Kitagawa 121SS Instruction Manual

Phosphine detector tube

Advertisement

Quick Links

★ READ CAREFULLY THIS INSTRUCTION MANUAL AND THE INSTRUCTIONS OF THE ASPIRATING
PUMP PRIOR TO USING THIS PRODUCT.
★ DO NOT DISCARD THIS INSTRUCTION MANUAL UNTIL ALL THE TUBES IN THIS BOX ARE USED
UP.
1. PERFORMANCE:
Measuring Range
and Sampling Time
Number of Pump Strokes
(*)Graduations on the detector tube are based on 1 pump stroke.
Colour Change
Detectable Limit
Operating Temperature
Aspirating Pump
1. THE DETECTOR TUBE CONTAINS CHEMICAL REAGENTS.
2. DO NOT TOUCH THESE REAGENTS DIRECTLY ONCE TUBES WERE BROKEN.
3. KEEP THE TUBES OUT OF THE REACH OF CHILDREN.
4. IF THE CONCENTRATION IS OVER THE FULL SCALE, THE HIGH
CONCENTRATION OF PHOSPHINE REMAINS IN THE ASPIRATING PUMP.
BE CAREFUL NOT TO BREATHE THE REMAINING GAS. AFTER MEASUREMENT,
THE REMAINING GAS IS PUSHED OUT FROM THE BOTTOM CASE OF THE
PUMP WHEN THE HANDLE IS PUSHED BACK AND PULLED. IN CASE OF THE
ABOVE, PUSH BACK AND PULL THE HANDLE WITHOUT CONNECTING THE TUBE
AT THE LOCAL EXHAUST VENTILATION DEVICE. REPEAT THIS OPERATION AT
LEAST FIVE TIMES IN ORDER TO REMOVE THE REMAINING GAS.
NOTICE
1. USE ONLY WITH PUMP MODELS AP-20, AP-20S, 400B, AP-1, AP-1S OR 400A. OTHERWISE,
CONSIDERABLE ERROR IN INDICATION MAY OCCUR.
2. BEFORE TESTING, CHECK THE ASPIRATING PUMP FOR LEAKS (REFER TO ITEM 8.
INSPECTION OF ASPIRATING PUMP). ANY PUMPS SHOWING SIGNS OF LEAKAGE SHOULD
BE CORRECTED BEFORE USE.
3. DO NOT USE THIS TUBE OUTSIDE THE STATED OPERATING TEMPERATURE RANGE.
4. STORE TUBES IN A COOL AND DARK PLACE (0-25 ℃/32-77°F), AND USE BEFORE
EXPIRATION DATE PRINTED ON TOP OF THE BOX.
5. PRIOR TO USE, READ CAREFULLY ITEM 9 USER RESPONSIBILITY.
6. READ THE CONCENTRATION IMMEDIATELY AFTER MEASUREMENT.
2. SAMPLING AND MEASUREMENT:
① Break both ends of the detector tube.
CAUTION SAFETY GLASSES AND GLOVES SHOULD BE WORN TO PREVENT INJURY
FROM SPLINTERING GLASS.
② Insert the detector tube into the aspirating pump securely as shown in Fig.1. (Arrow mark shall point to
the pump.)
③ Align the guide marks on the shaft and stopper of the aspirating pump.
④ Pull the pump handle at a full stroke until it locks and wait for 1 minute or until the completion of sampling is
INSTRUCTION MANUAL
PHOSPHINE DETECTOR TUBE
: 200 - 3000 ppm
400 - 6000 ppm
: 1 minute
30 seconds
: 1 (100 mL)
: White → Orange
: 5 ppm(1 pump stroke)
: 0 - 40 ℃ (32-104°F) (No correction is necessary.)
: Model AP-20, AP-20S, 400B, AP-1, AP-1S or 400A
1/2 (50 mL)
CAUTION
Fig.1
No.121SS

Advertisement

loading
Need help?

Need help?

Do you have a question about the 121SS and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Kitagawa 121SS

  • Page 1 INSTRUCTION MANUAL PHOSPHINE DETECTOR TUBE No.121SS ★ READ CAREFULLY THIS INSTRUCTION MANUAL AND THE INSTRUCTIONS OF THE ASPIRATING PUMP PRIOR TO USING THIS PRODUCT. ★ DO NOT DISCARD THIS INSTRUCTION MANUAL UNTIL ALL THE TUBES IN THIS BOX ARE USED 1.
  • Page 2 confirmed with the flow indicator of the pump (See descriptions about the flow indicator in the instruction manual of the pump). ⑤ On completion of sampling, read the scale at the maximum point of the stained layer. ⑥ When the concentration is over the scale range, 1/2 pump strokes can be used to determine concentrations of 400 to 6000 ppm.