Page 4
INTRODUCTION This Sper Scientific Dissolved Oxygen Meter Kit (Model SP-850081DOK) comes with dissolved oxygen (DO) probe (SP-850081DO) and also uses optional, interchangeable probes to read pH, millivolt (mV)/ oxygen reduction potential (ORP), conductivity, total dissolved solids (TDS), salinity, and temperature. Order only the probes required for your current parameters and add additional probes in the future.
Page 5
FEATURES • Reads DO, and has the option to read pH, mV/ORP, conductivity, TDS, salinity and temperature with optional probes • Automatic temperature compensation • Salinity and altitude compensation for DO measurements • Maximum and minimum values • Hold function •...
Page 7
SETUP Meter On and O 1. Press POWER to turn the meter on. 2. Press and hold POWER for 3 seconds and release to turn the meter off. Backlight The backlight will illuminate automatically when the meter is turned on. 1.
Page 8
Beeper 1. In Setup Mode, press SET until “bEEP” is displayed. 2. Press ▲ or ▼ to toggle between “Yes” (beeper is on) or “No” (disable beeper). 3. Press ENTER to save the selection. 4. Press SET to continue to the next setup parameter, or press ESC to exit Setup Mode.
Page 9
Select DO Altitude Compensation in Meters This feature is active in DO Mode only. 1. In Setup Mode, press SET until “HIGh” is displayed. 2. The default value is 0 meters. Press ▲ or ▼ to adjust the upper altitude compensation value in increments of 100 meters.
Page 10
Select pH Manual Temperature Compensation Value This feature is active in pH Mode only. 1. In Setup Mode, press SET until “t-SEt” is displayed. 2. The default value is 25° C / 77° F. Press ▲ or ▼ to adjust temperature compensation value.
Page 11
3. Follow the specific measurement procedures for the parameter in use. Parameter Page mV (ORP) Conductivity Salinity Data Hold Function 1. During measurement, press HOLD to freeze the reading on the display. “Hold” appears on the LCD. 2. Press HOLD to return to Normal Mode. Minimum and Maximum Values 1.
Page 12
pH MEASUREMENT (for use with SP-840016, SP-840049 & SP-840051 optional probes) Calibration The pH probe should be calibrated before first use. Recalibrate regularly to ensure highly accurate measurement. Note… Calibration should begin with pH 7.00 and then be followed by either pH 4.00 or pH 10.00.
Page 13
11. To clear the existing calibration data, select “CLr” from the calibration options. 12. Press ENTER to confirm the selection and return to Normal Mode. pH Probe Calibration with Manual Temperature Compensation 1. Plug and secure the pH probe into the pH socket on top of the meter.
Page 14
2. Plug the temperature probe into the TEMP socket on top of the meter. 3. Press POWER to turn the meter on. 4. Press FUNCTION to cycle through the available functions until “pH” is displayed. 5. Immerse the temperature probe into the solution to be measured.
Page 15
2. Press POWER to turn the meter on. 3. Press FUNCTION to cycle through the available functions until “OrP” is displayed. 4. Hold the electrode handle and immerse the sensing head completely into the solution to be measured. 5. Gently shake the electrode. 6.
Page 16
6. Press ENTER. The display will count down from 30 to 0 as calibration occurs. 7. The meter returns to Normal Mode. Do Measurement 1. Plug the DO probe (SP-850081DO) into the DO socket on top of the meter. 2. Press POWER to turn the meter on. 3.
Page 17
CONDUCTIVITY MEASUREMENT (for use with SP-850081C/TD or SP-850081C/SG optional probes) Conductivity Probe Calibration The conductivity probe should be calibrated before first use. Recalibrate the probe periodically to ensure accuracy. Note… One-point calibration should be executed at 1.413 mS. For multi-point calibration, execute 1.413 first and then proceed to other ranges.
Page 18
Conductivity Measurement 1. Plug the conductivity probe into the CD socket on top of the meter. 2. Press POWER to turn the meter on. 3. Press FUNCTION to cycle through the available functions until “Cd” is displayed. 4. Hold the electrode handle and immerse the sensing head completely into the solution to be measured.
Page 19
TDS Measurement 1. Plug the conductivity probe into the CD socket on top of the meter. 2. Press POWER to turn the meter on. 3. Press FUNCTION to cycle through the available functions until “tdS” is displayed. 4. Hold the electrode handle and immerse the sensing head completely into the solution to be measured.
Page 20
POWER SUPPLY Battery Power This meter uses one DC 9 V (006 P) heavy-duty alkaline battery. To install the battery before first use: 1. Unscrew and remove the screw on the back battery cover. 2. Slide o the battery cover in the direction of the arrows. 3.
Page 21
General Specifications Circuit Custom one-chip microprocessor LSI circuit Display 52 x 38 mm LCD Sample Rate ~ 1 second Operating < 85% RH Operating 0 - 50 °C Temperature 32 - 122 °F Power ~ DC 14 mA with backlight o Consumption ~ DC 26 mA with backlight on Power One DC 9V (006 P) heavy-duty alkaline battery...
Page 22
General Conductivity Temperature Automatic (ATC) from 0 to 60 °C (32 to Compensation 140 °F) with temperature compensation factor variable between 0 to 5.0% per °C. Probe 0 to 60 °C Operating Temperature Conductivity Measurement Range Accuracy 0 - 200.00 uS 200 uS 0.1 uS ±...
Page 23
Dissolved Oxygen Temperature Automatic (ATC) from 0 - 50 °C Compensation Salinity 0 - 50% salt Compensation Altitude 0 - 8900 meters Compensation (M.T.) Measurement Range Accuracy Dissolved 0 - 20.0 mg/L 0.1 mg/L ± 0.4 mg/L Oxygen Oxygen in Air 0 - 100.0% 0.1% O2 ±...
Page 24
Opening the meter to expose its electronics will break the waterproof seal and void the warranty. To obtain warranty service, Contact your Distributor: SCIGIENE CORPORATION Tel:416-261-4865 Email: support@scigiene.com The defective unit must be accompanied by a description of the problem and your return address.