SAFETY, ENVIRONMENTAL, AND REGULATORY INFORMATION
PXIe-5163
PXIe, 200 MHz, 1 GS/s, 14-bit PXI Oscilloscope
Read this document and the documents listed in the additional resources section about installation, configuration, and operation of this equipment before you
install, configure, operate, or maintain this product. Users are required to familiarize themselves with installation and wiring instructions in addition to
requirements of all applicable codes, laws, and standards.
Icons
Notice—Take precautions to avoid data loss, loss of signal integrity, degradation of performance, or damage to the model.
Caution—Take precautions to avoid injury. Consult the model documentation for cautionary statements when you see this icon printed on the model.
Cautionary statements are localized into French for compliance with Canadian requirements.
Shock Warning—Take precautions to avoid electrical shock.
Hot Surface—Take precautions to avoid physical burns.
Safety
Caution Observe all instructions and cautions in the user documentation. Using the model in a manner not specified can damage the model and
compromise the built-in safety protection. Return damaged models to NI for repair.
Attention Suivez toutes les instructions et respectez toutes les mises en garde de la documentation utilisateur. L'utilisation d'un modèle de toute autre
façon que celle spécifiée risque de l'endommager et de compromettre la protection de sécurité intégrée. Renvoyez les modèles endommagés à NI pour
réparation.
Caution All wiring must be insulated for the highest voltage used.
Attention Tout le câblage doit être isolé pour la plus haute tension utilisée.
Caution If the device has been in use, it may exceed safe handling temperatures and cause burns. Allow the device to cool before removing it from
the chassis.
Attention Si l'appareil a été utilisé, il peut avoir atteint des températures trop élevées pour être manipulé en toute sécurité, ce qui peut provoquer des
brûlures. Laisser l'appareil refroidir avant de le retirer du châssis.
Safety Voltage and Frequency
Connect only voltages that are below these limits.
1
Maximum input overload
50 Ω
1 MΩ
Notice Signals exceeding the maximum input overload may cause damage to the device.
Frequency
Measurement category
Caution Do not connect the PXIe-5163 to signals or use for measurements within Measurement Categories III or IV.
Attention Ne connectez pas le PXIe-5163 à des signaux et ne l'utilisez pas pour effectuer des mesures dans les catégories de mesure III ou IV.
1
Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
2
Derate above 500 kHz at 20 dB/dec until 5 MHz, then derate at 10 dB/dec.
|Peaks| ≤5 V, nominal
2
|Peaks| ≤250 V RMS,
nominal
200 MHz
CAT II
Need help?
Do you have a question about the PXIe-5163 and is the answer not in the manual?
Questions and answers