The operator must observe and/or act according to the information in order to avoid any mechanical or electrical Caution damage to the equipment. The operator must observe and/or act according to the information in order to avoid any personal injury. Warning 1526 Pre-installation Manual - AWP 2530...
Handle the crate with Never expose the crate to Do not stack the crates. Do not tilt the crate. indicated by the arrows care. water, or place it in a high- is always up humidity area. 1526 Pre-installation Manual - AWP 2530...
If width of the door is… Then… > 1850 mm No action is required as the crated unit can be transported (72.8") immediately to the installation site. > 1650 mm The Service Technician has to unpack the unit. (65") 1526 Pre-installation Manual - AWP 2530...
Page 10
Transportation Transporting the unit Pre-installation Manual - AWP 2530 1526...
2-6 for recommended cable type Installation kit Installation kit is included with the unit. It comprise the parts necessary to make the in - stallation. See Service Manual for installation part list. 1526 Pre-installation Manual - AWP 2530...
The recommended minimum free space around the unit is: 1200 mm (47.2") in front of the unit, • 800 mm (31.5") on the sides and at the back of the unit. • T33012 Pre-installation Manual - AWP 2530 1526...
Page 13
Machine dimensions and connections location 2 . 2 " ) . 6 " " ) 3 . 4 Light finisher exhaust Exposure exhaust Power supply Water inlet Dryer exhaust Compressed air Water outlet T11375 1526 Pre-installation Manual - AWP 2530...
All the exhaust must be connected to the appropriate location (directly to the outside or to the building exhaust installation). Make sure that the airflow is correct. Dryer, min. 100 m Light finisher, min. 400 m Pre-installation Manual - AWP 2530 1526...
Provide for additional cable protection, e.g. cable covers, if cable is exposed to heavier transport such as fork-lift trucks etc. The unit is Class 1 equipment. Therefore, the unit must be connected to earth to avoid electrical shocks. Please be aware of double pole/neutral fusing. 1526 Pre-installation Manual - AWP 2530...
The fuses must have a breaking capacity of min. 100 kA. If using automatic circuit breakers make sure that they are Type D. Power consumption Power consumption at … 230/400 VAC operation: approx. 15 kW Pre-installation Manual - AWP 2530 1526...
(cables, fuses, etc.) necessary to connect the unit to the mains? D. Is there a house electrician available? E. Are there any known problems in the building where the unit will be installed? If so, explain: ....................................................................................................................................................1526 Pre-installation Manual - AWP 2530...
Page 18
F. Can a hose (140 mm diameter) be connected to the exhaust system? G. Is there a solvent installation available (diameter 20 mm)? H. Availability of warm water for cleaning of unit? 4. Disposition of the various supplies and equipment on the installation site: Pre-installation Manual - AWP 2530 1526...