  • Page 1 Document No. 1788-57 BioSafe® Hybrid Cleanroom Installation © Copyright 2024 Terra Universal Inc. All rights reserved. • Revised Jul. 2024 Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
  • Page 2: Table Of Contents

    Wall Panel Installation ............................ 17 Concealed Suspended Ceiling Installation ..............................18 Coving Installation ............................20 Specifications ..............................21 Warranty ................................. 21 Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
  • Page 3: Introduction

    The information presented here is subject to change without notice. 1.0 Introduction This manual provides information on installing your Terra Universal BioSafe® Hybrid Cleanroom. Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
  • Page 4: Base & Window Installation

    Corners should be mitered and fixed  Channel should be laid up to location of door structural opening Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
  • Page 5  Place the top section of the recess base onto the bottom channel section Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
  • Page 6 Once located, the panels can be sealed to the base section if required by the specification. Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
  • Page 7 Slot the first wall panel into place and ensure it is plumb and level with the pre-formed corner once slotted into place Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
  • Page 8 Panel 3 Panel 1 Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
  • Page 9: Flush Window Installation

     Ensure adjacent panel is plumb and level Add the panel with the window already built in Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
  • Page 10 Fit one of the aluminum tongue sections into the top of the base channel, sealing as required (as per wall panels)  Fit the window panel onto the vertical and horizontal tongues Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
  • Page 11 Fit window header panel (which may be 3 or 4-sided depending on specification) on top of window panel and ensure all panels are plumb and level Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
  • Page 12 © Copyright 2017 Terra Universal Inc. All rights reserved. • Revised August 2017 • Document No. 1800-93  Once window panels are plumb and level, standard wall panel construction can be resumed Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
  • Page 13: Clamp Frame Door Installation

    For the door, form the “U” shape frame using the back profile, mect form the “U” using the front profile Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
  • Page 14 Ensure the panel adjacent to the door is plumb and level  Fit cleanroom door adjacent to panel as per manufacturers guidelines Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
  • Page 15 Cut 2 lengths of aluminum tongue to fit above the door frame to which the header panel will be fixed Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
  • Page 16  Fix door clamp frame to side and header panels using suitable fixings Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
  • Page 17: Wall Panel Installation

    After fixing the U frame into the wall panels, add the panel union to the wall panels over the U frame, and slide the door header panel to complete Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
  • Page 18: Concealed Suspended Ceiling Installation

    Fixings must be fitted with a maximum span dependent on ceiling loading as per specification and at a maximum of 48” intervals Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
  • Page 19 Cover strips should be attached with ¾” x 1/8” self-drilling, self-tapping screws at 1200 mm intervals 8" Hanging Plate Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
  • Page 20: Coving Installation

     These must be fitted prior to the coving facing being applied Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
  • Page 21: Specifications

    ISO 5 Materials Steel substrate walls uPVC wall coating 7.0 Warranty https://www.terrauniversal.com/warranty Thank you for ordering from Terra Universal! Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...