Operation; Introduction; Compensation Consideration; Calibration - HP 41941A Manual

Impedance probe kits
Table of Contents

Advertisement

3-1. INTRODUCTION

This section provides the information necessary to use the 41941A/B.
and Notes are given throughout, and they should be followed to insure operator
TIONS,
safety and servicaebility of the unit.
For detailed information on 4194A operation, refer to the
Manual .
3-2. COMPENSATION CONSIDERATIONS
has two compensation methods, calibration and offset. Calibration compensates
The
4194A
for measurement errors caused by the measurement instrument and the probe. Offset com­
pensates for measurement errors caused by residual impedance and stray admittance be­
tween the calibration plane and the

3-2-1. Calibration

Calibration is required to compensate for measurement errors due to the frequency charac­
teristics of the probe. When the probe is calibrated using the supplied calibration standards,
the calibration plane of the probe extends to the fringe of the pin's top plane as shown in
Figure 3-1.
The calibration plane is the point at which impedance can be correctly
measured.
SECTION 3

OPERATION

Note
OUT.
-----Calibration
Figure 3-1. 41941A/B's Calibration Plane
WARNINGS, CAU­
4
19
4 A
s Operation
'
Plane
3 -1

Advertisement

Table of Contents
loading

This manual is also suitable for:

41941b

Table of Contents