Authorized Persons ....................4 Disclaimer ......................5 Room Disinfecting with the Sanosil Q-Jet Fogger ..........5 Different ways to use the Sanosil Q-jet Compact ..........5 Important safety Directions (Read carefully before use) ........6 Product Description ..................7 Technical Data ......................7 Parts and Functions ....................
Sanosil Q-Jet Compact Introduction 1.1 Used Symbols Inside this manual or on the device there are Symbols having the following meaning: High risk! Disregard can lead to death or heavy injuries! Medium risk! Disregard can lead to death or heavy injuries!
Do always and only use suitable Sanosil disinfectants with a maximal hydrogen Peroxide content of 7.5%. The usage of any other disinfectant or accessories which are not specifically approved by Sanosil Ltd. is against the rules herein and can lead to heavy injuries as well as to damaging the device. The device is not meant for use as a stand-alone disinfection process but rather as an addition to regular cleaning and disinfection.
How to use the Sanosil Q-Jet Compact: Fogging of Sanosil Disinfectants (only use Sanosil ready for use products containing up to 7.5% hydrogen Peroxide) can be used for disinfecting hard, nonporous surfaces in rooms as an adjunct following regular cleaning and disinfecting of room surfaces.
To ensure effectiveness and safety, use only with Sanosil disinfectant. Using any other manufacturer’s disinfecting or cleaning product will result in serious injury, exposure and environmental damage. Sanosil Ltd. is not responsible for any injury or damage caused by using this device outside of the specific parameters detailed in this manual.
Sanosil Q-Jet Compact Product Description 3.1 Technical Data Power requirements: 220 - 240 V, 50Hz (fuse 6.3 A) Amperage: max. 5 A Operating temperature / 5°C to 35°C / max. 45% rH (before fogging) at higher humidity, the humidity must be lowered humidity: before start-up or a lower dosage of the active substance must be selected.
Sanosil Q-Jet Compact 3.7 Trolley (optional) • place the device on top of trolley • secure the device with the screw on the trolley CAUTION Risk of crushing! Hands and fingers can be crushed between device and trolley! Do always wear device on handle bars while setting the device on the trolley!
Sanosil Q-Jet Compact Operating Instructions 4.1 Warning WARNING Risk of Injuries! This device nebulizes chemical substances! Do not inhale these substances, always wear safety equipment when filling/emptying tank and pay attention to optical and acoustic warning signals. Read carefully all indications stated on the safety data sheet of the disinfection liquid and fully comply...
Please see table below for your convenience. Higher concentrations of 16 or 24ml/m may be required for some conditions. Please ask your Sanosil sales representatives for detailed info. • Please read and follow important safety directions on page 6 before use! 4.2.4...
Sanosil Q-Jet Compact 4.3 Set up Device 4.3.1 Calibration Initial use and frequently we recommend calibrating the device to make sure the tank filling level to be correct on the display. Find the calibration procedure in chapter 5, page 15 of this document.
Sanosil Q-Jet Compact 4.3.4 Switch on the device • Tip ON/OFF (3) to switch on the device 4.3.5 Setting up the device Please read and follow important safety directions on page 6 Display Important safety Directions (Read carefully before use) before use! •...
Sanosil Q-Jet Compact Procedure before re-entering the room 4.4.1 Right procedure before re-entering the room • Check with a hydrogen peroxide level sensor that the H -Level is below your local maximum permissible value or if not specified at least below 1 ppm.
Sanosil Q-Jet Compact 4.5 Emptying the Tank CAUTION! Risk of injuries! Do always wear gloves and safety googles when emptying the tank! • Before emptying the tank always turn OFF the device and disconnect the power supply plug! • Unscrew rear panel and remove it •...
Sanosil Q-Jet Compact Calibration, maintenance and troubleshooting WARNING Risk of injury! Do always switch off the device and disconnect the power supply cord/plug and empty the tank of the device fully before starting any cleaning or maintenance works! Maintenance works can only be carried out through well-trained personnel or professional electricians.
Sanosil Q-Jet Compact 5.2 Maintenance 5.2.1 Warning WARNING Risk of injury! Do always switch off the device and disconnect the power supply cord before starting any cleaning or maintenance works. Maintenance works can only be carried out through well-trained personnel.
Sanosil Q-Jet Compact Troubleshooting When asking for support with your agent: make sure in any case to fill in accurately manufactures Complaint Form before any contacting! Trouble Possible cause Solution • Check power supply • Ensure power supply to device Device does not work / does not •...
(e.g, air filter, etc.). Sanosil Ltd / CleanService Group Ltd. (manufacturer) will have the choice to repair the device, to replace it or to reimburse the actual paid amount. A warranty service does not extend the length of the warranty. Sanosil Ltd / CleanService Group Ltd.
Need help?
Do you have a question about the Q-Jet Compact 2 Series and is the answer not in the manual?
Questions and answers