NOVACARE ASX digital Junior User Manual

Anti-decubitus alternating pressure system

Advertisement

Quick Links

User Manual
ASX digital Junior
Anti-decubitus Alternating Pressure System

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ASX digital Junior and is the answer not in the manual?

Questions and answers

Summary of Contents for NOVACARE ASX digital Junior

  • Page 1 User Manual ASX digital Junior Anti-decubitus Alternating Pressure System...
  • Page 2: Table Of Contents

    1. General information Read the following instructions for use carefully and observe the safety and warning information. The novacare Alternating Pressure System ASX digital Junior is a system for treatment and pre- ® vention of ulcers/decubitus. As a medical aid, the product is mainly given to medical experts or professional users.
  • Page 3: Product Information

    The novacare ® alternating pressure system ASX digital Junior is suitable for use with decubitus ulcers of category I-IV according to EPUAP/NPUAP and is designed for a maximum body height of 130 cm and a maximum weight of 80 kg. For taller/heavier patients alternating pressure systems of the novacare ®...
  • Page 4 In the case of patients in pain or with a low pain threshold, the system should only be operated in CLP or static mode or a more suitable novacare system should be used.
  • Page 5: Installation & Use

    3,5 cm ® The medical products from novacare gmbh may only be used with the accessories provided for ® them. The accessories are available separately from novacare gmbh customer service. ® 4. Installation & Use Environmental conditions Storage...
  • Page 6 Key lock button With slight resistance, it should be possible to slide a flat hand between the vented cells and the person lying on them (= practical test). Correct settings if necessary. Note the key lock function. ASX digital Junior...
  • Page 7 45 minutes. Short-term low pressure, e.g. patient movement or cycle change, is only in- dicated with an optical signal, the low pressure indicator „LOW“ flashes. If the low pressure remains longer than five minutes, the acoustic alarm is also activated. www.novacare.org...
  • Page 8: Faults & Troubleshooting

    Do not use abrasive agents containing phenols or diethylene glycol monoethyl ether. Remove the cover of the system regularly and wash in the washing machine at up to 70°C with commercially available detergents and then dry gently at low temperatures. ASX digital Junior...
  • Page 9: Disposal & Environment

    Use in the nursing area (change of patient): Preparation in accordance with national regulatory requirements. Novacare gmbh expressly re- ® commends reprocessing in accordance with the recommendations of the German Robert Koch Institute on the hygiene requirements for the reprocessing of medical products.
  • Page 10: Abbreviations & Symbols

    IPXX Ingress Protection for electrical equipement maximum washing temperature, see section 6, Maintenance & Repair do not bleach Clothes dryer can be used. Maximum outlet temperature 60 ° C. do not iron professional chemical cleaning (normal process) ASX digital Junior...
  • Page 11: Warranty

    9. Warranty Novacare gmbh assumes a guarantee on every patient hoist in the novacare product range that ® ® is defective due to manufacturing and / or material defects within 24 months of purchase. If an examination by novacare gmbh shows that warranty claims have been made that relate ®...
  • Page 12 Bruchstrasse 48, 67098 Bad Dürkheim, Germany Tel. +49(0)6322-9565-0, Fax +49(0)6322-9565-65, www.novacare.org Misprints and errors are reserved. Color, material and construction changes possible.