Table 202: Guidance and manufacturers declaration
– electromagnetic immunity
TM
T he NeuroT rac
product is intended for use in the electromagnetic environment specified
below. T he customer or the user of the NeuroT rac
such an environment, and that precautions regarding that environment are heeded.
Immunity te st
IEC 60601
te st le ve l
Electrostatic
±6 kV contact
discharge (ESD)
±8 kV air
IEC 61000-4-2
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
TM
product should ensure that it is used in
Compliance
Ele ctromagne tic e nvironme nt–
le ve l
±6 kV contact
Floors should be wood, concrete or
±8 kV air
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30 %.
3 A/m
Power frequency magnetic fields should
be at characteristic levels of a typical
location in a typical commercial or
hospital environment.
15
guidance
Need help?
Do you have a question about the NeuroTrac MultiTENS and is the answer not in the manual?
Questions and answers