Summary of Contents for NOVACARE ASX digital 5 inch

  • Page 1 User Manual ASX digital 5 inch Anti-decubitus Alternating Pressure System...
  • Page 2: Table Of Contents

    1. General information Read the following instructions for use carefully and observe the safety and warning information. The novacare Alternating Pressure System ASX digital 5 inch is a system for treatment and preven- ® tion of ulcers/decubitus. As a medical aid, the product is mainly given to medical experts or professional users. Professional users are people who, due to their medical or comparable training, have sufficient knowledge of the product and its use.
  • Page 3: Product Information

    Intended use The novacare alternating pressure system ASX digital 5 inch is suitable for use with decubitus ulcers ® of category I-III according to EPUAP/NPUAP and is designed for a maximum weight of 160 kg. For...
  • Page 4 In the case of patients in pain or with a low pain treshold, the system should only be operated in CLP or static mode or a more suitable novacare® system should be used. Contraindications The system should not be used for patients with multiple trauma, fractures of the spine, pelvis and extremities, or intracranial injuries or concussion.
  • Page 5: Installation & Use

    Accessories The medical products from novacare gmbh may only be used with the accessories provided for ® them. The accessories are available separately from novacare gmbh customer service. ® 4. Installation & Use Environmental conditions...
  • Page 6 With slight resistance, it should be possible to slide a flat Key lock button hand between the vented cells and the person lying on them (= practical test). Correct settings if necessary. Note the key lock function. ASX digital 5 inch...
  • Page 7 If the pressure is too low for a short time, e.g. due to repositioning or cycle change, this is only in- dicated by a visual signal, the low pressure indicator "LOW" flashes. If the low pressure persists for more than five minutes, the acoustic alarm is also activated. www.novacare.org...
  • Page 8: Faults & Troubleshooting

    Do not use abrasive agents containing phenols or diethylene glycol monoethyl ether. Remove the cover of the system regularly and wash in the washing machine at up to 95°C with commercially available detergents and then dry gently at low temperatures. ASX digital 5 inch...
  • Page 9: Disposal & Environment

    Use in the nursing area (change of patient): Preparation in accordance with national regulatory requirements. Novacare gmbh expressly recom- ® mends reprocessing in accordance with the recommendations of the German Robert Koch Institute on the hygiene requirements for the reprocessing of medical products.
  • Page 10: Abbreviations & Symbols

    95 ° C do not bleach Clothes dryer can be used. Maximum outlet temperature 60 ° C. do not iron do not dry clean professional chemical cleaning (normal process) ASX digital 5 inch...
  • Page 11: Warranty

    9. Warranty Novacare gmbh assumes a guarantee on every patient hoist in the novacare product range that is ® ® defective due to manufacturing and / or material defects within 24 months of purchase. If an examination by novacare gmbh shows that warranty claims have been made that relate to ®...
  • Page 12 Bruchstrasse 48, 67098 Bad Dürkheim, Germany Tel. +49(0)6322-9565-0, Fax +49(0)6322-9565-65, www.novacare.org Misprints and errors are reserved. Color, material and construction changes possible.