  • Page 1 Wallaby warmer User Manual Infant Radiant Warmer Wallaby Effective date: 05 December 2024 Issue Number: 01 Version: 15 EN Wallaby. User manual 2265 Issue Date: 05 December 2024...
  • Page 2 Contents Company Information Warnings Introduction Safety Information Device Description Setting up Warmer Using Warmer Alarm System Cleaning Maintenance Disposal Specifications Explanation of Symbols Warranty Policy V-1-15-EN Wallaby. User manual...
  • Page 3 Do not leave the patient unattended under radiant warmer. Tel: +84 24 3766 6521 Fax: +84 24 3718 8050 Email: assistance@mtts-asia.com www.mtts-asia.com Logic s.r.l. Via Antonio Pigafetta 1 34147 Trieste, Italy SRN: IT-AR-000012301 V-1-15-EN Wallaby. User manual...
  • Page 4 • The presence of electro-surgical devices, shortwave or microwave equipment near this device The Wallaby Warmer device is intended to provide a controlled source of heat and maintain the body tem- could lead to electrical interference that negatively affects the function of the device.
  • Page 5 • The device is intended for use only by appropriately trained personnel and under the direction of qualified medical personnel who are familiar with currently known risks and benefits of radiant warmer use. Overhead Unit T-piece Resuscitator** Storage Shelf* IV Pole* Control Panel Side Panels Mattress Skin sensor X-ray Tray Storage Drawer* Cylinders Holder* Lockable Castors * Optional components ** Not included, illustration only V-1-15-EN Wallaby. User manual...
  • Page 6 Device Description Setting up Warmer Control Panel STEP 1 - Remove components from the box Wallaby Warmer functions are controlled by touch screen. Side Panel 54cm Base Side Panel 64cm Bed Mattress Stand Storage Shelves Bolt M10x80 Bolt M10x25 Stand plastic frame...
  • Page 7 STEP 7 - Attach 3 remaining Side Panels to Bed STEP 9 - Plug in and power up The connector plugs into the back of the Wallaby Warmer Control Unit and then directly into the power outlet in the wall.
  • Page 8 • A warmer cannot differentiate between hypothermia (low skin and low core temperature), and • The warmer cannot measure or control the baby’s skin temperature if the skin sensor is not fever (low skin and high core temperature). Ensure the clinical condition of the baby is regu- correctly positioned on the baby. larly reviewed. V-1-15-EN Wallaby. User manual...
  • Page 9 It may also be used to provide low power background heating when required. NOTE • The Wallaby Warmer does not adjust for patient temperature in PREWARM MODE. When a baby is placed under the warmer, either MANUAL or BABY MODE should be selected immediately. V-1-15-EN...
  • Page 10 The device will come back to original screen after 3 seconds. audible alarm will sound, and the heater power will be reduced to 25%. • If the ALARM MUTE button is now pressed, the Mode Screen flashing will be extinguished, the audible alarm silenced and the heater power will be restored at the set level. V-1-15-EN Wallaby. User manual...
  • Page 11 2. Turn the knob underneath the bed counter-clockwise to move the front of the bed down. 2. Touch MODE Selection panel to enter MODE Selection screen Baby bed is equipped with X-ray tray which is placed in the slot below it. 3. Press and hold the OFF symbol for 3 seconds V-1-15-EN Wallaby. User manual...
  • Page 12 • Dry all the surfaces after cleaning with a clean soft cloth or paper towel. DISCONNECT 25% of power alarm only • Ensure that no part of Wallaby Warmer or related components is immersed in any cleaning MANUAL Not active Not active Not active liquid or cleaning solution.
  • Page 13 • Ensure only approved replacement parts are used during service and maintenance procedures. 1. Clean Wallaby skin sensors with alcohol, or detergent or soap solution (maximum 2% in water), ensuring the manufacturer’s directions for use of the cleaning agent are followed. • Please contact an authorised MTTS representative for further assistance with any servicing or 2.
  • Page 14 1. Estimated lifetime of the Heater Element is 10,000 hours. 2. When the Wallaby Warmer main unit cannot detect Heater current (due to exceeded lifetime or Heater Element failure), error E3 is displayed. 3. Refer to the Technical Manual for Heater Element replacement instructions.
  • Page 15 10°C - 50°C / 0.1°C Setting range / resolution 34°C – 37.5°C / 0.1°C Temperature measurement Sensor accuracy / range ±0.3°C / 30°C – 40°C Power Cords North America (part 50-07302-01) C.E. (part 50-07305-01) U.K. (part 50-09273-01) V-1-15-EN Wallaby. User manual...
  • Page 16 MTTS authorized service provider can provide you with Consult instructions for use or the MTTS MAKES NO OTHER EXPRESS WARRANTY OR details. MTTS will not alter form, fit, or function of this Date of manufacture (YYYY-MM) electronic insturctions for use (ISO 15223-1)
  • Page 17 MTTS product has recurring failures, MTTS at THE LESSER OF THE PRICE YOU HAVE PAID FOR THE your sales receipt is the date of purchase unless MTTS or ing one of the following: Locate and contact your nearest MTTS service pro-...
  • Page 18 MEDICAL TECHNOLOGY TRANSFER AND SERVICES Co., LTD House No. 26, Alley 41, An Duong Vuong Street, Tay Ho District, Hanoi City, Vietnam SRN: VN-MF-000036669 Tel: +84 24 3766 6521 Fax: +84 24 3718 8050 Email: assistance@mtts-asia.com www.mtts-asia.com Logic s.r.l. Via Antonio Pigafetta 1 34147 Trieste, Italy...