Advertisement

Dolphin
CPAP
User Manual
2265
Effective date: April 2018
www.mtts-asia.com
Issue Number: 01 Version: 1.04EN
Dolphin CPAP. User manual
Issue Date: May 2018

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Dolphin CPAP and is the answer not in the manual?

Questions and answers

Summary of Contents for mtts Dolphin CPAP

  • Page 1 Dolphin CPAP User Manual 2265 Effective date: April 2018 www.mtts-asia.com Issue Number: 01 Version: 1.04EN Dolphin CPAP. User manual Issue Date: May 2018...
  • Page 2: Table Of Contents

    - Warnings (Pulse Oximeter) - Overview (Pulse Oximeter) - Setup (Pulse Oximeter) - Operation (Pulse Oximeter) - Alarm System (Pulse Oximeter) - Sensors & Patient Cables (Pulse Oximeter) - Maintenance (Pulse Oximeter) Specifications Explanation of Symbols Warranty Policy V.1.04.EN Dolphin CPAP. User manual...
  • Page 3: Company Information

    Tel: +84 24 3766 6521 www.masimo.com/patents.htm Fax: +84 24 3718 8050 Email: assistance@mtts-asia.com www.mtts-asia.com PROMETEO SRL Via Mantova 9 38122 Trento Italy V.1.04.EN Dolphin CPAP. User manual...
  • Page 4: Introduction

    This user manual is intended for healthcare professionals. The Dolphin CPAP provides precisely The Dolphin CPAP incorporates The Dolphin CPAP is to be operated by qualified personnel only. This manual, blended, heated and humidified gas to Masimo SET technology to provide ac- ®...
  • Page 5: Device Description

    Humidifier Mixed Gas Output PEEP Chamber Masimo Patient Cable Connector Use this connector to connect MTTS backup 12V power 12V UPS Connector supply Use this power inlet to connect the device to mains power. AC Power Inlet The power requirements for the device are listed on the back panel and in the Specifications section.
  • Page 6 °C °C information about the CPAP alarm system. cm H O > 14 cm H O > 14 Accessories Dolphin CPAP set is provided with following accessories (one piece of each): Part Order Code Component of Humidifier CPD-10800 Breathing Circuit...
  • Page 7 If the perfusion index is lower, Button Decrease Buttons. two green lights will rise and fall with the pulse. If the perfusion index is very low, a red light will flash with the pulse. V.1.04.EN Dolphin CPAP. User manual...
  • Page 8 Plastic Body Temperature sensor cable Screw Cap O-ring Heater wire (non removable) assembly with straight Screw Cap connector and humidifier temperature probe (T1) Metal Plug O-ring (non removable) Heater wire Metal Plug power cable (yellow) V.1.04.EN Dolphin CPAP. User manual...
  • Page 9 Temperature Probe Cable Connector (6-pin) PEEP Chamber can be used safely and effectively without the Silicone Cable, Heater Wire Power Cable Connector (4-pin) which is only used to ensure that the Spring is not lost V.1.04.EN Dolphin CPAP. User manual...
  • Page 10: Warnings (Cpap)

    Warnings (CPAP) The device should only be used with the humidifier, PEEP chamber, tubing, The Dolphin CPAP is to be operated by qualified personnel only. This manual, cables and connectors specified in this user manual. Use of components not accessory Directions for Use, all precautionary information, and specified in this user manual can result in serious injury to the patient or specifications should be read before use.
  • Page 11: Setup (Cpap)

    The fill lines indicate the amount of water that will be available for therapy once the humidifier is flipped right-side up. If leaking occurs, check that the parts are correctly assembled and not damaged. Discard and replace any damaged parts. V.1.04.EN Dolphin CPAP. User manual...
  • Page 12 Attach the screw cap by twisting it until tight. Place the assembled PEEP Chamber in the PEEP Chamber holder. Insert the spring into the screw cap in the orientation shown in the picture. V.1.04.EN Dolphin CPAP. User manual...
  • Page 13 (the direction of this part is not important). Then attach the silicone connector, wye connector, and expiratory breathing tube as shown in the pictures. V.1.04.EN Dolphin CPAP. User manual...
  • Page 14 Make sure the temperature probes are fully inserted. If they are not, it could result in a burn to the patient due to increased inspiratory gas temperature. V.1.04.EN Dolphin CPAP. User manual...
  • Page 15 Setup (CPAP) Setup (CPAP) Attach the power cord to the AC power inlet on the back of the Dolphin CPAP. 5. Attach the humidifier connection tube, gas hose, and power cord. Locate an AC power source with adequate power rating for the Dolphin CPAP. The...
  • Page 16: Operation (Cpap)

    The warm-up is complete when the temperature display value stops If a system failure is detected by the Dolphin CPAP at this time, an error flashing and the red light turns off. message will be indicated by the letter ‘E’ and a number on the display. If this happens, note the error number and contact your MTTS representative.
  • Page 17 T1/T2 button again. policy. If replacing, clean and disinfect the used humidifier according to °C °C the Reprocessing (CPAP) section. cm H O > 14 cm H O > 14 V.1.04.EN Dolphin CPAP. User manual...
  • Page 18: Alarm System (Cpap)

    Press this button to temporarily silence CPAP alarm sounds for 110 seconds. Press it again to The Dolphin CPAP must be reprocessed between uses with different patients. restore alarm sounds. Reprocess the Dolphin CPAP according to the Reprocessing (CPAP) section.
  • Page 19 MTTS itself off momentarily. anism. To force the refill mechanism to itself down.
  • Page 20: Reprocessing (Cpap)

    An alarm should sound and a flashing red light should be visible next to the heater wire image on the CPAP display. If either of these things does not happen, do not use the machine and contact your MTTS Autoclave representative.
  • Page 21 After rinsing, leave all of the parts to dry completely before proceeding to the next step. Process using step 4 only. Temperature Sensor Cable If you do not allow the parts to dry, the autoclave and chemical disinfection steps may not be effective. V.1.04.EN Dolphin CPAP. User manual...
  • Page 22 If distilled water is not available, boiled water that has been cooled to tepid and reuse of the glutaraldehyde solution. may be used. When disposing of the glutaraldehyde solution, follow manufacturer and local government recommendations. V.1.04.EN Dolphin CPAP. User manual...
  • Page 23 4. Temperature sensor cable cleaning and disinfection 5. Cleaning the outer surface of the Dolphin CPAP To clean the display panel or outer surface of the Dolphin CPAP housing, use a clean The following equipment is needed for cleaning: cloth moistened with a mild detergent solution to gently wipe the target surfaces.
  • Page 24: Maintenance (Cpap)

    Maintenance (CPAP) Maintenance (CPAP) The Dolphin CPAP includes a number of parts that must be replaced or serviced 1. Air filter replacement during the lifetime of the device. These parts include: Turn off the device and disconnect the power cord.
  • Page 25 CAL SEN message will illuminate as shown in the picture. The calibration may take °C °C up to two minutes. cm H O > 14 cm H O > 14 Turn off the device. V.1.04.EN Dolphin CPAP. User manual...
  • Page 26: Warnings (Pulse Oximeter)

    Warnings (Pulse Oximeter) Warnings (Pulse Oximeter) The Dolphin CPAP is to be operated by qualified personnel only. This manual, Do not use the device or Masimo oximetry sensors during magnetic resonance accessory Directions for Use, all precautionary information, and imaging (MRI) scanning. Induced current could potentially cause burns. The specifications should be read before use.
  • Page 27: Overview (Pulse Oximeter)

    ), 2,3-DPG, and fetal hemoglobin. Also, as blood gas the light, converts it into an electronic signal and sends it to the Dolphin CPAP for samples are usually taken over a period of 20 seconds (the time it takes to draw blood) calculation.
  • Page 28 The Masimo SET signal processing algorithm, Discrete Saturation ® ® Transform (DST) , reliably identifies the noise, isolates it and, using adaptive filters, cancels it. It then reports the true arterial oxygen saturation for display on the monitor. V.1.04.EN Dolphin CPAP. User manual...
  • Page 29: Setup (Pulse Oximeter)

    1. Plug in the Dolphin CPAP (if it is not already plugged in). 2. Connect the patient cable and sensor. Attach the power cord to the AC power inlet on the back of the Dolphin CPAP. Connect the patient cable to the Masimo Patient Cable Connector on the front of the device.
  • Page 30: Operation (Pulse Oximeter)

    If an error is detected by the Dolphin CPAP at this time, an error message will arrows to adjust the alarm limit. be indicated using ERR and a number on the display. If this happens, note the When the target number has been error number and contact your MTTS representative.
  • Page 31 OK button. /min Press the alarm silence button. Confirm that no alarm tones start for the next 110 seconds, and that the yellow light next to the alarm silence button is lit. /min V.1.04.EN Dolphin CPAP. User manual...
  • Page 32 (the patient cable connectors are not waterproof). Do not sterilize by irradiation, steam, or ethylene oxide. If the Dolphin CPAP loses mains power, the SpO and pulse rate alarm values Do not attempt to reprocess, recondition or recycle Masimo sensors or will be saved until the pulse oximeter display is next powered on.
  • Page 33 80s. When elevated levels The patient is in cardiac arrest or is in shock. of MetHb are suspected, laboratory analysis (CO-Oximetry) of a blood sample should be performed. V.1.04.EN Dolphin CPAP. User manual...
  • Page 34: Alarm System (Pulse Oximeter)

    Oximeter alarm values will be saved. If the Pulse Oximeter display is turned off using the power button or if there is a loss of mains power and the CPAP display is off, the Pulse Oximeter alarm values will return to default values. V.1.04.EN Dolphin CPAP. User manual...
  • Page 35 Wait for found pulse. perfusion is expected or evident. Pulse search (This search should occur whenever a sensor is first applied to a patient). Sensor not /min connected to Connect the sensor to the patient. patient /min V.1.04.EN Dolphin CPAP. User manual...
  • Page 36: Sensors & Patient Cables (Pulse Oximeter)

    For more information refer to the following table or contact your MTTS representative. Use only Masimo SET Before use of any sensor or cable, carefully read the sensor or cable Directions for sensors and sensor cables.
  • Page 37 Masimo SET patient cables These sensors are recommended because they are compatible with the Dolphin CPAP and are often appropriate for the Dolphin CPAP’s target patient Reusable patient cables of various lengths are available. Only use appropriate population. This is NOT a complete list of sensors compatible with the Dolphin Masimo oximetry cables for SpO measurements.
  • Page 38: Maintenance (Pulse Oximeter)

    Cleaning any of the described tests, discontinue its use and correct the problem before To clean the display panel or outer surface of the Dolphin CPAP housing, use a clean reconnecting the device to the patient. cloth moistened with a mild detergent solution to gently wipe the target surfaces.
  • Page 39: Specifications

    Volume available for vaporization Humidity chamber 20 mL Reservoir 1600 mL (maximum) Maximum operating pressure 7 kPa Heater plate overtemperature cutout 90 ± 5 °C 10.0 10.1 10.3 10.5 10.8 10.9 11.0 11.2 11.3 11.4 V.1.04.EN Dolphin CPAP. User manual...
  • Page 40 70 – 100% SaO with a resultant accuracy of 2.9% SpO . Contact your MTTS representative for testing specifica- tions. 2. Masimo SET technology with LNOP Adt sensors has been validated for motion accuracy in human blood studies on healthy adult male and female...
  • Page 41: Explanation Of Symbols

    Do not disassemble the device unless you General requirements for basic safety and essential performance - Collateral Serial number. are an MTTS trained technician or have been Standard: General requirements, tests and guidance for alarm systems in medical instructed to by qualified personnel.
  • Page 42: Warranty Policy

    FUNCTION IN CONDITIONS OF NORMAL USE. MTTS MAKES NO OTHER EXPRESS WARRANTY OR details. MTTS will not alter form, fit, or function of this to give you a refund of your purchase price or lease pay- EXCEPT AS INDICATED ABOVE, IN NO EVENT WILL MTTS...
  • Page 43 Warranty Card. The date on do not solve the problem, you can receive support by do- your sales receipt is the date of purchase unless MTTS or ing one of the following: Locate and contact your nearest MTTS service pro- your reseller informs you otherwise in writing.
  • Page 44 MEDICAL TECHNOLOGY TRANSFER AND SERVICES Co., LTD House No. 26, Alley 41, An Duong Vuong Street, Tay Ho District, Hanoi City, Vietnam Tel: +84 24 3766 6521 Fax: +84 24 3718 8050 Email: assistance@mtts-asia.com www.mtts-asia.com PROMETEO SRL Via Mantova 9...

This manual is also suitable for:

Dolphin

Table of Contents