PT4
Free Chlorine Equivalent
(FC
)
E TM
& Temperature Pen
ACCURATE
S
RELIABLE
S
Professional Water Analysis
• Accuracy of up to ±0.3 ppm FC
• Proprietary Sensor Design
• Calibration with pH and ORP
Standard Solutions
• Automatic Temperature Compensation
• Temperature Readout
• Waterproof
• Powered by one N Type Battery
(Included)
PACkAgE CoNTENTS:
PT4 Pocket Tester Pen - Battery Installed
Scoop
Soaker Cap
Pocket Clip
Holster
Lanyard
ORP Cleaning Paper
EDITED FOR USE
Minimum height 5 mm
SIMPLE
S
E
ULTRAPEN™ PT1
Conductivity, TDS, Salinity
& Temperature
ULTRAPEN™ PT2
pH & Temperature
ULTRAPEN™ PT3
oRP & Temperature
MADE IN USA
MADE IN USA
ORIGINAL

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ULTRAPEN PT4 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Myron L ULTRAPEN PT4

  • Page 1 ULTRAPEN ™ Free Chlorine Equivalent E TM ORIGINAL & Temperature Pen EDITED FOR USE ACCURATE Minimum height 5 mm RELIABLE SIMPLE Professional Water Analysis • Accuracy of up to ±0.3 ppm FC • Proprietary Sensor Design • Calibration with pH and ORP Standard Solutions •...
  • Page 2 INTRODUCTION Thank you for purchasing the Myron L ULTRAPEN™ PT4 FC Pen. This instrument is designed to be extremely accurate, fast, and simple to use in diverse water quality applications. Advanced features include automatic temperature compensation in calibration mode; highly stable microprocessor-based circuitry;...
  • Page 3: Operating Instructions

    LIVE measurement will time out after 5 minutes OR push and release the push button to turn the PT4 off at any time during LIVE measurement. IV. Calibration The Myron L Company recommends calibrating twice a month, depending on usage.
  • Page 4 NOTE: If the measurement is NOT within calibration limits for any reason, “Error” will display. Check to make sure you are using a proper Myron L Company pH Buffer or ORP Standard Solution. If the solution is correct, clean the sensor as described in Sensor Cleaning section on page 4.
  • Page 5 III. Sensor Cleaning (additional sensor cleaning methods at www.myronl.com) Cleaning the sensor: The Myron L Company recommends cleaning your sensor every two weeks, however this depends on application and frequency of use. Indications of a dirty sensor are slower and/or erroneous readings.
  • Page 6: Specifications

    Replacement FC Sensor (with instructions) OECP-25/-100 ORP Electrode Cleaning Paper, QTY 25 or QTY 100 For additional calibration kits available, contact the Myron L Company. MYRON L WARRANTY The PT4, excluding the FC sensor, has a one (1) year limited warranty.