Download Print this page

Consumable Materials And Preventative Maintenance - Verity Medical NeuroTrac MultiTENS Service Manual

Hide thumbs Also See for NeuroTrac MultiTENS:

Advertisement

Consumable Materials and Preventative Maintenance

WARNING! Only medically approved accessories should be used!
CONTROL UNIT
*
Wipe the surface once a week with a damp cloth or antiseptic wipe
*
Do not use cleaning sprays or alcohol-based cleaning solutions
*
Control unit disposal: please return to Verity Medical LTD or to the appointed
distributor.
ACCESSORIES
Battery:
*
To change the battery, open the battery door on the rear of the control unit by pressing
down on the raised rib pattern just below the belt clip. Lift the battery out of the
compartment. This is very easy and can be done by the user.
*
Check periodically for any discharge from the battery
*
Remove battery completely from unit if not in use for any extended period (typically
one week)
*
Low battery indicator of 6.9 volts shown on LCD display, when flashing change battery
for a new one.
*
Preferably use a PP3 alkaline battery
*
Battery disposal: please return to the supplier from whom you have purchased it.
Lead Wires:
*
The lead wires should be handled carefully and never stretched, as this can cause the
stimulation to function below normal standards or not at all
*
Examine lead wires before each treatment for loose connections or damage
*
Avoid stretching and twisting the lead wires
*
Store the lead wires carefully after each use
*
Lead wires Disposal: please return to the supplier from whom you have purchased
them.
Self-Adhesive Electrodes:
*
Check the short connectors have not become separated from the electrodes
*
Replace electrodes onto plastic film after use. If they drop onto the floor, debris will
adhere to conductive gel rendering the electrodes ineffective.
Electrode life can be considerably reduced by:
*
The type and condition of the skin
*
Deep seated moisturisers or make-up
For the Best Results:
*
Before each use cleanse the skin
*
After each use stick the pads on the shiny insert card and store in a cool and dry place,
such as the fridge (not freezer).
Caution: Static electricity may damage this product.