(User replaceable) Reference Junction (under Glass pH Glass Bulb) Electrode pH Sensor Temperature Protective Cap Sensor LO BATT °C Measurement key pH Meter ® MYRON L COMPANY Wrist/neck strap slot (user supplied) For detailed explanations, see Table of Contents 04-12-24...
Dimensions (LxWxH) 7.7x2.7x2.5 in. normal wear, abuse or tampering. This warranty is limited to the repair or 196x68x64 mm replacement of the pH1 only. The Myron L Company assumes no other Weight 10.8oz./310g responsibility or liability. Case Material pH Sensor Well Capacity 0.04 oz./1.2 ml...
(in order of preference): buffer solution or a damaged pH Sensor. Myron L Storage Solution, an almost saturated KCl solution, pH 4 buffer, (ref. Buffer Solutions, pg. 13) or at least a strong table salt solution. Not C.
Press to verify the pH calibration. (If the display reads 7.00, finger until reading agrees with buffer solution. skip the pH Zero Calibration and proceed to section b. pH Gain If the instrument will be used to read both acids and bases, Calibration.
Rinsing such liquids from the pH sensor C. pH Sensor Replacement well and refilling well with Myron L Storage Solution, a saturated KCl Order model RPG. When ordering, be sure to include the model and solution, pH 4 buffer, or a salty tap water, will extend the useful life.
A C C E S S O R I E S A. Buffer Solutions ® pH buffers are available in pH values of 4, 7 and 10. Myron L Company buffer solutions traceable NIST certified references and are color-coded for instant identification. They are also mold inhibited and accurate to within ±0.01 pH units @ 25°C.
Page 9
® (H ) ions will increase acidity, while an increase in (OH ) ions will increase D. The Myron L Integral pH Sensor Glass Surface alkalinity. The total concentration of ions is fixed as a characteristic of The sensor in the pH1 (figure at right)
XII. GLOSSARY N O T E S Logarithm - An arithmetic function. See pH Units, pg. 13. For details on specific areas of interest refer to Table of Contents.