eitan medical Sapphire Manual

Setup and mid - infusion actions
Hide thumbs Also See for Sapphire:

Advertisement

Quick Links

Sapphire
Setup and
Mid - Infusion Actions
The information provided here
applies to Sapphire SW revisions 15
onwards, and does not replace the
Sapphire
user manual
.
TM

Advertisement

loading
Need help?

Need help?

Do you have a question about the Sapphire and is the answer not in the manual?

Questions and answers

Summary of Contents for eitan medical Sapphire

  • Page 1 Sapphire ™ Setup and Mid - Infusion Actions The information provided here applies to Sapphire SW revisions 15 onwards, and does not replace the Sapphire user manual...
  • Page 2 Priming the Sapphire set • The Sapphire set can be primed via the pump, some sets can also be primed manually (non-PAV sets) • It is possible to prime with the pump before or after programming the infusion, in all delivery modes •...
  • Page 3 Setup Actions Changing the delivery mode Upon turn on, the Sapphire pump is automatically set to the last delivery mode used. To select a different delivery mode: • From the start up screen, select Options • From the options screen, select Delivery mode •...
  • Page 4: Mid-Infusion Actions

    Mid-Infusion Actions Locking and Unlocking the Pump The pump can be locked to avoid unintentional or unauthorized touching of the screen. Locking Unlocking the Pump the Pump No Code needed to unlock Code needed to unlock The pump can be unlocked with any code level, functions will be limited according to the code level entered.
  • Page 5 Mid-Infusion Actions Viewing Infusion Program During infusion the entire infusion program can be viewed on the Infusion values menu. To access infusion values during infusion: • Press View/Edit key • Select View system option • Press the infusion values row •...
  • Page 6 Mid-Infusion Actions Turning Off the Pump (Quit infusion or Resume) The pump can be turned off at any time by pressing the On/Off hard key and confirming on the screen. During infusion, the pump can retain the infusion status and resume it at a later time Turn off without resuming later...
  • Page 7 Mid-Infusion Actions PCA/PCEA/PIEB Giving a Clinician Bolus in PCEA PIEB infusions • Unlock the pump if it’s locked • Press View/Edit • Select C.bolus • Enter the High level code and press OK • Enter the bolus dose to be given and press OK •...
  • Page 8 2. Select Immediate Taper Down 3. Press OK to start taper down When the pump is set to High/Medium code, the user can adjust the taper down time Contact information Eitan Medical Headquarters support@eitanmedical.com +972.73.238.8888 Eitan Medical US customerservice@eitanmedical.com +1.877.541.9944 Eitan Medical UK customersupport.uk@eitanmedical.com...