Model :KY21-T10L Oxygen Concentrator Instruction manual Please read the instruction manual carefully before using the product The picture is for reference only, please refer to the actual product!...
CONTENTS Precautions Product features Main technical parameters Installation process Display panel and function operation Other features Maintenance Technical fault and overhaul Other issues...
1.Precautions Thank you for purchasing this product, please read the instruction manual carefully before using. The warning signs shown on this product are intended to let users understand the meaning of the signs and use this product correctly to prevent harm and property damage。The indication and meaning of warning signs are as follows:...
Page 4
▲Oxygen is one kind of combustion-supporting gases. WARNING When using it, do not smoke and keep away from fire, oil ,grease or any other flammable materials to avoid fire。 ▲It is not suitable to use and store the oxygen generator near heat sources, flammable and explosive materials, damp, corrosive gas, high and low temperature environ- ments...
▲Do not open the machine as unauthorized personnel. ▲Please seek for medical advice before using this machine. ▲The medical molecular sieve oxygen concentrator should not be used close to or attached with other equipment. If it must be used close to or attached to, it should be observed to verify that it can operate normally under that condition.
Page 6
Scope of application For families to produce oxygen-enriched air (93% oxygen) for oxygen-de- ficient people (oxygen therapy or alleviation of various discomfort caused by hypoxia). Prohibition: Patients with oxygen poisoning or oxygen allergy Product combination It is composed of an oxygen concentrator host, humidifier bottle, oxygen concentration status indicator, atomization component (optional accesso- ry), and remote control.
Page 7
Packing list ◎Main body(Including humidification bottle) ◎Atomization components ◎Remote control ◎Instruction manual ◎Power plug ※The atomization components (optional accessories) used with this product are all medical devices that have been approved for registration and licensing. Please use the accessories recommended or specified by our company.
3.Main Specifications Product name:Oxygen Concentrator Model number:KY21-T10L Oxygen production method:Pressure swing adsorption (PSA) Rated voltage:AC 230V/50Hz Rated power:580VA Oxygen Flow Rate:1-10L/min adjustable Oxygen Concentration:93% ± 3% Oxygen Output pressure:20kPa-70kPa Atomization rate:≥ 0.2mL/min The median diameter of fog particles:The median diameter of the atomized fog particles is 3.0 μ...
4.Installation process Schematic diagram of appearance Oxygen Outlet Humidifier Bottle Atomization Outlet LED Display Timing On/Off Flow- Flow+ Intake Window Switch Power Connector Exhaust Window...
Page 10
Using process ①Remove the humidification bottle Take out the humidification bottle as shown on the left ②Water injection Open the humidified water bottle at the top and inject pure water (cold boiled water or distilled water) to below the highest water level. Then put the humidification bottle back in place.
Page 11
⑤ Connect the oxygen tube Insert the large diameter end of the oxygen suction pipe into the "oxygen outlet" of the machine. For example,if purchased oxygen tube or mask-type oxygen inhalation component come with water trap, please open the water trap and pour out the water(Additional purchase required).
5.Display panel and function operation ① ② ③ Panel display 1. Timing display Timing 2. Oxygen concentration display mins Total 3. Oxygen flow display hours 4. Cumulative time display ④ ⑤ ⑥ ⑦ 5. Low oxygen concentration alarm display 6. Sleep status display 7.
Page 13
Flow - Botton Flow "-" key: Touch the flow "-" key to decrease the flow by one liter, and the minimum can be adjusted to 1L. Timing Button Touch the timing button, the timing will be decremented by 60 minutes, set in a one-way cycle, the highest setting value is 480 minutes;...
Sleep mode key In the working state, press the "Sleep" key, and after the "Sleep" icon on the display screen flashes three times, the display screen switches to sleep mode. When you press any key or touch any function key on the panel, the screen will be highlighted again.
Page 15
The measured value of the pressure sensor exceeds the set Pressure sensor failure threshold, the yellow indicator light is on, the alarm sounds intermittently, and the device is on standby. Compressor failure The working current of the compressor is abnormal, the yellow indicator light is on, the alarm sounds quickly and intermittently, and the equipment is on standby The working current of the cooling fan is abnormal, the yellow...
Page 16
Instruction 1、Before turning on the machine, unscrew the atomizing screw cap 2、Connect the atomization cup and the atomization port with the tube respectively. Pour the appropriate amount of medicine liquid into the atomization cup as needed or under the guidance of a professional doctor, and wear the atomizer as required by the picture.
7.Maintenance Clean the shell The surface of the cabinet is recommended to be wiped and cleaned at least once a month. Before cleaning, please cut off the power supply and wipe with a soft, slightly damp cloth soaked with disinfectant (medical alcohol) to prevent liquid from penetrating into the machine.
Page 18
Clean the oxygen tube and atomization components Should be washed once after each use. The nasal suction tip on the nasal oxygen inhaler should be cleaned and disinfected after each use. It can be soaked in 5% potassium permanganate solution for 5 minutes and then rinsed with clean water or wiped with alcohol.
Long-term idle processing Before leaving the oxygen concentrator for a long time, please do the following: 1、Remove the humidification bottle, pour out the water in the bottle to dry, and put it back in place. 2、Cut off the power and put away the power cord 3、Do not store the oxygen concentrator in a humid and dusty environ- ment, avoid direct sunlight and pollution.
Page 20
The oxygen tube is kinked or After turning on the blocked. When using the machine, the atomization function, there may Straighten out the oxygen machine runs be intermittent sound alarms, tube. normally and the which can be ignored when stopping the dagger. flow rate is small.
9.Other issues ▲The production date is detailed in the product inspection certificate. The normal use period is three years from the date of use of this product. ▲The service life of the whole machine must meet the normal working environment and normal maintenance conditions (excluding wearing parts, atomization accessories, etc.).
Page 22
Certificate of Conformance Product Name: Oxygen Concentrator KY21-T10L Model Number: Test Date: Inspector: The free service hotline: 400-830-2119 Shenzhen Deerjie International Development Co ., Ltd.
Page 23
Address: 301 , Building 39 , Dayun Software Town , No . 8288 Longgang Avenue ,Heao Community , Yuanshan Street , Longgang District , Shenzhen Tel:0755-89760988 Fax:0755-89760688 Email:service@dedakj-health.com Website: www.dedamed.com Service hotline: 400-830-2119 Manufacturer Address:Deda Industrial Park, Jianshe Avenue, Huaihua High-tech Zone, Hunan Province...
Need help?
Do you have a question about the KY21-T10L and is the answer not in the manual?
Questions and answers