Press any key for 3 seconds to enter the menus. (See below for details) 83850- 001-000 RF June 12, 2007 INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com...
1.6 Signal and Display Range It is important to distinguish between Signal Range (often called Span) and Display Range. 83850-001-000 RF June 12, 2007 INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com...
However, it cannot detect simple out of calibration conditions, which can only be detected by applying a gas standard. 83850-001-000 RF June 12, 2007 INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com...
Where GAS will show the target gas, and xxx is concentration of the gas. Units may be either ppm (parts per million) or %. 83850-001-000 RF June 12, 2007 INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com...
Press any key for 3 seconds to enter menu system Password Control 83850-001-000 RF June 12, 2007 INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com...
Page 9
Main Menu (and Relay 1 Settings sub-menu branch). Actuation Setpoint ACTUATE 00XX ppm (or %) 83850-001-000 RF June 12, 2007 INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com...
Page 10
Actuation setpoint is reached. A maximum of 60 minutes is possible, adjustable in 5-minute increments. Default is 00. Buzzer. BUZZER 83850-001-000 RF June 12, 2007 INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com...
Page 11
Note: An inverted response at the signal output can be achieved by setting CONC4MA higher than CONC20MA. 83850-001-000 RF June 12, 2007 INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com...
Page 12
CAL SPAN XXX ppm (or %) 83850-001-000 RF June 12, 2007 INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com...
Page 13
BAUDRATE 4800 Default is 4800 baud. Options: 600, 1200, 2400, 4800 and 9600 Change Password 83850-001-000 RF June 12, 2007 INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com...
(EIA-485), (e.g. Belden 9841). This is a twisted, shielded, balanced pair, 24 AWG, 120 Ohm. In order to prevent signal bounce-back and 83850-001-000 RF June 12, 2007 INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com...
If the gas flow was not sufficient, or the bayonet adapter was too loosely applied, the gas concentration in the sensor will stabilize at an incorrect low value. 83850-001-000 RF June 12, 2007 INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com...
LCD, return to factory for repair. Note: The unit performs a standard short beep at power-up. 83850-001-000 RF June 12, 2007 INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com...