description Cover ......... Updated revision level 2-1 .
  • Page 310 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS 4-10 ....Added acceptable range info., changed PVent value 5-27 ....... . . Updated paras. 5.7.61 and 5.7.65 5-36 .
  • Page 311 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS 4-1, 4-2, 4-4 ....Revised Figs 4-1, 4-2, 4-4 to reflect sw 1.20 4-19 .
  • Page 312 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS...