Generator Overview General Dimensions Unit Connections Unit Controls Air Connection Commissioning Connecting to the application Tubing Lengths Service Requirements Service Schedule Cleaning Troubleshooting Contact us SCIEX Support Page 2 Page 2 / 20 SCIEX Gas Generator N38 Operator Guide RUO-IDV-05-14029-B...
Users can refer to the contents page to find the relevant information. Review each of the following sections carefully. Thank you for selecting SCIEX to meet your gas generation needs, and should you require any further assistance or support do not hesitate to contact SCIEX.
If the user employs an operating procedure, item of equipment or a method of working which is not specifically recommended by SCIEX, the user must ensure that the equipment will not be damaged or become hazardous to persons or property.
Drain Outlet 1 × ¼” BSPP Pressure Gauges NOTE – These performance characteristics are valid only when paired with an approved SCIEX application. Refer to current application matrix for a list of supported products. General Dimensions cm (inches) W × D × H 42.8 ×...
Unpacking Although SCIEX takes every precaution with safe transit and packaging, it is advisable to fully inspect the unit for any sign of transit damage. Check ‘SHOCKWATCH’ and ‘TIP-N-TELL’ labels for signs of rough handling prior to unpacking. Any damage should be reported immediately to the carrier and SCIEX.
6. No. 8 Wall Plug × 7. No. 8 Screw × All of the generators output ports are located on the output panel at the side of the unit. Page 7 / 20 SCIEX Gas Generator N38 Operator Guide RUO-IDV-05-14029-B...
Refer to the drawing on the following page for the general dimensions of the unit. Minimum Operating Ambient Temperature: 5 °C (41 °F) Maximum Operating Ambient Temperature: 30 °C (86 °F) Page 8 / 20 SCIEX Gas Generator N38 Operator Guide RUO-IDV-05-14029-B...
428 mm / 16.8” 250 mm / 9.8” The generator must always be placed on a flat, level surface. Failure to do so will affect the performance of the generator. WARNING Page 9 / 20 SCIEX Gas Generator N38 Operator Guide RUO-IDV-05-14029-B...
Air Connection SCIEX Gas Generator N38 should be connected to a clean, dry, OIL - FREE source of compressed air. A minimum inlet pressure of 120 psig (8.3 barg) is required. Any doubts as to the suitability of your compressed air supply should be referred to SCIEX or any of their authorised partners.
In circumstances of no demand, the Generator remains operational and will provide Nitrogen immediately as demand resumes. Page 12 / 20 SCIEX Gas Generator N38 Operator Guide RUO-IDV-05-14029-B...
” O/D, ⁄ ” I/D). Tubing and fittings not supplied in the fittings kit. > 40 metres: Contact SCIEX with the relevant distance and we will calculate the flow resistance and the tubing size required. A combination of ”/ ⁄...
* Quote your generator serial number when ordering your Maintenance Kit to ensure selection of the appropriate kit for your generator. Note: To be serviced by manufacturer. Page 14 / 20 SCIEX Gas Generator N38 Operator Guide RUO-IDV-05-14029-B...
Under no circumstances should any solvents or abrasive cleaning solutions be used as these can contain fumes that could be harmful to the generator. CAUTION Care should be taken with Leak Detections Liquids. CAUTION Page 15 / 20 SCIEX Gas Generator N38 Operator Guide RUO-IDV-05-14029-B...
Troubleshooting Problem Possible Solution • Check pressure gauges are showing normal pressure. The mass spec is reporting low pressure. • Contact your Technical Assistance Centre. Page 16 / 20 SCIEX Gas Generator N38 Operator Guide RUO-IDV-05-14029-B...
SCIEX and its representatives maintain a staff of fully-trained service and technical specialists located throughout the world. They can answer questions about the system or any technical issues that might arise. For more information, visit the SCIEX website at sciex.com or contact us in one of the following ways: •...
Page 20
Any such use is intended only to designate those manufacturers’ products as supplied by SCIEX for incorporation into its equipment and does not imply any right and/or license to use or permit others to use such manufacturers’ and/or their product names as trademarks.
Need help?
Do you have a question about the Gas Generator N38 and is the answer not in the manual?
Questions and answers