NGI N35100 Series User Manual

Wide range bidirectional programmable dc power supply

Advertisement

Quick Links

ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING
User Manual
N35100 Series Wide Range Bidirectional
Programmable DC Power Supply
Hunan Next Generation Instrumental T&C Tech. Co., Ltd.
Version: V20240529

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the N35100 Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for NGI N35100 Series

  • Page 1 ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING User Manual N35100 Series Wide Range Bidirectional Programmable DC Power Supply Hunan Next Generation Instrumental T&C Tech. Co., Ltd. Version: V20240529...
  • Page 2: Table Of Contents

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING Contents 1 PREFACE .......................1 2 SAFETY INSTRUCTIONS ..................2 2.1 Safety Notes ..................... 2 2.2 Safety Symbols ....................2 3 INSPECTION & INSTALLATION ................3 3.1 Inspection ......................3 3.2 Connection to Power Cord ................3 3.3 Power-on Test ....................3 3.4 Connection of the DUT ..................
  • Page 3 ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING 5.5 SEQ Test ......................25 5.6 Car Waveform ....................27 5.6.1 ISO-16750-2 ..................27 5.6.2 LV124 ....................33 5.7 Charge ......................35 5.8 Dischage ......................36 5.9 RESI ........................ 37 5.10 CR ......................... 38 5.11 Protection ....................39 5.12 Application ....................
  • Page 4: Preface

    This manual is applied to N35100 series wide range bidirectional programming DC power supply, including installation, operation, specifications and other detailed information. The copyright of the manual is owned by NGI. Due to the upgrade of instrument, this manual may be revised without notice in future versions.
  • Page 5: Safety Instructions

    NGI shall not be liable for the results caused by the neglect of those instructions. 2.1 Safety Notes  Confirm the AC input voltage before supplying power.
  • Page 6: Inspection & Installation

    1. Check whether the instrument is damaged during transportation. If any severe damage to the package, please contact our authorized distributor or NGI. 2. Check accessories. 3. Check the whole instrument. If N35100 chassis is damaged or has abnormal operation, please contact our authorized distributor or NGI.
  • Page 7 ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING 1. Power-on Press POWER to switch on the power supply, if the power supply can not start normally, users should first check whether the power cord is connected, whether the power supply has been powered, whether the power switch has been turned on. Warnings When the power switch is off, some components inside the power supply may still carry a high voltage, to avoid the risk of electric shock, it is...
  • Page 8 ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING 6. Check whether the current displayed on the screen is close to the set current value. 7. Ensure that the power supply current can be adjusted from 0A to the maximum current value within the range.
  • Page 9: Connection Of The Dut

    3.4 Connection of the DUT 3.4.1 DUT connection cable N35100 series power supply does not come with a connection cable for the device under test, users need to choose their own wiring. The following points should be noted when selecting wiring: ·The maximum allowable current of the wires.
  • Page 10: 2/4-Wire Connection

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING 3.4.2 2/4-wire connection The rated output voltage of the N35100 series power supplies is greater than the safe voltage, and hazardous voltages may be present at the output and at the connection to the equipment under test. To protect the user from accidental exposure to hazardous voltages, ensure that the equipment under test and its connections are free of any accessible live parts.
  • Page 11 IP addresses of all N35100s are not duplicated. N35100 series power supply adopts UDP network communication mode, users can set the IP address, the default port number is 7000 and the default ID is 1.
  • Page 12: Product

    &absorb the power, and return power to the grid cleanly, so as to save the power consumption and reduce the space heat dissipation, which can greatly reduce the test cost. N35100 series provides high precision measurement and multiple testing functions, which can be widely used in new energy, automotive, energy storage, electric drive, battery simulation and other industries.
  • Page 13: Appearance & Dimension

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING 4.2 Appearance & Dimension 214mm(W)*500mm(D) (with shield)*44mm(H); Figure 4 Front Panel Dimension(mm) Figure 5 Side Dimension(mm) Figure 6 Rear Panel Dimension(mm)
  • Page 14: Front Panel Introduction

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING 4.3 Front Panel Introduction Figure 7 Front Panel Table 2 Number Name Function Power switch Power control Screen Displaying data Device model Displaying model number Buttons Operation mode Knob Move cursor to adjust value size Air outlet Exhaust outlet, cooling 4.3.1 Button&Knob...
  • Page 15 ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING Table 3 Button Definition [V-Set] To set voltage [I-Set] To set source current, load current [ESC] Return to previous menu [P-Set] To set source power, load power [Menu] Enter the main menu screen [Enter] To select a menu or confirm the operation [Shift] Compound function keys...
  • Page 16: Rear Panel Introduction

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING 4.4 Rear Panel Introduction Figure 9 Rear Panel Table 4 Name Icon Definition Output interface DC power supply output LAN port For remote control Sense interface Voltage sense RS232 interface Serial communication control device I/O Interface I/O Signal Interface RS485 interface...
  • Page 17: Ac Input Connection

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING 4.4.1 AC Input Connection Warning: Please confirm the AC input power and connect to correct AC power. Wrong AC power may cause serious damage to the instrument. Notes for AC power input connection:  Single phase input, 47Hz ~ 63Hz (Please refer to the voltage mark at the rear panel.) ...
  • Page 18: Lan Port

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING 13.3 21.1 26.7 33.6 42.4 53.5 67.4 0000 4.4.3 LAN Port The default connection to PC is via Ethernet. An Ethernet cable is provided as standard accessory in N35100 package. Steps for connection to PC: 1.
  • Page 19: Rs485/Can Interface

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING provided as standard accessory in N35100 package. Default baud rate for RS232 interface is 9600. Figure 10 RS232 Interface Table 6 Definition RXD, receive data TXD, transmit data GND, ground 4.4.5 RS485/CAN Interface Figure 11 RS485/CAN Interface...
  • Page 20: I/O Interface

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING Table 7 Definition Description RS485A RS485A RS485B RS485B CAN_FL CAN Power Supply GND CAN_FL CAN Power Supply GND CAN0_H CAN0_H CAN0_L CAN0_L CAN1_H CAN1_H(Parallel) CAN1_L CAN1_L(Parallel) GND_FL CAN Power Supply GND GND_FL CAN Power Supply GND CAN2_H CAN2_H(Debug)...
  • Page 21: Sense Terminal

    When N35100 is working, due to the parasitic resistance of the wire, the voltage at the load input terminal will be lower than the voltage at N35100 output terminal after the current flows through the wire, which will affect the output accuracy of N35100. N35100 series supports remote sense to compensate the voltage drop.
  • Page 22: Ground Terminal

    4.4.8 Ground Terminal The metal shielding of the N35100 series power supply is insulated from the internal live conductors, and the casing is normally uncharged. In order to prevent accidental occurrences from causing the shielding to become electrically charged, there is then a potential difference between the shielding and ground.
  • Page 23: Operation

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING 5 Operation N35100 series provides V/I mode, SEQ mode, SEQ Edit, Car Waveform, Charge mode, Discharge mode, SR, CR, Protection, Application, System, Factory Reset, Users can enter the required mode by pressing the corresponding button on the front panel.
  • Page 24: Interface

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING 5.1 Interface 3.2 inch LCD screen is applied for N35100, with resolution 820×320. Figure 16 Displayed Information Table 10 Definition Description ① Model Product specification model. Channel readback in current mode, including voltage, current and ②...
  • Page 25: Parameter Setting

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING 5.2 Parameter Setting 5.2.1 Numeric Input Numeric input is used to input digit, such as voltage, slew rate, protection threshold, IP address, etc. Figure 17 Numeric Edit Steps for numeric input: 1. Press [ ] or rotate the knob to move the cursor to the numeric input box.
  • Page 26: Source&Load State

    5.3 V/I 5.3.1 Source&Load State N35100 series products have two operating states: source and load. When used as a power supply, the energy provided by the device is in the source mode, and the current and power readback are positive. When used as a load, the device absorbs external energy and returns to the power grid in the load mode.
  • Page 27: Operation Steps

    Under V/I mode, users can set output voltage, output current and OVP value. N35100 will begin to output after setting is completed and is pressed. O n / O f f 5.4 SEQ Edit N35100 series supports SEQ Edit, with up to 10 sequence files and Max. 1000 steps per file.
  • Page 28: Seq Test

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING Figure 20 SEQ Edit There are 13 items can be set. When File No. and Step No. are changed, interface will be refreshed at the same time. Table 11 Parameter Function File No. To set the sequence file number, range 1-10 Total Steps To set the total test steps for the SEQ file, Max.
  • Page 29 ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING Figure 21 SEQ Operation Principle Methods to enter SEQ Test: S E Q Method 1: Press on the front panel directly. E n t e r → Select SEQ Test by Method 2: Press →...
  • Page 30: Car Waveform

    5.6 Car Waveform 5.6.1 ISO-16750-2 N35100 series have four types of analog waveform for vehicle electrical and electronic equipment immunity testing, which is convenient for customers to quickly call. The output pulse waveform is fully compliant with ISO-16750-2 international standard.
  • Page 31 ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING Select 24V car starting waveform Voltage/duration of the 24V waveform corresponding to LEVEL1 level 1 Voltage/duration of the 24V waveform corresponding to LEVEL2 level 2 Voltage/duration of the 24V waveform corresponding to LEVEL3 level 3 Load Dump Waveform Load-Dump Select decentralization load dump...
  • Page 32 ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING 5.6.1.1 Car Starting Voltage Short Drop Waveform This waveform simulates a momentary drop in supply voltage when starting a car. Figure 23 12V Waveform Figure 24 24V Waveform 5.6.1.2 Car Voltage Reset Test Waveform The simulated waveform is suitable for testing devices with a reset function.This voltage profile simulates the supply voltage dropping from Us min to 0.95 Us min at a rate of 5%, holding it for 5s, then rising to Us min and holding it for at least 10s for a...
  • Page 33 ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING Figure 25 Car Voltage Reset Test Waveform 5.6.1.3 Car Starting Waveform To simulate the effect of DC voltage with ripple under DC power supply when starting a car, the voltage curve in the t8 time period is upgraded from the original constant voltage straight line in the standard DIN40839 voltage waveform to a curve containing a superimposed AC voltage waveform with a frequency of 2Hz, and the curve should be selected according to the actual test requirements.
  • Page 34 ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING Figure 28 24V Built-in Waveform Setup Standard 5.6.1.4 Car Load Dump Characteristic Curve Inside most newer alternators, the magnitude of the load dump is suppressed (clamped) by the addition of a limiter diode. Load dump can occur due to corroded cables, poor contact or intentional disconnection from the battery while the engine is running.
  • Page 35 ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING Type Parameter Minimum Test Requirement a (V) 79≤U ≤101 151≤U ≤202 a (Ω) 0.5≤Ri≤4 1≤Ri≤8 10 pulses every 1 minute (ms) 40≤t ≤400 100≤t ≤350 (ms) – – If not otherwise agreed, please use a higher voltage level with higher internal resistance or a lower voltage level with lower internal resistance.
  • Page 36: Lv124

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING 79≤U ≤101 151≤U ≤202 Fixed at 35V, not Specified by *(V) customer (default: settable. 5 pulses every 1 minute (Ω) 0.5≤Ri≤4 1≤Ri≤8 (ms) 40≤t ≤400 100≤t ≤350 (ms) – – If not otherwise agreed, use a higher voltage level with higher internal resistance or a lower voltage level with lower internal resistance.
  • Page 37 ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING 5.6.2.1 E-04 Jump Start This waveform simulates the pulses generated during external startup of a car.The maximum experimental voltage is generated from the operating car and its raised chassis harness voltage. The waveform is shown below. Figure 31 E-04 Jump Start E-07 Supply voltage Slow Drop and Rise Simulates the waveform of slow drop and slow rise of the supply voltage, such as the...
  • Page 38: Charge

    6. Press "Shift "+"On/Off" on the front panel to trigger the waveform output. At this time, the status bar will display "Running". 5.7 Charge The N35100 series can be used to test batteries and overcapacity devices for charging under DC power. Figure 34 Charge Mode...
  • Page 39: Dischage

    When the charging function is switched on again, the charging capacity and charging time will be reset to zero and re-calculated. 5.8 Discharge The N35100 series can be used to test batteries and overcapacity devices for discharging under DC power. Figure 35 Discharge Mode...
  • Page 40: Resi

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING P-Discharge To set discharging power Judge the voltage value after discharging is completed, when the actual voltage reaches the cut-off voltage, the power V-End supply will automatically turn off the output. 0 means turn off this judgement, default is 0 Judge the capacity after discharging is completed, when the actual capacity reaches the cut-off capacity, the power supply...
  • Page 41 ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING output power of the power supply respectively, and the actual value depends on the load condition; R-Set is the analogue value of the internal resistance of the battery, corresponding to Eq. Ri. 5.10 CR Resistive mode is to adjust the input current according to the input voltage to discharge in constant resistance mode.
  • Page 42: Protection

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING 5.11 Protection On protection interface, users can set various protection parameters. When protection occurs, alarm information will be shown on the screen. Methods to enter Protection: Sh ift P R O T Method 1: Press first and then on the front panel directly.
  • Page 43 ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING Overvoltage protection delay time Overcurrent protection point Overcurrent protection Overcurrent protection delay time Over Power Protection Point Over power protection Over Power Protection delay time When the UVP setting is non-zero, it indicates that the function is turned on. If the device meets the following conditions at the same time, the UVP warning will appear: 1.
  • Page 44 ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING 2. The terminal voltage is greater than the set OVP point and the duration is greater than the protection delay time When OVP occurs, the device responds as follows: 1. Output of the device is switched off 2.
  • Page 45: Application

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING 2. The buzzer sounds once intermittently and three times continuously. 3. OTP will be displayed in black letters on a red background in the status bar. 5.12 Application Figure 39 Application1 Table 16 Item Description Time Shutdown Default OFF, ON for turning on.
  • Page 46 ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING Range 0.1~10s. After setting, press ON/OFF and Delay On then the power supply will delay the output accordingly. Range 0.1~10s. After setting, press ON/OFF and Delay Off then the power supply will delay the output accordingly.
  • Page 47: System

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING Corresponding interface, OFF/ONOFF/Clear can be selected, setting ONOFF Input Signal 3 indicates power output ON/OFF control, and setting Clear indicates clearing of power failure status. Set the trigger mode of Input Signal 3, there are two Control Mode modes, Trigger/Hold: Trigger means rising edge trigger, Hold means one action for one level change.
  • Page 48 ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING Figure 40 System Table 18 Item Function The default IP address is 192.168.0.123. The computer IP must be on the same network segment as the power supply, otherwise the connection will not be successful. Subnet Mask Default 255.255.255.0 Gateway...
  • Page 49: Factory Reset

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING 5.14 Factory Reset Method to enter Factory Reset: E n t e r Press →choose Factory Reset by M e n u →press E n t e r After selecting Confirm, please press to complete the setting. Note: After setting completed, it will take effect only after N35100 is restarted.
  • Page 50 ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING Figure 42 About Us...
  • Page 51: Application Software Installation & Configuration

    Please plug the Ethernet cable to PC Ethernet port and the other side to N35100 LAN port. After N35100 series is turn on, enter the system configuration interface to check the network IP. PC needs to keep the same network segment with N35100 in order to search the device.
  • Page 52 ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING Figure 44 Windows 7 settings Figure 45 Windows 7 settings Set Turn off the display and Put the computer to sleep to Never. Figure 46 Windows 7 settings...
  • Page 53 ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING  Windows 10 settings Click Start→Click Settings. Figure 47 Windows 10 settings Click System.
  • Page 54 ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING Figure 48 Windows 10 settings Click Power & sleep. Figure 49 Windows 10 settings Select Never for both options under Sleep.
  • Page 55: Network Ip Address Setting

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING Figure 50 Windows 10 settings 6.1.3 Network IP Address Setting The default IP of LAN port is 192.168.0.XXX (range from 0 to 255). Before operation, the computer IP should be assigned to the same network segment of N35100. But IP addresses should be different.
  • Page 56 ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING  Windows 7 Setting Click Start→Click Control Panel→Click Network and Sharing Center. Figure 51 Network IP Address Setting Click Change adapter settings. Figure 52 Network IP Address Setting...
  • Page 57 ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING Select the network→Right click and choose Properties. Figure 53 Network IP Address Setting Click Internet Protocol Version 4(TCP/IPv4) and fill the below information and press Figure 54 Network IP Address Setting...
  • Page 58 ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING Click Start→Input cmd. Input ping 192.168.0.123(default IP of N35100) and check if N35100 can communicate properly. Figure 55 Run Command If communicating properly, the below information will be reverted. Figure 56 Communication Test  Windows 10 Setting Click Start→Click Set→Click Network &...
  • Page 59 ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING Figure 58 Network IP Address Setting Select the network→Right click and choose Properties. Figure 59 Network IP Address Setting Click Internet Protocol Version 4(TCP/IPv4) and fill the below information and press...
  • Page 60 ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING Figure 60 Network IP Address Setting Click Start→Input cmd. Input ping 192.168.0.123(default IP of N35100) and check if N35100 can communicate properly. Figure 61 Run Command If communicating properly, the below information will be reverted. Figure 62 Communication Test...
  • Page 61: Application Software Installation And Uninstallation

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING 6.2 Application Software Installation and Uninstallation 6.2.1 Installation 1) Find the installation program "setup.exe" from the USB flash drive in accessory bag. 2) Make double-click on the file and begin installation. Figure 63 Program Installation 3) Click Next as prompted until the installation is completed.
  • Page 62: Uninstallation

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING Figure 64 Installation Completed 6.2.2 Uninstallation Methods for uninstallation: Method 1: Program uninstallation can be completed through Uninstall Program in Control Panel of the operating system, or by right-clicking the shortcut and selecting uninstall. Method 2: Find the setup program in your computer disk and delete.
  • Page 63 ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING It includes Online and Offline, Channel Config, System Config, History Data, graph. 2. Menu It includes Hardware Config, Channel Config. 3. Shortcut menu 4. Log Operation, fault and status information can be found in Log.
  • Page 64: Online And Scan

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING 6.3.1 Online and Scan Operation Steps: 1. Click Hardware Config. 2. Click Scan→Select 192.168.0.XXX network→Click OK→Click Save after the device is searched. Figure 67 Application Interface Figure 68 Hardware Configuration...
  • Page 65: Channel Configuration

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING 6.3.2 Channel Configuration Operation Steps: 1. Click Channel Config. 2. Change Channel: Click Channel→Input contents 3. Default Channel: Click Other→Set default value, Channel will be named from CH1 to CH99. 4. Multiple Devices Operation: Click Other→Select ON-All, quickly select all devices in the segment.
  • Page 66: Normal Mode

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING mode, output voltage, output current, output power. 4. Start Logging: Click the Start Logging to save data, the file data (.ndat format) will be automatically saved to the history data, users can choose to view in the History Data.
  • Page 67: Seq Test

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING Figure 71 Normal Mode 6.3.5 SEQ Test Operation steps: Click SEQ Test to edit SEQ file→click ON and OFF to execute or end file. As shown in the figure below: Figure 72 SEQ Test...
  • Page 68: Charge

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING Figure 73 SEQ Edit 6.3.6 Charge Click Charge to test charging voltage, current and to set end voltage, end current etc. →click Submit→click ON and OFF to start or stop test. As shown in the figure below:...
  • Page 69: Discharge

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING Figure 74 Charge 6.3.7 Discharge Click Discharge to test discharging voltage, current and to set end voltage, end time etc. →click Submit→click ON and OFF to start or stop test. As shown in the figure below: Figure 75 Discharge 6.3.8 CR Click CR to simulate resistance/current/power value→click Submit→click ON and...
  • Page 70: Car Wave

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING Figure 76 CR 6.3.9 Car Wave Car Wave test for selection of waveform. Level 1: Select the test execution standard; Level 2: Select the test content; Level 3: Select the standard value of the test content. After selecting, click Apply and ON to start test.
  • Page 71: Resistance Simulation

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING 6.3.10 Resistance Simulation In this mode, users can set parameters of V-Open, I-Max, P-Max, Internal R-Set. Click Submit and ON to start test. Figure 78 Resistance Simulation...
  • Page 72: Protection

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING 6.3.11 Protection Click Protection to set relevant parameters →click OK→close interface. As shown in the figure below: Figure 79 Protection 6.3.12 Output Parameter Click Protection to set relevant parameters →click OK→close interface. As shown in the figure below:...
  • Page 73: Maintenance And Self-Inspection

    Due to system upgrade or hardware problem, the device may break down. Please do the following necessary inspection to eliminate the troubles, which can save your maintenance and time cost. If the troubles cannot be recovered, please contact NGI. The inspection steps are as below.
  • Page 74: Main Technical Data

    ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING 8 Main Technical Data Warning 1. The measurement accuracy is identified by the following conditions: within one year after calibration, operation temperature between 18℃ and 28℃, and the relative humidity up to 80%. 2. Please prepare for half an hour before precision measurement. 3.
  • Page 75 ELECTRONIC SOLUTION PROVIDER FOR INTELLIGENT MANUFACTURING Resolution 10mΩ Setting Accuracy (Vin/Rset)*0.1%+0.1%IF.S. (23±5℃) Voltage Measurement Range 0-80.8V Resolution Accuracy(23±5℃) ≤0.02%+0.02%F.S. Temperature ≤30PPM/℃ Coefficient Current Measurement Range 0-55.55A Resolution Setting Accuracy ≤0.1%+0.1%F.S. (23±5℃) Temperature ≤50PPM/℃ Coefficient Output Protection Range 0~80,8V(101%F.S.) 0~55.55A(101%F.S.) 0~2525W(101%F.S.) Hardware Limit Range Voltage 105%F.S.
  • Page 76 Operating temperature: 0℃-40℃, storage temperature: -10℃~70℃ Operating Altitude <2000m, relative humidity: 5%-90%RH(non-condensing), Environment atmospheric Dimension(mm) 214mm(W)*422mm(D)*44.0mm(H) Net Weight 5.1kg Note 1: The above specifications are only for reference. For other specifications, please contact NGI. Note 2: All specifications are subject to change without notice.