Chison CBit 9 Instruction Manual

Digital color doppler ultrasound system
Table of Contents

Advertisement

Quick Links

Digital Color Doppler Ultrasound System
CBit 9
Instruction Manual
57-00907-00
Mar. 1
st
, 2022
V1.1
CHISON Medical Technologies Co., Ltd.
We reserve the right to make changes to this manual without prior notice.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CBit 9 and is the answer not in the manual?

Questions and answers

Summary of Contents for Chison CBit 9

  • Page 1 Digital Color Doppler Ultrasound System CBit 9 Instruction Manual 57-00907-00 Mar. 1 , 2022 V1.1 CHISON Medical Technologies Co., Ltd. We reserve the right to make changes to this manual without prior notice.
  • Page 2: Regulatory Requirement

    Essential Requirements of the Medical Device Directive. This manual is a reference for the CBit 9. Please verify that you are using the latest revision of this document. If you need the latest revision, contact your distributor.
  • Page 3 If you have any question or doubt, please contact CHISON's authorized service engineer. 4. CHISON’s Warranty only cover material and parts costs for repair, but do not cover any labor cost or onsite service cost at end user's side.
  • Page 4 Damage or loss caused by failure to meet the specified conditions for this system, such as inadequate power supply, improper installation or environmental conditions. d) Damage or loss caused by non-approved transportation by CHISON. e) Damage or loss due to use the system outside the region where the system was originally sold.
  • Page 5 9. This manual contains warnings regarding foreseeable potential dangers, but user shall always be alert to dangers other than those indicated as well. CHISON shall not be liable for damage or loss that results from negligence or from ignoring the precautions and operating instructions described in this operation manual.
  • Page 6 Digital Color Doppler Ultrasound System 16. Important data must be backed up on external memory media. CHISON shall not be liable for loss of data stored in the memory of this system caused by operator error or accidents. 17. Please put this operation manual with the system to ensure operator and manager can reach it at any time.
  • Page 7: Table Of Contents

    Digital Color Doppler Ultrasound System TABLE OF CONTENTS ..................1 IGITAL OLOR OPPLER LTRASOUND YSTEM ......................... 3 MPORTANT INFORMATION ........................1 HAPTER NTRODUCTION 1.1 System Overview ........................1 1.2 Contact Information ........................1 ........................3 HAPTER YSTEM AFETY 2.1 Safety Overview ......................... 3 2.2 Electrical Safety .........................
  • Page 8 Digital Color Doppler Ultrasound System ....................103 HAPTER ENERAL EASUREMENTS 5.1 Key for Measurement ......................103 5.2 Fast measurement ......................... 104 5.3 Measurement and Calculation ....................108 5.4 Edit measurement results ...................... 166 5.5 Report ............................ 167 ........................... 170 HAPTER RESET 6.1 Recall Preset ..........................
  • Page 9 Digital Color Doppler Ultrasound System 9.3 Safety Check .......................... 222 9.4 Troubleshooting ........................222 9.5 Service Responsibility ......................224 APPENDIX A T EC R ..............227 NFORMATION OF EPRESENTATIVE APPENDIX B S ..............228 YSTEM ONE RECOVERY UNCTION APPENDIX C MAXIMUM ACOUSTIC OUTPUT REPORT .............. 230 APPENDIX D T ............
  • Page 10: Digital Color Doppler Ultrasound System

    For additional information or assistance, please contact your local distributor or the appropriate support resource shown below: www.chison.com CHISON website Service Support CHISON Medical Technologies Co., Ltd. Tel: 0086-400-8878-020; 0086-0510-85311707 Fax: 0086-0510-85310726 E-mail: service@chison.com.cn Placing an Order CHISON Medical Technologies Co., Ltd.
  • Page 11 Digital Color Doppler Ultrasound System CHISON Medical Technologies Co., Ltd. Manufacturer No.3 Changjiang South Road, Xinwu District, Wuxi, 214028 Jiangsu, P.R. China US Agent Mr. Marco Mu, 2219 Rimland Drive,Suite 301,Bellingham, Barkley Villiage Bellingham,Washington,98226, UNITED STATES Phone: 360-3257028, Fax: 360-9253199, Email: us.agent@mid-link.net MID-LINK INTERNATIONAL CO., LTD...
  • Page 12: Important Information

    The system should be installed only by personnel authorized by the manufacturer. Do not attempt to install the system by yourself.  Only a CHISON's authorized service engineer can perform maintenance.  Only a qualified operator, or someone under qualified supervision, can use the system.
  • Page 13: Electrical Safety

     Do not put liquid on top of the main unit. NOTE: To dispose of this product properly, please contact the local CHISON’s Authorized Service Representative. 2.2 Electrical Safety Type of protection against electric s hock...
  • Page 14: Mode Of Operation

    Digital Color Doppler Ultrasound System Parts of probe intended to be immersed in normal use meet the requirements of watertight equipment (IPX7)  The IP Classification of System is Ordinary Equipment (IPX0) The Equipment is not suitable for use in the presence of a flammable anesthetic mixed with air (with oxygen or with oxide) Mode of operation ...
  • Page 15: Notice Against User Modification

    Locate the equipment as far away as possible from other electronic equipment.  Be sure to only use the cables provided by or designated by CHISON. Connect these cables following the installation procedures (e.g. wire power cord separately from signal cables).
  • Page 16: Labels

    Digital Color Doppler Ultrasound System 2.3 Labels Fig.2-1Rear panel label 2.3.1 Symbols on label WASTE OF ELECTRICAL AND ELECTRONIC This mark indicates that this product EQUIPMENT (WEEE): This symbol is used for contains a limited amount of hazardous substances Environment Protection, it indicates that the waste in the Chinese Standard GB/T 26572-2011 "Limited of electrical and electronic equipment must not be Requirements for Restricted Substances in...
  • Page 17 Digital Color Doppler Ultrasound System year. Do not use the following devices near this equipment: cellular phone, radio receiver, and Caution, consult accompanying documents. mobile radio transmitter, radio controlled toy, etc. This symbol advises the reader to consult the Use of these devices near this equipment could accompanying documents for important safety cause this equipment to perform outside the related information such as warnings and pre-...
  • Page 18 Digital Color Doppler Ultrasound System Power On: the main unit connects to the adapter with power supply, the battery start charge. Power off: the main unit disconnects to the adapter with power supply, the battery end charge. CAUTION: This Power Switch DOES ISOLATE Mains Supply.
  • Page 19: Patient Environmental Devices

    Digital Color Doppler Ultrasound System Please read the user handbook Before operation. Be careful, arrow area may Clamp hand ! In the process of push down the Forbidden to push the monitor to move monitor, it is forbidden to place items to this region, the machine.
  • Page 20 DO NOT touch patient and devices without IEC/EN 60601-1 approval to avoid the leakage current risk within the PATIENT ENVIRONMENT. Unapproved Devices CAUTION  DO NOT use unapproved devices.  If devices are connected without the approval of CHISON, the warranty will be INVALID.
  • Page 21 Digital Color Doppler Ultrasound System The system can’t be used with HF surgical equipment; otherwise the burns  to patient may occur.  Any device connected to this system must conform to one or more of the requirements listed below: -IEC standard or equivalent standards appropriate to devices.
  • Page 22: Biological Safety

    CHISON authorized engineer has checked the system and confirm the effectiveness and safety of the system and probes.
  • Page 23 Digital Color Doppler Ultrasound System and recognized efficacy for medical diagnosis, including use during human pregnancy, the American Institute of Ultrasound in Medicine herein addresses the clinical safety of such use: There are no confirmed biological effects on patients or instrument operators caused by exposures from present diagnostic ultrasound instruments.
  • Page 24: Scanning Patients And Education

    Digital Color Doppler Ultrasound System 2.6 Scanning Patients and Education The Track-3 or IEC60601-2-37 output display standard allows users to share the responsibility for the safe use of this ultrasound system. Follow these usage guidelines for safe operation:  In order to maintain proper cleanliness of the probes, always clean them between patients. ...
  • Page 25 Digital Color Doppler Ultrasound System  Maintain the shortest examination time necessary to produce a useful diagnostic result.  Do not hold the probe in a fixed position for any longer than is necessary. It should be removed from the patient whenever there is no need for real-time imaging or spectral Doppler acquisition. The frozen frame and Cine loop capabilities allow images to be reviewed and discussed without exposing the patient to continuous scanning.
  • Page 26: Understanding The Mi/Ti Display

    Digital Color Doppler Ultrasound System TI> 0.7 The overall exposure time of an embryo or fetus should be restricted in accordance with Table 2-2 below as a reference: Maximum exposure time (minutes) Table 2-2 Maximum recommended exposure times for an embryo or fetus ...
  • Page 27 Digital Color Doppler Ultrasound System Display Standard. For any diagnostic ultrasonic systems, Track-3 provides an Output Indices Display Standard. The diagnostic ultrasound systems and its operation manual contain the information regarding an ALARA (As Low As Reasonably Achievable) education program for the clinical end-user and the acoustic output indices, MI and TI.
  • Page 28 Digital Color Doppler Ultrasound System (cm) 0.9332 0.8128 0.7080 0.5957 0.8710 0.6607 0.5012 0.3548 0.8128 0.5370 0.3548 0.2113 0.7586 0.4365 0.2512 0.1259 0.7080 0.3548 0.1778 0.0750 0.6607 0.2884 0.1259 0.0447 0.6166 0.2344 0.0891 0.0266 0.5754 0.1903 0.0631 0.0158 I’=I*R Where I’ is the intensity in soft tissue, Iis the time-averaged intensity measured in water. Tissue Model: Tissue temperature elevation depends on power, tissue type, beam width, and scanning mode.
  • Page 29 Digital Color Doppler Ultrasound System TI is defined as the ratio of the In Situ acoustic power (W.3) to the acoustic power required to raise tissue temperature by 1° C (W deg ), TI = W.3/W deg . Three TIs corresponding to soft tissue (TIS) for abdominal; bone (TIB) for fetal and neonatal cephalic;...
  • Page 30: Operator Control Features

    Digital Color Doppler Ultrasound System Display and Report in Different Mode Located on the upper middle section of the system display monitor, the acoustic output display provides the operator with real-time indication of acoustic levels being generated by the system. For B-Scan Mode Only display and report MI, and start from 0.4 if maximum MI >...
  • Page 31 Digital Color Doppler Ultrasound System are detailed under the bioeffect portion of each control in the Optimizing the Image chapter. Always observe the Acoustic Output display for possible effects. Best practices while scanning HINTS: Raise the Acoustic Output only after attempting image optimization with controls that have no effect on Acoustic Output, such as Gain and TGC.
  • Page 32: Chapter 3 Preparing The System For Use

    Digital Color Doppler Ultrasound System Chapter 3 Preparing the System for Use 3.1 Site Requirements 3.1.1 Operation Environmental Requirements The following environmental conditions are within system tolerances for operation: Temperature: 10º C ~ 40º C Relative Humidity: 30%~75%, non-condensing Atmosphere Pressure: 700hPa ~ 1060hPa Strong radiation sources or powerful electromagnetic waves (e.g.electro-magnetic waves from radio broadcasting) may result in image ghosting or noise.
  • Page 33: Transport And Storage Environmental Requirements

    Digital Color Doppler Ultrasound System • reorient or relocate the affected device(s). • increase the separation between the equipment and the affected device. • power the equipment from a source different from that of the affected device. • consult the point of purchase or service representative for further suggestions. 3.1.2 Transport and Storage Environmental Requirements The following environmental transport and storage conditions are within system tolerances: Temperature:...
  • Page 34 Digital Color Doppler Ultrasound System NOTE Please follow the outlined power requirements. Only use power cords that  meet the system guidelines—failure to follow these procedures may result in system damage. Line power may vary in different geographic locations. Refer to the ...
  • Page 35 Temperature: 23± 5 , Humidity: 65± 20 6) Upon receipt of the CBit 9 and before first time usage, it is highly recommended that the customer performs one full discharge/charge cycle. If the battery has not been used for over 2 months, the customer is recommended to perform one full discharge/charge cycle.
  • Page 36 Digital Color Doppler Ultrasound System CAUTION: Do not immerse the battery in water or allow it to get wet. Do not use or store the battery near sources of heat such as a fire or heater. Do not use any chargers other than those recommended. Do not put the battery into a fire or apply direct heat to it.
  • Page 37: System Specifications

    Digital Color Doppler Ultrasound System 3.2 System Specifications 3.2.1 Console Overview Fig. 3-1 a: Console Overview The following pictures show the system in different views.
  • Page 38 Digital Color Doppler Ultrasound System LCD monitor Touch Screen Probe Holder Control panel DVD RW Fig. 3-1 b: Front Side View 2 USB ports ECG port Probe connectors Fig. 3-1 c: Right Side View...
  • Page 39: Physical Specification

    Digital Color Doppler Ultrasound System Fig. 3-1 d: Back Side View 3.2.2 Physical Specification Dimensions of main unit (approx.): 806mm(Depth)*543mm(Width)*1423mm(Height) Net weight of main unit (approx): 52kg (no probe included)
  • Page 40: Key System Features

    Digital Color Doppler Ultrasound System 3.2.3 Key System Features Display B, B/M, M, B/B, 4B, 2D Steer, CFM, PD, DPD, PW, B/BC, Triplex, Quadplex, CW, Free  Steering M, TDI, Color M, Curved Panoramic, Trapezoid Image, Compound, SRA, Elastography, Stress Echo, Super Needle, 4D, Virtual HD, Sono Crystal, FHI, AIO. ...
  • Page 41 Digital Color Doppler Ultrasound System time interval, pulsatility index and resistance index. Built in battery. Turn on /off the green power switch to charge /incharge the battery. ...
  • Page 42: Configuration Of The System

    Digital Color Doppler Ultrasound System Configuration of the System Mode CBit 9 B mode Standard B/M Mode Standard M Mode Standard Dual Mode Standard Quad Mode Standard 2D Steer Option CFM Mode Standard CPA Mode Standard DPD Mode Option PW Mode...
  • Page 43 Digital Color Doppler Ultrasound System Compound Standard Standard Elastography Option Stress Echo Option LV Tracking Option Strain Option S-flow Option Standard Auto LGC Option Option Intergrated hard disk Option Human Body Mark Standard Auto IMT Option Free NT Option Super Needle Option General measurement package Option...
  • Page 44 Digital Color Doppler Ultrasound System TCD measuremen package Option Breast measurement package Option Standby Option DICOM 3.0 Option 4D Software package Option Virtual HD Option X-Contrast Option Option Q-image Option Q-flow Option Q-beam Option Option D3C60L Option D7L40L Option D12L40L Option D7L40L-REC Option...
  • Page 45 Digital Color Doppler Ultrasound System D5C20L Option M8L40L Option D10L30L Option D8L50L Option Biopsy Option Probe connector 3.2.4 Accessories Linear probe: D7L40L 4.0-15.0MHz Convex probe: D3C60L 2.0-6.8MHz Linear probe: D12L40L 7.0-15.0MHz Linear probe: D7L40L-REC 4.0-15.0MHz Trans-vaginal transducer: D7C10L 4.0-15.0MHz Trans-vaginal transducer: D6C12L 4.0-12.0MHz Phased array probe: D3P64L 1.5-5.3MHz Phased array probe: D5P64L 2.0-8.0MHz...
  • Page 46 Digital Color Doppler Ultrasound System Micro convex probe: D6C15L 4.0-12.0MHz Volume transducer: V4C40L 2.0-6.8MHz Micro convex probe: D5C20L 4.0-10.7MHz Micro convex probe: D3C20L 2.0-6.8MHz Intra-operative transducer: D10L30L Signal crystal linear reansducer: M8L40L 4.0-10.7MHz 4.0-15.0MHz Liner probe: D8L50L 5.0-16.0MHz 3.2.5 I/O ports ...
  • Page 47: System Positioning & Transporting

    Digital Color Doppler Ultrasound System 3.2.6 DVD/CD-RW The supported DVD/CD disk information and compatibility of DVD/CD-RW see the following table: DS-8ACSH Compatibility DOS 6.xx, XP / 2003 / Vista / Win7 / Win8 / Win8.1 / Linux OS PCAV DVD+R 8X maximum PCAV DVD-R...
  • Page 48: When Moving The System

    Digital Color Doppler Ultrasound System  Unplug the power cord (if the system is plugged into wall outlet).  Disconnect all cables from off-board peripheral devices (external printer, etc.) from the console. NOTE: To prevent damage to the power cord, DO NOT pull excessively on the cord or sharply bend the cord while wrapping it.
  • Page 49: Powering The System

    Digital Color Doppler Ultrasound System carefully to prevent damage from vibration.  Avoid unpaved roads, excessive speeds, and erratic stops or starts. 3.4 Powering the System 3.4.1 Acclimation Time After being transported, the unit requires one hour for each 2.5 º C increment if its temperature is below 10 º...
  • Page 50 Digital Color Doppler Ultrasound System WARNING  To avoid risk of fire, the system power must be supplied from a separate, properly rated outlet.  Under no circumstances should the AC power plug be altered, changed, or adapted to a configuration rated less than specified. Never use an extension cord or adapter plug.
  • Page 51: Power Off

    Digital Color Doppler Ultrasound System NOTE:  After powering off the system, please wait for more than 3 minutes to power on again.  When the system is powered on, for safety reason, please avoid moving the system 3.4.4 Power Off To power off the system: ...
  • Page 52: Selecting Probes

    Digital Color Doppler Ultrasound System NOTE: Power off the system if you will not use the system for a long period of  time (including storage/transportation condition), and you should not allow the system in standby status, otherwise the batteries will be out of power and permanently damaged.
  • Page 53 Digital Color Doppler Ultrasound System NOTE: Inspect the probe before and after each use for damage or degradation to the housing, strain relief, lens, seal and connector. DO NOT use a probe that appears damaged until its functional and safe performance is verified.
  • Page 54: Deactivating The Probe

    Digital Color Doppler Ultrasound System Deactivating the Probe When deactivating the probe, the probe is automatically placed in a standby mode. To deactivate a probe:  Ensure the system is in “Transducer Selection” interface. If necessary, press the PROBE-key to return. ...
  • Page 55: Optional Installation

    Digital Color Doppler Ultrasound System 3.6 Optional installation 3.6.1 Connect the printer 1) It needs three cables: Remote cable, Video signal cable, Power cord. See picture in Fig.3-3. 2) Connect the remote cable to remote port on the back panel of ultrasound system. See picture in Fig.3-3.
  • Page 56 Digital Color Doppler Ultrasound System 3.6.2 Set the system for Video Printer CAUTION: Please confirm the video printer is turned on and connected well with the main unit, then you can do below setting 1) Press the key to enter "System Settings" interface, select "General",click“Keyboard”. See picture in Fig.3-4.
  • Page 57: User Interface Control

    Digital Color Doppler Ultrasound System  You can't print the system information. 3.6.3 Connect the PC printer 1. Place the printer smoothly. 2. Connect the printer to the system. 3. Set the print manager. Please see more information in 7.8. 4.
  • Page 58 Digital Color Doppler Ultrasound System  Annotation  Measurement and calculation package  File management and image archiving  DICOM setting  User defined preset  Clip image saving  Patient data entry  Q-image  Space compound imaging  Freeze/Unfreeze ...
  • Page 59 Digital Color Doppler Ultrasound System 3.7.1 Control Panel and Alphanumeric Keyboard Fig. 3-5: Overview of Control Panel and Alphanumeric Keyboard See layout of the Control Panel and Alphanumeric Keyboard in the above figure. The main function of each key is introduced in the following table. ...
  • Page 60 Digital Color Doppler Ultrasound System TGC Sliders  Input comment by alphanumeric keyboard.  : Finish the exam of current patient; delete the information of current patient.  (Body mark):Enter to the Bodymark function.  Enter to the function. ARROW mark ...
  • Page 61 Digital Color Doppler Ultrasound System  One key to full screen, press the key to enlarge the image to full screen.  Press the key to adjust X –CONTRAST. X -CONTRAST  In B mode, reverse the 2D image orientation 180 degrees. ...
  • Page 62 Digital Color Doppler Ultrasound System Activate/turn off CPA Mode (also named as Color Power Angio mode).  Press the CPA-key to turn on the CPA mode if the system is in B mode;  Press the CPA-key for second time to turn off CPA and return to the previous mode (either B-mode or duplex Doppler) ...
  • Page 63 Digital Color Doppler Ultrasound System  Rotate it to adjust the probe direction on the body mark status.  In B and CFM mode, rotate it to zoom the image. Press twice to enter multipleen largements.  If the probe supports CW mode, e.g. a phased array probe, press CW-key starts CW mode.
  • Page 64 Digital Color Doppler Ultrasound System SYSTEM SETTING  Press this key to enter REPORT function. REPORT  Print the images when the printer is working. PRINT2  Print the images when the printer is working. PRINT1  Confirm the command entry; ...
  • Page 65 Digital Color Doppler Ultrasound System Menu time to select the item and adjust the parameters. Press MENU-knob for third time to exit from current item. Rotate the MENU- knob to select the item.  Press to change Calc package.  Press to change menus of different modes. Change ...
  • Page 66 Digital Color Doppler Ultrasound System controls. The system will display the appropriate menus for the selected Mode and functions. There two soft-Menu controls, user can press Menu-knob on the control panel or touch the parameters on the touch screen to set the value. Press the menu to adjust the...
  • Page 67: Chapter 4 Imaging

    Digital Color Doppler Ultrasound System Chapter 4 Imaging 4.1 General Description  How to begin an exam  How to select a probe and an application.  How to optimize the image  The operations after getting the image: adding annotation and body mark, storing and recalling the image 4.2 Beginning an Exam Begin an exam by entering new patient information.
  • Page 68: Patient Data Entry

    Digital Color Doppler Ultrasound System Fig.4-1 4.2.2 Patient Data Entry Press the PATIENT-key to display the Patient interface. Fig.4-2 Method of entering the patient information: 1) Move the cursor to input area and input the patient information through the keyboard. Move...
  • Page 69: Display Interface

    Digital Color Doppler Ultrasound System trackball to exchange between each option. E.g., patient name, birthday (birthday result will be automatic calculated after entering the age), age (age result will be automatic calculated after entering the birthday), sex, weight, height.etc. 2) Click “OK” key to save the data after entering the appropriate information, then the system will go back to B mode.
  • Page 70 Digital Color Doppler Ultrasound System Display current network situation. It will display the IP address after connecting to the network. Display the status of current removable disks. If exist removable disks, click this icon can quickly enter into the storage manager interface and do the operation for disks. Fig.4-4 Display space size of all the drivers.
  • Page 71: Optimizing The Image

    Digital Color Doppler Ultrasound System Fig.4-5 Show the present state of charge and connection of adapter. 4.3 Optimizing the Image 4.3.1 Image Parameters Display Meanin Meaning Meaning Meaning Freq Freq Sweep Frequency Frequency Freq. Frequency Speed Dynami c range Frame rate G:Gain D:Dynamic range...
  • Page 72: Scanning Modes

    Digital Color Doppler Ultrasound System Color Acoustic Doppler /direction power Angle power angle P:Persistence Display Sample depth volume gate 4.3.2 Scanning Modes The system can support the following modes:   Quadplex Mode B Mode   CW mode B/M Mode ...
  • Page 73: Dual Mode

    Digital Color Doppler Ultrasound System Fig.4-6 B Mode 4.3.4 Dual Mode In B mode, press key. System will display the current image at the left side of the screen, Press key second time, system will freeze the image displayed at the left side and activate the image displayed at the right side at the same time.
  • Page 74 Digital Color Doppler Ultrasound System NOTE: There is only one image could be activated at one time.Quad display is also available for the Color Mode and PD Mode. 4.3.6 B/BC Mode In active color mode, turn the [B/BC] item on to display a real B mode image at the right side of the screen and active Color mode image at the left side of the screen.
  • Page 75: Cfm Mode

    Digital Color Doppler Ultrasound System Fig.4-8 B/M Mode M-mode Exam Procedure:  Get a good B-mode image. Survey the anatomy and place the area of interest near the center of the B-mode image.  Press the M-knob, move the Trackball to position the M cursor over the area that you want to display in M-mode.
  • Page 76 Digital Color Doppler Ultrasound System CFM is a Doppler mode intended to add color-coded qualitative information concerning the relative velocity and direction of fluid motion within the B-mode image. CFM is useful to see flow in a broad area. It allows visualization of flow in the CROI, whereas Doppler mode provides spectral information in a smaller area.
  • Page 77 Digital Color Doppler Ultrasound System Fig.4-9 CFM Mode CFM Scanning Hints: PRF: increase/decrease the PRF on the color bar. Imaging of higher velocity flow requires increased velocity scale values to avoid aliasing Wall Filter: affect low flow sensitivity versus motion artifact Color Map: allow you to select a specific color map.
  • Page 78 Digital Color Doppler Ultrasound System Fig.4-10 CPA Mode Press CPA -key to enter into the CPA mode and then CPA light is on. Direction PD mode In CPA mode, press MENU-knob to pop up the CPA Menu. Rotate MENU-knob to select the DPD Mode and press MENU-knob to enter into DPD mode.
  • Page 79 Digital Color Doppler Ultrasound System Fig.4-11 PW Mode PW mode Exam Procedure:  Get a good B-mode image. Press C-knob to help locate the vessel you wish to examine.  Press PW-knob to display the sample volume cursor and gate. ...
  • Page 80: Doppler Mode Scanning Hints

    Digital Color Doppler Ultrasound System key to toggle the Trackball function between Sample Volume Gate position and size. Press the UPDATE-key after the Sample Volume Gate is defined to activate the Spectral Doppler mode. Press the UPDATE-key for second time to toggle back to 2D (B or Color) update and deactivate the Spectral Doppler.
  • Page 81 Digital Color Doppler Ultrasound System Triplex menu on to enter the Triplex active mode.  While in B mode, press CPA-key and D-knob access the B+Doppler+CPA mode, turn Triplex menu on to enter the Triplex active mode. Activating Quadplex mode ...
  • Page 82: Trapezoidal Mode

    Digital Color Doppler Ultrasound System 4.3.13 Trapezoidal Mode Trapezoidal image is available for linear probes. In B mode, press menu knob to pop up B Menu, turn Trapezoidal Mode menu on to enter to Trapezoidal Mode. 4.3.14 Free Steering M mode (option) Free steering M mode is only available for phased array probe.
  • Page 83: Tdi Mode

    Digital Color Doppler Ultrasound System Fig.4-13 CW Mode 4.3.17 TDI Mode TDI mode is tissue Doppler mode, which is intended to provide information of low-velocity tissue motion, specifically for cardiac movement. Only phased array probe is available for TDI function. Fig.4-14 TDI Mode Press TDI-key to enter into TDI mode.
  • Page 84 Digital Color Doppler Ultrasound System internal forces. Based on the recorded signals, it analyzes, estimates and displays the strain of biological tissue. In B mode, turn elastography menu on can enter into elastography mode. The Elastography screen: Fig.4-15 Elastography Mode Change the size and the position of the Elastography box by using the trackball.
  • Page 85 Digital Color Doppler Ultrasound System 3. Use other keys 4.3.20 Image Optimization of B Mode Soft menu adjustment: Menu Function Q-Image Adjust the parameter of the Q-Image. Turn on the SRA function. Turn on the compound function. SRA and Compound can’t work at the Compound same time.
  • Page 86: Other Adjustments

    Digital Color Doppler Ultrasound System 2D Steer Adjust the steer angle of B image Elastography Turn on Elastography Center Line Turn on the center line Utility Post Processing, Slide Show, Demo Image, Protocal Shortcut adjustment: Shortcut(Action) Menu Menu function SK2(rotate) Frequency Adjust frequency of the probe.
  • Page 87 Digital Color Doppler Ultrasound System TGC with 8 adjustable slides: Slide the set of slider bars can change the depth gain of 2D images. TGC returns signal amplifier to correct for the attenuation caused when adding depth. TGC balance the image and make the density of echo uniformly distributed in images. Similarly the enlargement of every sliding channel area is also different.
  • Page 88 Digital Color Doppler Ultrasound System 4.3.22 Color Flow Map(CFM) Mode image optimization Soft menu adjustment: Menu Function Color Map Adjust the color map Color Invert Realize invert of blood flow Q-beam Turn on Q-beam function Q-flow Turn on Q-flow function Persistence Adjust the value of persistence Color Mode...
  • Page 89 Digital Color Doppler Ultrasound System CFM gain control. Rotate C- knob after C lights is on .Adjust the color gain and adjust range is from 0 to 255. Sampling box. Moves sampling box through trackball to the area which you are interested and press the ENTER-key to confirm the position of sampling box then adjust the size of box trough trackball and press ENTER-key to confirm.
  • Page 90 Digital Color Doppler Ultrasound System Other adjustments: CPA gain control Rotate C knob after CPA light is on then adjust the gain and the range is from 0 to 255. Sampling box. Moves sampling box through trackball to the area which you are interested and press the ENTER- key to confirm the position of sampling box then adjust the size of box trough trackball and press ENTER- key to confirm.
  • Page 91 Digital Color Doppler Ultrasound System Color Map Change the color map QuickAngle Turn on the Quick Angle function Auto Cal Turn on Auto Cal function Shortcut adjustment: Shortcut(action) Menu Menu Function SK2(rotate) Freq. Adjust the probe launch frequency SK3(rotate) Baseline Adjust the baseline position SK4(rotate)...
  • Page 92 Digital Color Doppler Ultrasound System 4.3.25 Continuous Wave Doppler Imaging Optimization(CW) Soft menu adjustment: Menu Function 2D Map Adjust the color of spectrum Spectrum Enhance Adjust the brightness of the spectrum Dynamic Adjust the dynamic range of the CW spectrum Audio Adjust from 1% to 100% Wall Filter...
  • Page 93 Digital Color Doppler Ultrasound System 4.3.26 ECG (option) The ECG module is a device that provides the 3 lead ECG signal acquisition for cardiac application. It is not intent for the ECG diagnostic purpose as in the 12-lead module. In the cardiac application, the ECG trace is displayed on the bottom of the screen.
  • Page 94 Digital Color Doppler Ultrasound System 4.3.28 Stress Echo (option) It works with a phased array probe. Fig.4-16 Press the Stress Echo buttion on the touch panel to enter the Stress Echo. Fig.4-17 When the Stress Echo turns on, select the Default 4x4 template to start the study. User can also open the Timer1 for the study if needed.Save the cine for the study.
  • Page 95: After Capturing The Image

    Digital Color Doppler Ultrasound System Fig.4-18 Press OK to exit the Protocol Analyze. User also can edit the Template, define the Number levels and Number projection views. Fig.4-19 4.4 After Capturing the Image 4.4.1 Adding Comments Comments means input the words or symbols on an ultrasound image to bring attention, notate or communicate information observed during the examination.
  • Page 96: Adding Body Mark

    Digital Color Doppler Ultrasound System image, cine review image, real-time image, frozen image. You can input characters as comments; add custom comment bar; insert the pre-defined comments from the comment library. Input characters key will display “︱”on the screen, and then system will enter into the comment Press status.
  • Page 97: Saving Still Images

    Digital Color Doppler Ultrasound System screen. Press the bodymark that want to be added, move trackball after adding the bodymark image and then adjust the position of probe. Rotate SK5 knob to adjust the probe direction. Press ENTER-key to confirm after finishing the adjustment. Press UPDATE-key and move the Trackball to change the position of the body mark.
  • Page 98 Digital Color Doppler Ultrasound System Fig.4-20 Fig.4-21 Play/Pause: Press this key to Play or Pause cine playback. Play Speed: Rotate the SK2 knob to adjust the playback speed of cine. Start Pos: Rotate the SK3 knob to set the frame of start position.
  • Page 99 Digital Color Doppler Ultrasound System End Pos: Rotate the SK4 knob to set the frame of final position. Reset Range: Reset the range of cine playback, the default range of playback is 1-maximum frame of current cine. Manual playback cine: Press cursor key and back to recall status then manual playback through trackball.
  • Page 100: Browse Images

    Digital Color Doppler Ultrasound System Delete the file. 4.4.5 Browse Images Press key to enter into the EasyView interface. Fig.4-23 Info: press this key can check the information of current patient. Report: press this key to enter into the report interface. Close: click this key and choose to close information of patient.
  • Page 101: Files Management

    Digital Color Doppler Ultrasound System 4.4.6 Fast Storage Setting: Enter into System Settings. Press General Menu and choose Normal Sub Menu to do fast storage setting. Fig.4-24 After confirming time, back to examination. For example, choose 5 seconds in the fast storage setting. The system begins to save cine in the 5 seconds after press If you choose Manual in the fast storage setting, you should press for two times.
  • Page 102 Digital Color Doppler Ultrasound System Fig.4-25 Search: Can search the patient files after enter a search term. Search will be more accurate when input more conditions. Information: View the selected patient information. Report: select a patient and click this item, patient report information will appear. Backup:Click this item will store the database of this patient to the portable storage medium Recover: Click this item can recover the images or video from USB to the system Send to:Select the content and click, can send files to other medium.
  • Page 103 Digital Color Doppler Ultrasound System Easy View:Select patient and click it to enter into easy view. Cancel:Click “Cancel” to exit. 4.4.8 Network Storage Function Before you use this function, please refer to Section 7.5 to set network storage parameters. Click button to enter into the Easy Review interface, select the picture you want to send, click to enter into the export interface, select the network drive and click "OK"...
  • Page 104 Digital Color Doppler Ultrasound System If the DICOM function is not open, the system setting interface will not display DICOM page. Please make sure the DICOM function is open before you use DICOM function. Fig.4- 27 Add: Press this icon to add a DICOM service link Modify: Select the DICOM link which need to be modified.
  • Page 105 Digital Color Doppler Ultrasound System Fig.4-28 Service: Click the drop list to select storage function. AE title: Name the server at will. Host IP: Set the IP of the service host. Port: Make the port of the work station the same as the port of the server. Alias: The name of current link.
  • Page 106: Dicom Print

    Digital Color Doppler Ultrasound System Batch mode: In batch mode, when the patient finishes his exam, press End key to end the current exam. All the saved images will be saved as DCM format and sent to the server automatically. Follow: In follow mode, during the exam, press key to save the image.
  • Page 107: Dicom Worklist

    Digital Color Doppler Ultrasound System Port: Make the port of the work station the same as the port of the server. Alias: The name of current link. Time out: time delay. Mode: batch, follow, manual selectable. Batch: all the saved images in current exam will be sent to the server and printed automatically in one time.
  • Page 108 Digital Color Doppler Ultrasound System Fig.4-30 Service: Click the drop list to select worklist function. AE title: Name the server at will. Host IP: Set the IP of the server. Port: Make the port of the work station the same as the port of the server. Alias: The name of current link.
  • Page 109 Digital Color Doppler Ultrasound System Fig.4-31 Press the keyboard button shaped as “ ”, and press “Worklist”. Press “Search” to start a worklist update by the matching key specified in Worklist Configuration, all the matched items will be queried in worklist table. User also can add additional for matching keys, which are “PatientID”, “PatientName”, “Accession#”...
  • Page 110 Digital Color Doppler Ultrasound System Fig.4-32 Service: Click the drop list to select MPPS function. AE title: Name the server at will Host IP: Set the IP of the server Port: Make the port of the work station the same as the port of the server. Alias: The name of current link Time out: time delay Applying the selected worklist item will result in automated creation of an MPPS instance.
  • Page 111 Digital Color Doppler Ultrasound System DICOM SR Storage Fig.4-33 Service: Click the drop list to select Storage function. Select the DICOM SR which below the Time Out item. AE title: Name the server at will. Host: Set the IP of the server. Port: Make the port of the work station the same as the port of the server.
  • Page 112: Chapter 5 General Measurements

    Digital Color Doppler Ultrasound System Chapter 5 General Measurements Introduction Measurements and calculations derived from ultrasound images are intended to supplement other clinical procedures available to the physician. The accuracy of measurements is not only determined by system accuracy, but also by the use of proper medical protocols by the operator. When appropriate, make sure to note any protocol associated with a particular measurement or calculation.
  • Page 113: Fast Measurement

    Digital Color Doppler Ultrasound System  【ENTER】 During the measurement, the functions of 【ENTER】key are as follows: When cursor is on the menu, press the key to choose the options and start the measurement. During the measurement, press the key to anchor the start point and end point. ...
  • Page 114 Digital Color Doppler Ultrasound System 5.2.3 Fast measurement in B mode Distance 1. In B Cine mode, press the DIST-key to bring up the first white plus sign (“+”) cursor. 2. Use the Trackball and the ENTER-key to anchor the start point of the desired distance to be measured.
  • Page 115 Digital Color Doppler Ultrasound System 3. Move the Trackball and press ENTER-key to set start point 4. The second point marker is displayed by a line connecting the first and second points. Use the Trackball to lengthen the line in order to change the diameter and press the ENTER-key to fix the length.
  • Page 116 Digital Color Doppler Ultrasound System finish the envelope of spectrum automatically. The cursor“+”will appear on the screen. 2. Move the trackball to choose a start point of one cycle, press the Enter-key to confirm. 3. The second cursor“+”will appear on the screen automatically, move the trackball to the end point of current cycle, and press the ENTER-key to set.
  • Page 117: Measurement And Calculation

    Digital Color Doppler Ultrasound System M Time 1. In M Cine mode, press the DIST-key or TRACE-key to enter M fast measurement. The cursor“+” on a line will appear on the screen. 2. Press MTime on the touch screen. 3. Move the trackball to move the cursor, press Enter-key to fix it. 4.
  • Page 118 Digital Color Doppler Ultrasound System 5.3.1 Measurement in B mode Press Calc-Key to enter into measurement package menu. Press Change-Key to change measurement package. 5.3.1.1 General Measurement in B mode Fig.5- 1 1. Distance 1) Move the trackball to move the cursor to the “distance” item. Press Enter-key to select it and a mark “+”will appear on the screen 2) Move the trackball to anchor the first mark and press Enter-key to set it.
  • Page 119 Digital Color Doppler Ultrasound System 2) Rotate the trackball to anchor the first mark and press Enter-key to fix it. 3) The second mark will appear on the screen. Rotate the trackball to change the length of the long axis of the ellipse. Press Enter-key to confirm the length. 4) Rotate the trackball to change the length of another axis of the ellipse and press Enter-key to fix it.
  • Page 120 Digital Color Doppler Ultrasound System 7. Volume(3 distance) Through 3 distance measurement, calculate the volume of object Volume=distance1×distance2×distance3×3.1415926/6 8. Volume(1distance1ellipse) Through 1distance measurement and 1 ellipse measurement, calculate the volume of object. Volume=distance ×axis A × axis B×3.1415926/6 9. Volume(2 ellipse) Through 2 ellipse measurement, calculate the volume of the object.
  • Page 121 Digital Color Doppler Ultrasound System 5.3.1.2 Cardiology Measurement in B mode Fig.5-2 1. Teichholz Calculation formula for volume  LVID   LVIDd  LVID   LVIDs <<Items to be measured>>...
  • Page 122 Digital Color Doppler Ultrasound System Meas. item Description Meas. Tool LVIDd [cm] Left ventricular internal diameter at end diastole Line LVIDs [cm] Left ventricular internal diameter at end systole Line << Items to be calculated >> Calc. item Description Calc. formula Please refer to <<Calculation formula EDV [ml] End-diastolic left ventricular Volume...
  • Page 123 Digital Color Doppler Ultrasound System ri: Radius of the i-th circle LVLd:Left ventricular long-axis diameter at end diastole LVLs:Left ventricular long-axis diameter at end systole <<Items to be measured>> Meas. item Description Meas. Tool EDV(A4C) [ml] End-diastolic volume (A4C) Measurement (trace) ESV(A4C) [ml] End-systolic volume (A4C ) Measurement (trace)
  • Page 124 Digital Color Doppler Ultrasound System r2i:Radius of i-th ellipse for 2CH LVLd2: Length of the left-ventricular long axis at end diastole for 2CH. LVLs2: Length of the left-ventricular long axis at end systole for 2CH . r4i:Radius of i-th ellipse for 4CH LVLd4: Length of the left-ventricular long axis at end diastole for 4CH.
  • Page 125 Digital Color Doppler Ultrasound System ESV(A4C) [ml] End-systolic volume (A4C) Measurement (trace) << Items to be calculated >> Calc. item Description Calc. formula End-diastolic left ventricular Please refer to <<Calculation formula EDV[ml] volume for volume>> Please refer to <<Calculation formula ESV[ml] End-systolic left ventricular volume for volume>>...
  • Page 126 Digital Color Doppler Ultrasound System systole Measurement (Distance)" Left ventricular short-axis area at the level of Refer to "Area/Circumference LVAMd[cm the mitral valve at end diastole Measurement (Area)" LVAMs [cm Left ventricular short-axis area at the level of Refer to "Area/Circumference the mitral valve at end systole Measurement (Area)"...
  • Page 127 Digital Color Doppler Ultrasound System Meas. item Description Meas. Tool IVSd[cm] lnterventricular septal thickness at end diastole Line LVIDd[cm] Left ventricular short-axis diameter at end diastole Line LVPWd[cm] Left ventricular posterior wall thickness at end diastole Line IVSs[cm] lnterventricular septal thickness at end systole Line LVIDs[cm] Left ventricular short-axis diameter at end systole...
  • Page 128 Digital Color Doppler Ultrasound System <<Items to be measured>> Meas. item Description Meas. Tool LVLd[cm] Left ventricular long-axis length at end diastole. Line LVLs[cm] Left ventricular long-axis length at end systole. Line LVAMd[cm] Left ventricular short-axis area at the level of the mitral valve at Area end diastole.
  • Page 129 Digital Color Doppler Ultrasound System Meas. item Description Meas. Tool LVLd[cm] Left ventricular long-axis length at end diastole. Line LVLs[cm] Left ventricular long-axis length at end systole. Line LVAMd[cm] Left ventricular short-axis area at the level of the mitral valve at Area end diastole.
  • Page 130 Digital Color Doppler Ultrasound System Calc. item Description Calc. formula LAD/AOD LAD/AOD AOD/AOD 9. Aortic Valve <<Items to be measured>> Meas. item Description Meas. Tool AV Diam[cm] Aortic valve diameter Line AV Area[cm Aortic valve area Ellipse 10. Pulmonary Valve <<Items to be measured>>...
  • Page 131 Digital Color Doppler Ultrasound System LVOT Area[cm Left ventricular outflow tract area Ellipse 13. RVOT <<Items to be measured>> Meas. item Description Meas. Tool RVOT Diam[cm] Right ventricular outflow tract diameter Line RVOT Area[cm Right ventricular outflow tract area Ellipse 14.
  • Page 132 Digital Color Doppler Ultrasound System Calc. item Description Calc. formula LV Mass(g) Left ventricular mass 1.04*[(LVPWd+IVSd+LVIDd) -LVIDd ]-13.6 LV Mass index Left ventricular mass index LV Mass/BSA A-L(Area-Length) <<Items to be measured>> Meas. item Description Meas. Tool LVAd sax Epi[cm Left ventricular epicardial short axis area at the Area level of the papillary muscle tips at end diastole...
  • Page 133 Digital Color Doppler Ultrasound System From the widest long axis radius to the semi-major a[cm] Line axis tips From the widest long axis radius to the semi-major d[cm] Line axis tips of mitral valve <<Items to be calculated>> Calc. item Description Calc.
  • Page 134 Digital Color Doppler Ultrasound System SV=0.785*AV Diam * AV Diam*abs(VTI AV SV(ml) Aortic valve Stroke Volume PV HR Pulmonary valve Heart Rate HR=VTI/Mean SV=0.785*AV Diam * AV Diam*abs(VTI PV SV Pulmonary valve Stroke Volume PV CO Pulmonary valve Cardiac output CO=SV*HR/1000 The ratio of Pulmonary flow volume Qp/Qs...
  • Page 135 Digital Color Doppler Ultrasound System IVC Ins / IVC Exp The ratio of IVC Ins and IVC Exp IVC Ins/IVC Exp 19.RA/LA <<Item to be measured>> Description Meas. method Meas. item name RA[cm Right Atrium Measurement (Area) LA[cm Left Atrium Measurement (Area) <<Items to be calculated>>...
  • Page 136 Digital Color Doppler Ultrasound System 5.3.1.3 Urology Measurement in B mode Fig.5-3 Meas. item mark unit Meas. method name Refer to "Distance Measurement (Distance)". Pre Length Pre Width Refer to "Distance Measurement (Distance)". Pre Height Refer to "Distance Measurement (Distance)". Pre Vol.= (π/6) * Pre Length* Pre Width* Pre Height Pre Vol.
  • Page 137 Digital Color Doppler Ultrasound System Height Refer to "Distance Measurement (Distance)" V = (π/6) * PV Length* PV Width* PV Height Volume PPSA PPSA=0.12×PV PSAD ng/ ml PSAD = SPSA[ng]/PV[ml] Note: SPSA must be input by manual in patient interface when you select the urology exam type Length Refer to "Distance Measurement (Distance)
  • Page 138 Digital Color Doppler Ultrasound System Vessel Dis Vessel Dis Refer to "Distance Measurement (Distance) 5.3.1.4 Pediatric Measurement in B mode Fig.5-4 Hip is used for evaluating the abnormity of the cotyle. In order to make measurement, user has to draw three lines on the image responding to the anatomy. The system will calculate the angles automatically.
  • Page 139 Digital Color Doppler Ultrasound System between D1 and D3(acute angle). 5.3.1.5 Vascular Measurement in B mode Fig.5-6 For CCA and ICA, user also can measure the thickness of intimae. IMT: measure the thickness of intimae manual. IMT Auto: get a good B image for intimae, and then click this measurement. Set start position and end position, besure to contain the intimae in the box, the system will calculate the thickness of intimae.
  • Page 140 Digital Color Doppler Ultrasound System Fig.5-7...
  • Page 141 Digital Color Doppler Ultrasound System Fig.5-8 Meas. item mark unit Meas. Method and calc. formula name Refer to “Area/Circumference Measurement (Area)” A Out ² Refer to “Area/Circumference Measurement (Area)” (Rt)/(Lt) A In ² StA% StA%= (A Out- A In)/ A Out *100% D Out Refer to "Distance Measurement (Distance) D In...
  • Page 142 Digital Color Doppler Ultrasound System (Rt)/(Lt) The same The same as above as above The same as above (Rt)/(Lt) Vertebral A The same The same The same as above (Rt)/(Lt) as above as above INT IIL The same The same as above as above The same as above...
  • Page 143 Digital Color Doppler Ultrasound System (Rt)/(Lt) PERON The same The same as above as above The same as above (Rt)/(Lt) DRPED The same The same The same as above (Rt)/(Lt) as above as above 5.3.1.6 GYN Measurement in B mode Fig.5- 9 Meas.
  • Page 144 Digital Color Doppler Ultrasound System Width Refer to "Distance Measurement (Distance) " PV = (π/6) * L* W* H Volume ENDO endometrium Refer to "Distance Measurement (Distance) " OV_Volume LengthWidthHeigh Refer to "Distance Measurement (Distance) " tVolume Refer to "Distance Measurement (Distance) " Refer to "Distance Measurement (Distance) "...
  • Page 145 Digital Color Doppler Ultrasound System Fig.5- 10 Meas. item name Mark Unit Meas. Method and calc. formula Refer to "Distance Measurement (Distance)” week/day EDD = current date + (280days – average UGA) mm/dd/yyyy Refer to "Distance Measurement (Distance) week/day EDD = current date + (280days – average UGA) mm/dd/yyyy Refer to "Distance Measurement (Distance)”...
  • Page 146 Digital Color Doppler Ultrasound System Refer to "Distance Measurement (Distance)” week/day AC(ellipse) mm/dd/yyyy Refer to "Distance Measurement (Distance)” week/day mm/dd/yyyy Refer to "Distance Measurement (Distance)” week/day mm/dd/yyyy Humerus Refer to "Distance Measurement (Distance)” Humerus week/day mm/dd/yyyy Refer to "Distance Measurement (Distance)” week/day mm/dd/yyyy Refer to "Distance Measurement (Distance)
  • Page 147 Digital Color Doppler Ultrasound System week/day mm/dd/yyyy Refer to "Distance Measurement (Distance)” APTD Refer to "Distance Measurement (Distance)” APTD week/day mm/dd/yyyy Refer to "Distance Measurement (Distance)” week/day mm/dd/yyyy Refer to "Distance Measurement (Distance)” ULNA Refer to "Distance Measurement (Distance)” ULNA week/day mm/dd/yyyy Tibia...
  • Page 148 Digital Color Doppler Ultrasound System Refer to "Distance Measurement (Distance)” Refer to "Distance Measurement (Distance)” Refer to "Distance Measurement (Distance)” week/day mm/dd/yyyy Refer to "Distance Measurement (Distance)” Refer to "Distance Measurement (Distance)” LVent Refer to "Distance Measurement (Distance)” LVent Refer to "Distance Measurement (Distance)” Refer to "Distance Measurement (Distance)”...
  • Page 149 Digital Color Doppler Ultrasound System D Out Refer to "Distance Measurement (Distance) D In Refer to "Distance Measurement (Distance) StD% StD%= (D Out- D In)/ D Out *100% Vessel Area cm² Refer to Area(Ellipse) Vessel Dis Refer to "Distance Measurement (Distance) Refer to "Distance Measurement (Distance)”...
  • Page 150 Digital Color Doppler Ultrasound System above as above 5.3.1.8 Carotid Measurement in B mode Fig.5-11 Meas. item name Mark Unit Meas. Method and calc. formular Refer to "Area Measurement (Ellipse)” A Out ² Refer to "Area Measurement (Ellipse)” A In ²...
  • Page 151 Digital Color Doppler Ultrasound System (Lt/Rt) above above The same as The same as The same as above above above (Lt/Rt) The same as The same as The same as above above above (Lt/Rt) Vertebral A The same as The same as The same as above above above...
  • Page 152 Digital Color Doppler Ultrasound System Refer to "Distance Measurement (Distance)” Height Refer to "Distance Measurement (Distance)” Width Refer to "Area Measurement (Trace)” A Out Refer to "Area Measurement (Trace)” A In ² StA% StA%=(A Out-A In)/ A Out*100% Artery Refer to "Distance Measurement (Distance)” D Out Refer to "Distance Measurement (Distance)”...
  • Page 153 Digital Color Doppler Ultrasound System Refer to "Distance Measurement (Distance)” D Out Refer to "Distance Measurement (Distance)” D In StD% StD%=(D Out-D In)/ D Out*100% Refer to "Area Measurement (Trace)” Vessel Area ² Refer to "Distance Measurement (Distance)” Vessel Dis 5.3.1.10 Small Parts Measurement in B mode Refer to general measurement in B mode.
  • Page 154 Digital Color Doppler Ultrasound System Meas. item name mark unit Meas. Method and calc. Formula Refer to “Area/Circumference Measurement (Area)” A Out ² Refer to “Area/Circumference Measurement (Area)” A In ² StA% StA%= (A Out- A In)/ A Out *100% D Out Refer to "Distance Measurement (Distance) (Rt)/(Lt)...
  • Page 155: General Measurement In M Mode

    Digital Color Doppler Ultrasound System (Rt)/(Lt) as above as above PCOA The same The same The same as above as above as above (Rt)/(Lt) The same The same The same as above as above as above (Rt)/(Lt) Vertebral A The same The same The same as above as above...
  • Page 156 Digital Color Doppler Ultrasound System Rotate the trackball to select the “M distance” item in the menu. A start cursor“+” will appear on the screen. Move the cursor through rotating the trackball and press Enter-key to fix the first point. The second cursor will appear.
  • Page 157 Digital Color Doppler Ultrasound System Refer to time measurement in general measurement in M mode. 3. Slope Refer to slope measurement in general measurement in M mode. 4. HR Refer to HR measurement in general measurement in M mode. 5. Left Ventricle Cube Calculation formula for volume: EDV ...
  • Page 158 Digital Color Doppler Ultrasound System Cardiac output [L/min] CO=SV x HR /1000 Ejection fraction EF=SV/ EDV Fractional shortening FS=(LVIDd-LVIDs)LVIDd Stroke Volume Index SI=SV/BSA Cardiac Output Index CI=CO/BSA Teichholz <<Calculation formula for volume>>  LVID   LVIDd  LVID  ...
  • Page 159 Digital Color Doppler Ultrasound System FS [%] Fractional Shortening FS=(LVIDd-LVIDs)/LVIDd Stroke Volume Index SI=SV/BSA Cardiac Output Index CI=CO/BSA Gibson Calculation formula for volume:      LVID LVIDd <<Items to be measured>> Meas. item Description Meas. Tool LVLd[cm] Left ventricular long-axis length at end diastole.
  • Page 160 Digital Color Doppler Ultrasound System <<Items to be measured>> Meas. item Description Meas. Tool IVSd[cm] lnterventricular septal thickness at end diastole M Line LVPWd[cm] Left ventricular posterior wall thickness at end diastole M Line 6. Mitral Valve <<Items to be measured>> Meas.
  • Page 161 Digital Color Doppler Ultrasound System 7. Aortic Valve <Items to be measured>> Meas. item name Description Meas. Method AOD[cm] Aorta diameter at diastole Refer to distance measurement in M mode Ao Sinus Aortic sinus diameter Refer to distance measurement in M mode Diam[cm] Ao Asc Diam[cm] Aorta ascendens diameter...
  • Page 162 Digital Color Doppler Ultrasound System E-F Time(s) E-F Distance(cm) MV A-C Int time A-C Int slope(cm/s) A-C Int Time(s) A-C Int Distance(cm) 9. Pulmonary Valve <<Items to be measured>> Meas. item name Description [Unit] RVPEP[s] Right ventricular pre ejection phase RVET[s] Right ventricular ejection time A wave Amp(cm)...
  • Page 163 Digital Color Doppler Ultrasound System 11.LV Mass <<Items to be measured>> Meas. item name Description IVSd lnterventricular septal thickness at end diastole LVIDd Left ventricular short-axis diameter at end diastole LVPWd Left ventricular posterior wall thickness at end diastole 12.TAPSE <<Items to be measured>>...
  • Page 164 Digital Color Doppler Ultrasound System 2) Move the cursor “+” to the place where need to be measured and press Enter-key to fix it. 3) The value of velocity and pressure will appear on the screen. 4) Repeat step 1) to 3) to measure the next point. 2.
  • Page 165 Digital Color Doppler Ultrasound System Fig.5- 16 Peak Velocity or Doppler frequency  Peak velocity over one cardiac cycle (Vpk) Doppler Time Distance  Time distance between two cursors in ms. The invert is the heart beat rate if the two cursors are at one cardiac cycle period.
  • Page 166 Digital Color Doppler Ultrasound System 5. Manual trace Automatically measure the Peak Systolic Velocity, VTI, end diastolic velocity, HR, Time, minimum velocity, PI, and RI after manually tracing of the curve is finished. NOTE: Manual trace requires the user to trace two peak point of two cycles.
  • Page 167 Digital Color Doppler Ultrasound System 8. Area The same as “area-trace” in general measurement in B mode. 9. ICA/CCA Measure ICA and CCA separately and get the ratio value of ICA/CCA 10. Volume Flow <<Items to be measured>> Meas. item name Description [Unit] Meas.
  • Page 168 Digital Color Doppler Ultrasound System name E Vel E-wave flow velocity [cm/s] Refer to "Velocity Measurement'. A Vel A-wave flow velocity [cm/s] Refer to "Velocity Measurement'. Pressure half time [ms] Refer to " Velocity Measurement". E Dur E-wave duration [ms] Refer to "Time Measurement (Time)".
  • Page 169 Digital Color Doppler Ultrasound System Fig.5-17 2) Measure [MV V Trace]. Fig.5-18 3) Measure [MV Diameter Fig.5-19 2. AV Measurement of aortic valve function is performed by using 2-D-mode and Doppler images. <<Measurement items>> Meas. Item name Description [Unit] Meas. method LVOT Vel PW-mode LV out tract flow velocity Refer to "Velocity Measurement'.
  • Page 170 Digital Color Doppler Ultrasound System LVOT PG PW-mode LV out tract pressure Refer to "Velocity Measurement'. gradient [mmHg] LVOT V General Meas. Item name Description [Unit] Meas. method Refer to “Full trace” LVOT VTI Aortic valve VTI [m] Refer to “Full trace” LVOT VM Aortic valve mean velocity [cm/s] Refer to “Full trace”...
  • Page 171 Digital Color Doppler Ultrasound System Fig.5- 20 2) Measure [LVOT Diam] Fig.5- 21 3. TV Measurement of tricuspid valve function is performed by using Doppler images. <<Measurement items>> Meas. Item name Contents Meas. Method TV VTI Tricuspid valve VTI [m] Refer to "Manual Trace Measurement "...
  • Page 172 Digital Color Doppler Ultrasound System Refer to "Velocity Measurement” TV Vel Tricuspid valve velocity [cm/s] Tricuspid valve pressure gradient TV PG Refer to " Velocity Measurement' [mmHg] <<Measurement points>> 1) Measure [TV VEL Trace] Fig.5-22 2) Measure [TV VEL] Fig.5-23 4.
  • Page 173 Digital Color Doppler Ultrasound System Pulmonary valve maximum PV VP Refer to "Full Trace Measurement ". velocity [cm/s] PV MPG Pulmonary valve mean Refer to "Full Trace Measurement ". pressure gradient [mmHg] Refer to “ Velocity Measurement” PV PPG Pulmonary valve maximum pressure gradient [mmHg] Refer to "...
  • Page 174 Digital Color Doppler Ultrasound System Fig.5-24 2) Measure [PVV Trace] Fig.5-25 3) Measure [ET],[ACT], and [PEP] Fig.5-26 4) Measure [PV Diam]...
  • Page 175: Edit Measurement Results

    Digital Color Doppler Ultrasound System Fig.5-27 5.3.3.3 Urology Measurement in PW mode The measurement method and calculation items are the same as general measurement in PW mode. 5.3.3.4 Pediatric Measurement in PW mode The measurement method and calculation items are the same as general measurement in PW mode. 5.3.3.5 Small parts Measurement in PW mode The measurement method and calculation items are the same as general measurement in PW mode.
  • Page 176: Report

    Digital Color Doppler Ultrasound System To move result position, operations are as follows: 1. In the measurement status, rotate SK2-key to choose “Move result position”. 2. Press SK2-key, and use trackball to move the position of measurement results. 3. Press Enter-key to confirm it. To reset result position, operations are as follows: 1.
  • Page 177 Digital Color Doppler Ultrasound System Choose the image:Click the image can add the image to the image area in the report, and click the image in the report to remove it. Preview: preview the whole report and its format before printing Print: print the current report, please make sure that the printer is normal working.
  • Page 178 Digital Color Doppler Ultrasound System Fig.5-30 Input Template: Input annotation from templates. Fig.5-31...
  • Page 179: Chapter 6 Preset

    Digital Color Doppler Ultrasound System Chapter 6 Preset 6.1 Recall Preset a) Press button to choose the probe type. b) Choose a clinical application. c) Choose a preset. The default choice of the system is DEFAULT preset. After entering into the system, if you want to change preset and do not want to quit the current interface, operate as follows: 1.
  • Page 180: Manage Preset

    Digital Color Doppler Ultrasound System Rotate SK4 to choose user defined. Press SK4 for editing user name. (Just user 1 to user 5 can be edited). Press SK5 to save preset. 6.3 Manage Preset Press key to enter into system setting interface. Select preset page. Add or delete the current preset.
  • Page 181 Digital Color Doppler Ultrasound System Export: export system preset. Plug in USB, press the Export key, system automatically derive the preset to U disk. OK: exit system setting interface...
  • Page 182: Chapter 7 System Setting

    Digital Color Doppler Ultrasound System Chapter 7 System Setting Press the key to enter system setting interface. User can do user-defined setting. 7.1 General settings Normal Settings: Set the hospital information, date-time, language etc. Fig.7-1 Hospital Hospital; Department Input hospital name, department name Information To set Screen Saver ON or Screen Saver...
  • Page 183: Keyboard Settings

    Digital Color Doppler Ultrasound System Export to network media Always show STC Curve Image Option Select according to the requirement. Always show LGC Curve Set Date Time Set Date and time (need password) Date-Time Set Regional Set Time Zone, Data Format, Time Format Color Map Set Color Click the Configure key to set the color map...
  • Page 184: Key Config

    Digital Color Doppler Ultrasound System FOOT SWITCH1 Choose the function for User1 Function Setting FOOT SWITCH2 Choose the function for User2 P1 Key Choose the function for P1-key Back to Real- Time Scan End Exam Option Back to Worklist Select according to the requirement Back to Patient Information Trackball Option...
  • Page 185 Digital Color Doppler Ultrasound System Fig.7-3 Set the function for the 0~9 keys of Alphanumeric Keyboard. Select the Key Function that want to set, then select the Function for the selected key. Fig.7-4 Set the function for the 0~9 keys of Alphanumeric Keyboard. Select the Key Function that want to set, then select the Measure for the selected key.
  • Page 186: Function Config

    Digital Color Doppler Ultrasound System Function Config: Fig.7-5 Items Description Para Info Related Select to display TIB in B/C mode and show para setting name Focus Setting Select it to change the focus by trackball in B Mode or focus position changed by depth Patient Info Display Select to display Gender and Age...
  • Page 187: Measurement

    Digital Color Doppler Ultrasound System 7.2 Measurement Set the measurement formula of measurement. Edit Settings: Click the measurement page, and click on the Edit page, then users can edit custom configuration measurement. Fig.7-6 1. Package Name: Choose the different examinations. 2.
  • Page 188 Digital Color Doppler Ultrasound System General Settings: Click the General page, and you can do some general settings of measurement items Fig.7-7 Items Description Configuration Select according to the requirement Font size Set to default font size of the measurement results One: one cycle to measure heart rate in the Doppler envelope calculation Two: two cycles to measure heart rate in the Doppler envelope...
  • Page 189: Comment

    Digital Color Doppler Ultrasound System OB Table Setting: Click OB Table page, you can set the formula of the OB measurement items for measuring fetal weight formula. Fig.7-8 Set measurement Item: choose a measurement project, gestational age table or developmental table, and then choose a formula, click the information button, you can view the current formula.
  • Page 190: Library Page

    Digital Color Doppler Ultrasound System Fig.7-9 Search option Search comment or Input search function Abbreviations Select the comment item, it will search notes when searching Select abbreviations, it will search abbreviations when searching Font size Font size Set the font size of notes Library Page: Edit and delete the annotation in the comment library.
  • Page 191 Digital Color Doppler Ultrasound System Fig.7-10 Comment Libs Provide the comment notes in different exam types. Comment UserLibs List the comment notes have been chosen from Libs and added by user. Establish the new exam type: click New near the drop-down list on the right to establish the new exam type.
  • Page 192: Bodymark

    Digital Color Doppler Ultrasound System edit the selected comment note. Click Up, Down, Top, Button to adjust the sequence of the comment note selected. Click OK to save the edit and exit the system settings interface. 7.4 BodyMark Fig.7-11 This function is to preset body marks in each preset or application of probes. Libs: provide different body marks of different presets or applications of probes UserLibs:list the body mark chosen from the Libs.
  • Page 193: Report

    Digital Color Doppler Ultrasound System Click to add the selected item in the body mark library into User Body Mark. Click to add all items in the body mark library into User Body Mark. Click Delete icon to delete the selected item on the right. Click Delete All icon to delete all items on the right.
  • Page 194: Template Page

    Digital Color Doppler Ultrasound System Fields: Display the content of the selected section. All: Click this button to add “available section “to " selected section. Add: Select the section that need to be added, then click this button, add selected section "selected section".
  • Page 195: General Page

    Digital Color Doppler Ultrasound System Add Folder: Press this icon to add a new Folder and name it. Add Item: Press this icon to a new Item under current folder, and name it. Delete: Press this icon to delete the template, which you selected. Save: Save the changes.
  • Page 196: Admin

    Digital Color Doppler Ultrasound System 7.6 Admin Fig.7-15 NOTE: If the HIPPA function is not open, the system setting interface will not display Admin page. Please make sure the HIPPA function is open before you use this function. Enable Account Control: Select the “Enable Account Control” to enable Account Control.Unselect the “Enable Account Control”...
  • Page 197: Network

    Digital Color Doppler Ultrasound System Fig.7-16 User Name: Select the user name which want to login. Password: Input the password for user name. Login: Click the icon to login the system. Emergency: Login as “Emergency User”, no password required.The Emergency user only can review and manager the patient information that was created by Emergency User.
  • Page 198 Digital Color Doppler Ultrasound System Fig.7-17 Network Settings: Network Adapter Display the existing network cards on this system. The user can select the network card for current use. Obtain an IP address When this option is selected, the system will obtain an IP automatically address automatically Use the following IP address...
  • Page 199 Digital Color Doppler Ultrasound System Port Input the port information Network Connection Status Check the network connection status. Connected The system is connected to the network successfully. Unconnected The system is not connected to the network. Network Information Display the current network information Config Configuration information: Click this button to display the network configuration information...
  • Page 200 Digital Color Doppler Ultrasound System Fig.7-18 Click “Advanced Sharing” Fig.7-19 Select “Share this folder”, and click “Permissions” to modify user permissions.
  • Page 201 Digital Color Doppler Ultrasound System Fig.7-20 Select “Full Control”, “Change”, “Read” for “Allow”, click “OK” to complete modification. Click Network Storage page, enter the network storage settings page. Then input shared directory name and IP address, and click Add. Fig.7-21 Shared Dir: The shared directory name should be same as the shared folder name in target computer.
  • Page 202: System

    Digital Color Doppler Ultrasound System IP address: the IP address of the computer with the shared directory. User name: input the name of computer account. Password: input the password of computer account. Ping: Click it test the communication between the shared PC and the system. Add: add a network storage service.
  • Page 203 Software Lease: Key Input It is time to the data time, the system cannot be entered for use. User can decode by pressing Key Input key and input the code which gets from CHISON's authorized service engineer. Log: Export Log Plug in U disk to export log files which saved on system automatically for engineers to analyze.
  • Page 204 Digital Color Doppler Ultrasound System Fig.7-23 Operating methods: insert U disk with installation driver to the system, click “Install Driver”. Fig.7-24 Select the driver, click “Install”, the system will jump out installation guide. Complete the installation by following up the guide. The steps are as below:...
  • Page 205 Digital Color Doppler Ultrasound System Fig.7-25 Click “Yes”. Fig.7-26 Select “Dynamic Mode”, click “Install”.
  • Page 206 Digital Color Doppler Ultrasound System Fig.7-27 Being installed automatically. Fig.7-28 Click “Finish” to finish the printer installation. Add:Click “Add”, complete adding printer according to operation guide.
  • Page 207 Digital Color Doppler Ultrasound System Fig.7-29 Click “NEXT”. Fig.7-30 Select “Local printer attached to this computer”, and Click “Next”.
  • Page 208 Digital Color Doppler Ultrasound System Fig.7-31 Select “Use the following port”, and click “Next”. Fig.7-32 Select the printer, and click "Next".
  • Page 209 Digital Color Doppler Ultrasound System Fig.7-33 Select “Keep existing driver (recommended)”, and Click “Next”. Fig.7-34 Select “Yes”, and Click "Next".
  • Page 210 Digital Color Doppler Ultrasound System Fig.7-35 Select “Yes” to print a test page, and click “Next” to continue. Fig.7-36 Click “Finish”.
  • Page 211: Chapter 8 Probes

    Digital Color Doppler Ultrasound System Chapter 8 Probes 8.1 General Description Fig.8-1: Convex Probe Overview The probes provide high spatial and contrast ultrasound imaging. These probes operate by pulsing sound waves into the body and listening to the returning echoes to produce high-resolution brightness mode, and a real time display.
  • Page 212: Inspecting Probes

    Before and after each use, inspect carefully the probe’s lens, cable, casing, and connector. Look for any damage that would allow liquid to enter the probe. If any damage is suspected, do not use the probe until it has been inspected and repaired/replaced by a CHISON's authorized service engineer.
  • Page 213 Digital Color Doppler Ultrasound System Fig.8-1 Recommended materials for cleaning and disinfecting probe Solution Origin Qualified Use Active Ingredients Purpose CIDEZYME Soak enzyme Clean Epizyme Rapid(Rapid Soak enzyme Clean Multi-Enzyme cleaner ANIOSYME DD1 Soak enzyme Clean neodisher mediclean Soak enzyme Clean MetriZyme Soak...
  • Page 214: Probe Safety

    Before inserting the connector into the probe port, inspect the probe connector pins. If a pin is bent, DO NOT use the probe until it has been inspected and repaired/replaced by a CHISON's authorized service engineer.  Electrical leakage checks should be performed on a routine basis by CHISON's authorized service engineer. Mechanical hazard:...
  • Page 215: Special Handling Instructions

    Digital Color Doppler Ultrasound System A defective probe or excess force can cause patient injury or probe damage:  Observe depth markings and do not apply excessive force when inserting or manipulating endocavitary probe.  Inspect probes for sharp edges or rough surfaces that may injure sensitive tissue. ...
  • Page 216: Probe Handling And Infection Control

    Digital Color Doppler Ultrasound System Sterilant Exposure from Probe handle to Patient: DO NOT allow the sterilant to contact the patient. Only immerse the probe to its specified level. Ensure that no solution has entered the probe’s handle before scanning the patient. If sterilant comes into contact with the patient, refer to the sterilant’s instruction manual.
  • Page 217 Digital Color Doppler Ultrasound System so could damage the system. Perform Cleaning probe after each use  Disconnect the probe from the ultrasound console and remove all coupling gel from the probe by wiping with a soft cloth and rinsing with flowing water. ...
  • Page 218 Digital Color Doppler Ultrasound System gown if there is a risk of splashing or splatter. Disinfecting the probes: After each use, please disinfect the probes. Ultrasound probes can be disinfected using liquid chemical germicides. The level of disinfection is directly related to the duration of contact with the germicide.
  • Page 219 CAUTION: Please do not use any gel or other materials which are not provided by CHISON. Un-authorized gel, lubricants and other materials may corrode probes and other parts of the device, for example the keyboard. This may reduce the safety and effectiveness of the system and probes, and may also reduce the life time of the systems and probes.
  • Page 220: Planned Maintenance

    Digital Color Doppler Ultrasound System  Methanol, ethanol, isopropanol, or any other alcohol-based product.  Mineral oil  Iodine  Lotions  Lanolin  Aloe Vera  Olive Oil  Methyl or Ethyl Parabens (para hydroxybenzoic acid)  Dimethylsilicone Planned maintenance The following maintenance plan is suggested for the system and probes to ensure optimum operation and safety.
  • Page 221 Digital Color Doppler Ultrasound System AIUM outlines cleaning the endocavitary transducer: Guidelines for Cleaning and Preparing Endocavitary Ultrasound Transducers Between Patients From AIUM Approved June 4, 2003 The purpose of this document is to provide guidance regarding the cleaning and disinfection of transvaginal and transrectal ultrasound probes.
  • Page 222 Digital Color Doppler Ultrasound System High-Level Disinfection - Destruction/removal of all microorganisms except bacterial spores. Mid-Level Disinfection - Inactivation of Mycobacterium Tuberculosis, bacteria, most viruses, fungi, and some bacterial spores. Low-Level Disinfection - Destruction of most bacteria, some viruses and some fungi. Low-level disinfection will not necessarily inactivate Mycobacterium Tuberculosis or bacterial spores.
  • Page 223 Digital Color Doppler Ultrasound System Other agents such as quaternary ammonium compounds are not considered high level disinfectants and should not be used. Isopropanol is not a high level disinfectant when used as a wipe and probe manufacturers generally do not recommend soaking probes in the liquid. The FDA has published a list of approved sterilants and high level disinfectants for use in processing reusable medical and dental devices.
  • Page 224 Digital Color Doppler Ultrasound System modification of this protocol. These guidelines take into account possible probe contamination due to a disruption in the barrier sheath. In summary, routine high-level disinfection of the endocavitary probe between patients, plus the use of a probe cover or condom during each examination is required to properly protect patients from infection during endocavitary examinations.
  • Page 225: Probe Operation Instructions

    Digital Color Doppler Ultrasound System 8.3 Probe Operation Instructions For details on connecting, activating, deactivating, disconnecting, transporting and storing the probes, see Section 3.7 “Probes” in Chapter 3. 8.3.1 Scanning the Patient In order to assure optimal transmission of energy between the patient and probe, a conductive gel must be applied liberally to the patient where scanning will be performed.
  • Page 226 Digital Color Doppler Ultrasound System  Load some ultrasound gel into condom.  Take the condom with one hand, and put the probe head into the condom.  Fasten the condom on the end of the probe handle.  Confirm the integrity of the condom, and repeat the above steps to the condom if any damage to the condom is found.
  • Page 227 Digital Color Doppler Ultrasound System condom is one time usable).  When cleaning the endo-cavitary probe, it is important to be sure that all surfaces are thoroughly cleaned. Disinfecting To keep the effectiveness of the disinfection solutions, a thoroughly cleaning must be done to the probe before the disinfecting, make sure no residues remain on the probe.
  • Page 228: Chapter 9 System Maintenance And Troubleshooting

    9.2 System Care and Maintenance The system is a precise electrical device. To ensure the best performance and operation of the system, observe proper maintenance procedures. Contact the local CHISON’s Authorized Service Representative for parts or periodic maintenance inspections. CHISON will provide circuit diagrams, component part lists, descriptions, calibration instructions to assist to SERVICE PERSONNEL in parts repair.
  • Page 229 To avoid electric shock hazard, do not remove panels or covers from console. This servicing must be performed by CHISON's authorized service engineer. Failure to do so could serious injury. If any defect is observed or malfunctions occur, do not operate the equipment but inform CHISON’s authorized service engineer for information.
  • Page 230 Digital Color Doppler Ultrasound System  Do NOT use a glass cleaner that has a hydrocarbon base (such as Benzene, Methyl Alcohol or Methyl Ethyl Ketone) on monitors with the filter (anti-glare shield). Hard rubbing will also damage the filter. NOTE: When cleaning the screen, make sure not to scratch the LCD.
  • Page 231: Safety Check

    Digital Color Doppler Ultrasound System cleansers because they will damage the cabinet. No further maintenance, such as lubrication, is required.  For more information, see the Printer's Operation Manual.  9.3 Safety Check To ensure the system work normally, please make a maintenance plan, check the safety of the system periodically.
  • Page 232 Digital Color Doppler Ultrasound System 3) Adjust the image parameters, e.g. Gain, Dynamic range. No OB calculation package Select the OB application before scanning. menu PRINT-key doesn’t work 1) Check if the approved printer is connected; 2) Check if the printer power is on; 3) Check printer connection;...
  • Page 233: Service Responsibility

    Digital Color Doppler Ultrasound System 9.5 Service Responsibility If users install, use and maintain the system fully according to CHISON exchange the momanual, operation manual and service manual, then main unit has a life time of 5 years and probes have life time of 5 years after ex-work (except that the 4D probe life time is 18 months).
  • Page 234 Digital Color Doppler Ultrasound System REFERENCE: 1. AIUM: Medical Ultrasound Safety, American Institute of Ultrasound in Medicine, Laurel, MD, 1994. NOTE: As of this writing an update of this publication is being finalized by the AIUM. AIUM: Acoustic Output Labeling Standard for Diagnostic Ultrasound Equipment: A Standard for How Manufacturers Should Specify Acoustic Output Data, Revision 1, American Institute of Ultrasound in Medicine, Laurel, MD, 2008.
  • Page 235 Digital Color Doppler Ultrasound System 11. Medical Electrical Equipment standard IEC 60601-1, IEC60601-1-2, IEC 60601-2-37 12. Diagnostic Ultrasound Physics and Equipment, edit by P. R. Hoskins, in 2003...
  • Page 236: Appendix A The Information Of Ec Representative

    Digital Color Doppler Ultrasound System APPENDIX A The Information of EC Representative Shanghai International Holding Corp. GmbH (Europe) Add: Eiffestrasse 80, 20537 Hamburg, Germany Tel: 0049-40-2513175 Fax: 0049-40-255726 E-mail: antonsissi@hotmail.com shholding@hotmail.com...
  • Page 237: Appendix B System One-Key-Recovery Function

    Digital Color Doppler Ultrasound System APPENDIX B System one-key-recovery Function This system has one-key-recovery function. Uses can use this function to recovery system when the system has problems. Detailed operations are as follows: Press button to boot the system. Press <sys> button when the system enter into the following boot screen. When the recovery system starts, the following interface will pop up.
  • Page 238 Digital Color Doppler Ultrasound System Click “Yes” to confirm and start system recovery. The system will start to backup ultrasound system data, after finishing the backup, the system will transfer to the recovery procedure automatically. Click “No” to cancel system recovery. After click the “...
  • Page 239: Appendix C Maximum Acoustic Output Report

    Digital Color Doppler Ultrasound System APPENDIX C MAXIMUM ACOUSTIC OUTPUT REPORT Transducer Model: D3C60L Operating Mode: B Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.34 r,α – 149.33 149.33 77.21 77.21 –...
  • Page 240 Digital Color Doppler Ultrasound System Transducer Model: D3C60L Operating Mode: B+C Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 0.93 r,α – 144.78 144.78 50.69 50.69 – Acoustic Parameters – 6.15 –...
  • Page 241 Digital Color Doppler Ultrasound System Transducer Model: D3C60L Operating Mode: PW Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 0.81 r,α – 210.22 198.89 73.17 70.12 4.08 Acoustic Parameters 6.35 4.40 4.40 pii,α...
  • Page 242 Digital Color Doppler Ultrasound System Transducer Model: D3C60L Operating Mode: B+M Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.51 r,α – 246.23 246.23 60.87 60.87 4.16 Acoustic Parameters 5.08 5.60 5.60 pii,α...
  • Page 243 Digital Color Doppler Ultrasound System Transducer Model: D3C60L Operating Mode: Elasticity imaging Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.47 r,α – 160.12 160.12 84.93 84.93 – Acoustic Parameters – 6.55 –...
  • Page 244 Digital Color Doppler Ultrasound System Transducer Model: D7L40L Operating Mode: B Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 2.39 r,α – 30.34 30.34 16.94 16.94 – Acoustic Parameters – 2.00 –...
  • Page 245 Digital Color Doppler Ultrasound System Transducer Model: D7L40L Operating Mode: B+C Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.39 r,α – 46.25 46.25 26.37 26.37 – Acoustic Parameters – 1.65 –...
  • Page 246 Digital Color Doppler Ultrasound System Transducer Model: D7L40L Operating Mode: PW Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.59 r,α – 40.13 46.59 27.16 31.28 1.75 Acoustic Parameters 1.83 1.70 1.70 pii,α...
  • Page 247 Digital Color Doppler Ultrasound System Transducer Model: D7L40L Operating Mode: B+M Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.91 r,α – 65.49 66.13 33.87 35.41 1.78 Acoustic Parameters 1.82 1.65 1.65 pii,α...
  • Page 248 Digital Color Doppler Ultrasound System Transducer Model: D7L40L Operating Mode: Elasticity imaging Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 2.24 r,α – 36.48 36.48 20.11 20.11 – Acoustic Parameters – 2.00 –...
  • Page 249 Digital Color Doppler Ultrasound System Transducer Model: D10L30L Operating Mode: B Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 2.42 r,α – 28.59 28.59 11.13 11.13 – Acoustic Parameters – 2.00 –...
  • Page 250 Digital Color Doppler Ultrasound System Transducer Model: D10L30L Operating Mode: B+C Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.53 r,α – 35.21 35.21 17.42 17.42 – Acoustic Parameters – 1.80 –...
  • Page 251 Digital Color Doppler Ultrasound System Transducer Model: D10L30L Operating Mode: PW Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.73 r,α – 39.69 43.81 15.15 18.56 1.89 Acoustic Parameters 1.94 1.80 1.80 pii,α...
  • Page 252 Digital Color Doppler Ultrasound System Transducer Model: D10L30L Operating Mode: B+M Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 2.11 r,α – 55.46 54.94 23.98 25.26 1.88 Acoustic Parameters 2.00 2.00 2.00 pii,α...
  • Page 253 Digital Color Doppler Ultrasound System Transducer Model: D10L30L Operating Mode: Elasticity imaging Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 2.12 r,α – 28.61 28.61 10.70 10.70 – Acoustic Parameters – 2.00 –...
  • Page 254 Digital Color Doppler Ultrasound System Transducer Model: D12L40L Operating Mode: B Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.39 r,α – 12.33 12.33 3.55 3.55 – Acoustic Parameters – 1.80 –...
  • Page 255 Digital Color Doppler Ultrasound System Transducer Model: D12L40L Operating Mode: B+C Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.18 r,α – 32.45 32.45 15.56 15.56 – Acoustic Parameters – 1.80 –...
  • Page 256 Digital Color Doppler Ultrasound System Transducer Model: D12L40L Operating Mode: PW Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 0.54 r,α – 54.14 54.14 28.41 28.41 1.80 Acoustic Parameters 1.80 3.00 3.00 pii,α...
  • Page 257 Digital Color Doppler Ultrasound System Transducer Model: D12L40L Operating Mode: B+M Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.21 r,α – 45.61 44.56 32.20 31.23 1.80 Acoustic Parameters 1.80 1.80 1.80 pii,α...
  • Page 258 Digital Color Doppler Ultrasound System Transducer Model: D12L40L Operating Mode: Elasticity imaging Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.22 r,α – 15.24 15.24 7.11 7.11 – Acoustic Parameters – 1.80 –...
  • Page 259 Digital Color Doppler Ultrasound System Transducer Model: D6C12L Operating Mode: B Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.46 r,α – 30.33 30.33 16.37 16.37 – Acoustic Parameters – 2.50 –...
  • Page 260 Digital Color Doppler Ultrasound System Transducer Model: D6C12L Operating Mode: B+C Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 0.91 r,α – 61.25 61.25 32.06 32.06 – Acoustic Parameters – 3.35 –...
  • Page 261 Digital Color Doppler Ultrasound System Transducer Model: D6C12L Operating Mode: PW Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.11 r,α – 36.28 39.56 21.16 24.41 1.77 Acoustic Parameters 1.93 2.40 2.40 pii,α...
  • Page 262 Digital Color Doppler Ultrasound System Transducer Model: D6C12L Operating Mode: B+M Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.72 r,α – 34.26 38.33 21.60 22.42 1.78 Acoustic Parameters 2.07 2.30 2.30 pii,α...
  • Page 263 Digital Color Doppler Ultrasound System Transducer Model: D6C12L Operating Mode: Elasticity imaging Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.77 r,α – 25.42 25.42 12.28 12.28 – Acoustic Parameters – 2.50 –...
  • Page 264 Digital Color Doppler Ultrasound System Transducer Model: D7C10L Operating Mode: B Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 0.99 r,α – 16.65 16.65 10.17 10.17 – Acoustic Parameters – 2.25 –...
  • Page 265 Digital Color Doppler Ultrasound System Transducer Model: D7C10L Operating Mode: B+C Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 0.96 r,α – 50.24 50.24 27.95 27.95 – Acoustic Parameters – 2.46 –...
  • Page 266 Digital Color Doppler Ultrasound System Transducer Model: D7C10L Operating Mode: PW Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.07 r,α – 55.58 58.44 38.25 40.55 1.73 Acoustic Parameters 1.99 2.60 2.60 pii,α...
  • Page 267 Digital Color Doppler Ultrasound System Transducer Model: D7C10L Operating Mode: B+M Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 0.98 r,α – 50.22 51.53 30.14 30.58 1.75 Acoustic Parameters 2.29 2.40 2.40 pii,α...
  • Page 268 Digital Color Doppler Ultrasound System...
  • Page 269 Digital Color Doppler Ultrasound System Transducer Model: D7C10L Operating Mode: Elasticity imaging Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 0.81 r,α – 31.23 31.23 20.19 20.19 – Acoustic Parameters – 2.25 –...
  • Page 270 Digital Color Doppler Ultrasound System Transducer Model: D3P64L Operating Mode: B Index label Below Below surface surface surface surface Maximum index value Index component value at Z 1.63 r,α 142.23 142.23 142.23 101.61 101.61 – Acoustic Parameters – 4.50 – pii,α...
  • Page 271 Digital Color Doppler Ultrasound System Transducer Model: D3P64L Operating Mode: B+C Index label Below Below surface surface surface surface Maximum index value Index component value at Z 1.35 r,α 65.22 65.22 65.22 45.26 45.26 – Acoustic Parameters – 3.37 – pii,α...
  • Page 272 Digital Color Doppler Ultrasound System Transducer Model: D3P64L Operating Mode: PW Index label Below Below surface surface surface surface Maximum index value Index component value at Z 1.25 r,α 90.02 90.02 90.02 63.32 63.32 3.05 Acoustic Parameters 5.35 5.92 5.92 pii,α...
  • Page 273 Digital Color Doppler Ultrasound System Transducer Model: D3P64L Operating Mode: B+M Index label Below Below surface surface surface surface Maximum index value Index component value at Z 1.64 r,α 125.42 129.20 129.20 77.78 81.48 2.85 Acoustic Parameters 2.92 2.55 2.55 pii,α...
  • Page 274 Digital Color Doppler Ultrasound System Transducer Model: D3P64L Operating Mode: CW Index label Below Below surface surface surface surface Maximum index value Index component value at Z 0.20 r,α 66.58 70.00 70.00 48.87 50.21 5.40 Acoustic Parameters 6.75 6.80 6.80 pii,α...
  • Page 275 Digital Color Doppler Ultrasound System Transducer Model: D3P64L Operating Mode: CFM-M Index label Below Below surface surface surface surface Maximum index value Index component value at Z 1.57 r,α 44.28 48.55 48.55 26.47 28.51 6.15 Acoustic Parameters 6.14 6.15 6.15 pii,α...
  • Page 276 Digital Color Doppler Ultrasound System Transducer Model: D3P64L Operating Mode: TDI Index label Below Below surface surface surface surface Maximum index value Index component value at Z 1.35 r,α 71.23 71.23 71.23 55.51 55.51 – Acoustic Parameters – 3.35 – pii,α...
  • Page 277 Digital Color Doppler Ultrasound System...
  • Page 278 Digital Color Doppler Ultrasound System Transducer Model: D5P64L Operating Mode: B Index label Below Below surface surface surface surface Maximum index value Index component value at Z 1.75 r,α 75.12 75.12 75.12 74.43 74.43 – Acoustic Parameters – 2.20 – pii,α...
  • Page 279 Digital Color Doppler Ultrasound System Transducer Model: D5P64L Operating Mode: B+C Index label Below Below surface surface surface surface Maximum index value Index component value at Z 1.31 r,α 47.13 47.13 47.13 46.21 46.21 – Acoustic Parameters – 2.30 – pii,α...
  • Page 280 Digital Color Doppler Ultrasound System Transducer Model: D5P64L Operating Mode: PW Index label Below Below surface surface surface surface Maximum index value Index component value at Z 1.00 r,α 25.71 29.45 29.45 25.55 28.88 1.68 Acoustic Parameters 1.79 1.95 1.95 pii,α...
  • Page 281 Digital Color Doppler Ultrasound System Transducer Model: D5P64L Operating Mode: B+M Index label Below Below surface surface surface surface Maximum index value Index component value at Z 1.87 r,α 50.24 52.11 52.11 49.48 50.55 1.78 Acoustic Parameters 1.98 2.10 2.10 pii,α...
  • Page 282 Digital Color Doppler Ultrasound System Transducer Model: D5P64L Operating Mode: CW Index label Below Below surface surface surface surface Maximum index value Index component value at Z 0.47 r,α 54.00 55.58 55.58 53.31 54.21 3.88 Acoustic Parameters 4.08 4.00 4.00 pii,α...
  • Page 283 Digital Color Doppler Ultrasound System Transducer Model: D5P64L Operating Mode: CFM-M Index label Below Below surface surface surface surface Maximum index value Index component value at Z 1.22 r,α 11.90 16.00 16.00 11.78 15.78 2.88 Acoustic Parameters 3.00 3.18 3.18 pii,α...
  • Page 284 Digital Color Doppler Ultrasound System Transducer Model: D5P64L Operating Mode: TDI Index label Below Below surface surface surface surface Maximum index value Index component value at Z 1.61 r,α 42.22 42.22 42.22 41.28 41.28 – Acoustic Parameters – 2.32 – pii,α...
  • Page 285 Digital Color Doppler Ultrasound System Transducer Model:M8L40L Operating Mode: B Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.86 r,α – 35.14 35.14 17.24 17.24 – Acoustic Parameters – 2.05 – pii,α...
  • Page 286 Digital Color Doppler Ultrasound System Transducer Model: M8L40L Operating Mode: B+C Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.25 r,α – 52.49 52.49 26.42 26.42 – Acoustic Parameters – 1.65 –...
  • Page 287 Digital Color Doppler Ultrasound System Transducer Model: M8L40L Operating Mode: PW Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.44 r,α – 38.89 38.89 20.25 20.25 1.66 Acoustic Parameters 1.74 1.65 1.65 pii,α...
  • Page 288 Digital Color Doppler Ultrasound System Transducer Model: M8L40L Operating Mode: B+M Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.73 r,α – 47.00 47.00 24.14 24.14 1.76 Acoustic Parameters 1.80 1.90 1.90 pii,α...
  • Page 289 Digital Color Doppler Ultrasound System Transducer Model: M8L40L Operating Mode: Elasticity imaging Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 2.21 r,α – 28.54 28.54 13.79 13.79 – Acoustic Parameters – 2.05 –...
  • Page 290 Digital Color Doppler Ultrasound System Operating Mode: B Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 2.08 r,α – 32.56 32.56 17.10 17.10 – Acoustic Parameters – 1.95 – pii,α – 6.14 6.14 6.14...
  • Page 291 Digital Color Doppler Ultrasound System Operating Mode: B+C Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.25 r,α – 50.21 50.21 26.46 26.46 – Acoustic Parameters – 1.65 – pii,α – 6.36 6.35 6.35...
  • Page 292 Digital Color Doppler Ultrasound System Operating Mode: PW Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.39 r,α – 33.23 38.44 18.99 21.26 1.45 Acoustic Parameters 1.65 1.65 1.65 pii,α – 6.32 6.32 6.30...
  • Page 293 Digital Color Doppler Ultrasound System Operating Mode: B+M Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.71 r,α – 46.36 46.69 30.58 30.77 1.77 Acoustic Parameters 1.86 2.45 2.45 pii,α – 6.05 6.18 6.05...
  • Page 294 Digital Color Doppler Ultrasound System Operating Mode: Elasticity imaging Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.49 r,α – 45.21 45.21 20.52 20.52 – Acoustic Parameters – 1.95 – pii,α –...
  • Page 295 Digital Color Doppler Ultrasound System Transducer Model: V4C40L Operating Mode: B Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.40 r,α – 123.40 123.40 63.05 63.05 – Acoustic Parameters – 3.20 –...
  • Page 296 Digital Color Doppler Ultrasound System...
  • Page 297 Digital Color Doppler Ultrasound System Transducer Model: V4C40L Operating Mode: B+C Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 0.94 r,α – 46.52 46.52 27.93 27.93 – Acoustic Parameters – 5.27 –...
  • Page 298 Digital Color Doppler Ultrasound System Transducer Model: V4C40L Operating Mode: PW Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 0.78 r,α – 85.28 86.92 47.73 48.55 4.02 Acoustic Parameters 3.98 3.20 3.20 pii,α...
  • Page 299 Digital Color Doppler Ultrasound System Transducer Model: V4C40L Operating Mode: B+M Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.29 r,α – 124.55 126.91 69.12 69.46 3.78 Acoustic Parameters 5.03 4.10 4.10 pii,α...
  • Page 300 Digital Color Doppler Ultrasound System Transducer Model: V4C40L Operating Mode: Elasticity imaging Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.65 r,α – 89.25 89.25 58.20 58.20 – Acoustic Parameters – 3.20 –...
  • Page 301 Digital Color Doppler Ultrasound System Transducer Model: D6C15L Operating Mode: B Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.93 r,α – 58.79 58.79 42.86 42.86 – Acoustic Parameters – 2.70 –...
  • Page 302 Digital Color Doppler Ultrasound System Transducer Model: D6C15L Operating Mode: B+C Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.72 r,α – 55.60 55.60 43.10 43.10 – Acoustic Parameters – 1.80 –...
  • Page 303 Digital Color Doppler Ultrasound System Transducer Model: D6C15L Operating Mode: PW Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.58 r,α – 66.36 66.36 55.13 55.13 1.80 Acoustic Parameters 1.80 2.85 2.85 pii,α...
  • Page 304 Digital Color Doppler Ultrasound System Transducer Model: D6C15L Operating Mode: B+M Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.79 r,α – 60.22 63.54 50.18 52.28 2.00 Acoustic Parameters 2.00 3.55 3.55 pii,α...
  • Page 305 Digital Color Doppler Ultrasound System Transducer Model: D6C15L Operating Mode: Elasticity imaging Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.67 r,α – 44.68 44.68 35.71 35.71 – Acoustic Parameters – 2.70 –...
  • Page 306 Digital Color Doppler Ultrasound System Transducer Model: D3C20L Operating Mode: B Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.26 r,α – 105.23 105.23 79.82 79.82 – Acoustic Parameters – 4.95 –...
  • Page 307 Digital Color Doppler Ultrasound System Transducer Model: D3C20L Operating Mode: B+C Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 0.47 r,α – 56.52 56.52 37.10 37.10 – Acoustic Parameters – 3.90 –...
  • Page 308 Digital Color Doppler Ultrasound System Transducer Model: D3C20L Operating Mode: PW Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 0.59 r,α – 75.44 78.68 40.13 42.25 3.88 Acoustic Parameters 3.48 5.60 5.60 pii,α...
  • Page 309 Digital Color Doppler Ultrasound System Transducer Model: D3C20L Operating Mode: B+M Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.22 r,α – 98.55 98.55 51.69 51.69 3.48 Acoustic Parameters 4.25 4.65 4.65 pii,α...
  • Page 310 Digital Color Doppler Ultrasound System Transducer Model: D3C20L Operating Mode: Elasticity imaging Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.08 r,α – 108.94 108.94 67.54 67.54 – Acoustic Parameters – 4.95 –...
  • Page 311 Digital Color Doppler Ultrasound System Transducer Model: D5C20L Operating Mode: B Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.88 r,α – 58.40 58.40 38.97 38.97 – Acoustic Parameters – 2.45 –...
  • Page 312 Digital Color Doppler Ultrasound System Transducer Model: D5C20L Operating Mode: B+C Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.36 r,α – 72.30 72.30 57.46 57.46 – Acoustic Parameters – 2.85 –...
  • Page 313 Digital Color Doppler Ultrasound System Transducer Model: D5C20L Operating Mode: PW Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.55 r,α – 78.88 78.88 53.44 53.44 – Acoustic Parameters – 2.35 2.35 pii,α...
  • Page 314 Digital Color Doppler Ultrasound System Transducer Model: D5C20L Operating Mode: B+M Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.95 r,α – 65.45 69.24 48.53 51.43 – Acoustic Parameter – 2.05 2.05 pii,α...
  • Page 315 Digital Color Doppler Ultrasound System Transducer Model: D5C20L Operating Mode: Elasticity imaging Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.75 r,α – 46.58 46.58 30.31 30.31 – Acoustic Parameters – 2.45 –...
  • Page 316 Digital Color Doppler Ultrasound System Transducer Model: D8L50L Operation Mode: B Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 2.43 r,α – 32.16 32.16 17.31 17.31 – Acoustic Parameters – 1.88 –...
  • Page 317 Digital Color Doppler Ultrasound System Transducer Model: D8L50L Operation Mode: B+C Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 1.54 r,α – 45.32 45.32 22.17 22.17 – Acoustic Parameters – 1.68 –...
  • Page 318 Digital Color Doppler Ultrasound System Transducer Model: D8L50L Operation Mode: PW Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 2.06 r,α – 51.28 54.59 35.11 37.28 1.75 Acoustic Parameters 1.83 1.73 1.73 pii,α...
  • Page 319 Digital Color Doppler Ultrasound System Transducer Model: D8L50L Operation Mode: B+M Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 2.02 r,α – 61.22 65.12 29.58 31.26 1.84 Acoustic Parameters 1.88 1.68 1.68 pii,α...
  • Page 320 Digital Color Doppler Ultrasound System Transducer Model: D8L50L Operation Mode: Elasticity imaging Index label Below Below surface surface surface surface – Maximum index value Index component value at Z 2.16 r,α – 38.44 38.44 20.19 20.19 – Acoustic Parameters – 1.88 –...
  • Page 321: Display Accuracy And Acoustic Measurement Uncertainties

    Digital Color Doppler Ultrasound System Display Accuracy and Acoustic Measurement Uncertainties According to IEC60601-2-37 and NEMA UD-3 2004, the display accuracy and acoustic measurement uncertainties are summarized in the table below. Display accuracy of MI is ± 20%, and TI is ± 40% or <0.2, if MI, TI below 0.5. Measurement Uncertainty Item (Percentage, 95% Confidence Value...
  • Page 322: Appendix D Transducer Maximum Surface Temperature

    The calculation of the expanded uncertainty is based on the ISO Guide to the Expression of uncertainty in measurement. Three transducer samples have been tested and the confidence coefficient is at 95%, the value of t.975 is 4.30. The measurement data were obtained under the test conditions employed at CHISON. Maximum Maximum...
  • Page 323: Appendix E Guidance And Manufacturer's Declaration

    1. Guidance and manufacturer’s declaration – electromagnetic emissions The CBit 9 is intended for use in the electromagnetic environment specified below. The customer or the user of the CBit 9 should assure that it is used in such an environment. Electromagnetic environment...
  • Page 324 (60 % dip in UT) (60 % dip in UT) operation during power mains interruptions, it is supply for 5 cycles for 5 cycles recommended that the input lines 70 % UT 70 % UT CBit 9 be powered from an uninterruptible power...
  • Page 325 Digital Color Doppler Ultrasound System (30 % dip in UT) (30 % dip in UT) supply or a battery. 61000-4-11 for 25 cycles for 25 cycles <5 % UT <5 % UT (>95 % dip in UT) (>95 % dip in UT) for 5 sec for 5 sec Power...
  • Page 326 RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the CBit 9 is used exceeds the applicable RF compliance level above, the CBit 9 should be observed to verify normal operation.
  • Page 327 Digital Color Doppler Ultrasound System disturbances are controlled. The customer or the user of the CBit 9 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the CBit 9 as recommended below, according to the maximum output power of the communications equipment.
  • Page 328: Appendix F Measurement Results Summary

    Digital Color Doppler Ultrasound System APPENDIX F Measurement Results Summary Measurement Useful Range Accuracy Distance Full Screen <± 5% Circumference: Full Screen <± 5% trace method, ellipse method Area: Full Screen <± 10% trace method, ellipse method Volume Full screen <±...
  • Page 329 CHISON CHISON Medical Technologies Co., Ltd. No.3 Changjiang South Road, Xinwu District, Wuxi, 214028 Jiangsu, P.R. China...