Caution damage to the equipment. The operator must observe and/or act according to the information in order to avoid any personal injury. Warning 1505 Pre-installation Manual - AWP 1116 PD HQ...
Never expose the crate to Do not stack the crates. Do not tilt the crate. indicated by the arrows care. water, or place it in a high- is always up humidity area. 1505 Pre-installation Manual - AWP 1116 PD HQ...
> 2400 mm No action is required as the crated unit can be transported (97.2") immediately to the installation site. > 2100 mm The Service Technician has to unpack the unit. (82.7") 1505 Pre-installation Manual - AWP 1116 PD HQ...
Page 10
Transportation Transporting the unit Pre-installation Manual - AWP 1116 PD HQ 1505...
2-6 for recommended cable type Installation kit Installation kit is included with the unit. It comprise the parts necessary to make the in - stallation. See Service Manual for installation part list. 1505 Pre-installation Manual - AWP 1116 PD HQ...
The recommended minimum free space around the unit is: 1500 mm (59.1") in front of the unit, • 800 mm (31.5") on the sides and at the back of the unit. • T11418 Pre-installation Manual - AWP 1116 PD HQ 1505...
Dimensions and connections location Dimensions and connections location . " . " . " . " . " . " b l e p l y i r h t e r T 419 1505 Pre-installation Manual - AWP 1116 PD HQ...
Exhaust All the exhaust must be connected to the appropriate location (directly to the outside or to the building exhaust installation). Make sure that the airflow is correct. Exhaust, min. 200 m Pre-installation Manual - AWP 1116 PD HQ 1505...
The unit is Class 1 equipment. Therefore, the unit must be connected to earth to avoid electrical shocks. Please be aware of double pole/neutral fusing. 1505 Pre-installation Manual - AWP 1116 PD HQ...
The fuses must have a breaking capacity of min. 100kA. If using automatic circuit breakers make sure that they are Type D. Power consumption Power consumption at … 230/400 VAC operation: approx. 26 kW Pre-installation Manual - AWP 1116 PD HQ 1505...
(cables, fuses, etc.) necessary to connect the unit to the mains? D. Is there a local electrician available? E. Are there any known problems in the building where the unit will be installed? If so, explain: ....................................................................................................................................................1505 Pre-installation Manual - AWP 1116 PD HQ...
Page 18
F. Can a hose (100 mm diameter) be connected to the exhaust system? G. Is there a water installation available (diameter 1/2")? H. Availability of warm water for cleaning of unit? 4. Disposition of the various supplies and equipment on the installation site: Pre-installation Manual - AWP 1116 PD HQ 1505...