TABLE OF CONTENTS: 1. Product Description 2. Product Features 3. Raw Material 4. Micropipette Operation 5. Tip Ejection 6. Pipetting Technique 7. Calibration 9. Error Calculation 10. Maintenance 11. Short Term Service 12. Long Term Service 13. Disassembly & Re-assembly 14.
Easy tip ejection: Product Description These micropipettes are designed with An in-built tip-ejector is designed ergonomic and intuitive handling, ensuring high to eject tips with optimal force. precision with its µAir™ technology. They are available in Single Channel Fixed and Variable Volume ranges.
Pipetting Technique Reverse Technique: 1. To aspirate the liquid in the tip, press the plunger to the second stop and immerse the pipette tip vertically in the liquid. 2. Slowly release the plunger while the tip is immersed. The liquid will be aspirated into the pipette tip.
Page 5
4. Adjust the volume back to the nominal volume of the pipette by rotating the plunger. 5. Perform a few pipetting operations to check the resulting volume. If the readings on the balance are still outside permissible limits, please perform the calibration procedure again. Single Channel Variable Volume Micropipettes The Single Channel Variable Volume Micropipette is available in 9 unique volume...
Multichannel Micropipettes Short Term Service The Multichannel Micropipette is available in 6 The micropipette must be checked at the unique volume ranges from 0.5 µl to 300 µl. beginning of each day for dust and dirt on the outer surface. Particular attention must Volume be paid to the Tip Cone.
Page 7
Step1: Screwback the tip cone to the piston assembly by rotating it in the clockwise direction. Step 2: Place the upper ejector housing and the lower ejector housing on the piston assembly and snap them back into place. Step 3: Assemble the ejector housing to the upper housing by rotating in the clockwise direction.
Step 2: Screw the piston and tip cone assembly C1. Disassembly: together at the piston holder. (For multichannel micropipettes) Step1: Hold the pipette and pull out the ejector bottom Step 3: Push the ejector bottom into the main The Tip cone area will be exposed assembly.
Performance Optimizatio Autoclaving Activity Action Pipette with a consistent autoclaving them at 121°C (252°F) at 15 psi rhythm, pressure and speed. for a minimum of 20 minutes. Tip Size and Fit if needed. After autoclaving, the micropipette must be cooled to room temperature for at least two hours.
Package Shelf Hanger The micropipettes are shipped in specially designed packages containing the following items. The Micropipette Operation Manual High Precision Micropipette Single and Multichannel (Variable & Fixed Volume) Caution These micropipettes are designed to allow easy in-lab service. If you would prefer to Calibration Tool have us or your local representative for servicing your instrument, please make sure...